Product presentation

- Back mounted hanging hole
- LCD display screen
- LED indicator light
- Start stop button
- Serial number
- Light sensor
- Power supply+data interface
- Built-in sensors
- Activate/Flight Mode Button
- External sensor
LCD display

- Work status
- Function Description
- Flight mode
- Signal status
- Battery level
- Light and vibration alarm signs
- Time and number of records
- Temperature and humidity values
Note: To ensure data reliability, please do not enable recording when the battery level is less than or equal to 20% (
between 10% and 20%)
Selection table
(standard light and vibration sensors)
| Model | Probe type | Measuring range | Accuracy |
|
Glog 5 T |
Built in temperature |
-30℃~70℃ |
±0.5℃ |
|
Glog 5 TH |
Built in temperature and humidity |
-30℃~70℃ 0%RH~100%RH | ±0.5℃
±5%RH |
|
Glog 5 TE |
External temperature+internal temperature |
-40℃~85℃ |
±0.5℃ |
|
Glog 5 THE |
External temperature and humidity+internal temperature |
External: -40℃~85℃ 0%RH~100%RH
built-in: -30℃~70℃ |
±0.5℃
±5%RH |
|
Glog 5 TLE |
External ultra-low temperature+internal temperature | External: -200℃~150℃ built-in: -30℃~70℃ | ±0.5℃(-40℃~85℃)
±1℃(-100℃~150℃) ± 2 ℃ (other) |
| Glog 5 CO₂ | External ul-CO₂ & temperature
& humidity built-in temperature |
-30℃~70℃ 0%RH~100%RH
0%~100vol% |
±0.5℃
±5%RH 1vol%+3vol% measured |
Technical parameter
| Vibration range | 0g~16g |
| Range of illumination intensity | 0~52000Lux |
| Resolving power | 0.1°C/0.1%RH/0.1g/1Lux |
| Key | Double button design |
| Indicator light | Red and green, red and blue LED indicator lights |
| Display | Broken code display screen |
| Positioning method | LBS+GPS (optional) |
| Storage capacity | 60000 (pre shadow data) 32000 (normal recording+rear shadow data) |
| Shadow data | Front+Rear |
| Record interval | 1min~24h; Default: 5 minutes |
| Upload interval | 1min~24h adjustable Default: 60 minutes |
| Data upload method | 4G+2G |
| Freight initiation method | Button, platform, timer |
| Freight stopping method | Button, platform, full memory |
| Repeat startup | 3 times (within the shelf life) |
| Flight Mode | Button, timer, electronic fence |
| Alarm Mode | Overlimit and low battery |
| Battery type | 3.0V disposable lithium manganese battery 4800mAh (CR14505 3B6H) |
| OTA upgrade | Upgrades can be made through data management software and platforms |
| Waterproof grade | IP65 (built-in) |
| Work environment | -30 ℃~70℃, 0% RH~100% RH (non condensing) |
| Storage environment | 15~30℃,20~75%RH |
| Specification and size | 103 x 61.3 x 30 (mm) |
| Battery life | Record once every 5 minutes in a 25 ℃ environment, upload every 30 minutes, and use the temperature and humidity recorder for a total of 60 days; |
| Shelf life | (Usage time+Storage time) 1 year |
Technical parameter
Add device
Login platform website: http://www.i-elitech.com, or scan the code to download and install the APP . Register, log in, and add devices according to the prompts.
Activation
Press and hold the
button for 5s, and the screen will display normally, indicating that the device has been activated.
Note: Clicking on the button and the screen displaying normally indicates that the device has been activated;
Start Record
Press and hold the
button for 5s, and the screen will display a
symbol, indicating successful startup recording; If the
indicator flashes, it will enter a delayed countdown state, and after the delay ends, it will automatically start recording;
Stop recording
Press and hold the
button for 5s, and the screen will display a
sign indicating that the recording has been successfully stopped;
Data export
The device is connected to the computer USB interface through a data cable and will automatically generate PDF+CSV format data reports. You can copy the generated report to your computer for saving.
Note: The keys are invalid during the data export process
Flight Mode
- Enable: Press and hold the
button for 5s, and the screen will display a
symbol indicating that the airplane mode has been successfully activated; - Close: Press and hold the
button for 5s. If the screen
icon disappears, it indicates that the airplane mode has been turned off;
Note: To enable and disable other flight modes, please operate through the platform or app.
Freight Management Operation Instructions
- After logging into the platform or app, click on the ‘Freight Management’ menu in the main menu;
- Click on the ‘Configuration’ menu on the homepage, and sequentially create and add ‘User Management’, ‘Carrier’, ‘Location’, ‘Alarm Presets’, and’ Freight Template ‘;
- Click on the ‘Freight Management’ menu on the homepage to create and manage freight;
Note: For detailed operation manuals, please refer to the operating instructions on the platform or APP;
LED indicator light description

LCD display instructions

Safety instructions and precautions
Safety instructions
- To ensure that you install and use this product correctly, please read and strictly comply with the following terms:
Battery
- Please use the original battery and do not use other batteries to prevent damage to the equipment or other malfunctions. Do not disassemble, squeeze, hit, or heat the battery privately, and do not place it in a fire, otherwise it may cause battery explosion and fire.
Equipment
- Do not use this equipment in flammable and explosive gas environments, otherwise it may cause explosions or fires. During the use of the equipment, if the equipment emits a burnt or other odor, the power supply should be immediately cut off and the manufacturer or supplier should be contacted in a timely manner.
Precautions and maintenance requirements
- If the equipment is not used for a long time, it should be removed and stored in a dry and cool environment in the packaging box.
- Users are not allowed to make any unauthorized changes to this device. Any unauthorized changes may affect the accuracy of the device or even damage it.
- Do not place the device outdoors for use to avoid instrument short circuits, burns, and other malfunctions caused by adverse weather conditions such as rain and lightning.
- When the recorder is offline (no data upload), please confirm the networking status of the device;
- Please use within the measurement range of the recorder
- Please do not forcefully impact this recorder
- Sensors should be calibrated regularly to ensure their accuracy meets usage requirements
- The measured values of the recorder may be affected by the following factors:
Temperature error:
The stable time in the measurement environment is too short; Close to heat sources, cold sources, or directly exposed to sunlight.
Humidity error
The stable time in the measurement environment is too short; Long term exposure to steam, water mist, water curtains, or condensation environments.
Contaminated
Being in a dusty or other polluted environment
- If the sensor is found to be blocked or contaminated, it should be wiped with a dust-free cloth dipped in clean water. It is strictly prohibited to use high molecular weight chemical solvents such as alcohol for wiping.
- If scrapped, it must be strictly handled in accordance with the “Regulations on the Management of Recycling and Disposal of Waste Electrical and Electronic Products”.
FAQs
What should I do if the battery level is less than 20%?
To ensure data reliability, do not enable recording when the battery level is less than or equal to 20%.
How do I know if the device is activated?
Clicking on the activation button and seeing the screen display normally indicates that the device has been activated.
How can I export data from the device?
Connect the device to the computer via USB to generate PDF+CSV format data reports automatically.
Documents / Resources
![]() |
Elitech Glog 5 Real Time Single Use IoT Data Logger [pdf] User Manual Glog 5 T, Glog 5 TH, Glog 5 TE, Glog 5 THE, Glog 5 TLE, Glog 5 CO, Glog 5 Real Time Single Use IoT Data Logger, Glog 5, Real Time Single Use IoT Data Logger, Single Use IoT Data Logger, Data Logger, Logger |