Motic BA210LED Digital Microscope Instruction Manual
Thank you for choosing a Motic Digital Microscope. Motic is constantly improving its instruments and adapting them to the requirements of modern research techniques and testing methods. This may involve modification to the mechanical as well as optical components of our digital microscope. Therefore, all descriptions and illustrations in this instruction manual, including all specifications, are subject to change without prior notice.
FCC Statement
Any Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation.
FCC Radiation Exposure Statement
This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator and your body. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the receiving antenna, Increase the separation between the equipment and receiver, Connect the equipment into an outlet on a circuit different from that to which the receiver is connected, Consult the dealer or an experienced radio/TV technician for help.
Correct operation ensures the digital microscope's long-term proper usage. Contact your local Motic distributor for maintenance needs or contact Motic directly through their website www.motic.com.
Infinity Corrected Optics
In this optical concept, the light beams are parallel after leaving the objective in the direction of the eyepieces. A second optical element, the tube lens (normally located in the eyepiece tube), is used to converge the parallel beams, resulting in an intermediate image. The intermediate image is focused by the eyepieces to provide the real image for visual observation. The implementation of a tube lens gives the opportunity to minimize chromatic aberrations and other "optical defects". Further, in "Infinity Optics", the distance between the objective and tube lens is not as strictly fixed as in the (historically older) “Finite Optics” of 160mm tube length. This allows additional optical components to be inserted between the objective and tube head. Fluorescence attachments, discussion bridges, eye level risers, and other options can be added without affecting the image quality. In general, "Infinity Optics" provides flexibility and the opportunity to add additional optional features.
Conventional Digital Microscope
The conventional digital microscope has a two-stage magnification system. There are two lens systems, the objective and the eyepiece, mounted at opposite ends of a body tube. The objective forms an enlarged real image of the object being examined and is called the intermediate image. The intermediate image is further enlarged by the eyepiece and is seen as a virtual image of the intermediate image. The eye can examine this final image, situated at infinity. The total magnification of the digital microscope is determined by the focal lengths of the objective and eyepiece, and can be calculated as objective magnification X eyepiece magnification. For instance, 40x objective X 10x eyepiece = 400x magnification.
Digital Microscope Terminology
- Abbe Condenser: A two-lens sub-stage condenser located below the stage of a digital microscope, functioning to collect light and direct it onto the object being examined. Its high numerical aperture makes it particularly suited for use with most medium- and high-magnification objectives.
- Aperture, Numerical (N.A.): The numerical aperture is an important factor determining the efficiency of the condenser and objective. It is represented by the formula: (N.A. = ηsinα), where η is the refractive index of a medium (air, water, immersion oil etc.) between the objective and the specimen or condenser, and α is half of the maximum angle at which light enters or leaves the lens from or to a focused object point on the optical axis.
- Cover Glass Thickness: Transmitted light objectives are designed to image specimens that are covered by a thin cover glass (cover slip). The thickness of this small glass piece is now standardized at 0.17 mm for most applications.
- Diaphragm, Condenser: A diaphragm that controls the effective size of the condenser aperture. A synonym for the condenser illuminating aperture diaphragm.
- Diopter Adjustment: The adjustment of the eyepiece of an instrument to provide accommodation for the eyesight differences of individual observers.
- Depth of Focus: The axial depth of the space on both sides of the image plane within which the image is sharp. The larger the N.A. of the objective, the shallower the depth of focus.
- Field of View (F.O.V.): That part of the image field, which is imaged on the observer's retina, and hence can be viewed at any one time. The field of view number is now one of the standard markings of the eyepiece.
- Filter: Filters are optical elements that selectively transmit light. They may absorb part of the spectrum, or reduce overall intensity, or transmit only specific wavelengths.
- Immersion Oil: Any liquid occupying the space between the object and digital microscope objective. Such a liquid is usually required by objectives of 3-length or less.
- Magnification: The number of times by which the size of the image exceeds the original object. Lateral magnification is usually meant. It is the ratio of the distance between two points in the image to the distance between the two corresponding points in the object.
- Micrometer: A metric unit of length measurement = 1x10⁻⁶ meters or 0.000001 meters.
