LN2 Memory Loc USB Data Logger
SPECIFICATIONS
Range: –200 to 105.00°C
Accuracy: ±0.25°C
Resolution: 0.01°C (0.1°F)
Temperature
Sampling Rate: 10 seconds
Memory Capacity: 1,048,576 points
USB Download Rate: 180 readings per second
Battery: 2 AAA (1.5V)
SETTING THE TIME/DATE
1. Slide the DISPLAY switch to the DATE/TIME position,
the thermometer will display the time-of-day and date.
Adjustable parameters are Year->Month->Day->Hour-
>Minute->12/24 hour time.
2. Press the SELECT button to enter the setting mode.
3. Subsequently press the SELECT button to select
which parameter to adjust. The selected parameter will
flash once selected.
4. Press the ADVANCE button to increment the selected
parameter.
5. Hold the ADVANCE button to continuously “roll” the
selected parameter.
6. Press the EVENT DISPLAY button to toggle between
Month/Day (M/D) and Day/Month (D/M) display
modes.
If no button is pressed for 15 seconds while in the setting
mode, the thermometer will exit the setting mode.
Changing the position of the DISPLAY switch while in the
setting mode will save the current settings.
VIEWING THE TIME-OF-DAY/DATE
To view the time-of-day/date, slide the DISPLAY switch to
the DATE/TIME position.
SELECTING THE UNIT OF MEASURE
To select the desired unit of temperature measure (°C or
°F), slide the UNITS switch to the corresponding position.
SELECTING THE TEMPERATURE PROBE CHANNEL
Slide the PROBE switch to either position ‘1’ or position ‘2’
to select the corresponding probe channel P1 or P2.
All temperature readings displayed will correspond with
the selected probe channel.
Note: Both probe channels are sampled and monitored
continuously regardless of the selected probe channel.
MINIMUM AND MAXIMUM MEMORY
The minimum temperature stored in memory is the
minimum temperature measured since the last clear of
min/max memory. The maximum temperature stored in
memory is the maximum temperature measured since the
last clear of min/max memory.
MINIMUM AND MAXIMUM TEMPERATURE
VALUES ARE NOT PROGRAMMABLE.
Minimum and maximum temperature values are stored
individually for each probe channel P1 and P2. Both
channels are monitored continuously regardless of the
selected probe channel.
VIEWING MIN/MAX MEMORY
1. Slide the PROBE switch to select temperature probe
channel to be displayed.
2. Slide the DISPLAY switch to the MIN/MAX position.
3. The thermometer will display the current, minimum,
and maximum temperatures for the selected probe
channel.
4. Press the EVENT DISPLAY button to display the
minimum temperature with the corresponding date and
time of occurrence.
5. Press the EVENT DISPLAY button a second time to
display the maximum temperature with the corresponding
date and time of occurrence.
6. Press the EVENT DISPLAY button to return to the
current temperature display.
No button press for 15 seconds while viewing the minimum
or maximum event data will trigger the thermometer
to return to the current temperature display.
CLEARING THE MIN/MAX MEMORY
1. Slide the PROBE switch to select the temperature
probe channel to be cleared.
2. Slide the DISPLAY switch to the MIN/MAX position.
3. Press the CLEAR SILENCE ALM button to clear the
current minimum and maximum temperature readings.
Alarms
Alarm high and low limits can be set individually for each
probe channel (P1 and P2).
SETTING ALARM LIMITS
1. Slide the switch to the ALARM position. Then slide the
PROBE switch to select the probe channel for which
alarms will be set.
Each digit of the alarm value is set individually:
Low Alarm Sign (Positive/Negative) -> Low Alarm
Hundreds/Tens -> Low Alarm Ones -> Low Alarm Tenths
-> High Alarm Sign (Positive/Negative) -> High Alarm
Hundreds/Tens -> High Alarm Ones -> High Alarm Tenths.
2. Press the SELECT button to enter the setting mode.
The LOW ALM symbol will flash.
3. Press the SELECT button to select the digit to adjust.
Each subsequent press of the SELECT button will
move to the next digit. Digit will flash while selected.
4. Press ADVANCE button to increment the selected
digit.
