Drive Full Electric Low Height Bed

Model # 15005L

Owner's Manual and Assembly Instructions

This manual must be given to the Owner/User of this bed and should be read carefully before putting this product into use. Keep manual in an accessible location for future reference.

NEW BREAK RESISTANT PANELS

Click here to view detailed product information and pricing. Call 800-397-5899 to order today!

TABLE OF CONTENTS

NOTE: DO NOT operate this product without first reading and understanding this user manual. Damage or injury may result from improper use of this product.

Packaging / Handling

Unpacking

  1. Check cartons for any obvious damage to packaging or contents. If damage is evident, contact your local Drive Medical dealer/provider.
  2. Remove any loose packaging from the cartons.
  3. Carefully remove all components from the cartons.

The Two Cartons will contain the following components:

Inspection

  1. Examine each component carefully for damage, scratches, dents or any other visual damage.
  2. Carefully remove Tie Wraps holding power cord, hand pendant and drive shaft.
  3. Inspect power cord and plug for cuts or damage.
  4. Inspect the junction box and motors for damage, loose connections, cuts or damage to the wires.
  5. Ensure motor connections to junction box are connected correctly and secure.

Storage

  1. If the bed is not assembled immediately, retain cartons and packaging for storage until assembly is completed.
  2. Store the packaged bed in a dry area.
  3. DO NOT place other objects on top of the packaged bed cartons.

Assembly Instructions

1. Confirm that all parts are present:

A visual representation shows the bed frame components labeled A through I, including the head and foot springs, head and foot boards, drive shaft, locking pins, hand crank, casters, and transport straps.

Caution: DO NOT attempt to operate bed controls until completion of assembly as damage to bed components and/or personal injury may occur.

Assembly Instructions (Continued)

2. With the Head and Foot Spring positioned upright, connect Head Spring by aligning with foot spring and swinging both ends away from each other.

A visual representation shows the head and foot springs being connected.

3. Optional: Insert lock pin through connected sleep surface and secure with spring clip.

A visual representation shows a lock pin being inserted.

4. Connect Head Spring and Foot Spring Pull Tube to provide the up/down motion/adjustment of the Head and Foot Spring Sections:

Visual representations show the connection of the pull tubes to the lift arms and securing with pins and clips.

Assembly Instructions Cont.

5. Install the casters by inserting the stem of the caster into the leg receptacle. It is recommended to install one locking caster and one non-locking caster on each bed end. The locking casters should be installed diagonally from each other. The 15005L Low Height bed gives you the option of installing the casters on the inboard non-adjustable legs or the outboard adjustable legs. To achieve the full low height feature, install the casters on the inboard non-adjustable legs.

A visual representation shows a caster being inserted into a leg receptacle, with labels for 'Leg Receptacle' and 'Stem'. Two images show the head and foot boards with casters attached.

6. Carefully lower the bed frame flat onto the floor. Attach the Head and Foot Boards by tilting the Boards toward your body and aligning mounting pins located on the Spring Sections with the Hooks on the Head/Foot Boards.

NOTE: The Head Board is taller than the Foot Board, however, they are interchangeable with each other.

A visual representation shows a person attaching the head board to the bed frame.

7. Assemble Hi/Lo Drive Shaft and adjust to the proper length so it will fit between the Head Board drive pin and Hi/Lo Motor drive pin at the foot end.

Visual representations show the drive shaft being connected to the motor drive pin and the head board drive pin.

Assembly Instructions Cont.

8. Release the spring loaded shaft on the hi/low motor and insert the shaft into the gear box on the footboard.

A visual representation shows the shaft being inserted into the gearbox.

9. Make sure the slot on the spring loaded shaft engages the roll pin on the gear box.

A visual representation shows the shaft engaged with the roll pin.

10. Raise the Head Spring and Foot Spring to release tension between the springs and connect Head and Foot spring fabric by squeezing together and attaching with spring links.

A visual representation shows hands connecting the spring fabric with spring links.

Operation

Hand Control

The Model #15005L Low Height Bed comes with a 3 Function hand held Pendant.

A visual representation shows the 3-function hand pendant with labeled buttons.

Caster Lock and Floor Lock Feature

When the bed is in its lowest height position, the Casters will come in contact with the floor and the bed will roll freely. Engaging the Caster Locks will help hold the bed in place.

When the bed height is raised, the legs will come in contact with the floor and the casters will be raised off the floor providing a stable base for the bed frame.

A visual representation shows the bed in its lowest position with casters on the floor, and in a raised position with the legs on the floor.

Bed Rail Installation

The Deck Frame contains labels that provide the location for installing Half Rails and Full Rails.

Visual representations show labels indicating where to install half and full rails, and images of half rails and full rails being attached to the bed frame.

