EMShape Neo Plus Body Sculpt Machine 9500W User Manual
Welcome
Welcome to your EMShape Neo Plus Body Sculpt Machine 9500W.
User Notice
Thank you for choosing the MSHAPE BEAUTY EMShape Neo® Plus Body Sculpt Machine 5 in 1. This machine combines high-tech essence, fashionable style, and innovative functions. We hope to contribute to your beauty. Before using the instrument, please read this user manual carefully and keep it in a safe place for future use. This user guide is an operation guide, not a product guarantee.
MSHAPE BEAUTY reserves the right to interpret and modify printing errors, software upgrades, and product improvements in this manual at any time. These changes will be directly incorporated into the new version of the user manual without notice. This user manual is compiled by MSHAPE BEAUTY; all rights reserved. Privacy is subject to investigation.
Preface
Thank you for choosing the MSHape Beauty EMShape Neo Plus Body Sculpt Machine 9500W. The EMShape Neo Plus is engineered to perform body sculpting more effectively than any alternative equipment available in the global beauty marketplace. The machine's sleek design and high-end specifications guarantee maximum efficiency and fast, impressive benefits. Let your spa's clientele indulge in a spa treatment that offers premium results.
Harness the power of the latest HIEMT technology and synchronized RF to deliver efficient and effective muscle building and fat reduction results. The procedure is 100% non-invasive, safe, and painless, with no downtime.
Thank you again, and please enjoy your body sculpting equipment!
I. Product Information & Specification
1.1 Product Information
This machine uses the latest HIEMT technology + synchronized RF to deliver efficient & effective muscle building + fat reduction. This process triggers the production of new collagen chains, thereby increasing muscle support and volume.
HIEMT (high-intensity electromagnetic technology) uses strong focused electromagnetic waves to stimulate and train the targeted muscle tissue. During a 30-minute treatment session, focused magnetic waves penetrate up to 8cm of muscle tissue, forcing the muscles to contract involuntarily many thousands of times. Such supra-maximal contractions are not achievable by normal exercise. The body naturally responds to this supra-maximal exercise by rebuilding muscle tissue, making the body more toned and stronger. As with normal exercise, there is a large amount of natural fat decomposition, resulting in body slimming.
1.2 Product Specification
- Voltage: 100 V/ 110 V/ 220 V (Voltage selected according to your region)
- Technologies: High-intensity EMS, RF, EMT
- Modes: Stretching to warm up, deepen exercise, activate muscle, strengthen muscle, strong fat burning, relief
- Pulse width: ~310µs ± 10%
- Cooling system: Air cooling
WARNING! RF energy on this machine is VERY STRONG and the temperature can become VERY HOT. To make the treatment comfortable and effective, please always start with a low level of RF.
Training is strongly recommended. MSHAPE BEAUTY is not responsible for any adverse effect that should occur due to misuse of the equipment by the operator.
1.3 List of Accessories
Item | Quantity | Description/Image Placeholder |
---|---|---|
Main Unit | 1 | [Main Unit Diagram] |
Flat Handle | 2 | [Flat Handle Diagram] |
Curved Handle | 2 | [Curved Handle Diagram] |
Pelvic Floor Training Cushion | 1 | [Pelvic Floor Cushion Diagram] |
Belt Straps | 4 | [Belt Straps Diagram] |
Power Cord | 1 | [Power Cord Diagram] |
II. Product Appearance
II.1 Front View
[Diagram showing the front of the machine with labels: Curved Handle, LCD Screen, Flat Handle]
II.2 Back View
[Diagram showing the back of the machine with labels: A Handle Socket, B Handle Socket, C Handle Socket, D Handle Socket, Power Switch, Power Cord Socket]
II.3 Side View
[Diagram showing the side of the machine]
II.4 Accessories Application
Apply to abdomen and buttocks: [Diagram showing handles and straps for torso application]
Apply to legs and arms: [Diagram showing handles for limb application]
III. Warnings
III.1 Precautions Before/During/After Use
- Use at least one hour after meals. ⏰ DO NOT use when feeling full after eating.
- Before using the instrument, remove any metal jewelry you are wearing. ?? DO NOT touch metal objects during the treatment.
- Before starting the operation, touch the probe vertically to the skin, then turn on the switch of the main unit and adjust the intensity of the operation mode. ? Start with the lowest level, and then increase the intensity sequentially according to the user's skin tolerance.
III. Warnings (Continued)
- ??️ DO NOT eat during machine operation. Try to eat some protein after one hour of session.
- ? DO NOT use if there are open wounds or any skin infections such as eczema or dermatology issues.