- Nanometer: A unit of length in the metric system equal to 10⁻⁹ meters.
- Phase-contrast (microscopy): A form of microscopy, which converts differences in object thickness and refractive index into differences in image amplitude and intensity.
- Real Viewfield: The diameter in millimetres of the object field. Real Viewfield = Eyepiece Field of View / Objective Magnification. For example BA210: Eyepiece field of view = 20mm, Objective magnification = 10X, Diameter of the object field = 20/10 = 2.0mm.
- Resolution: The result of displaying fine details in an image.
- Resolving Power: A measure of an optical system's ability to produce an image which separates two points or parallel lines on the object.
- Total Magnification: The total magnification of a digital microscope is the individual magnifying power of the objective multiplied by that of the eyepiece.
- Working Distance: This is the distance between the objective front lens and the top of the cover glass when the specimen is in focus. In most instances, the working distance of an objective decreases as magnification increases.
- X-axis: The axis that is usually horizontal in a two-dimensional coordinate system. In microscopy, the X-axis of the specimen stages is considered that which runs left to right.
- Y-axis: The axis that is usually vertical in a two-dimensional coordinate system. In microscopy, the Y-axis of the specimen stages is considered that which runs front to back.
1. Setting Up the Instrument
Avoid placing the instrument in locations exposed to direct sunlight, dust, vibration, high temperature, and high humidity, and where it is difficult to unplug the power supply cord.
1.1 Operating Environment
- Indoor use
- Altitude: Max 2000 meters
- Ambient temperature: 15°C to 35°C
- Maximum relative humidity: 75% for temperature up to 31°C, decreasing linearly to 50% relative humidity at 40°C
- Supply voltage fluctuations: Not to exceed ±10% of the normal voltage.
- Pollution degree: 2 (in accordance with IEC60664)
- Installation / Overvoltage category: 2 (in accordance with IEC60664)
- Air pressure: 75kPa to 106 kPa
- Avoid frost, dew, percolating water, and rain.
1.2 Verifying Input Voltage
- The automatic voltage selection works with a broad range of settings. However, always use a power cord that is rated for the voltage used in your area and that has been approved to meet local safety standards. Using the wrong power cord could cause fire or equipment damage.
- If using an extension cord, use only a power supply cord with a protective earth (PE) wire.
- In order to prevent electric shock, always turn the power switch on the power supply off before connecting the power cord.
1.3 Mechanical Stage
- Remove specimen holder for fast hand scanning of slides.
- Left hand and right hand operation stages are available.
1.4 Specimen Holder
- Attach the specimen holder using the two mounting holes.
1.5 Objectives
- Lower the stage completely. Screw the objectives into the revolving nosepiece so that clockwise rotation of the nosepiece brings the next higher magnification objective into position.
1.6 Condenser
- Raise the stage by turning the coarse focus knob.
- Raise the condenser carrier by turning the condenser focus knob.
- Insert the condenser into the mount with the aperture scale facing forward towards the user.
- Secure with the condenser clamp screw.
- Turn the condenser focus knob to raise the condenser as far as it will go.
1.7 Eyepiece Tube
- Loosen the eyepiece clamp screw. Insert the round dovetail mount on the eyepiece tube into the round dovetail mount on the digital microscope arm. Tighten the eyepiece tube clamp screw to secure the eyepiece tube in place.
1.8 Eyepieces
- Use the same magnification eyepieces for both eyes.
- To lock the eyepiece, insert each eyepiece completely into the eyepiece sleeve and tighten the clamp screws.
- Twist the eyepiece (anti-clockwise or clockwise) with 20~30 degrees and pull the eyepieces gently out when removing the eyepiece.
1.9 Filters
- Remove the collector cover and place the filter in the filter holder located around the field lens. Screw the collector cover, taking care that dust, dirt, and fingerprints do not get on the filter and the field lens.
- Filter Selection:
- Filter: ND2 (T=50%) Function: For brightness adjustment in photomicrography
- Filter: ND4 (T=25%) Function: For routine microscopy and photomicrography
- Filter: ND16 (T=6.25%) Function: For phase contrast and contrast adjustment with black and white film
- Filter: Blue filter (colour balance filter) Function: For routine microscopy and photomicrography
- Filter: Green interference (546nm) Function: For phase contrast and contrast adjustment with black and white film
- Filter: HE (didymium filter) Function: For colour photomicrography of HE stained specimen with tungsten type film
- A diffuser is built into the base of the digital microscope.
