M3xxx Series Firmware Update Instructions

Ludlum Measurements, Inc.

501 Oak Street, P.O. Box 810
Sweetwater, Texas 79556
325-235-5494, FAX: 325-235-4672

M3xxx Series Firmware Update Instructions

General

This document specifically addresses updating firmware on the Model 3003, 30031, 3014, 3078, and 3078i. If you have any questions, please contact your local sales representative or Ludlum Measurements, Inc. at 800-622-0828 before proceeding.

The newest generation of Model 3000 series instruments (Gen 2) has greater capabilities than its predecessors. With this greater capability comes the need to store more information than ever before. As a result, new procedures are needed to update instrument firmware in the field.

Warning

The Ludlum Lumic Firmware Updater is free to customers and may be used to update the firmware on particular instruments to add or change currently available features. Changes may or may not be reversible. While Ludlum Measurements has taken great care to develop firmware updates that will not affect current instrument calibration, it is still recommended that all firmware updates be performed prior to instrument recalibration. If you have any concerns before proceeding, contact your local sales representative or Ludlum Measurements, Inc. at 800-622-0828.

Procedure

Equipment Required:

  • Ludlum Lumic Firmware Updater (available for free from the Ludlum Measurements, Inc. website: Lumic Firmware Updater Installer)
  • USB cable A male to A male (Ludlum P/N 22-9150)
  • Text Editor

Steps:

  1. Install the latest version of Ludlum Lumic Firmware Updater.
  2. Turn on the device to be updated.
  3. Connect the USB cable between the device and computer.
  4. Wait until the USB driver has been successfully installed. (This may take a short time, or up to 5 minutes.)
  5. Open the Lumic Firmware Updater and select the "Model 3xxx Series" from the 'Connected instrument' box.
  6. Select File > Save Parameters to a File and name the parameter file something unique. (There will be a pause as it connects and retrieves all of the appropriate information.)
  7. Open the comma separated value file, that was just created, using a text editor and search for the following parameters and record the results:
    • Dev_FW (e.g., F.2.1.1000, F.3.0.1084)
    • Dev_HW[0]_Name (e.g., 5519-001, 5519-305)

    Note: Leave the file open as it will need to be edited. The file may contain parameters such as Dev_Cnt_AudioMode, Dev_Cnt_DisplayMode, Dev_Det_Retain, Dev_DIPSw[0], Dev_HW[0]_Name, Dev_HW[1]_Name, etc.

    Description of Figure 1: Save Parameters to a File: A screenshot of the Lumic Firmware Updater software showing the 'File' menu open, with 'Save Parameters to a File' highlighted.

    Description of Figure 2: Comma Separated Value File Name: A screenshot of a file save dialog box, showing a file named '25010274 Parameters49630N16' being saved as a 'Comma Separated Value File'.

  8. If the Dev_FW value determined in step 7(a) is 2.0.xxxx or less, stop now and contact your sales representative or Ludlum Measurements, Inc. before proceeding.
  9. Edit the file from step 7:
    • Save the parameter file a second time with a new name that will help indicate that the file has been edited. (Upon completion, two copies of the same file will be saved on the computer; however, this file will now be edited.)
    • Return to the first line of the file.
    • Search for 'Dev_' to find the beginning of the device parameters, skipping the detector parameters ('DetMem[x]_...') in the process.
    • Delete all remaining lines of the file, leaving only the list of parameters beginning with 'DetMem[x]_' as well as time and date.
    • Save the file.

    Description of Figure 3: Device Firmware Number: A snippet of text showing example device parameters, including Dev_FW, Dev_HW[0]_Name, and other configuration settings.

  10. From the Ludlum Measurements website (Ludlum Measurements Website), download the latest version of firmware associated with your instrument and save the desired new firmware file to your computer. Please note that the value determined in Step 7 is the printed circuit board number that will be needed when downloading the appropriate firmware. (e.g. 5519-305_M3XXX_A-F.3.0.1084.lmzip)
  11. Select 'Browse' on the main screen and select the new firmware file to be loaded from step 10. Once the desired firmware is selected, click 'Open'. The 'Firmware Info' section will display details such as Instrument Model, Board, Model Code, Version, and Build.
  12. Select the 'Update' button on the main screen.
    • Once the file has finished loading, the instrument will restart and the device will disappear from 'Connected instruments' on the Lumic Firmware Updater screen.
  13. When the instrument reappears on the Lumic Firmware Updater screen, select the "Model 3xxx Series" from the 'Connected instruments' box.
  14. Select File > Load parameters from a file.
  15. Select the parameter file saved in step 9 (the edited file) and select 'Open'.

    Note: No error messages should pop up. If they do, it may be an indication of a problem with the saved data or the instrument. For instance, the previously saved value may now be outside of the available parameter limits for the latest firmware release and the parameter will need to be reconfigured before proceeding. Stop and contact your local sales representative or Ludlum Measurements, Inc. if you have any concerns.

  16. Review parameters and instrument operation including performing a simple check source test to confirm instrument operability.
Models: 3003, 3003i, Multi Detector Radiation Meters, Detector Radiation Meters, Radiation Meters, 3003, Meters

PDF preview unavailable. Download the PDF instead.

Firmware-Update-Instructions M3003-M3014-M3078 Adobe PSL 1.2e for Canon Canon iR-ADV 4235 PDF

Related Documents

Preview Ludlum Model 4525 Generation 5 Installation Manual
Comprehensive installation guide for Ludlum Measurements, Inc. Generation 5 radiation portal monitor systems, including models 4525-7000, 4525-10500, and 4525-14000. Covers packaging, site preparation, anchoring, leveling, wiring, and diagrams.
Preview Ludlum Measurements Model 43-37 Series Gas Proportional Detector Specifications
Detailed specifications for the Ludlum Measurements Model 43-37 and 43-37-1 series gas proportional detectors, designed for alpha and beta survey applications.
Preview Ludlum Model 4525 Generation IV Series Radiation Portal Monitor Installation Manual
This installation manual provides comprehensive guidance for setting up Ludlum Model 4525 Generation IV Series Radiation Portal Monitors, covering site preparation, hardware installation, and software configuration for effective radiation detection in vehicle traffic.
Preview Ludlum Model 30-6 Extended Reach microR Survey Meter Datasheet
Detailed specifications and features for the Ludlum Model 30-6 Extended Reach microR Survey Meter, a rugged, low-weight gamma detector system designed for radiation measurement.
Preview Ludlum Model 42-5 Neutron Ball Cart Manual and Specifications
Comprehensive manual detailing the specifications, operating procedures, calibration methods, and parts list for the Ludlum Model 42-5 Neutron Ball Cart, a radiation detection instrument.
Preview Ludlum Model 297 Signal Splitter User Manual
User manual for the Ludlum Model 297 Signal Splitter, detailing its introduction, getting started guide, specifications, recycling information, and diagrams.
Preview Ludlum Model 9-7 Ion Chamber Display Unit: Features, Specifications, and Detectors
Detailed information on the Ludlum Model 9-7 Ion Chamber Display Unit, including its features, technical specifications, compatible detectors, and available options for radiation measurement.