- ❤️ DO NOT use if there are heart conditions (such as with arrhythmias), especially those with pacemakers.
- DO NOT use if diagnosed with cancer.
III. Warnings (Continued)
- ? DO NOT use if diagnosed with infectious disease.
- ? DO NOT use if pregnant, menstruating, conceiving, or breastfeeding.
- ? DO NOT use if you have undergone major surgery. Diabetic and cerebrovascular disease (stroke patients) must avoid.
- ? DO NOT use if you have metal on any part of the body.
IV. Treatment Position
The position of the handles during treatment must be the same as shown in the images below for every area of the body:
Abdomen: [Image showing handle placement on the abdomen]
Hips: [Image showing handle placement on the hips]
Shoulder: [Image showing handle placement on the shoulder]
Leg: [Image showing handle placement on the leg]
Arm: [Image showing handle placement on the arm]
Notes:
- Be careful to avoid positions of ribs and bones.
- When operating the abdomen, select one or two handles according to the size of the customer's abdomen.
- It is recommended to treat one part for 30 minutes.
- Before operation, metal objects worn on the body must be removed, and metal objects on clothes must also be avoided.
- Before operation, fasten the bandage, insert the handle into the bandage and fix it, and pay attention to the close position of the working head.
- Adjust the energy intensity before starting the operation. The treatment starts from low intensity and gradually improves, and slowly improves according to the comfort of customers.
- Note: If the customer doesn't feel anything during the operation, the energy intensity can be appropriately strengthened, and the degree of strengthening can be asked whether the customer accepts it.
V. Mode Selection & Operation
a. Mode Selection
I. Function and Mode Selection
- On the interface, select the type of treatment to be used on the function selection interface.
- There are two General Mode Categories: AUTO and MANUAL mode. Select which one to use. (Figure 1)
- On setting the information interface, select the appropriate settings according to gender, age range, and fat thickness. (Figure 2)
[Figure 1: Interface showing AUTO and MANUAL mode selection options]
[Figure 2: Interface showing user information settings for gender, age, and fat thickness]
Click the icon shown above to start working.
V. Mode Selection & Operation
b. AUTO Mode
[Figure 3: Screenshot of the AUTO mode interface showing various parameters like RF intensity, Level Selection, Treatment Time, Handle Intensity]
II. AUTO MODE
- When using AUTO mode, there is a choice of TREATMENT LEVELS (please refer to page 13). Select a TREATMENT LEVEL according to the patient's condition. When in doubt, start with M1 to be conservative.
- The handle strengths A and B can be adjusted ranging from 8% (weak) to 100% (strong). Set the handle strength before the start of treatment. The treatment starts from low intensity and should be gradually increased according to the patient's tolerance.
- A B: Handle intensity of A and B, click [➖] and [➕] to adjust intensity.
- C D: Handle intensity of C and D, click [➖] and [➕] to adjust intensity.
- AB-RF: RF intensity of AB handle, click [➖] and [➕] to adjust intensity.
- CD-RF: RF intensity of CD handle, click [➖] and [➕] to adjust intensity.
- Set the working time. The default working time of one treatment is 30 minutes. This can be adjusted from 5 minutes to 60 minutes.
- Click the Start button to start the treatment. Click the Pause button if you need to pause the treatment.
NOTE: Adjust the settings when the device is in Pause mode.
V. Mode Selection & Operation
c. Manual Mode
[Figure 1: Interface showing AUTO and MANUAL mode selection options]
[Figure 4: Screenshot of the MANUAL mode interface showing frequency and intensity controls]
III. MANUAL MODE
- On the interface, select Manual mode.
- On setting the information interface (Figure 2), select the appropriate settings according to gender, age range, and fat thickness.
- Adjust the frequency. Please see the next page for frequency setting.
V. Mode Selection & Operation
c. Manual Mode (Continued)
4. Set the frequency; adjustment settings are below:
- F1: Frequency 1, the intensity range is 3Hz (weak) to 200Hz (strong)
- F2: Frequency 2, the intensity range is 3Hz (weak) to 200Hz (strong)
- F3: Frequency 3, the intensity range is 3Hz (weak) to 200Hz (strong)
NOTE: It's recommended that F1 be set to 50Hz, F2 to 5Hz, and F3 to 150Hz, to achieve a balance of fat reduction and muscle building.
[Figure 5: Screenshot of the MANUAL mode interface showing frequency and intensity controls with specific settings]
5. Before the start of treatment, set the frequency parameters according to the required frequency.
- F1 frequency works for 5 minutes,
- F2 frequency works for 1 minute,
- F3 frequency works for 5 minutes, and
- F1, F2, and F3 frequencies are cycled in sequence.