2.0 Power Cord
Connect the socket of the power cord to the AC inlet on the rear of the base of the digital microscope. Plug in the other end of the cord to an AC outlet with ground conductor.
2. Usage of Digital Microscope Components
2.1 Coarse and Fine Focusing
- Focusing is carried out with the coarse and fine focus knobs at the left and right of the digital microscope stand.
- The direction of vertical movement of the stage corresponds to the turning direction of the focus knobs.
- One rotation of the fine focus knob moves the stage 0.2mm. The graduation on the fine focus knob is 2 microns.
- Never attempt either of the following actions, since doing so will damage the focusing mechanism: Rotate the left and right knob while holding the other. Turning the coarse and fine focus knobs further than their limit.
2.2 Coarse Focus Torque Adjustment
- To increase the torque, turn the torque adjustment ring located behind the left-hand coarse focus knob in the direction indicated by the arrow. To reduce the torque, turn the ring in the direction opposite to that indicated by the arrow.
2.3 Stage Upper Limit Stop Adjustment
- (Upper Stage Limit is preset at the factory; please only adjust if necessary)
- The Stage Upper Limit stop marks the stage position at which the specimen is in focus, i.e., by restricting the movement of the coarse focus knob.
- With the specimen in focus, turn the stage upper limit stop knurled ring clockwise until it reaches the stop.
- When the stage upper limit stop is in position, the stage cannot be raised from that position. However, the fine focus knob can move the stage regardless of the limit but will only lower the stage.
- Lower the stage by using the coarse focus knob anticlockwise.
2.4 Beam Splitter Lever
- The beam splitter lever of the trinocular eyepiece tube can be used to select the amount of light distributed between the trinocular eyepiece tube and the vertical phototube.
- When the beam splitter lever is pushed in until it reaches the limit, 100% of the light enters the observation tube. When the beam splitter lever is pulled out to the limit, the ratio of light entering the observation tube and phototube will be 20:80.
2.5 Interpupillary Distance Adjustment
- Before adjusting the interpupillary distance, bring a specimen into focus using the 10x objective.
- Adjust the interpupillary distance so that both the right and left field of view become one.
- This adjustment will enable the user to observe the specimen with both eyes.
2.6 Diopter Adjustment
- Diopter adjustment compensates for the differences in vision between the left and right eyes. In addition to making observation through both eyes easier, this adjustment also reduces the extent to which focusing is lost when the objective magnification is changed. In particular, this occurs when a low magnification objective is used.
- The left eyepiece has a separate focusing provision to compensate for slight differences in the focusing of each eye.
- Using the right eye only and viewing through the right-hand eyepiece, adjust the focus with the digital microscope fine or coarse adjustment until the image of the specimen is at its sharpest.
- Using the left eye only and viewing through the left-hand eyepiece with its independent diopter-focusing ring, focus until the specimen image is at its sharpest.
- The digital microscope should now be ready for binocular viewing.
2.7 Condenser (Sourced Focused (Critical) Illumination)
- Critical illumination relies on using the sub-stage condenser to produce a focused image of the homogeneous light source in the plane of the specimen in order to achieve an even illumination condition over the entire field of view.
- To have the light source focused in the plane of the specimen, an imaging disc with concentric circles (and with its matte surface facing toward the digital microscope base) is placed on the field lens and is focused on the specimen plane. This is achieved by moving the condenser up or down with the condenser focus knob.
- The correct vertical setting of the condenser remains unaltered when changing magnifications.
- Because the light source is imaged onto the specimen, both the specimen and the light source are said to be in the field plane. The condenser iris diaphragm controls the N.A. of the system and is therefore said to be located in an aperture plane of the digital microscope.
2.8 Use of Aperture Diaphragm
- The condenser aperture diaphragm is provided for adjusting the numerical aperture (N.A.) of the illuminating system of the digital microscope; it decides the resolution of the image, contrast, depth of focus, and brightness.