Note: The negative sign will flash if the sign is negative;
no symbol will flash if the sign is positive. Press the
ADVANCE button to toggle the sign while it is selected.
If no button is pressed for 15 seconds while in the setting
mode, the thermometer will exit the setting mode.
Changing the position of the DISPLAY switch while in the
setting mode will save the current settings.
VIEWING THE ALARM LIMITS
1. Slide the PROBE switch to select the probe channel
alarm limits to be displayed.
2. Slide the DISPLAY switch to the ALARM position.
ENABLING/DISABLING ALARMS
1. Slide the ALARM switch to the ON or OFF position to
enable or disable the alarms.
2. Alarms are enabled for both probe channels P1 and
P2 while the switch is set to ON. Alarms are disabled
for both probe channels P1 and P2 while the switch is
set to OFF.
3. The alarms cannot be configured to enable individual
channels P1 or P2 only.
ALARM EVENT HANDLING
An alarm event will trigger if the alarm is enabled and a
temperature reading is recorded below the low alarm set
point or above the high alarm set point.
When an alarm event triggers, the thermometer buzzer
will sound, and the LED for the alarming temperature
channel will flash (P1 or P2). If the alarming probe channel
is selected, the LCD symbol will flash signaling which
set point was breached (HI ALM or LO ALM).
An active alarm may be cleared by either pressing the
CLEAR SILENCE ALM button or disabling the alarm functionality
by sliding the ALARM switch to the OFF position.
Once an alarm is cleared, it will not re-trigger until after
the temperature returns to within the alarm limits.
Note: If an alarm event is triggered and returns to within
the alarm limits before being cleared, the alarm event will
remain active until it is cleared.
VIEWING ALARM EVENT MEMORY
- Slide the PROBE switch to select the probe channel
alarm data to be displayed. - Slide the DISPLAY switch to the ALARM position. The
current temperature, low alarm limit, and high alarm
limit will display. - Press the EVENT DISPLAY button. The thermometer
will display the alarm limit, date, and time of the most
recent alarm out of range condition. The symbol ALM
OUT will display to signal the date and time displayed
indicate when the temperature when out of tolerance. - Press the EVENT DISPLAY button a second time.
The thermometer will display the alarm limit, date, and
time of the most recent alarm event returning to within
the alarm limits. The symbol ALM IN will display to
signal the date and time displayed indicate when the
temperature returned to within tolerance. - Press the EVENT DISPLAY button to return to the
current temperature display.
No button press for 15 seconds while viewing the alarm
events will trigger the thermometer to return to the current
temperature display.
Note: If no alarm event has occurred for the selected
probe channel, the thermometer will display “LLL.LL” on
each line.
DATA LOGGING OPERATION
Thermometer will continuously log temperature readings
for both probe channels into permanent memory at
one-minute intervals. The total memory capacity is
1,048,576 data points. Each data point contains the
temperature reading for P1, the temperature reading for
P2, and the date and time of occurrence.
Note: All data stored in Celsius (°C) and date format is
MM/DD/YYYY.
The thermometer will also store the most recent 10 alarm
events. Each alarm event data point contains the probe
channel which alarmed, the alarm set point that was
triggered, the date and time the channel reading went
out of range, and the date and time the channel reading
returned to within range.
VIEWING THE MEMORY CAPACITY
Slide the MEM VIEW switch to the ON position.
The first line will display the current percentage of memory
full. The second line will display the number of days
remaining before memory is full. The third line will display
the logging interval (one minute).
Note: The MEM symbol will become active on the display
when the memory is 95% full.
SETTING THE LOGGING INTERVAL
1. Slide the MEM VIEW switch to the ON position. The
first line will display the current percentage of memory
full. The second line will display the number of days
remaining before memory is full at the current logging
interval. The third line will display the current logging
interval.
2. To increment the logging interval, press the ADVANCE
button. The minimum logging interval is one minute
(0:01). The maximum logging rate is 24 hours (24:00).
Once 24 hours is selected, the next subsequent press
of the ADVANCE button will return to one minute.
