Milwaukee pH51 Waterproof pH Tester

User Manual and Specifications

Specifications

Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy (@25 °C) ±0.1 pH
Calibration Manual, at 2 points through trimmers
pH electrode MA73600
Environment 0 to 50 °C; 100% RH
Battery type 3 x 1.5V, alkaline
Battery life more than 1500 hours of continuous use
Dimensions 165 x 30 x 30 mm
Weight 80 g

Accessories

MA73600 Replaceable pH electrode
M10004B pH 4.01 buffer, 20 ml sachet, 25 pcs.
M10007B pH 7.01 buffer, 20 ml sachet, 25 pcs.
M10010B pH 10.01 buffer, 20 ml sachet, 25 pcs.
MA9015 Electrode storage solution, 230 ml
MA9016 Electrode cleaning solution, 230 ml
M10000B Electrode rinse solution, 20 ml sachet, 25 pcs.

Certification

Milwaukee Instruments conform to the CE European Directives.

C RoHS compliant

Disposal of Electrical & Electronic Equipment: Do not treat this product as household waste. Hand it over to the appropriate collection point for the recycling of electrical and electronic equipment.

Disposal of waste batteries: This product contains batteries. Do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.

Please note: proper product and battery disposal prevents potential negative consequences for human health and the environment. For detailed information, contact your local household waste disposal service or go to www.milwaukeeinstruments.com (USA & CAN) or www.milwaukeeinst.com.

Recommendation

Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any modification introduced by the user to the supplied equipment may compromise the meter's performance. For your and the meter's safety, do not use or store the meter in a hazardous environment. To avoid damage or burn, do not perform any measurement in microwave ovens.

Warranty

This instrument is warranted against defects in materials and manufacturing for a period of 2 years from the date of purchase. The electrode is warranted for 6 months. This warranty is limited to repair or free-of-charge replacement if the instrument cannot be repaired. Damage due to accidents, misuse, tampering, or lack of prescribed maintenance is not covered by warranty. If service is required, contact your local Milwaukee Instruments Technical Service. If the repair is not covered by the warranty, you will be notified of the charges incurred. When shipping any meter, make sure it is properly packaged for complete protection.

Milwaukee Instruments reserves the right to make improvements in design, construction, and appearance of its products without advance notice.

Conditioning

Remove the cap and immerse the pH51 in MA9015 storage solution for 2 hours to activate the electrode.

Calibration

Unscrew the battery compartment cap on the top of the meter and perform one of the following two-point calibration procedures, according to the required application:

Acid calibration (for measurements below pH 7):

Alkaline calibration (for measurements above pH 7):

Replace the battery compartment cap.

Operation

Electrode Replacement

The electrode can be easily replaced in the following way:

Battery Replacement

To replace the batteries, unscrew the battery compartment cap and replace all three 1.5V batteries, paying attention to their polarity. Make sure the gasket is in place before screwing back the cap.

Batteries should only be replaced in a non-hazardous area using the battery type specified in this instruction manual.

Models: pH51 PRO Waterproof pH Tester, pH51 PRO, Waterproof pH Tester, pH Tester, Tester

PDF preview unavailable. Download the PDF instead.

Milwaukee-pH51-Manual Adobe InDesign 15.1 (Macintosh) Adobe PDF Library 15.0

Related Documents

Preview Milwaukee MW12 Digital Phosphate Tester User Manual & Specifications
Comprehensive user manual for the Milwaukee MW12 Digital Phosphate Tester. Includes specifications, operation instructions, battery replacement, warranty, and compliance information for accurate phosphate measurement.
Preview Milwaukee MW11 Digital Total Chlorine Tester User Manual
User manual for the Milwaukee MW11 Digital Total Chlorine Tester, detailing specifications, operation, preliminary examination, functional description, and battery replacement procedures.
Preview Milwaukee MI405 PRO Ammonia Medium Range Photometer User Manual
User manual for the Milwaukee MI405 PRO Ammonia Medium Range Photometer, detailing specifications, measurement procedures, on-sample reading indicators, battery replacement, accessories, certification, recommendations, warranty, display codes, and error messages.