- Stopping down will lower the resolution and brightness but increase the contrast and depth of focus.
- An image with appropriate contrast in most cases can be obtained with an aperture diaphragm closed down to 2/3 of the maximum value.
- To adjust the aperture diaphragm: adjust the condenser aperture diaphragm ring referring to the condenser aperture scale, or by observing the diaphragm image visible on the exit pupil inside the eyepiece tube, or by using a centering telescope after removing one of the eyepieces and focusing on the aperture diaphragm.
- When swinging out the top lens of the condenser for low magnification objectives, open the condenser aperture diaphragm completely.
3. Photomicrographic Procedure
- To ensure vibration-free operation, set the digital microscope on a sturdy vibration-free table or a bench with a vibration-proof device.
- Pull the optical path selection lever of the trinocular eyepiece tube all the way out to the limit; the ratio of light entering the observation tube and phototube will be 20:80.
- For the same total magnification, select a combination of the highest possible objective magnification and lowest possible projection lens magnification to achieve the utmost image definition and contrast.
- To ensure optimal illumination, check the position and centring of the lamp and position of the condenser.
- Select a blue filter for routine application. An additional colour-compensating filter can also be used depending on the colour rendition.
- Adjusting the field diaphragm is important for the purpose of limiting extraneous light that may cause flare and lower the contrast. Stop down the diaphragm to achieve an illuminated area slightly larger than that of the field of view.
- A change of depth of focus, contrast, and resolution of image is attainable with an aperture setting that is 2/3 of the objective N.A.
- For specific photomicrographic procedures, refer to the manual of the specific camera being used.
4. Using Oil Immersion Objectives
- Oil immersion objectives are labelled with the additional engraving “Oil” and are to be immersed in oil between the specimen and the front of the objective.
- The immersion oil supplied by Motic is synthetic, non-fluorescing, and non-resining oil, with a refractive index of 1.515.
- Normally, cover glass must be used with oil immersion objectives, with a few exceptions. Deviations from thickness are not important as a layer of immersion oil acts as compensation above the cover glass.
- The small bottle of oil supplied with every immersion objective facilitates the application of the oil to the cover slip.
- Remove any air bubbles in the nozzle of the oil container before use.
- Immersion oil must be used sparingly. After the examination, the oil should be wiped off the objective with a lens cleaning tissue and the residual film removed with soft cloth moistened with petroleum benzene or absolute alcohol.
- Locate the field of interest with a lower magnification objective. Swing the objective out of the light path and add one drop of immersion oil over the site of the specimen.
- Swing in the oil immersion objective. There should be a small column of oil from the cover slip to the objective lens. Use the fine focus to make the image sharp.
- Freedom from air bubbles must be ensured. To check for air bubbles, remove an eyepiece, fully open the field and aperture diaphragms, and look at the exit pupil of the objective within the eyepiece tube. Air bubbles are recognized by the presence of a surrounding black ring. Bubbles may often be dislodged by moving the slide to and fro or by slightly rocking the revolving nosepiece back and forth. If not successful in clearing the bubbles, then the oil must be wiped off and replaced with a fresh drop.
5. Assembling the Digital Microscope
5.1 Preparation
Before installing the software, please ensure the following:
- Plug in the digital microscope's power cord and switch the unit on.
- Check that you have a Motic software U-disk with your digital microscope.
- Your computer should meet the Minimum System Configuration as listed below.
Minimum System Configuration:
- CPU: Intel based 1.4GHz or better for PC or G4 for Mac
- HDD: 1GB available disk space
- RAM: 512MB or more
- Display Card: 32MB display memory
- Operating System: Windows XP or higher / Mac OSX
5.2 Image Adjustment
- Ensure that you pull the Beam Splitter all the way out until you hear it click into place. This will allow the camera chip inside the head to be exposed to the light from the digital microscope.
- Then you can start the live image transmission by clicking on the Camera icon (shortcut Ctrl M).
- If you do not see an image, please click on "Auto Exposure". For further details, please consult the software Quick Start Guide contained as a PDF file in the U-disk.
- Bring the image into focus by adjusting the focusing control.