Mattress Retainer Adjustment

Before a mattress is laid on the bed, flip up the mattress guard to reduce the movement of the mattress.

A visual representation shows the mattress guard being flipped up.

Manual Hi/Lo Adjustment Crank

In the event of a loss of power or Hi/Lo motor malfunction; the Hi/Lo feature can be manually adjusted using the Hand Crank. The Hand Crank can be operated in two locations: directly into the Head or Foot board gear box or from the top of the Head or Foot Board access port.

WARNING: Manual Hi/Lo Adjustment Crank MUST BE REMOVED before the bed is used. Failure to remove the crank may cause damage to the bed and/or personal injury to those in the room.

A visual representation shows the manual hi/lo adjustment crank being inserted into the bed frame.

Maintenance & Safety Checks

Drive recommends the following maintenance and cleaning procedures be conducted between users.

ELECTRONICS

BED FRAME & SLEEPING SURFACE

CLEANING

NOTE: This bed is NOT compatible with wash down tunnels.

LUBRICATION & MECHANICAL

WARRANTY

During the warranty period, defective items will be repaired or replaced at Drive's option at no charge.

WARNINGS

GENERAL WARNINGS

DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner's manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.

Read all the instructions before using the bed.

Refer to the owner's manuals for beds and rails for additional product and safety information.

After any adjustments, repair or service and before use, make sure all attaching hardware is tightened securely.

After the bed has been assembled, always test to make sure that all sections of the bed are properly and securely in place before using.

ALWAYS use caster locks except when moving the bed.

Body weight should be evenly distributed over the surface of the bed. DO NOT lay, sit or lean in such a way that your entire body weight is placed only on raised head or foot sections of the bed. This includes when assisting the user in repositioning or transferring in or out of bed.

Check all parts for shipping damage and test before using. In case of damage, DO NOT use. Contact a qualified technician for further instruction.

Close supervision is necessary when this bed is used by or near children or people with disabilities.

DO NOT let any individual underneath the bed or in between the raised bed frame components at anytime.

DO NOT permit more than one person on/in the bed at any time.

DO NOT use outdoors.

DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner's manuals, service manuals or instruction sheets supplied with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may occur.

For Dealers Only - Set-up and Assembly Instructions are in the front of this manual. These procedures must be performed by a qualified technicians only.

WARNINGS

The March 2006 version of the FDA's bed safety guidelines are published by Hospital Bed Safety Workgroup. The latest revision of this document is available at http://www.fda.gov.

If the unit is not working properly, call a qualified technician to examine the unit and repair it.

Drive Medical products are specifically designed and manufactured for use in conjunction with Drive Medical accessories. Accessories designed by other manufacturers have not been tested by Drive Medical and are not recommended for use with Drive Medical products.

Keep all moving parts, including the main frame, mattress deck (head and foot springs/sections) and all drive shafts free of obstruction (i.e. blankets/sheets, heating blankets/pads, tubing, wiring, etc. and other types of products using electric cords which may get tangled around the bed, side rails or legs) during operation of the bed.

Keep the product a minimum of 12 inches away from any direct heat source.

Make sure head and foot springs/sections are connected securely to the universal bed ends before use.

Physically challenged individuals who cannot prevent themselves from rolling/climbing out of this bed may require alternative safe means of restraint.

Procedures other than those described in this manual must be performed by a qualified technician.

Risk of injury to persons - DO NOT place video equipment such as televisions or computer monitors on bed.

NEVER allow patients to use trapeze or traction units as a total individual weight support.

Trapeze units are to be used only for assisting the patient in repositioning or transferring into or out of bed.

Trapeze units must be positioned on a universal bed end as near as possible to the center point of the bed end.

The bed is not designed to be used as a patient transport. When transporting a patient, use an approved patient transport. Otherwise, injury or damage may result.

The bed may be equipped with locking casters. When transferring into or out of the bed, always lock the locking caster(s). Inspect for correct locking action on caster locks before using bed. Even with casters properly locked, some flooring surfaces such as tile or wood, will allow the bed to move under some conditions. Use on surfaces such as these must be evaluated by the care provider.

The initial set up of this bed must be performed by a qualified technician.

WARNINGS

The total weight limit of the Drive Medical 36 inch (91.4cm) wide Manual/Semi-Electric/Full Electric bed is 450 pounds (204 kg) which includes the weight of the occupant, all accessories such as mattress, side/assist rails, trapeze, any personal effects, etc.

Example: If total weight of mattress and all accessories is 100 lbs, then total occupant weight can be no more than 350 lbs.

Use this bed only for its intended use as described in these instructions. DO NOT use attachments not recommended by the manufacturer.

After raising/lowering the head/foot end of the bed, check the distance between the bottom of the bed rail and the mattress. If there is excessive distance between the bottom of the bed rail and the mattress in which individuals may become entangled, adjust the height of the bed rail (if applicable), or provide alternative means of patient protection.

