ANENG 621A Intelligent Digital Multimeter User Manual

Touch screen intelligent digital multimeter

1. Safety Information

Warning: To fully understand the function of the meter and ensure safe operation, please read and follow this manual carefully. Special care should be taken when using this meter; improper use may cause electric shock or damage to the meter. Follow usual safety regulations and take effective safety precautions.

The instrument meets the safety requirements of IEC-61010-1, IEC-61010-2-030, IEC-61010-2-032) electronic measuring instruments, belongs to secondary pollution, and the overvoltage standard is CAT III 600V. Please follow the safety operation guide and use the meter safely.

1.1 Safety Instructions

1.2 Electrical Symbols

1.3 Maintenance

2. Description

2.1 Part Name

1 Non-contact voltage detection sensing area

2 Type-c charging cable port

3 LCD touch button

4 Display

5 Input socket

6 The power button

(Diagram of the multimeter with numbered parts is described here)

2.2 LCD Touch Display

The LCD displays various symbols and readings:

Symbol/Reading Description
AC & DC Indicates AC or DC measurement.
)) Connection/Disconnection indication
AUTO Automatic range mode
Automatic shutdown indication
? Low battery
H Reading hold state
V, A Volt (voltage), ampere (current)
Ω, kΩ, MΩ Ohm kilohm and megohm (resistance)
Hz Hertz
DIODE
nF, uF, mF Microfarad
°C, °F Centigrade and Fahrenheit
NCV Non-contact voltage detection
Live The line of fire test

3. Technical Indicators

Gauges shall be specified for a period of one year, between 18°C and 28°C, with relative humidity less than 75% condition.

3.1 Comprehensive Indicators

  1. Automatic selection of measuring function and range.
  2. Full range overload protection.
  3. The maximum voltage between the measuring end and the ground is 1000V DC or 750V AC.
  4. Working height: maximum 2000m.
  5. Display: LCD.
  6. The maximum display value is 9999 digits.
  7. Polarity indication: automatic indication, '-' indicates negative polarity.
  8. Superview display: 'OL' or '-OL'.
  9. Sampling time: about 3 times/second.
  10. Unit display: with function, electric quantity unit display.
  11. Automatic shutdown time: 15 minutes.
  12. Power supply: 3.7V/600mAh lithium battery.
  13. Battery undervoltage indicator: LCD ? symbol.
  14. Temperature coefficient: less than 0.1X accuracy /°C.
  15. Operating temperature: 0° C-40 °C.
  16. Storage temperature: -10°C- 50°C.

3.2 Technical Indicators

3.2.1 Alternating current

Range Resolution Accuracy
999.9mA 1mA ± (1.0% reading + 8 digits)
10A 0.01A

Minimum input current 20mA AC current. Maximum input current 10A AC current. Frequency range: 40HZ-1000Hz.

3.2.2 DC current

Range Resolution Accuracy
999.9mA 1mA ± (1.0% reading + 5 digits)
10A 0.01A

Minimum input current 20mA DC current. Maximum input current 10A DC current.

3.2.3 DC voltage

Range Resolution Accuracy
999.9mV 0.1mV ± (0.5% reading + 3 digits)
9.999V 0.001V
99.99V 0.01V
999.9V 0.1V

In AUTO mode, the input voltage ranges from 0.5V to 1000V DC. The input voltage ranges from 0.0mV to 999.9mV DC in mV mode.

Note: In the mV range, with the input open, the meter may have a pulsating reading. This is normal because of the high sensitivity of the meter when connected to the pen. When you measure the circuit, you get a true measurement.

3.2.4 AC voltage

Range Resolution Accuracy
999.9mV 0.1mV ±(0.8% reading + 5 digits)
9.999V 0.001V
99.99V 0.01V
999.9V 0.1V

In AUTO mode, the input voltage ranges from 0.5V to 1000V AC. In mV mode, the input voltage ranges from 0.0mV to 999.9mV AC. Frequency rate pattern: 40HZ-1000Hz.

3.2.5 Line connection test

Range Resolution Explain
•)) 0.1 Ω If resistance of the line being measured is less than 50 Ω, the buzzer in the instrument will make continuous alarming sounds.

Overload protection: 600V DC or AC (valid value).

3.2.6 Resistance

Range Resolution Accuracy
999.9Ω 0.1 Ω ± (0.8% reading + 3 digits)
9.999kΩ 0.001 kΩ
99.99kΩ 0.01 kΩ
999.9kΩ 0.1 kΩ
9.999MΩ 0.001MΩ
99.99MΩ 0.01 MΩ

Input protection: maximum 600V DC or AC valid value.