- With a 4X objective selected on the digital microscope, ensure that the image on the screen is in focus and in the center of the screen.
- Using a 40X objective next, adjust the fine focusing to ensure that the observation is on the correct focal plane.
- Please note that when changing from 4X to 40X, the image should be Parfocal as well as Parcentered on screen.
5.3 Calibration
- Put the calibration slide (with four black dots) shipped with the digital microscope on the stage and using a 10X objective, focus on one of the dots so that the entire dot can be seen on the computer screen. Please note the diameter of the dot you are looking at.
- Click on the Calibration Wizard in the Motic Application Program and follow the on-screen instructions to calibrate the digital microscope.
NOTE: Calibration needs to be done at least once and on at least one objective (Motic recommends the 10X). When the computer screen or the screen resolution is changed, the calibration needs to be done again. Please read the software Quick Start Guide for further information.
6. Troubleshooting Table
As you use your digital microscope, you may occasionally experience a problem. The troubleshooting table below contains the majority of frequently encountered problems and their possible causes.
6.1 Optical
Problem | Possible Cause |
---|---|
Vignetting or uneven brightness in the field of view or field of view only partially visible | Lamp not installed properly Condenser not mounted correctly Condenser is set too low Aperture diaphragm closed too far Revolving nosepiece not clicked into position Trinocular eyepiece tube optical path selector lever in intermediate position Filter not in placed in properly |
Dust or dirt in the field of view | Aperture diaphragm closed too far Condenser is set too low Dust or dirt on specimen surface Dust or dirt on field lens, filter, condenser or eyepiece |
Poor image (low contrast or resolution) | Condenser is set too low Aperture diaphragm closed too far No cover glass Too thick or thin cover glass Immersion oil not used with oil immersion lens Air bubbles in immersion oil Specified immersion oil not used Immersion oil on dry objective Greasy residue on eye lens Incorrect illumination |
Uneven focus | Specimen holder not fixed securely on stage Specimen not secured in position Specimen tilted on stage surface |
Focusing is not possible with high magnification objectives | Slide is upside down Cover glass is too thick |
High magnification objectives strike the specimen when changing over from low to high magnification | Slide is upside down Cover glass is too thick |
Insufficient parfocality of objectives | Eyepiece tube diopter not adjusted Eyepiece tube diopter not adjusted |
No cohesion of binocular image | Magnification or field of view of left and right eyepieces differ Interpupillary distance not adjusted Eyepiece tube diopter not adjusted |
Eye strain or fatigue | Interpupillary distance not adjusted Diopter adjustment not made Field of view of left and right eyepiece differ Inadequate illumination |
6.2 Electrical
Problem | Possible Cause |
---|---|
Lamp does not light | Power supply not plugged in Lamp not installed Lamp burnt out |
Inadequate brightness | Specified lamp not being used Specified lamp not being used |
Lamp blows out immediately | Connectors are not securely connected |
Lamp flickers | Lamp near end of service life Lamp not securely plugged into socket |
7. Care and Maintenance
7.1 Do Not Disassemble
- Disassembly may significantly affect the performance of the instrument and may result in electric shock or injury and will void the terms of the warranty.
- Never attempt to dismantle any parts other than described in this manual. If you notice any malfunction, contact your nearest Motic representative.
7.2 Cleaning the Digital Microscope
7.2.1 Lenses and Filters
- To clean lens surfaces or filters, first remove dust using an air blower. If dust still persists, use a soft/clean brush or gauze.
- A soft gauze or lens tissue lightly moistened with the mixture of alcohol and ether (ratio: alcohol : 3 and ether: 7) should be used to remove grease or fingerprints.
- Use only a mixture of alcohol and ether (ratio : alcohol : 3 and ether: 7) to remove immersion oil from objective lenses.
- Because the mixture of alcohol and ether (ratio : alcohol : 3 and ether: 7) is highly flammable, be careful handling around open flame.
- Do not use the same area of gauze or lens tissue to wipe more than once.
7.2.2 Cleaning of Painted or Plastic Components
- Do not use organic solvents (thinners, alcohol, ether, etc.). Doing so could result in discolouration or in the peeling of paint.
