partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors logo

partech OxyTechw2 RDO Pro-X Dissolved Oxygen Sensor WaterWatch2 Monitors

partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors product

OxyTechw2 RDO Sensor
The OxyTechw2 RDO Sensor measures dissolved oxygen and temperature in a variety of applications. Partech’s RDO sensor is and accurate. The stainless steel housing provides corrosion resistance. A replaceable sensor cap ensures quality data.

Standard Sensors

  • Dimensions: 203 mm (7.9) long x 47 mm (1.8) diameter
  • Weight: 0.8 kg (1.76 lbs) including 5m of cable
  • Accuracy: Range
  • Cable Length: 5 or 10 meters standard
  • Service Requirement: No routine servicing

RDO Sensor – Maintenance

How to Replace the RDO Cappartech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 02

  1. The monitor displays a warning when the sensor cap is close to expiration.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 03
  2. Remove the nose cone from the probe.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 04
  3. Pull the RDO cap away from the probe.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 05
  4. Align the flat edge on the tip of the probe with the flat edge inside the cap. Slide the cap in place.

RDO Sensor – Programming

Using the Salinity Settingspartech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 06

  1. Select Sensor Config from the main menu and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 07
  2. Choose Set Salinity and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 08
  3. Select the appropriate default or choose User Defined. Press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 09
  4. Use.the.arrow.keys.to. set the value. Press OK when finished.

Setting Barometric Pressurepartech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 10

  1. Select Sensor Config from the main menu and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 11
  2. Select Set Baro Pressure and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 12
  3. Use the arrow keys to set the value and press OK when finished.

Restoring Default Settingspartech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 13

  1. Select Sensor Config from the main menu and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 14
  2. Choose Restore Defaults and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 15
  3. The monitor displays sensor info. Press OK when the restore is complete.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 16
  4. When the restore is complete, press OK.

RDO Sensor – Calibration

Dissolved Oxygen Sensor – 100% Open Air Calibration Calibrate the DO sensor using the saturation method described below. See the manual for the Partech 750w? Monitor to get more information about calibration.

  1. partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 17Press Menu on the keypad to display the main menu.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 18
  2. Select Measurement Config and press the OK key.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 19
  3. Choose DO Saturation and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 20
  4. Highlight Calibrate and press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 21
  5. The monitor displays sensor info. Press OKpartech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 22
  6. Calibrate the sensor in open air free from drafts. Press OK.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 23
  7. Wait for a saturation percentage of roughly 90%. Then press OK.

General Info
The DO sensor measures dissolved oxygen and temperature in a variety of applications. Partech’s DO sensor is rugged and accurate. The stainless steel housing provides corrosion resistance. A replaceable sensor cap ensures quality data.partech -Oxy-Techw2 -RDO -Pro-X- Dissolved- Oxygen -Sensor- WaterWatch2 -Monitors 011-970-498-1500
www.in-situ.com

Documents / Resources

partech OxyTechw2 RDO Pro-X Dissolved Oxygen Sensor WaterWatch2 Monitors [pdf] Installation Guide
OxyTechw2 RDO Pro-X Dissolved Oxygen Sensor WaterWatch2 Monitors, OxyTechw2 RDO, Pro-X Dissolved Oxygen Sensor WaterWatch2 Monitors, Oxygen Sensor WaterWatch2 Monitors, Sensor WaterWatch2 Monitors, WaterWatch2 Monitors

References

Leave a comment