Once patient assessment concludes that the patient's condition increases the chance of entrapment, the bed MUST be in the flat position when left unattended.

Proper patient assessment and monitoring, and proper maintenance and use of equipment is required to reduce the risk of entrapment. Variations in bed rail dimensions, and mattress thickness, size or density could increase the risk of entrapment. Visit the FDA website at http://www.fda.gov to learn about the risks of entrapment.

Replacement mattresses and bed side rails with dimensions different than the original equipment supplied or specified by the bed frame manufacturer are not interchangeable. Variations in bed side rail design, width and thickness or firmness of the mattress could cause/contribute to entrapment. Use only authorized Drive Medical replacement parts and/or accessories otherwise the warranty is void. Drive Medical will not be responsible for any damage or injury that may result.

To reduce the risk of entrapment make certain that the bed rail cross braces DO NOT exceed the width of the mattress. Mattress MUST fit bed frame and side rails snugly to reduce the risk of entrapment.

When operating this bed, ALWAYS ensure that the individual utilizing the bed is positioned properly within the confines of the bed. DO NOT let any extremities protrude over the side or between the bed rails when performing these functions.

Drive Medical recommends that the mattress be centered on the bed frame. Otherwise, individuals may become entangled between the bed rail and the bed frame.

Mattress MUST fit bed frame and assist rails snugly to reduce the risk of entrapment.

SAVE THESE INSTRUCTIONS

NOTE: Updated versions of this manual are available on www.DriveMedical.com.

WARNINGS

BED RAIL WARNINGS

Although bed rails are not rated to any specific weight limitation, the bed rails may become deformed or broken if excessive side pressure is exerted on the bed rails. The bed rail is not an assist rail for getting into or out of bed. DO NOT use the bed rails as push handles when moving the bed.

Always test to make sure that the side rails are properly and securely in place before using the bed.

DO NOT install the optional bed rails without reading and understanding all of the instructions in the instruction sheet that accompanies the bed rail kit.

DO NOT use the side rails as push handles for moving the bed.

These bed rails are intended to prevent an individual from inadvertently rolling out of bed. DO NOT use for restraint purposes.

When used with a manual/electrical bed, the bed rails DO NOT fall under any weight limitations. Bed rails can be deformed or broken if excessive side pressure is exerted on the bed rails. These bed rails are used for the purpose of preventing an individual from inadvertently rolling out of bed. The bed rails are NOT intended nor may be used for restraint purposes. If an individual is capable of injuring himself/herself, a physician or a healthcare professional should be consulted for alternative means of safe restraint.

ELECTRICAL WARNINGS

A safety feature of this product includes protection against overheating caused by excessive or extended periods of operation. Depending on the duration, this includes multiple or repeated adjustments or the use of multiple functions at one time. To ensure trouble free operation, always allow a slight pause between multiple adjustments and avoid pressing more than one function button at a time. If thermal protection activation should occur, the bed will not respond to pendant commands. Given this situation, release the pendant button and allow the bed unit to sit for several minutes. This will allow the protection function time to reset and restore bed function. Depending on severity of the initial overheating, this could take up to 30 minutes.

Allow a slight pause between adjustments and avoid pressing multiple buttons at the same time. If pendant buttons are depressed too rapidly or multiple buttons are pressed at the same time, the desired feature may not activate. Simply release the pendant button, permit a slight pause and then activate the next operation.

ALWAYS remove manual crank(s) before performing electronic functions. Otherwise, the crank(s) will turn when the motor is on and could cause personal injury or damage to the bed.

Always unplug the bed from the electrical outlet before cleaning.

DO NOT open assemblies such as the motors, pendant, junction boxes or gear boxes. No user serviceable parts are inside. Only qualified technicians are permitted to repair these parts. If unqualified individuals perform any work on these beds, the warranty is void.

WARNINGS

DO NOT place pendant under or between objects. This may unintentionally press the buttons and may cause injury or damage.

Electronic equipment may be influenced by Radio Frequency Interference (RFI). Caution should be exercised with regard to the use of portable communications equipment in the area around such equipment. If RFI causes erratic behavior, unplug the electric bed IMMEDIATELY. Leave unplugged while transmission is in progress.

Ensure all cables and cords are routed such that they will not become entangled or pinched. Otherwise damage or injury may result.

If a button on the pendant sticks or does not release and/or one of the functions will not stop moving. Unplug the power cord from the electrical supply outlet and take the bed out of service for repair.

If a liquid is spilled in or around the electric bed, unplug the electric bed before cleaning. Clean up the spill and allow the electric bed or the area around the electric bed to dry thoroughly before using the electric controls again.

Inspect the covering of the bed's control panel and the patient control panel to assure that the covering is not cracked or damaged.

