E+E Sigma 05: Modular Sensor Platform
Your partner in sensor technology.
Overview
The Sigma 05 is a modular platform for intelligent probes, supporting analog outputs and an optional display. Its plug-in interchangeable probes make it suitable for harsh environments. The Sigma 05 is available with a polycarbonate or die-cast aluminum housing.
Flexibility: Various Probes and Measurement Targets
The Sigma 05 features an RS485 bus supporting up to three E+E plug-and-play probes with Modbus RTU protocol. Measurement values can be output via two freely selectable analog outputs and an optional graphic display.
Plug and Play
The Sigma 05 automatically detects E+E plug-and-play probes and independently performs hub configuration according to the rules. This simplifies probe replacement.
Configuration and Adjustment
The free PCS10 product setting software allows for easy configuration of the Sigma 05, including assignment of measurement values, setting of limit values, display layout, analog output scaling, and adjustment of connected probes.
Examples of Plug-and-Play Sensors with Sigma 05
Polycarbonate Enclosure Sigma 05 with Oil-Moisture Sensor MOP301
Die-cast Aluminum Enclosure Sigma 05 with CO2, Humidity, and Pressure Sensor EE872
Die-cast Aluminum Enclosure Sigma 05 with Air Velocity and Temperature Sensor EE680
Polycarbonate Enclosure Sigma 05 with Humidity and Temperature Sensor HTP501 for temperatures up to 120 °C (248 °F)
Features
External Mounting Holes
- Mounting with cover closed
- Protection against site contamination
- Easy and quick installation
Backlit Graphic Display
- User-configurable display layout
- Simultaneous display of up to 3 measured values
- Real-time status information
- Display of measured values within pre-set ranges
[Knockout for 1/2" Conduit Fitting (US)]
Bayonet Nut
Opens/closes with a 1/4 turn
Flush Display
Flush surface for easy cleaning, preventing dirt accumulation
M12 Probe Connection
- Connection for up to 3 probes
- Direct connection to Sigma 05 or via M12 cable (up to 10m)
Enclosure
- IP65/NEMA 4 (X)
- Durable for harsh environments
- Polycarbonate or die-cast aluminum housing
PCB
- Freely selectable voltage or current output
- USB-C service interface
- LED for status indication
- Optimized component placement on the underside of the PCB for mechanical protection during installation
Test report according to DIN EN 10204-2.2
Sigma 05 Plug-and-Play Probes
Connecting plug-and-play probes allows the Sigma 05 to function as a interchangeable modular sensor.
EE872: Probe for CO2, Humidity, Temperature, Pressure. More info at: www.epluse.com/ee872.
EE072: Humidity and Temperature probe. More info at: www.epluse.com/ee072.
EE074: Temperature probe. More info at: www.epluse.com/ee074.
EE671: Air velocity probe. More info at: www.epluse.com/ee671.
EE680: Air velocity probe for laminar flow monitoring. More info at: www.epluse.com/ee680.
MOP301: Oil-moisture probe measuring up to 120 °C (248 °F). More info at: www.epluse.com/mop301.
HTP501: Humidity and Temperature probe measuring up to 120 °C (248 °F). More info at: www.epluse.com/htp501.
DNV Certification (Optional)
The Sigma 05 modular sensor platform can be optionally DNV certified when used with the MOP301 oil-moisture probe and HTP501 humidity and temperature probe. Please refer to the respective probe datasheets for details.
The Sigma 05 modular sensor platform is available with DNV certification (optional) for humidity and oil-moisture applications.
Reference Probe
The functionality and accuracy of the Sigma 05 can be checked by connecting an E+E reference probe (Modbus RTU) instead of a standard sensor probe. This reference probe provides fixed values for various measurement parameters and is supplied with an individual test report. For more information, refer to the Reference Probe Modbus RTU Quick Guide at www.epluse.com/sigma05.
Field Loop Calibration
E+E sensor platforms are designed for modularity, allowing for easy field loop calibration or adjustment as required by industries adhering to FDA regulations. Using extension cables, the sensor probe can be connected to a portable calibrator without disconnecting the Sigma 05 main unit.
[Illustration showing EE072 humidity probe connected to Humor 20 humidity calibrator, and EE074 temperature probe connected to a dry block calibrator.]