VIEWING UNIQUE DEVICE ID NUMBER
1. Slide the MEM VIEW switch to the ON position.
2. Press EVENT DISPLAY button. The second and third
lines will display the first eight digits of the ID number.
3. Press the EVENT DISPLAY button a second time. The
second and third lines will display the last eight digits
of the ID number.
4. Press the EVENT DISPLAY button to return to the
default display.
DOWNLOADING STORED DATA
Note: The USB download will not occur if battery LCD
symbol is active. Plug the supplied AC adapter into
the thermometer to provide suffi cient power for USB
operation.
1. The data can be downloaded directly to a USB Mass
Storage Flash Drive. To begin download, insert
USB fl ash drive into USB port located on left side of
thermometer.
2. The download will begin upon insert. LED P1 will turn
on to indicate the download has begun. Wait up to 60
seconds for LED to turn on after insertion of the drive.
Flash drives with larger amounts of data stored on the
drive prior to downloading will take longer to begin the
download.
3. Once download process is complete, LED P1 will
turn off. Do not remove USB drive until process is
complete.
4. The data transfer rate is approximately 180 data points
per second.
Note: DO NOT leave the USB Mass Storage Flash Drive
inserted into the unit. Insert, DOWNLOAD, and then
remove. The device cannot continuously write to a USB.
REVIEWING STORED DATA
The downloaded data is stored in a comma-delimited
CSV fi le on the fl ash drive. The fi lename naming convention
is “D1D2D3D4D5D6D7R1.CSV” where D1 through
D7 are the last seven digits of the thermometer’s unique
ID number and R1 is the revision of the fi le starting with
the letter “A”.
If more than one fi le is written from the same thermometer
to the USB fl ash drive, the revision letter will be incremented
in order to preserve the previously downloaded
files.
The data fi le can be opened in any software package
supporting comma-delimited fi les including spreadsheet
software (Excel) and text editors.
The fi le will contain the thermometer unique ID number,
the most recent ten temperature events, and all stored
temperature readings with date and time stamps.
Note: All data stored in Celsius (°C) and date format is
MM/DD/YYYY.
DISPLAY MESSAGES
If no buttons are pressed and LL.LL appears on the display,
this indicates that the temperature being measured
is outside of the temperature range of the unit, or that the
probe is disconnected or damaged.
BENCH STAND
The unit is supplied with a bench stand located on the
back. To use the bench stand, locate the small opening at
the bottom back of the unit. Place your fi ngernail into the
opening and fl ip the stand out. To close the stand, simply
snap it shut.
BATTERY REPLACEMENT
To replace the battery, remove the battery cover, located
on the back of the unit by sliding it down. Remove the
exhausted batteries and replace with two (2) new AAA
alkaline batteries. Insert the new batteries with the proper
polarity as indicated by the illustration in the battery
compartment. Replace the battery cover.
Replacing the batteries WILL clear the minimum/maximum
memories and high/low alarm settings. However,
replacing the batteries WILL NOT clear the time-of- day/
date settings or stored temperature data.
STATIC SUPPRESSOR INSTALLATION
Static generated, radio frequency can affect any cable
through the air or by physical contact. To protect against
radio frequency, install a suppressor onto your thermometers
cable to absorb radio frequency as follows:
Lay the cable along the center of the
suppressor with the connector to your left.
Loop the right end of the cable under the
suppressor and back up again laying the cable along
the center of the suppressor.
Carefully, snap the two halves together with the looped
cable routed through the center.
❹
This completes the installation of the suppressor.
WARRANTY, SERVICE, OR RECALIBRATION For warranty, service, or recalibration, contact:
TRACEABLE® PRODUCTS 12554 Old Galveston Rd. Suite B230
Webster, Texas 77598 USA
Ph. 281 482-1714 • Fax 281 482-9448
E-mail support@traceable.com • www.traceable.com
Traceable® Products are ISO 9001:2015 Quality Certifi ed by DNV and ISO/IEC 17025:2017 accredited as a Calibration Laboratory by A2LA.
Cat. No. 6458 / 6459
Traceable® is a registered trademark of Cole-Parmer. Memory-Loc™ is a trademark of Cole-Parmer.
©2020 Traceable® Products. 92-6458-00 Rev. 2 072425