Keep all electrical cords away from heated or hot surfaces.

Keep the cord away from heated surfaces.

NEVER operate if the unit has a damaged cord or plug. If it is not working properly, call a qualified technician for examination and repair.

NEVER operate this bed if it has a damaged cord or plug, if it is not working properly, if it has been dropped, damaged, or dropped into water. Return the bed to a service center for examination and repair.

On Full Electric beds, the Hi/Lo crank MUST be removed before the bed is used. Failure to remove the crank may cause damage or personal injury.

Refer servicing to qualified personnel only. Grounding reliability depends upon a properly grounded wall outlet.

The pendant and power cords must be routed and secured properly to ensure that the cords DO NOT become entangled, pinched and/or severed during operation of the electric bed.

Unplug from the outlet before servicing.

Unplug the power cord from its power source before performing any maintenance on the manual/electric bed. DO NOT unplug the power cord from the junction box. Damage to cord will result.

Use caution when disconnecting the pendant. DO NOT press pendant buttons.

WARNINGS

When bed is not to be used for an extended period, unplug electric bed from the wall outlet.

When using an extension cord, use only a three wire extension cord having at least 16 AWG (American Wire Gauge) wire and the same or higher electrical rating as the device being connected. Use of improper extension cord could result in a risk of fire and electric shock. Three to two prong adapters should not be used. Use of three to two prong adapters can result in improper grounding and present a shock hazard to the user.

When using nasal or masked type oxygen administering equipment, the oxygen or air tubing MUST be routed and secured properly to ensure that the tubing does not become entangled and/or severed during normal operation of the bed.

DO NOT use near explosive gases.

Possible fire hazard when used with oxygen administering equipment other than nasal or masked type.

Use masked or nasal type oxygen administering equipment only in conjunction with this bed. The use of ANY other type of oxygen administering equipment can result in a fire hazard.

Notes:

PDF preview unavailable. Download the PDF instead.

owners-manual-drive-medical-15005l-full-electric-low-height-bed Adobe PDF Library 10.0.1

Related Documents

Preview Delta 1000 Ultra Light Plus Semi Electric Bed: Assembly and Operating Manual
Comprehensive assembly and operating manual for the Drive Medical Delta 1000 Ultra Light Plus Semi Electric Bed (Item # 15030). Covers parts identification, step-by-step assembly for two methods, bed rail installation, height adjustment, maintenance, safety warnings, and technical specifications.
Preview Drive Full-Electric Lightweight Bariatric Bed 15300LW Owner's Manual
This comprehensive owner's assembly and operating manual provides detailed instructions for the Drive Full-Electric Lightweight Bariatric Bed, model 15300LW. It covers safety precautions, assembly, operation, maintenance, and technical specifications for this medical bed.
Preview Drive Medical Delta 1000 Ultra Light Plus, Full Electric Bed Owner's Assembly and Operating Manual
Assembly and operating instructions for the Drive Medical Delta 1000 Ultra Light Plus, Full Electric Bed. Includes parts list, assembly steps, motor installation, bed rail installation, optional height adjustment, storage, maintenance, safety checks, specifications, and warnings.
Preview Drive Delta 1000 Ultra Light Plus Full Electric Bed Owner's Assembly and Operating Manual
Assembly, operation, maintenance, and safety instructions for the Drive Delta 1000 Ultra Light Plus Full Electric Bed (Item # 15033). Includes detailed steps, warnings, and specifications.
Preview Drive Medical Hospital Beds 15005LP, 15004P, 15003P: Owner's Assembly and Operating Manual
This manual provides comprehensive assembly, operating, and safety instructions for Drive Medical's 15005LP Full Electric, 15004P Semi-Electric, and 15003P Manual Hospital Beds. It includes detailed technical specifications, maintenance procedures, troubleshooting guidance, and important safety warnings.
Preview Drive Medical Delta 1000 Ultra Light Plus Semi Electric Bed Assembly and Operating Manual
Comprehensive assembly and operating instructions for the Drive Medical Delta 1000 Ultra Light Plus Semi Electric Bed (Item # 15030), covering parts identification, assembly steps, maintenance, safety precautions, specifications, and warnings.
Preview Drive Delta 1000 Ultra Light Plus Full Electric Bed Assembly and Operating Manual
This manual provides assembly and operating instructions for the Drive Delta 1000 Ultra Light Plus Full Electric Bed, item # 15033. It includes detailed steps for assembly, maintenance, safety checks, and specifications.
Preview Delta 1000 Ultra Light Plus Semi Electric Bed: Owner's Assembly and Operating Manual
Comprehensive guide for assembling, operating, and maintaining the Drive Medical Delta 1000 Ultra Light Plus Semi Electric Bed (Item # 15030), including safety warnings and specifications.