Specifications
Enclosure Material: Polycarbonate / Die-cast Aluminum
[Diagram showing dimensions of the enclosure.]
Polycarbonate enclosure without display, height 46mm (1.81").
Technical Data
Probe Connection
- Max. number of connectable sensor probes: 3
- Max. number of measured parameters: 5 (2 analog outputs, 3 display values)
- Electrical Connection: 5-pin M12x1 socket
Digital
- Probe Interface: RS485
- Protocol: Modbus RTU
- Default Settings: 9600 Baud, 8 databits, parity even, 1 stop bit
- Baud Rates: 9600, 19200, 38400, 57600, 76800, 115200
- Measured Value Data Format: FLOAT32, INT16
For detailed communication settings, refer to the user manual and the Modbus application note at www.epluse.com/sigma05.
Output
Analog
Choice of 2 output options:
- 0-1V, 0-5V, 0-10V, 4-20mA, 0-20mA
- Accuracy at 20°C (68°F): 0.02% FS for 0-10V and 0-20mA
General
- Power Supply: Class III (USA & Canada: Class 2 power supply required)
- Max. Supply Current: 0.5 A
- Electrical Connection: Max. 2.5 mm² screw terminals
- Cable Gland: M16x1.5 cable diameter Ø4...8 mm (0.16...0.31")
- Configuration Interface: USB-C port
- Operating and Storage Conditions: 0...95% RH (non-condensing), 700...1200 mbar
- Operating Temperature: -40...+60 °C (-40...+140 °F)
- Storage Temperature: -20...+50 °C (-4...+122 °F)
Enclosure
- Material: Polycarbonate / Aluminum AI 383
- Protection Rating: IP65/NEMA 4 (X) for both
- Display: UL94 V-0 for without display, UL94 HB for with display
Electromagnetic Compatibility
- EN 61326-1:2013, EN 61326-2-3:2013
- FCC Part15 Class A, ICES-003 Class A
- Industrial Environment
Distribution and Conformity Certification
- CE, UK CA
- DNV-CG-0339
- DNV Certificate No. TAA00003FA
Configuration and Adjustment
Download the PCS10 product setting software for free at www.epluse.com/pcs10.
Refer to the "Accessories" datasheet for compatible cables/connectors.
DNV certification: Refer to the order guide.
Ordering Guide
Hardware configuration | Description | Code | Sigma05- |
---|---|---|---|
Certification | No DNV certification | No code | |
With DNV certification¹⁾ | AP2 | AP2 | |
Enclosure Material | Polycarbonate (PC) | HS1 | HS1 |
Die-cast Aluminum AI 383 | HS3 | HS3 | |
Display | Without display | D0 | |
With display | D2 | D2 | |
Analog Outputs | 0-1 V | GA1 | GA1 |
0-5 V | GA2 | GA2 | |
0-10 V | GA3 | GA3 | |
0-20 mA | GA5 | GA5 | |
4-20 mA | GA6 | GA6 | |
Unit | Metric (SI) | U1 | U1 |
Non metric (US/GB) | U2 | U2 |
Example Order: SIGMA05-AP2HS1D2GA6U1
Item | Code | Description |
---|---|---|
Certification | AP2 | With DNV certification |
Enclosure Material | HS1 | Polycarbonate (PC) |
Display | D2 | With display |
Analog Output | GA6 | 4-20 mA |
Accessories
For more details, please refer to the "Accessories" datasheet.
Description | Code |
---|---|
Configuration adapter for Modbus | HA011018 |
Product setting software (Free download: www.epluse.com/pcs10) | PCS10 |
Connection cable M12-M12 (unshielded) | |
L = 2 m (6.6 ft) | HA010813 |
L = 5 m (16.4 ft) | HA010814 |
L = 10 m (32.8 ft) | HA010815 |
Modbus RTU reference probe | HA010406 |
Power supply adapter 100-240V AC ~ 24V DC (No DNV approval) | V03 |
PC connection USB cable (USB-A to USB-C) | HA010327 |
M12 Y adapter | HA030204 |
Assembly M12x1 cable connector, 5-pin socket (female) | HA010708 |
Assembly M12x1 cable connector, 5-pin plug (male) | HA010706 |
M12 connector protective cap (female) | HA010781 |
M12 connector protective cap (male) | HA010782 |