- For stubborn dirt, moisten a piece of gauze with diluted detergent and wipe clean.
- For plastic components, only moisten a piece of gauze with water and wipe clean.
7.3 Disinfecting the Digital Microscope
- Follow the standard procedures for your laboratory.
7.4 When Not in Use
- When not in use, cover the instrument with a vinyl dust cover and store in a place low in humidity where mould is not likely to form.
- Store the objectives, eyepieces, and filters in a container or desiccator with drying agent.
- Proper handling of the digital microscope will ensure years of trouble-free service.
- If repair becomes necessary, please contact your Motic agency or their Technical Service directly.
7.5 LED Replacement
Note: If equipment is used in a manner not specified by the manufacturer, the warranty may be void. To avoid getting wet, do not use the digital microscope near water.
- Unscrew four hexagonal screws retaining the base plate.
- Disconnect the LED connection cables from the power supply printed circuit board.
- Loosen the LED board locating ring, take away the LED board locating ring.
- Install the new LED. Take out a qualified LED groups with power and PIN, nest the pressure ring and reverse action steps as above.
- At the time of installation, the LED group centre should be adjusted to the center of the collector as possible, which is called the axis centre.
- When installing the bulb, do not touch the glass surface of the bulb with bare fingers. Doing so will cause fingerprints, grease, etc., to burn onto the bulb surface, reducing the illumination provided by the bulb. If the surface is contaminated, wipe it clean using lens tissue.
7.6 Bulb Replacement
The lamp and the lamphouse become very hot during and after a period of operation. Risk of burn – Do not touch the lamp during or immediately after period of operation. Make sure the lamp has cooled sufficiently before attempting to replace the lamp.
7.6.1 LED Module
- In order to prevent electric shock, always turn the power switch off and unplug the power cord before installing or replacing the bulb.
- Place the digital microscope on its back and pull back the lamp house cover plate.
- Firmly insert the LED module into the socket pinholes until it reaches the limit.
- Close the lamp house cover plate and secure until it snaps into position.
8. Warning Labels
The following warning labels (or symbols) are found on the digital microscope; study the meaning of the warning labels (or symbols) and always use the equipment in the safest possible manner.
Warning Label / Symbol | Explanation |
---|---|
| | Indicates that the main switch is ON. |
O | Indicates that the main switch is OFF. |
~ | Indicates alternating current. |
! | CAUTION! Risk of danger. Please consult documentation in all cases where this symbol is used. |
Do not pick the digital microscope up from the bottom during equipment operation.
Proper handling of the digital microscope will ensure years of trouble-free service. If repair becomes necessary, please contact your Motic agency or their Technical Service directly.
Motic Hong Kong Limited (Hong Kong)
Unit 2002, L20, Tower Two, Enterprise Square Five, 38 Wang Chiu Road, Kowloon Bay, Kowloon, Hong Kong
Tel: 852-2837 0888 Fax: 852-2882 2792
Motic Instruments Inc. (Canada)
130-4611 Viking Way, Richmond, B.C., V6V 2K9 Canada
Tel: 1-877-977 4717 Fax: 1-604-303 9043
Motic Deutschland GmbH (Germany)
Christian-Kremp-Strasse 11 D-35578 Wetzlar, Germany
Tel: 49-6441-210 010 Fax: 49-6441-210 0122
Motic Europe (Spain)
C. Les Corts 12, Pol. Ind. Les Corts. 08349 Cabrera de Mar, Barcelona, Spain
Tel: 34-93-756 6286 Fax: 34-93-756 6287
Website: http://www.motic.com
E-mail: info@motic.com.hk
Motic China Group., Ltd. (China)
Motic Building, Torch Hi-Tech Industrial, Development Zone, Xiamen P.R.C.
Tel: 86-0592-562 7866 Fax: 86-0592-562 7855
© 2007-2017 Motic China Group Co., Ltd. All rights reserved. Motic is a registered trademark and service mark of Motic China Group Co., Ltd. Microsoft Windows logo is a registered trademark of Microsoft Corporation. All other trademarks are the property of their respective owners.
Design Change: The manufacturer reserves the right to make changes in instrument design in accordance with scientific and mechanical progress, without notice and without obligation.