3.2.7 Diode

Range Resolution Accuracy
Diode test ► 1 V 0.001 V Testing current: about 1mA; Open circuit voltage: about 2.8V. The display shows an approximation of the diode forward pressure drop.

Input protection: the maximum value is 600V DC or AC.

3.2.8 Capacitance

Range Resolution Accuracy
9.999nF 0.001nF ± (10%reading+ 40 digits)
999.9nF 0.1nF ±(2.5%reading+20 digits)
9.999uF 0.001 u F
99.99uF 0.01 u F
999.9uF 0.1 u F
9.999mF 1 u F
99.99mF 0.01mF

Input protection: the maximum value is 600V DC or AC.

3.2.9 The temperature

Range Resolution Accuracy
°C 1°C -40°C~ 1000°C ± (1% reading +3°C)
°F 1°F - 40°F ~ 1832°F ± (1% reading +3°F)

Accuracy does not include error of thermocouple probe. Overload protection 250V DC or AC (valid value).

3.2.10 Frequency

3.2.10.1 Pass grade A:
Range Resolution Accuracy
100.0 Hz 0.1Hz ± (1.0%reading+ 5digits)
1000Hz 1Hz

Measuring range: 40HZ-1000Hz. Input signal pattern: > 3A AC current (Effective Value).

3.2.10.2 Through V gear:
Range Resolution Accuracy
100.0 Hz 0.1Hz ± (1.0%reading+ 5digits)
1000Hz 1Hz

Measuring range: 40HZ-1000Hz. Input signal pattern: > 0.8V AC current (Effective value).

4. Operating Instructions

4.1 Reading Hold

In the process of measurement, if you need to hold the reading, touch the "H" key; the display value will be locked. Touch the "H" key again to remove the reading hold.

4.2 Flashlight

Touch the "?" button to turn on the lighting function of the flashlight barrel, which will be automatically turned off after about 1 minute.

4.3 Automatic Shutdown

4.4 Preparation before measurement

4.5 Non-contact Voltage Detection (NCV)

(Diagram showing NCV detection is described here)

Note:

  1. Voltage may still exist even if there is no alarm indication. Do not rely on contactless voltage detectors to see if there's a voltage on the wire. Probe operation may depend on socket design, insulation thickness, and different types of factors such as the impact.
  2. In NCV mode, the meter does not measure voltage, resistance, and current at the same time.

4.6 AC-DC Current Measurement

Note: When the meter is in diode gear, the pen is inserted into the 10A input hole and the meter will not enter current gear.

4.7 DC Voltage Measurement

Insert the black stylus into the COM jack and the red stylus into the V jack. Connect the markers together. The voltage source or load ends are measured and read on the LCD.

AUTO automatic measurement mode: When the measured signal > 0.5V, the meter will display the current measured DC voltage value. When the signal being measured is < 0.5V, the meter will default to the resistance value and display the internal resistance value of the measured signal. Touch the "mV" key to switch to the DC mV voltage level. Connect the pen to both ends of the voltage source or load for measurement, reading on the LCD.

DC mV measurement mode: when the measured signal < At 600.0mV, the meter will display the current measured DC voltage value. When the measured signal > At 999.9mV, the meter will display "OL".

(Diagram showing DC voltage measurement is described here)

4.8 AC Voltage Measurement

Insert the black stylus into the COM jack and the red stylus into the V jack. Connect the pen to both ends of the voltage source or load for measurement, reading on the LCD.

AUTO automatic measurement mode: when the measured signal > At 0.5V, the meter will display the current measured AC voltage value. When the signal being measured is < 0.5V, the instrument will default to the resistance value, display the measured signal internal resistance value. Touch the "mV" key to switch to AC mV voltage. Connect the pen to both ends of the voltage source or load for measurement, reading on the LCD.

AC mV measurement mode: when the measured signal < At 600.0mV, the meter will display the current measured AC voltage value. When the signal is measured > At 999.9mV, the meter will display "OL".

(Diagram showing AC voltage measurement is described here)

4.9 Resistance Measurement

Warning: When measuring the impedance on the line, make sure that the circuit is disconnected and the capacitor on the circuit discharges completely.

