OHAUS i-D33P75SD Bench Scale
Quick Start Guide
1. Introduction
This quick start guide contains the installation, operation, and maintenance instructions for the i-D33P75SD bench scale. Please read it completely before installation and operation.
1.1. Safety Precautions
Definition of Signal Warnings and Symbols
Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring the safety notes may lead to personal injury, damage to the instrument, malfunctions, and false results.
- WARNING: For a hazardous situation with medium risk, possibly resulting in severe injuries or death if not avoided.
- CAUTION: For a hazardous situation with low risk, resulting in damage to the device or the property or in loss of data, or minor or medium injuries if not avoided.
- ATTENTION: For important information about the product. May lead to equipment damage if not avoided.
- NOTE: For useful information about the product.
Warning Symbols
[General hazard] [Explosion hazard] [Electrical shock hazard]
Safety Precautions
CAUTION: Read all safety warnings before installing, making connections, or servicing this equipment. Failure to comply with these warnings could result in personal injury and/or property damage. Retain all instructions for future reference.
- Before connecting power, verify that the equipment's input voltage range and plug type are compatible with the local AC mains power supply.
- Do not position the equipment such that it is difficult to reach the power connection.
- Only connect the power cord to a compatible grounded electrical outlet.
- Only use a power cord with a rating that exceeds the specifications on the equipment label.
- Make sure that the power cord does not pose a potential obstacle or tripping hazard.
- Operate the equipment only under ambient conditions specified in these instructions.
- Do not operate the equipment in hazardous or unstable environments.
- Do not allow liquids to enter the equipment.
- Use only approved accessories and peripherals.
- Disconnect the equipment from the power supply when cleaning.
- Service should only be performed by authorized personnel.
WARNING: Never work in an environment subject to explosion hazards! The housing of the instrument is not gas tight. (Explosion hazard due to spark formation, corrosion caused by the ingress of gases).
WARNING: Electrical shock hazards exist within the housing. The housing should only be opened by authorized and qualified personnel. Remove all power connections to the unit before opening.
2. INSTALLATION
2.1. External Connections
RS232 Connection
Connect the optional RS232 cable to the RS232 connector on the rear housing of the i-DT33P indicator. Please refer to table 2-1 for the definition of each pin if needed.
Description of RS232 Connector: The rear of the indicator features an RS232 connector. The connector has 9 pins.
Pin | Connection |
---|---|
1 | N/C |
2 | TXD |
3 | RXD |
4 | N/C |
5 | GND |
6 | N/C |
7 | N/C |
8 | N/C |
9 | N/C |
Rechargeable Battery Power (Option)
Allow the battery to charge for 12 hours before using the scale on battery power. The instrument can be operated during charging, and the battery is protected against overcharging. For maximum operating time, the battery should be charged only at room temperature.
3. OPERATION
3.1. Overview of Display and Controls
Description of Control Panel: The control panel features several buttons for operation: On/Zero, Print, Function, and Tare. The display shows weight, units, and status messages.
Button | Primary Function (Short Press) | Secondary Function (Long Press) | Menu Function (Short Press) |
---|---|---|---|
On/Zero | If the terminal is off, press to power on. | If the terminal is on, press to set the zero point. | Off: If the terminal is on, press to power off. |
Sends the current value to the RS232 port. | Units: Changes the weighing unit. | Mode: Changes the application mode. The i-D33P75SD scale should be used in "Weigh" mode for materials testing. | |
Function | Initiates an application mode. | Back: Moves back to the previous menu item. | Menu: Enters the user menu. |
Tare | Performs a tare operation. | Shows a tare value in application modes. | Exit: Exits the user menu. Aborts a calibration in progress. Exits when displaying totalization result or under and over value in check mode. |
Yes | Accepts the current setting on the display or selects a sub-menu or menu item. | ||
No | Advances to the next menu or menu item. Rejects the current setting on the display and advances to the next available one. |
Notes on Button Presses:
- Short Press: press less than 1 second.
- Long Press: press and hold for more than 3 seconds.
4. CALIBRATION MENU
Enter the calibration menu C.A.L to perform calibrations.
4.1. Initial Calibration
When the scale is operated for the first time, a zero and span calibration are recommended to ensure accurate weighing results.
Max Capacity (kg) | Mass (kg) | Max Capacity (lb) | Mass (lb) |
---|---|---|---|
75 kg | 75 kg | 150 lb | 150 lb |
Note: Calibration unit used must match the calibration unit selected in menu option Capacity Unit.
4.2. Zero Calibration [ZErO]
Zero calibration uses one calibration point. The zero calibration point is established with no weight on the scale. Use this calibration method to adjust for a different static load without affecting the span or linearity calibration.
Calibration procedure (when the calibration unit is kg)
- Long press the Menu button until you see MENU.
- Release the button and wait for the display to show C.A.L. Press the Yes button.
- The display shows ZErO. Press the Yes button.
- 0 kg and the calibration unit are flashing on the display. With no weight on the pan, press the Yes button to establish the zero point.
- The display shows ----, and then -DONE- when the Zero calibration is finished.
Note: If zero calibration has failed or if after 40 seconds the calibration is still not successful, CAL E is displayed for 3 seconds and the previous calibration data will be restored.
Then the display shows SPAN. Press the Exit button to exit.
4.3. Span Calibration [SpaN]
Span calibration uses one calibration point. It is established with a calibration mass placed on the scale. Span calibration should be performed after zero calibration.
Calibration procedure (when the calibration unit is kg)
- Long press the Menu button until you see MENU.
- Release the button and wait for the display to show C.A.L. Press the Yes button.
