☑ LUDLUM MEASUREMENTS, INC.

Model 30-6 Extended Reach microR Survey Meter

Features

Visuals

Main Product Image: A full view of the Ludlum Model 30-6 Extended Reach microR Survey Meter, featuring a long adjustable pole, a detector head at the end of the pole, and a digital display unit mounted on the pole. The display unit is shown in detail in a separate inset image.

Close-up of Pole and Handle: An inset image showing a close-up of the adjustable pole and handle mechanism of the Model 30-6 survey meter, highlighting its construction and adjustability.

Meter Face Detail: An inset image displaying the Model 30 meter face. It features a digital screen showing a reading of '0', several control buttons, and indicator lights.

Specifications

Part Number: 48-4195

Detector: EJ-212 plastic scintillator, 15.2 x 2.5 cm (6 x 1 in.) (D x L)

Sensitivity (¹³⁷Cs gamma): Typically 150 kcpm per µSv/h (1500 cpm per µR/hr)

Energy Response: Energy dependent

Linearity: Reading ±10% of actual value

LCD Display: 3 digit LCD with large 13.4 mm (0.53 in.) digits, low-battery indicator, MAX, ALARM, MUTE

Backlight: Built-in ambient light sensor automatically activates low-power LED backlight, unless internal dipswitch is set to continuous-on (will reduce battery life). Alarm light intensity varies based on ambient light levels.

Range: 0.00 to 2000 µR/hr; 0.00 to 20 µSv/h

User Controls:

Response Time: User-selectable from 1 to 60 seconds, auto-response rate FAST or SLOW, or Fixed FAST and SLOW

Alarms: Count rate, exposure/dose, and count alarm setpoints adjustable over the display range

Sigma Audio Alarm: Instrument takes a baseline reading at startup. If subsequent readings exceed two sigma above baseline, a steady pulse pattern will be heard.

Fault Detection: Overload protection, Loss of Count protection, Over-Range, Calibration Expiration Lockout

Dead Time Correction: Employs first and second order corrections for extended performance

Operating Voltage: 700 to 1100 Vdc

Audio: Greater than 75 dB at 0.6 m (2 ft), approximately 4 kHz

Click Audio Divide: x1, x10, x100, x1k

Power: Two alkaline or two rechargeable "AAA" batteries

Battery Life: Approximately 100 hours of operation, 24 hour low battery warning

Construction: Display: High-impact plastic with water-resistant rubber seals; Detector Housing and Pole: Aluminum with powder-coat finish

Temperature Range: -20 to 50 °C (-4 to 122 °F), may be certified for operation from -40 to 65 °C (-40 to 150 °F)

Environmental Rating: NEMA rating of 3x or IP rating of 52 with audio hole seal option (hole seal option will reduce maximum volume by approximately 10 dB)

Overall Length: Adjustable from 111 to 161 cm (43.6 to 63.4 in.), with 11 different length settings

Overall Weight: 2.9 kg (6.5 lb)

Contact Information

Ludlum Measurements, Inc.

P.O. Box 810, Sweetwater, Texas 79556

Web: http://www.ludlums.com

Tel: 800-622-0828 / 325-235-5494

Fax: 325-235-4672

Email: sales@ludlums.com

Note: Specifications subject to change without notification. Ludlum Measurements, Inc. is not responsible for errors or omissions.

Document Date: Sep 2025

PDF preview unavailable. Download the PDF instead.

M30-6 Adobe PDF Library 17.0 Adobe InDesign 20.5 (Windows)

Related Documents

Preview Ludlum Measurements Model 43-37 Series Gas Proportional Detector Specifications
Detailed specifications for the Ludlum Measurements Model 43-37 and 43-37-1 series gas proportional detectors, designed for alpha and beta survey applications.
Preview Ludlum Model 4525 Generation IV Series Radiation Portal Monitor Installation Manual
This installation manual provides comprehensive guidance for setting up Ludlum Model 4525 Generation IV Series Radiation Portal Monitors, covering site preparation, hardware installation, and software configuration for effective radiation detection in vehicle traffic.
Preview Ludlum Model 4525 Generation 5 Installation Manual
Comprehensive installation guide for Ludlum Measurements, Inc. Generation 5 radiation portal monitor systems, including models 4525-7000, 4525-10500, and 4525-14000. Covers packaging, site preparation, anchoring, leveling, wiring, and diagrams.
Preview Ludlum Model 9-7 Ion Chamber Display Unit: Features, Specifications, and Detectors
Detailed information on the Ludlum Model 9-7 Ion Chamber Display Unit, including its features, technical specifications, compatible detectors, and available options for radiation measurement.
Preview M3xxx Series Firmware Update Instructions for Ludlum Measurements Instruments
This document provides step-by-step instructions for updating the firmware on Ludlum Measurements M3xxx Series instruments, including models 3003, 30031, 3014, 3078, and 3078i, using the Lumic Firmware Updater software.
Preview Ludlum Model 42-5 Neutron Ball Cart Manual and Specifications
Comprehensive manual detailing the specifications, operating procedures, calibration methods, and parts list for the Ludlum Model 42-5 Neutron Ball Cart, a radiation detection instrument.
Preview Ludlum Model 297 Signal Splitter User Manual
User manual for the Ludlum Model 297 Signal Splitter, detailing its introduction, getting started guide, specifications, recycling information, and diagrams.