Note:

  1. When the input is open, the LCD will display the "OL" overrange status.
  2. If the resistance being measured is higher than 10MΩ, the meter may take several seconds to stabilize the reading, which is normal for high resistance readings.

(Diagram showing resistance measurement is described here)

4.10 The Line of Fire Test

4.11 Capacitance Measurement

Note:

  1. When measuring large capacitors, a fixed time is required to stabilize readings.
  2. When measuring polar capacitors, pay attention to the corresponding polarity to avoid damaging the instrument.

4.12 Temperature Measurement

Touch "SEL" key to switch TEMP file, thermocouple red plug into the end of °C/°F, black plug into the COM jack. When the reading is stable, the temperature can be read directly from the display screen.

Note: The maximum measurement temperature of type K thermocouple is 250°, instantaneous measurement up to 300°C.

4.13 Diode Test

Note: The meter displays an approximation of the forward voltage drop of the diode. If the pen is connected inversely or the pen is open, the LCD displays "OL".

(Diagram showing diode test is described here)

4.14 Line Connection Test

Warning: Danger of electric shock. When testing the circuit on and off, make sure the circuit is disconnected and the capacitor on the circuit is completely discharged.

Note: If the meter is open or the resistance of the measured line is greater than 999.9Ω, "OL" is displayed.

(Diagram showing line connection test is described here)

5. Maintenance

5.1 Type - c Line Charging

Warning: During the charging process, the meter pen should be removed from the measuring circuit to avoid the danger of electric shock.

If the "?" symbol appears, it means to charge.

5.2 Replace the Test Pen

Warning: When replacing the stylus, you must replace the same or the same grade of the table. Pen meter must be in good condition, pen grade: 1000V 10A.

Note: If the insulation layer of the pen is damaged, such as exposed wire of the wire, the pen must be replaced.

6. The Attachment

(Information about attachments like 'A pair of', 'A copy of the' is noted here)

Models: FS8233Pro, Touch Screen Intelligent Digital Multimeter, FS8233Pro Touch Screen Intelligent Digital Multimeter

File Info : application/pdf, 18 Pages, 1.20MB

PDF preview unavailable. Download the PDF instead.

ANENG621A User Manual Corel PDF Engine Version 18.0.0.448 CorelDRAW X8

Related Documents

Preview ANENG M127/M128 Multimeter User Manual
Comprehensive user manual for the ANENG M127 and M128 portable smart multimeters, detailing features, functions, and operating instructions.
Preview ANENG M107 Mini Multimeter: Features, Specifications, and Usage Guide
Discover the ANENG M107, a compact and intelligent smart digital multimeter. This guide covers its automatic measurement capabilities, AC/DC voltage and current testing, resistance measurement, NCV detection, flashlight, innovative storage, and detailed functional parameters.
Preview ANENG ST181 Mini Digital Clamp-On Multimeter Operation Manual
Operation manual for the ANENG ST181 Mini Digital Clamp-On Multimeter, covering safety, panel description, specifications, and detailed instructions for various electrical measurements including voltage, current, resistance, capacitance, frequency, and temperature.
Preview ANENG 683 Digital Multimeter User Manual and Specifications
Comprehensive user manual for the ANENG 683 digital multimeter, covering safety instructions, technical specifications, operating procedures, maintenance, and package contents. Learn how to accurately measure voltage, current, resistance, capacitance, and temperature.
Preview ANENG AN113E Digital Multimeter User Manual - Features, Specifications, and Operation
Comprehensive user manual for the ANENG AN113E digital multimeter. Learn about its features, safety information, specifications, measurement instructions for voltage, current, resistance, and more.
Preview Aneng A3008 Pro: Smart Pen Digital Multimeter with TrueRMS and NCV
Discover the Aneng A3008 Pro, a compact and intelligent pen-style digital multimeter. Featuring TrueRMS, NCV detection, and a wide range of measurement capabilities including voltage, resistance, capacitance, and frequency. Ideal for professionals and hobbyists.
Preview Mini Digital Clamp-On Multimeter Operating Instructions
This document provides operating instructions, safety information, and specifications for the Mini Digital Clamp-On Multimeter. It covers measurements for AC/DC voltage, DC/AC current, resistance, audible continuity, diode, temperature, and Non-Contact ACV detection.
Preview ANENG A3002 High Precision Pen Multimeter - Features and Specifications
Explore the ANENG A3002, a portable digital pen multimeter offering high precision, non-contact voltage detection, and a bright HD screen. Learn about its features, measurement capabilities, and after-sales support.