F&P AIRVO 2

Disinfection Kit Manual

BEFORE YOU START

This User Manual is intended for healthcare professionals.

Read this User Manual including all warnings. Failure to do so may result in injury or increased risk of infection. Keep this manual in a safe place for future reference.

OTHER REFERENCES

1. OVERVIEW

The 900PT600 Disinfection Kit is used to achieve high-level disinfection of the AIRVO 2 / myAIRVO 2 humidifier outlet elbow. This manual itself describes the recommended reprocessing procedures to be carried out between patients, including cleaning, high-level disinfection using the supplied disinfection tube and filter, and correct storage.

WARNINGS

PACKAGE CONTENTS (900PT600)

AVAILABLE SEPARATELY

2. REPROCESSING PROCEDURES

The AIRVO 2 humidifier must be cleaned and disinfected between patients. During use, the patient interface, heated breathing tube, water chamber and outlet elbow at the heated breathing tube connection port may become contaminated. It is particularly important that the outlet elbow at the heated breathing tube connection port is cleaned and disinfected properly. This should take place as soon as possible after use.

Note: Do not clean the left-hand chamber port.

The following steps must be carried out between patients:

Procedure Description
Cleaning Clean outlet elbow and exterior surfaces with recommended cleaning agents, then wipe to remove any residue.
High-level Disinfection Carry out disinfection cycle with disinfection tube.
Storage Store AIRVO 2 appropriately.

In addition, consider replacing the air filter at the rear of the AIRVO 2 if a highly infectious patient is known to have used the unit.

Standard aseptic techniques to minimize contamination should be followed when handling the unit and accessories. This includes proper hand-washing, avoiding hand contact with connection ports, safe disposal of the used consumables and suitable storage of the unit after cleaning and disinfection. The use of protective gloves is strongly recommended.

The AIRVO 2 and its accessories are not compatible with washer-disinfector processes.

3. CLEANING

This cleaning step aims to remove any contamination from the outlet elbow (between the right-hand chamber port and the heated breathing tube connection port), before the disinfection cycle is carried out. It is important to carry out this cleaning step first to ensure that the disinfection cycle will be effective. In addition, it is important to clean the other outside surfaces of the unit.

Note: Do not clean the left-hand chamber port, as this can damage the non-return valve located at this port.

THE FOLLOWING EQUIPMENT IS REQUIRED:

WARNINGS

The products above have been tested by Fisher & Paykel Healthcare. If none of these are available, other cleaning agents may be used as long as they are pH-neutral, non-abrasive, non-toxic and non-corrosive. Do not use any cleaning agents incompatible with polycarbonate plastic (including, but not limited to, ammonia, ammonium hydroxide, caustic soda, iodine, methanol, methylated spirits, turpentine and alkaline bleaches such as sodium hypochlorite). Always remove any residue with a clean, damp, lint-free cloth.

PREPARE THE AIRVO 2 FOR CLEANING

CLEAN THE OUTLET ELBOW

You must physically clean the outlet elbow that connects the right-hand chamber port to the heated breathing tube connection port.

Note: Do not clean the left-hand chamber port, as this can damage the non-return valve located in this port.

There are two options for cleaning the elbow:

USING THE CLEANING SPONGE-STICK
USING A DISPOSABLE CLOTH

WIPE THE OUTSIDE SURFACES OF THE AIRVO 2

4. HIGH-LEVEL DISINFECTION

This step involves running the AIRVO 2 disinfection cycle, using the red disinfection tube and blue disinfection filter from the AIRVO 2 Disinfection Kit 900PT600. This cycle should be carried out at the recommended ambient temperature and humidity levels (see below).

This cycle has been validated to achieve high-level disinfection, by heating the outlet elbow at the heated breathing tube connection port to at least 87 °C (189 °F) for at least 30 minutes.

To ensure that these conditions are met for every disinfection cycle, the AIRVO 2 logs time and temperature (from two independent sensors) throughout the cycle. A successful disinfection cycle is assured only when the countdown reaches zero.

THE FOLLOWING EQUIPMENT IS REQUIRED:

OPERATING CONDITIONS:

WARNINGS

1. CONNECT THE RED DISINFECTION TUBE TO THE AIRVO 2

One end of the disinfection tube has a blue plastic connector. Lift the sleeve and slide the connector onto the unit. Push the sleeve down to lock.

Connect the red end of the disinfection tube to the left-hand chamber port as shown and push fully into position.

2. FIT THE BLUE FILTER

Fit the blue filter to the right hand chamber port as shown. Push it firmly into position.

3. START THE DISINFECTION CYCLE

Connect the power cable to the mains/utility power socket and start the disinfection cycle by holding the On/Off button for 3 seconds until the unit beeps.

WARNING: Disconnect any oxygen supply from the unit before starting the disinfection cycle.

4. WARM-UP AND CALIBRATION

The unit will first go through a series of calibration checks for approximately 2 minutes, during which a warm-up symbol will be displayed on screen.

5. DURING DISINFECTION

After a successful calibration, the disinfection cycle begins and runs for 55 minutes.

Press the Mode button at any time to view a graph of the current disinfection cycle. The graph shows:

If no button has been pushed for 1 minute, a screensaver will be launched.

6. AFTER DISINFECTION

After a successful disinfection, the display will reach zero as shown and an audible information signal will sound. This display will alternate with the current disinfection cycle number. See Section 6, “Disinfection History", for more information.

