Bolus From Your Smartphone
The instructions below are provided as a reference tool for patients or caregivers who are already familiar with the use of an insulin pump and with insulin therapy in general and have already installed the t:connect mobile app on their smartphone and paired it with their t:slim X2 insulin pump. Not all screens are shown and some screens may vary depending on your smartphone model. For more detailed information on the operation of the t:connect mobile app, please refer to the appropriate section in the User Guide.
Step 1: Navigate to Bolus
From the Navigation bar, tap Bolus.
Visual Description: A smartphone screen showing the t:connect mobile app dashboard with a navigation bar at the bottom. The 'Bolus' icon in the navigation bar is highlighted.
Step 2: Enter Carb Information
Tap 0 grams to enter the carbs for the user's bolus. Note: If this area reads "units," the carb feature is turned off in the active Personal Profile.
Visual Description: A smartphone screen showing the bolus input screen. The 'Carbs' field is displayed as '0 grams' and is ready for input.
Step 3: Enter Glucose Information
Tap + mg/dL, use the keypad to enter the user's blood glucose (BG) value, and then tap Done. Note: If certain advanced features are being used, the current Dexcom G6 continuous glucose monitoring (CGM) reading may auto-populate.
Visual Description: A smartphone screen showing the bolus input screen. The 'Glucose' field is displayed as '53 mg/dL' and is ready for input. A numeric keypad is shown below for data entry.
Step 4: Calculate and Adjust Bolus
Use the numeric keypad to enter desired value. Tap Done to close keypad.
If needed, tap the calculated units to manually adjust the recommended dose. Tap Next (iOS) or the Right Arrow (Android) to continue. Note: If a bolus is calculated that is larger than your Max Bolus setting, the bolus will be reduced.
Visual Description: A smartphone screen showing the bolus calculation screen. It displays the calculated 'Correction Bolus' and 'Food Bolus' in units. The 'Next' button is visible.
Step 5: Add Correction Bolus (if applicable)
If a BG is entered that is above the user's target the option to add a correction bolus will appear. To accept adding a correction bolus, tap Yes; otherwise, tap No.
Visual Description: A smartphone screen asking "Add Correction Bolus?" with the current glucose level above the target. Options to select 'Yes' or 'No' are presented.
Step 6: Reduce Bolus (if applicable)
If a BG is entered that is below the target but above 70 mg/dL, the option to reduce the bolus amount will appear. To accept that reduction tap Yes; otherwise, tap No.
Visual Description: A smartphone screen showing an option to reduce the bolus amount due to a glucose reading below the target but above 70 mg/dL.
Step 7: Verify and Confirm Bolus
Verify the dose and tap Confirm (iOS) or the Checkmark (Android) to continue. To make changes or view calculations, tap Back (iOS) or the Back Arrow (Android). Note: Calculations above are based on preset insulin-to-carb ratios and correction factors, which may be set in Personal Profiles.
Visual Description: A smartphone screen displaying the final bolus details for confirmation, including the total units to be delivered. A 'Confirm' or 'Checkmark' button is visible.
Step 8: Deliver Bolus
Tap Deliver Bolus. The Bolus Confirmed screen will appear to confirm delivery has started on both the insulin pump and the t:connect mobile app. Note: Bolus delivery will not start without an active Bluetooth® wireless connection. Note: Use your smartphone's security feature to confirm the bolus request.
Visual Description: A smartphone screen indicating "Bolus Confirmed" and showing that delivery has started.
Bolus in Progress
A BOLUS IN PROGRESS bar will appear until the entire bolus has been delivered.
To cancel the undelivered portion of the bolus, tap X on the home screen of the pump or Dashboard of the t:connect mobile app. Then tap Yes to confirm cancellation.
Visual Description: A smartphone screen showing a "BOLUS IN PROGRESS" bar, indicating that insulin delivery is ongoing.
Important Notes and Compatibility
- Note: If the app loses Bluetooth connection you can check status directly on the pump, and the bolus can be cancelled, if needed, directly from the pump.
- Warning: DO NOT ignore symptoms of high or low glucose. If the readings on your t:connect mobile app do not match your symptoms, check your pump display and confirm that it has an established Bluetooth connection with your smartphone.
- Compatible Devices: A compatible smartphone is required to initiate a bolus on the t:slim X2 insulin pump with the t:connect mobile app. Please visit tandemdiabetes.com/mobilesupport or the App Guide for information on compatible devices and operating systems.
- Recommended Smartphone Setup: See the Getting Started and Troubleshooting tips in the App Guide (Settings > Help > App Guide) for the use and setup of your smartphone. Important topics also covered include managing operating system updates and supported devices.
- Bolus Feature Unavailable: You must rely on your pump for all therapy decisions if your smartphone is not compatible, becomes lost or damaged, or loses Bluetooth connectivity with your pump for any reason.
Important Safety Information
RX ONLY. The t:slim X2 insulin pump with interoperable technology is an alternate controller enabled (ACE) pump that is intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in people requiring insulin. The pump is able to reliably and securely communicate with compatible, digitally connected devices, including automated insulin dosing software, to receive, execute, and confirm commands from these devices. The pump is indicated for use in individuals six years of age and greater. The pump is intended for single patient, home use and requires a prescription. The pump is indicated for use with NovoLog or Humalog U-100 insulin. Users of the pump must: be willing and able to use the insulin pump and all other system components in accordance with their respective instructions for use; test blood glucose levels as recommended by their healthcare provider; demonstrate adequate carb-counting skills; maintain sufficient diabetes self-care skills; see healthcare provider(s) regularly; and have adequate vision and/or hearing to recognize all functions of the pump, including alerts. The t:slim X2 pump must be removed before MRI, CT, or diathermy treatment. Visit tandemdiabetes.com/safetyinfo for additional important safety information.
The feature set available within the t:connect mobile app is dependent on the pump software version and the compatible smartphone's model and operating system (OS).
The Display and Data Upload feature set provides a secondary display of pump and continuous glucose monitoring (CGM) information, including display of your pump alerts and alarms, and enables wireless uploading of pump and CGM data to the Tandem cloud through an internet or wireless data connection. Standard carrier data rates may apply.
The Bolus Delivery plus Display and Data Upload feature set additionally allows users to request, stop, and cancel a bolus on the pump from the t:connect mobile app. At least one smartphone security setting must be enabled to utilize the Bolus Delivery feature of the t:connect mobile app.
WARNING: Always rely on your pump to make therapy decisions when using a smartphone that is incompatible with the Bolus Delivery feature.
PRECAUTIONS: Always rely on your pump to make therapy decisions if using a smartphone that is incompatible, the smartphone is lost or damaged, or the smartphone loses Bluetooth® Connectivity with your pump. Important pump alerts and alarms are only received as app notifications when enabled and the app is either active or running in the background.
© 2022 Tandem Diabetes Care, Inc. All rights reserved. Tandem Diabetes Care, t:connect, and t:slim X2 are either registered trademarks or trademarks of Tandem Diabetes Care, Inc. in the United States and/or other countries. Dexcom and Dexcom G6 are registered trademarks of Dexcom, Inc. in the United States and other countries. The Bluetooth wordmark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Tandem Diabetes Care is under license. Apple and Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc. Google Play and the Google Play logo are trademarks of Google LLC. All third-party marks are the property of their respective owners. ML-1008363_B
Tandem Diabetes Care
Diabetes Care
(877) 801-6901
tandemdiabetes.com
t:connect Mobile App
An advanced way to monitor your diabetes
Download on the App Store [App Store Link]