6. The handle strengths A and B can be adjusted ranging from 8% (weak) to 100% (strong). Set the handle strength before the start of treatment. The treatment starts from low intensity and should be gradually increased according to the patient's tolerance.
- A B: Handle intensity of A and B, click [➖] and [➕] to adjust intensity.
- C D: Handle intensity of C and D, click [➖] and [➕] to adjust intensity.
- AB-RF: RF intensity of AB handle, click [➖] and [➕] to adjust intensity.
- CD-RF: RF intensity of CD handle, click [➖] and [➕] to adjust intensity.
7. Set the working time. The default working time of one treatment is 30 minutes. This can be adjusted from 5 minutes to 60 minutes.
8. Click the Start button to start the treatment. Click the Pause button if you need to pause the treatment.
NOTE: Adjust the settings when the device is in Pause mode.
VI. Mode Selection & Operation
d. Pelvic Floor Cushion
1. Align the handle with the socket on the back of the instrument and insert it; insert the power cord into the rear socket of the device, turn ON the power switch, and the device will start immediately.
2. Place a cushion on the chair, sit on the center, and relax.
[Diagram showing the Pelvic Floor Cushion with ergonomic design, support protector, and placement on hip/thigh areas]
NOTE: Use MANUAL mode when using the Pelvic Floor Cushion. The recommended setting for F1: 110 Hz, F2: 5 Hz, and F3: 140 Hz.
VII. Treatment Levels in AUTO Mode
MODE | STAGE 1 | STAGE 2 | STAGE 3 | STAGE 4 | STAGE 5 | STAGE 6 |
---|---|---|---|---|---|---|
M1 | 3 Hz/1 minute | 140 Hz/5 minutes | 4 Hz/1 minute | 200 Hz/5 minutes | 4 Hz/1 minute | 160 Hz/5 minutes |
M2 | 5 Hz/1 minute | 200 Hz/5 minutes | 5 Hz/1 minute | 80 Hz/5 minutes | 5 Hz/1 minute | 140 Hz/5 minutes |
M3 | 5 Hz/1 minute | 200 Hz/5 minutes | 5 Hz/1 minute | 160 Hz/5 minutes | 5 Hz/1 minute | 140 Hz/5 minutes |
M4 | 4 Hz/1 minute | 160 Hz/5 minutes | 4 Hz/1 minute | 200 Hz/5 minutes | 4 Hz/1 minute | 160 Hz/5 minutes |
M5 | 5 Hz/1 minute | 200 Hz/5 minutes | 5 Hz/1 minute | 160 Hz/5 minutes | 5 Hz/1 minute | 200 Hz/5 minutes |
Training Level Descriptions:
- M1: Basic training for beginners (suitable for people who hardly exercise, effective basic training for weak muscle groups).
- M2: Adaptation training (suitable for primary exercisers who have not exercised for a long time, and also suitable for obese people with less muscles).
- M3: Preliminary training (suitable for primary exercisers with insufficient muscle or slightly higher body fat, also suitable for slightly obese people).
- M4: Advanced training (suitable for intensive endurance training, suitable for all fitness groups and regular exercisers, and also suitable for obese people. As the training frequency further increased, the load is gradually increased to increase the strength and dimensionality of muscle contraction. For shaping).
- M5: Professional training (suitable for intensive training, suitable for regular exercisers and athlete groups, and also suitable for obese people. Due to increase in training frequency and high intensity of action, the fat burning ability of muscles is improved, and fat is consumed at high frequency).
VIII. Troubleshooting Solutions
Machine Error | Cause | Troubleshoot |
---|---|---|
Machine not working | 1. Power plug not inserted 2. Switch not ON 3. Power fuse failure | 1. Check the power 2. Check the switch 3. Replace the fuse |
Handle not working | Check if the connection is loose | Disconnect and reconnect the handle |
IX. Maintenance
VII.1 Regular Use
- To make sure that the machine will last longer, after each 30 minutes of using the machine, allow it to rest/cool down for 5 minutes before starting the treatment again.
- Check the power socket, confirm that there is a reliable connection to the ground.
- Check if the power cord is damaged before use; do not use if damaged.
VII.2 Cleaning
- Perform regular cleaning to ensure sanitation and proper function.
- The outside body of the machine can be cleaned using a damp cloth or alcohol wipes.
- Clean only when the machine is unplugged and disconnected from the power source.