7. SWITCH OFF THE AIRVO 2

Switch off the unit and unplug from the mains/utility power socket. The unit is now ready for storage.

TROUBLESHOOTING

For full troubleshooting instructions, please refer to the ‘Troubleshootingʻ section in the AIRVO 2 / myAIRVO 2 Technical Manual. If problem still persist, please contact your Fisher and Paykel Healthcare representative.

Problem Action
Fault E42 If a problem occurs during disinfection, the display will flash and the unit will start beeping. Switch the unit off and then restart. If the problem persists, note the fault code and contact your Fisher & Paykel Healthcare representative.
Disconnect oxygen alarm If oxygen is detected during disinfection, the “Disconnect oxygen" alarm will be generated and the disinfection cycle will be stopped. Disconnect the oxygen, turn the unit off and restart to restart disinfection.
Check for blockages alarm If the silicone flaps of the non-return valve, found inside the chamber inlet port, has been displaced, the unit may generate a "Check for blockages" alarm. Return the silicone flaps to the correct position using a non-sharp tool, or replace the non-return valve with 900PT911. The 'Check for blockages' alarm may also be generated if the unit is operated at altitudes exceeding 2,000 m (6000 ft).
Check tube Ensure that both the disinfection tube and the disinfection filter are connected correctly and securely onto the AIRVO 2 chamber ports. Turn the unit off and restart. If problems persist, use a new disinfection tube.

5. STORAGE

After the disinfection cycle, it is imperative that the unit is stored appropriately. This involves using the Clean Storage Cover (900PT603) or an alternative method which will keep the outlet elbow free from contamination while being stored between patients.

USING THE CLEAN STORAGE COVER (900PT603)

ALTERNATIVE STORAGE METHOD

If the Clean Storage Cover is not used, the alternative storage method is to leave the disinfection tube and filter attached to the AIRVO 2. Store the unit (with disinfection tube and filter attached) in an area suitable for medical devices, which is clean, dry and dust-free.

6. DISINFECTION HISTORY

The disinfection cycle number is a unique number that can be recorded for keeping a log of disinfection cycles.

AFTER EACH DISINFECTION CYCLE

The first time that the disinfection cycle is successfully completed, “#1” will be displayed. This number will increment by 1 after each successful disinfection cycle. #7 indicates that seven successful disinfection cycles have been carried out on this AIRVO 2.

AT AIRVO 2 STARTUP

When the AIRVO 2 is switched on for patient use, the following is displayed onscreen:

If this is the first switch-on after a successful disinfection cycle: A green traffic light will be displayed. This AIRVO 2 is safe for use on a new patient.

If this is NOT the first switch-on after a successful disinfection cycle: An orange traffic light will be displayed. This AIRVO 2 is NOT safe for use on a new patient.

7. SYMBOL DEFINITIONS

PDF preview unavailable. Download the PDF instead.

900PT600 AIRVO 2 Disinfection Manual PDFium PDFium

Related Documents

Preview Fisher & Paykel Optiflow+ Tracheostomy Direct Connection OPT970 User Manual
This document provides instructions for the Fisher & Paykel Optiflow+ Tracheostomy Direct Connection OPT970, a patient interface for delivering humidified respiratory gases to tracheostomy patients. It includes setup, intended use, warnings, cautions, and usage guidelines for both hospital and home environments.
Preview F&P Airvo 3 User Manual: High Flow Therapy Device Guide
This user manual provides comprehensive instructions for the Fisher & Paykel Healthcare Airvo 3, a device designed for high flow respiratory therapy. It covers device setup, operation, safety guidelines, troubleshooting, and technical specifications for healthcare professionals.
Preview Fisher & Paykel AIRVO 2 Humidifier User Manual
User manual for the Fisher & Paykel AIRVO 2 humidifier, providing instructions on setup, operation, maintenance, and troubleshooting for healthcare professionals.
Preview Fisher & Paykel AIRVO 2 Humidifier User Manual
Comprehensive user manual for the Fisher & Paykel AIRVO 2 humidifier, detailing setup, operation, maintenance, and troubleshooting for healthcare professionals.
Preview F&P Airvo 2: Respiratory Support and Airway Hydration System
Discover the F&P Airvo 2, an integrated system for Optiflow tracheostomy high flow therapy, offering advanced humidification, respiratory support, and airway hydration for patients.
Preview Fisher & Paykel Healthcare Homecare and Long Term Care Facilities Catalogue
This catalogue from Fisher & Paykel Healthcare presents a comprehensive range of respiratory support products for homecare and long-term care facilities. It details various CPAP masks, including nasal, full face, nasal pillows, and oral types, along with CPAP machines, humidifiers, and breathing circuits. Product specifications, features, and spare parts are listed for each item, covering models such as Evora, Eson, Zest, FlexiFit, Vitera, Simplus, Forma, Brevida, Pilairo, Opus, Oracle, SleepStyle, myAirvo, F&P 810 System, and F&P 550 System.
Preview Fisher & Paykel AIRVO 2 Humidifier User Manual
User manual for the Fisher & Paykel AIRVO 2 humidifier, providing instructions on setup, operation, maintenance, and troubleshooting.
Preview myAIRVO 2 User Manual: Humidified Respiratory Therapy Device
Comprehensive user manual for the Fisher & Paykel Healthcare myAIRVO 2, detailing setup, operation, cleaning, maintenance, and technical specifications for humidified respiratory therapy.