Palintest -logoLumiso Expert
Quick Start Guide 

Palintest Lumiso Expert Benchtop Kit-

Registration and technical support

Thank you for choosing a Palintest Lumiso
Expert Photometer. Please register your instrument now at www.palintest.com/products/lumiso-expert or scan the QR Code.

Palintest -qr

There, you will also find invaluable support for your new instrument, including:

  • User Manual
  • Supporting documents for all test methods
  • Firmware updates
  • Palintest Connect cloud based data storage and management

Getting started

Onscreen instructions are available for all test methods.

Palintest Lumiso Expert Benchtop Kit-fig1

Achieving accurate results

Blanking

  • Blanking is critical and sets the zero level for each test.
  • Usually, the blank is a sample without any reagents added.
  • The onscreen instructions show blanking as the first few steps, however, the “blank” can be prepared in a separate tube and blanking carried out just prior to the  measurement step.

Best practice when testing

Rinse equipment thoroughly with deionised water or the water being tested.

Palintest Lumiso Expert Benchtop Kit-fig2

Measure volumes accurately as shown (or use a pipettor or syringe).

Palintest Lumiso Expert Benchtop Kit-fig4

Use the correct Palintest reagents and observe all safety information supplied with them.

Palintest Lumiso Expert Benchtop Kit-fig5

Check for bubbles and remove as shown.

Palintest Lumiso Expert Benchtop Kit-fig6

Align tubes as shown. Always use the tube cap as it excludes light from the optics.

Palintest Lumiso Expert Benchtop Kit-fig7

Keep instrument clean. Dry all tubes before placing in the optical cell.

Palintest Lumiso Expert Benchtop Kit-fig8

Verify instrument calibration
Check Standards are available for your Lumiso Expert. Use these regularly to confirm that the instrument is within calibration.

Palintest Lumiso Expert Benchtop Kit-fig9

Palintest -iconwww.palintest.comPalintest -logoA Halma company
©Palintest Limited 2022. All rights reserved.
Registered office: Palintest House, Kingsway, Team Valley,
Tyne & Wear, England NE11 0NS.
Registered in England, Number 1204118
ZIINST7210 V1

Documents / Resources

Palintest Lumiso Expert Benchtop Kit [pdf] User Guide
Lumiso Expert Benchtop Kit, Expert Benchtop Kit, Benchtop Kit

References

Leave a comment

Your email address will not be published. Required fields are marked *