- Short press the No button to navigate until you see SPAN. Press the Yes button.
- The calibration point and the unit are flashing on the display based on the scale capacity and unit set in the menu (e.g., 030.000 kg). If you do not need to change the calibration point, skip to step 6.
- To change the calibration point:
- Short press the No button several times until the desired number appears. Note: press the Back button can decrease the digit.
- Short press the Yes button to accept the number and move on to the next digit.
- Repeat the process until all the digits are correct.
- Press the Yes button to accept calibration point. It is flashing on the display.
- Place a calibration mass of the specified weight on the pan and press the Yes button.
- The display shows ----, and then -DONE- when the calibration is finished.
- Then the display shows LIN. Press the Exit button to exit.
Notes:
- If calibration has failed, CAL E is displayed for 3 seconds and the previous calibration data is restored.
- If after waiting for 40 seconds the calibration is still not successful, CAL E is displayed for 3 seconds and the previous calibration data is restored.
4.4. Linearity Calibration [LIN]
Linearity calibration uses 3 calibration points. The full calibration point is established with a weight on the pan. The mid calibration point is established with a weight equal to half of the full calibration weight on the pan. The zero calibration point is established with no weight on the pan. The full calibration and mid calibration points can be altered by users during the calibration procedure to reflect actual weights available for calibration.
Calibration procedure (when the calibration unit is kg)
- Long press the Menu button until you see MENU.
- Release the button and wait for the display to show C.A.L. Press the Yes button.
- Short press the No button several times to navigate until you see LIN. Press the Yes button.
- 0 kg and the calibration unit are flashing on the display. With no weight on the pan, press the Yes button to establish the zero point. The display shows ----.
- Then the first calibration point and the unit are flashing on the display based on the scale capacity and unit you set in the setup menu (e.g., 015.000 kg). If you do not need to change the calibration point, skip to step 7.
- To change the calibration point:
- Short press the No button several times until the desired number appears. Note: press the Back button can decrease the digit.
- Short press the Yes button to accept the number and move on to the next digit.
- Repeat the process until all the digits are correct.
- Press the Yes button to accept calibration point. It is flashing on the display.
- Place a calibration mass of the specified weight on the pan and press the Yes button. The display shows ----.
- The second calibration point and the unit are flashing on the display based on the capacity and unit you set in the setup menu (For example, 030.000 kg).
- If calibration has failed, CAL E is displayed for 3 seconds and the previous calibration data is restored.
- If after waiting for 40 seconds the calibration is still not successful, CAL E is displayed for 3 seconds and the previous calibration data is restored.
- Repeat step 6 and 7.
- The display shows ----, and then -DONE- when the Linearity calibration is finished.
- After that the display shows DONE. Press the Exit button to exit.
Note:
4.5. Calibration Test [C.test]
A calibration test printout can be generated if a printer is attached. For more information refer to section 4.2.6 in the Defender 3000 Indicators (i-DT33) manual.
End Cal [End]
When End is displayed, press the Yes button to exit this menu and advance to the next sub-menu, or press the No button to advance to the first menu item in this sub-menu.
5. MAINTENANCE
CAUTION: DISCONNECT THE UNIT FROM THE POWER SUPPLY BEFORE CLEANING.
5.1. Cleaning
The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
WARNING: Electric Shock Hazard. Disconnect the equipment from the power supply before cleaning. Make sure that no liquid enters the interior of the instrument.
Attention: Do not use solvents, harsh chemicals, ammonia or abrasive cleaning agents.
5.2. Troubleshooting
For technical issues contact an Authorized Ohaus Service Agent. Please visit our website www.ohaus.com to locate the OHAUS office nearest you.
6. TECHNICAL DATA
6.1. Specifications
Equipment Ratings: | |
---|---|
Altitude: | 2,000m |
Operating temperature: | -10°C to 40°C |
Humidity: | Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative humidity at 40°C. |
Electrical supply: | 100 - 240V~, 0.5A, 50/60Hz |
Voltage fluctuations: | Mains supply voltage fluctuations up to ±10% of the nominal voltage. |
Overvoltage category (Installation category): | II |
Pollution degree: | 2 |
Model Specifications
Model | i-D33P75SD |
---|---|
Capacity x Readability | 75 kg x 0.02 kg 150 lb x 0.05 lb |
Weighing units | Kilogram, Gram, Pound, Ounce, Pound: Ounce |
Modes | Weighing, Counting, Check, Totalization |
Stabilization time | Within 2 seconds |
Zero range | To capacity by subtraction |
Maximum Overload | 150% |
Battery Operating Time | Dry cell battery: up to 150 hours continuous use with backlight off Rechargeable battery (optional): up to 75 hours continuous use with backlight off |
Operating Temperature Range | -10°C to 40°C/14°F to 104°F |
Indicator to Base Cable Length | 1.8m / 72" (uncoiled) |
Display | 45 mm / 1.8 in digit height LCD display with 4-color backlight |
Power cord length | 1.9m / 75" (fully extended) |
Scale Base Size | 316mm L x 280mm W x 45mm H / 12.4" L x 11" W x 1.77" H |
Indicator Size | 252 x 176 x 80 mm / 9.9 x 6.9 x 3.1 in |
Power | 100-240 VAC / 50-60 Hz internal power supply |
Net Weight | 4.5kg / 9.9lb |
Shipping Weight | 9.2kg / 20.3lb |
Description of Scale Base Overall Dimension: The scale base has a length of 316mm (12.44 inches) and a width of 280mm (11.02 inches). The height is 45mm (1.77 inches).