MW11 Digital Total Chlorine Tester

User Manual

Specifications

Range0.00 to 3.50 ppm
Resolution0.01 ppm
Accuracy±0.03 ppm ±3% of reading @ 25 °C
Typical EMC Dev.±0.01 ppm
Light SourceLight Emitting Diode @ 525 nm
Light DetectorSilicon Photocell
MethodAdaptation of USEPA method 330.5. The reaction between total chlorine and the DPD reagent causes a pink tint in the sample.
Environment0 to 50°C (32 to 122 °F) max. 95% RH non-condensing
Battery Type1 x 1.5V AAA
Auto-Shut OffAfter 5 min. of non-use and 2 min. after reading.
Dimensions87.5 x 63.7 x 35.2 mm (3.4 x 2.5 x 1.4")
Weight65 g (2.29 oz.)

Optional Accessories

Reagent Sets

Other Accessories

Certification & Disposal

Milwaukee Instruments conform to the CE European Directives.

Disposal of Electrical & Electronic Equipment: Do not treat this product as household waste. Hand it over to the appropriate collection point for the recycling of electrical and electronic equipment.

Disposal of waste batteries: This product contains batteries. Do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.

Please note: proper product and battery disposal prevents potential negative consequences for human health and the environment. For detailed information, contact your local household waste disposal service or go to www.milwaukeeinstruments.com (USA & CAN) or www.milwaukeeinstruments.eu.

Recommendation

Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any modification introduced by the user to the supplied equipment may compromise the tester's performance. For your and the tester's safety, do not use or store the tester in a hazardous environment. To avoid damage or burn, do not perform any measurement in microwave ovens.

Warranty

This tester is warranted against defects in materials and manufacturing for a period of 2 years from the date of purchase. This warranty is limited to repair or free of charge replacement if the tester cannot be repaired. Damage due to accidents, misuse, tampering, or lack of prescribed maintenance is not covered by warranty. If service is required, contact your local Milwaukee Instruments Technical Service. If the repair is not covered by the warranty, you will be notified of the charges incurred. When shipping any tester, make sure it is properly packaged for complete protection.

Milwaukee Instruments reserves the right to make improvements in design, construction, and appearance of its products without advance notice.

Preliminary Examination

Thank you for choosing Milwaukee Instruments! Remove the tester and accessories from the packing and examine it carefully. For further support, please contact your closest Milwaukee Instruments office.

Each MW11 tester is supplied complete with:

Functional Description

The tester has the following components:

Operation

  1. Press the [On/Off button] to turn the tester on. All segments are displayed for a few seconds followed by "Add", "C1" with "Press" blinking.
  2. Fill the cuvette with 10 mL of unreacted sample and replace the cap. Place the cuvette into the tester and close the cap.
  3. Press the [On/Off button]. When the display shows "Add", "C.2" with "Press" blinking, the tester is zeroed.
  4. Remove the cuvette, unscrew the cap, and add the content of one packet of MI524-0 reagent. Replace the cap and shake gently for 20 seconds. Replace the cuvette into the tester.
  5. Press and hold the [On/Off button]. The display will show a countdown prior to the measurement (e.g., 2:29). Alternatively, wait 2 minutes and 30 seconds and press the [On/Off button].
  6. When the timer ends, the tester will perform the reading. The tester directly displays the concentration of total chlorine in ppm (e.g., 1.52).

The tester automatically turns off 2 minutes after reading.

Battery Replacement

To save the battery, the tester shuts down after 5 minutes of non-use and 2 minutes after reading. A fresh battery lasts for a minimum of 5000 measurements, depending on the light level.

When the battery capacity is under 10%, the tester displays "bAt" at start up.

When the battery is drained, the tester displays "bAd" then "bAt" message.

Replace the battery with a new one and restart the tester.

To replace the battery, follow these steps:

  1. Press the [On/Off button] to turn the tester off.
  2. Turn the tester upside down and use a screwdriver to unfasten the screw and remove the battery cover.
  3. Remove the old battery and replace it with a fresh 1.5V AAA battery, inserting the negative end first.
  4. Replace the battery cover, fasten, and tighten the screw.
Models: MW11 Digital Total Chlorine Tester, MW11, MW11 Chlorine Tester, Digital Total Chlorine Tester, Chlorine Tester, Digital Chlorine Tester, Tester

File Info : application/pdf, 2 Pages, 172.56KB

PDF preview unavailable. Download the PDF instead.

ISTMW11 04 21

References

PDFium PDFium

Related Documents

Preview Milwaukee MW12 Digital Phosphate Tester User Manual & Specifications
Comprehensive user manual for the Milwaukee MW12 Digital Phosphate Tester. Includes specifications, operation instructions, battery replacement, warranty, and compliance information for accurate phosphate measurement.
Preview Milwaukee MI405 PRO Ammonia Medium Range Photometer User Manual
User manual for the Milwaukee MI405 PRO Ammonia Medium Range Photometer, detailing specifications, measurement procedures, on-sample reading indicators, battery replacement, accessories, certification, recommendations, warranty, display codes, and error messages.
Preview Milwaukee pH51 Waterproof pH Tester User Manual and Specifications
User manual and specifications for the Milwaukee pH51 Waterproof pH Tester, including conditioning, calibration, operation, electrode replacement, and battery replacement instructions. Features include a pH range of 0.0 to 14.0 pH and a 2-year warranty.