Milwaukee MI405 PRO Ammonia Medium Range Photometer

Dear Customer, Thank you for choosing a Milwaukee Instruments product. This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter.

Specifications

ParameterValue
Range0.00 to 9.99 mg/L (NH3-N)
Resolution0.01 mg/L
Precision±0.30 mg/L @5.00 mg/L
Light SourceBlue LED 466 nm
Light DetectorSilicon Photocell and 466 nm narrow band interference filter
MethodAdaptation of Nessler method
Environment0 to 50 °C (32 to 122 °F), 100% RH max
Battery Type9 volt (1 pc)
Auto-Shut offAfter 10' of non-use
Dimensions192 x 104 x 52 mm (7.5 x 4.1 x 2")
Weight380 g

Measurement Procedure

  1. Turn the meter on by pressing the [ON/OFF button].
  2. When the LCD displays "- - -", the meter is ready.
  3. Fill the cuvet with 10 mL of sample, up to the mark, and replace the cap.
  4. Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove.
  5. Press [ZERO button] and "SIP" will blink on the display.
  6. After a few seconds the display will show "-0.0-". The meter is now zeroed and ready for measurement.
  7. Remove the cuvet. Open it.
  8. Add 4 drops of NH3-1 (10 drops in case of seawater analysis) and swirl gently.
  9. Add 4 drops of NH3-2 (10 drops in case of seawater analysis). Replace the cap and swirl gently.
  10. Immediately reinsert the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove.
  11. Press and hold [READ button] for 3 seconds and the display will show the countdown prior to the measurement. Alternatively, wait for 3 minutes and 30 seconds and just press [READ button]. In both cases "SIP" will blink during measurement.
  12. The instrument directly displays the concentration in mg/L of ammonia nitrogen (NH3-N). To convert the reading to mg/L of ammonia (NH3), multiply by a factor of 1.216.

On Sample Reading

Battery Replacement

Battery replacement must only take place in a non-hazardous environment. Simply rotate the battery cover on the back of the meter. Detach the battery from the terminals and attach a fresh 9V battery while paying attention to the correct polarity. Insert the battery and replace the cover.

Accessories

Certification

Milwaukee Instruments conform to the CE European Directives. Disposal of Electrical & Electronic Equipment: Do not treat this product as household waste. Hand it over to the appropriate collection point for the recycling of electrical and electronic equipment. Disposal of waste batteries: This product contains batteries. Do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling. Please note: proper product and battery disposal prevents potential negative consequences for human health and the environment. For detailed information, contact your local household waste disposal service or go to www.milwaukeeinstruments.com (USA & CAN) or www.milwaukeeinst.com.

Recommendation

Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any modification introduced by the user to the supplied equipment may compromise the meter's performance. For your and the meter's safety, do not use or store the meter in a hazardous environment. To avoid damage or burn, do not perform any measurement in microwave ovens.

Warranty

These instruments are warranted against defects in materials and manufacturing for a period of 2 years from the date of purchase. This warranty is limited to repair or free of charge replacement if the instrument cannot be repaired. Damage due to accidents, misuse, tampering, or lack of prescribed maintenance is not covered by warranty. If service is required, contact your local Milwaukee Instruments Technical Service. If the repair is not covered by the warranty, you will be notified of the charges incurred. When shipping any meter, make sure it is properly packaged for complete protection.

Guide to Display Codes

Error Messages

On Zero Reading

Other Errors

Models: MI405 PRO Ammonia Medium Range Photometer, MI405, PRO Ammonia Medium Range Photometer, Medium Range Photometer, Photometer

File Info : application/pdf, 2 Pages, 381.57KB

PDF preview unavailable. Download the PDF instead.

ISTMI405 07 20

References

Adobe PDF Library 15.0 Adobe InDesign 15.1 (Windows)

Related Documents

Preview Milwaukee MW12 Digital Phosphate Tester User Manual & Specifications
Comprehensive user manual for the Milwaukee MW12 Digital Phosphate Tester. Includes specifications, operation instructions, battery replacement, warranty, and compliance information for accurate phosphate measurement.
Preview Milwaukee MW11 Digital Total Chlorine Tester User Manual
User manual for the Milwaukee MW11 Digital Total Chlorine Tester, detailing specifications, operation, preliminary examination, functional description, and battery replacement procedures.
Preview Milwaukee pH51 Waterproof pH Tester User Manual and Specifications
User manual and specifications for the Milwaukee pH51 Waterproof pH Tester, including conditioning, calibration, operation, electrode replacement, and battery replacement instructions. Features include a pH range of 0.0 to 14.0 pH and a 2-year warranty.