SCT-1100 Quick Setup Guide
Key Functions
Key | Function | Key | Function | Key | Function |
---|---|---|---|---|---|
ZERO | In setup: scroll forward through parameters In numeric input: decreases digit to be modified |
MODE | In setup: quickly position the first step of a menu In numeric input: selects digit to be modified, from left to right |
C ON/OFF | Turn the instrument on/off In setup: exit a step without confirming the setting In numeric input: clears the present value |
TARE | In setup: scroll back through the functions In numeric input: increases digit to be modified |
In setup: enter into a parameter or to confirm a setting In numeric input: confirms the entry made |
Setup Menu
- Press the ON/OFF key (C) to turn the instrument on.
- Press the mode key (►) when the firmware version displays to enter the setup menu.
- Press the print key (◄◄) in the setup menu to view/select the parameter choices.
Parameter Selection and Configuration:
Parameter | Description |
---|---|
EYPE | Press and [up/down arrows] to set channel type (e.g., ch, trAn57, or dEP .ch). |
nChan | Press and [up/down arrows] to select the number of channels. |
ChAn | Press and [up/down arrows] to select the channel to calibrate. |
du.dec | Press and [up/down arrows] to set decimals and minimum division. |
CAPAC | Press and [up/down arrows] to set the scale capacity. |
CEL.CAP | Press and [up/down arrows] to set the total capacity of the load cells. |
CEL.SEN | Press and [up/down arrows] to set the mV/V sensitivity of the load cell. |
DEAD .Ld | Press and [up/down arrows] to set the dead load present on the load cell. |
0.CAL16 | Quick calibration of zero. Press and [up/down arrows] to set the zero value. For new acquisition of zero, with scale empty. Establishes new zero reference. |
SPAN | Apply calibration weight. Press and [up/down arrows] to enter applied calibration weight. |
Adc.uu | Check the load cell signal in millivolts. |
An.out | Select channel to be configured. |
RONAH | Set the analog output on net or gross weight. |
Ao2Er | Set the maximum value of analog output. |
Aon in | Set the analog output when scale shows zero weight. |
485 | Set the minimum value of analog output. |
SEE .Add | Set the key to be emulated by the input. |
bAud | Press and [up/down arrows] to set baud rate. |
Calibration Information:
Default Factory Calibration: The instrument has a default calibration with a capacity of 10,000 kg, load cell sensitivity of 2.000mV/V, and 1 division.
Theoretical Calibration: Use parameters EYPE, nChan, chan, du.dec, CAPAC, CEL.CAP, CEL.SEN, DEAD.Ld to perform a theoretical calibration.
Calibration Using Known Weight: Use parameters EYPE, nChan, chan, du.dec, CEL.CAP, 2Ero, SPAn to perform a standard calibration with a known weight. For parameters 2Ero & SPAn, press and wait for ERNOT, STORE?. Then press [enter key] to accept. For parameter SPAn, apply test weight before pressing [enter key].
Setpoint Programming
In the weigh mode, press and hold [enter key] to access output's setpoint programming.
Parameter | Description |
---|---|
outPut | Select channel to be configured. |
norni | Set the output normal state. |
FunC | Setpoint on the gross weight. |
Gro5 | Disabled. |
net | Setpoint on the net weight. |
Wiring Diagram
The document includes a detailed wiring diagram illustrating the connections for Load Cells 1, 2, 3, and 4, as well as power supply, analog output, serial port, inputs, outputs, and relays.
Pin Assignments:
- VE 12-24 VDC Power Supply: Pin 1 (+VDC +12-24 VDC), Pin 2 (GND 0 VDC (GND))
- Inputs and Outputs (Optoisolated Inputs Positive Logic (12-24 VDC, 5-20 mA max)): Pin 3 (COM Common Output), Pin 4 (IN1 Input 1), Pin 5 (IN2 Input 2)
- Relays: Pin 6 (RL1 Relay 1), Pin 7 (RL2 Relay 2), Pin 8 (COM Common Relay)
- Analog Output: Pin 9 (+ +20 mA), Pin 10 (- -0 mA (GND)), Pin 11 (V+ +10 V), Pin 12 (V- 0 V (GND))
- Serial Port: RS-485 (Pin 13 (A) 485 + Line, Pin 14 (B) 485 - Line), RS-232 (Pin 15 TX Transmission, Pin 16 RX Reception, Pin 17 GND Ground)
- Load Cell 1: Pins 18 (SIG+ Signal +), 19 (SIG- Signal -), 20 (SEN+ Sense +), 21 (SEN- Sense -), 22 (EXC+ Excitation +), 23 (EXC- Excitation -)
- Load Cell 2: Pins 24 (SIG+ Signal +), 25 (SIG- Signal -), 26 (EXC+ Excitation +), 27 (EXC- Excitation -)
- Load Cell 3: Pins 28 (SIG+ Signal +), 29 (SIG- Signal -), 30 (EXC+ Excitation +), 31 (EXC- Excitation -)
- Load Cell 4: Pins 32 (SIG+ Signal +), 33 (SIG- Signal -), 34 (EXC+ Excitation +), 35 (EXC- Excitation -)
Note: When settings are complete, press C until the indicator displays SAUEP. Press [enter key] to save setup and return to weigh mode. Pressing any other key exits the setup and discards changes.