Ludlum Model 9-7 Ion Chamber Display Unit

Part Number: 48-3689

Introduction

The Ludlum Model 9-7 Ion Chamber Display Unit is a digital readout and control unit that can be connected to any compatible ion chamber (sold separately) to provide remote monitoring in high-range beta and gamma fields. This instrument is a direct replacement for the discontinued Eberline Model RO-7, and will work with any of the existing RO-7 series detectors, extensions, and cables.

The gamma measurement range is 0.001 R/hr to 19.99 kR/hr, covered by four separate ion chamber detectors. These ion chambers can be connected either directly to the instrument or via rigid extensions and/or flexible cables to achieve the length required. The detectors will remain functional with extensions/cables up to 152 m (500 ft) in length. The Model 9-7 automatically recognizes which detector is connected so no recalibration or adjustment is required when switching detectors.

The instrument display and controls are very similar to the original RO-7 to minimize any procedural changes or re-training. The backlit LCD displays three easily-readable digits with a floating decimal point to indicate the measurement along the R/hr scale. When connected to the highest range detector, a separate kR/hr indicator is prominently illuminated to warn the operator of the higher scale in use. Controls include a rotary switch for selecting instrument on/off and battery status, an LCD backlight switch (lamp on/off), and a zero adjust knob.

Also Available: Model 9-7-1 (PN 48-3756), a version that displays measurements in Sv/h units.

Note: The Model 9-7 is not RoHS compliant.

Features

Specifications

COMPATIBLE DETECTORS: Sold separately. Accommodates Ludlum Model 9-7 (see below) and Eberline RO-7 detectors.

DISPLAY: 3½-digit LCD with 1.3 cm (0.5 in.) digits, floating decimal point, and negative sign; backlit illumination controlled by switch.

INDICATORS: Battery OK (green lamp ?), kR/h range active (red lamp ?).

RESPONSE TIME: Approximately 2.5 seconds to 90% of final reading.

CONTROLS: OFF, BAT TEST, ON rotary power switch; LAMP ON-OFF toggle switch (for display backlight); ZERO ADJUST knob.

POWER: Two "D" cell batteries, approximately 160 hour battery life.

TEMPERATURE RANGE: -20 to 50 °C (-4 to 122 °F).

HUMIDITY: 0 to 90%, non-condensing.

CONSTRUCTION: Aluminum housing with beige powder coat, includes carrying handle and folding tripod stand.

SIZE (H x W x L): 24.1 x 9.7 x 23.6 cm (9.5 x 3.8 x 9.3 in.), excluding detector.

WEIGHT: 1.6 kg (3.5 lb), including batteries.

Detectors (available separately)

ModelPart NumberRangeResolution
9-7-BLD47-3693-10.001 to 1.999 R/hr0.001 R/hr (1.0 mR/hr)
9-7-BHL-1*47-3882-10.01 to 19.99 R/hr0.01 R/hr (10 mR/hr)
9-7-BM47-36940.1 to 199.9 R/hr0.1 R/hr (100 mR/hr)
9-7-BH47-36960.01 to 19.99 kR/hr0.01 kR/hr (10 R/hr)

* Note: The Model 9-7 requires modification to use the 9-7-BHL-1 detector.

Options (other cable lengths are available)

Visual Descriptions

Image 1: A Ludlum Model 9-7 Ion Chamber Display Unit is shown. It is a beige, rectangular device with a carrying handle and a folding tripod stand. A detector is connected via a cable. The front panel features a rotary control knob, a toggle switch, and a backlit LCD display. The unit is labeled "LUDLUM MEASUREMENTS INC. MODEL 9-7 ION CHAMBER".

Image 2: Two smaller images are displayed. The left image shows the control panel with a rotary switch labeled "OFF", "BAT TEST", "ON", and a toggle switch labeled "LAMP ON-OFF", along with a "ZERO ADJUST" knob. The right image shows the digital display reading "01.3" with indicators for "kR/hr" and "R/hr".

Image 3: A Ludlum Model 9-7 Ion Chamber Display Unit is shown connected to a detector. The unit is beige, with a handle and tripod stand. This image is captioned "Model 9-7 with Detector (available separately)".

Image 4: A close-up of a "Model 9-7-BM Mid-Range Ion Chamber" detector is shown. It is a cylindrical device with a silver-colored metal housing and a black tip, connected to a cable.

Image 5: An "Underwater Housing for Model 9-7 Detector" is depicted. It is a silver-colored cylindrical unit with a cable attached, standing upright on a black base.

PDF preview unavailable. Download the PDF instead.

M9-7 Adobe PDF Library 17.0 Adobe InDesign 20.4 (Windows)

Related Documents

Preview Ludlum Model 30-6 Extended Reach microR Survey Meter Datasheet
Detailed specifications and features for the Ludlum Model 30-6 Extended Reach microR Survey Meter, a rugged, low-weight gamma detector system designed for radiation measurement.
Preview Ludlum Measurements Model 43-37 Series Gas Proportional Detector Specifications
Detailed specifications for the Ludlum Measurements Model 43-37 and 43-37-1 series gas proportional detectors, designed for alpha and beta survey applications.
Preview Ludlum Model 4525 Generation IV Series Radiation Portal Monitor Installation Manual
This installation manual provides comprehensive guidance for setting up Ludlum Model 4525 Generation IV Series Radiation Portal Monitors, covering site preparation, hardware installation, and software configuration for effective radiation detection in vehicle traffic.
Preview Ludlum Model 42-5 Neutron Ball Cart Manual and Specifications
Comprehensive manual detailing the specifications, operating procedures, calibration methods, and parts list for the Ludlum Model 42-5 Neutron Ball Cart, a radiation detection instrument.
Preview M3xxx Series Firmware Update Instructions for Ludlum Measurements Instruments
This document provides step-by-step instructions for updating the firmware on Ludlum Measurements M3xxx Series instruments, including models 3003, 30031, 3014, 3078, and 3078i, using the Lumic Firmware Updater software.
Preview Ludlum Model 4525 Generation 5 Installation Manual
Comprehensive installation guide for Ludlum Measurements, Inc. Generation 5 radiation portal monitor systems, including models 4525-7000, 4525-10500, and 4525-14000. Covers packaging, site preparation, anchoring, leveling, wiring, and diagrams.
Preview Ludlum Model 297 Signal Splitter User Manual
User manual for the Ludlum Model 297 Signal Splitter, detailing its introduction, getting started guide, specifications, recycling information, and diagrams.