Installation Guide for partech models including: OxyTechw2 RDO Pro-X Dissolved Oxygen Sensor WaterWatch2 Monitors, OxyTechw2 RDO, Pro-X Dissolved Oxygen Sensor WaterWatch2 Monitors, Oxygen Sensor WaterWatch2 Monitors, Sensor WaterWatch2 Monitors, WaterWatch2 Monitors
Partech OxyTechw² Rugged Dissolved Oxygen Sensor, RDO, P230470 | USABlueBook.com
Partech OxyTechw² Rugged Dissolved Oxygen Sensor, RDO, P230469 | USABlueBook.com
File Info : application/pdf, 3 Pages, 1.41MB
DocumentDocumentOxyTechw2 RDO Sensor General Info The.RDO.sensor.measures.dissolved.oxygen.and.temperature.in .a.variety.of.applications. Partech's.RDO.sensor.is.rugged.and.accurate..The.stainless.steel housing.provides.corrosion resistance..A.replaceable.sensor.cap.ensures.quality.data.. STANDARD SENSORS DIMENSIONS WEIGHT OxyTechw² RDO Sensor 203 mm (7.9") long x 47 mm (1.8") diameter 0.8 kg (1.76 lbs) inc 5m of cable ACCURACY RANGE CABLE LENGTH ±0.1 mg/l (0-20 mg/l) ±2% (20-60 mg/l) 0-60 mg/l; 0-600% Saturation -5°C to 50°C (23°F to 122°F) 5 or 10 metres standard SERVICE REQUIREMENT No routine servicing RDO Sensor - Maintenance How to Replace the RDO Cap Old cap New cap The monitor displays a.warning.when.the. sensor cap is close to expiration. Remove the nose cone from the probe. Pull the RDO cap away from the probe. Align.the.flat.edge.on. the tip of the probe with.the.flat.edge. inside the cap. Slide the cap in place. 1-970-498-1500 1 www.in-situ.com RDO Sensor - Programming Using the Salinity Settings Select Sensor Config from the main menu and press OK. Choose Set Salinity and press OK. Setting Barometric Pressure Select the appropriate Use.the.arrow.keys.to. default or choose User set the value. Press OK Defined. Press OK. when.finished. Select Sensor Config from the main menu and press OK. Select Set Baro Pressure and press OK. Restoring Default Settings Use.the.arrow.keys.to. set the value and press OK.when.finished. Select Sensor Config from the main menu and press OK. Choose Restore The monitor displays Defaults and press OK. sensor info. Press OK. When the restore is complete,.press.OK. 1-970-498-1500 2 www.in-situ.com RDO Sensor - Calibration 1-970-498-1500 3 www.in-situ.com