WTW VARION® Plus Digital ISE Combination Sensor for Ammonium and Nitrate

Reagent-free measurement of Ammonium and Nitrate ion-selectively with automatic compensation of Potassium/Chloride using the VARION®Plus 700 IQ.

Information on application possibilities and areas of use is available on our website.

Product Overview

The WTW VARION® Plus is a digital ion-selective combination sensor designed for the precise measurement of Ammonium and Nitrate in water. It features reagent-free operation and automatic compensation for Potassium or Chloride ions, enhancing accuracy and simplifying usage.

Physical Characteristics

The sensor has a minimum immersion depth of 50 mm. The main body of the sensor is approximately 392 mm in length. Key connection points are indicated at 48, 55, and 40 mm along the sensor, with a SACIQ socket connection.

Technical Specifications

FeatureAmmonium MeasurementNitrate Measurement
Measuring PrinciplePotentiometricPotentiometric
Maximum Sensor ConfigurationCommon reference electrode, two measuring electrodes, one compensation electrode
Integrated Measuring Chains: ReferenceVARION®Plus RefVARION®Plus Ref
Measuring ElectrodeVARION®Plus NH4, VARION®Plus KVARION®Plus NO3, VARION®Plus Cl
Measurement Ranges/ResolutionNH4-N: 1 ... 2,000 mg/l / 1 mg/l;
0.1 ... 100.0 mg/l / 0.1 mg/l
NH4+: 1 ... 2,580 mg/l / 1 mg/l;
0.1 ... 129.0 mg/l / 0.1 mg/l
K+: 0.1 ... 1,000 mg/l / 0.1 mg/l
NO3-N: 1 ... 1,000 mg/l / 1 mg/l;
0.1 ... 100.0 mg/l / 0.1 mg/l
NO3-: 5 ... 4,500 mg/l / 5 mg/l;
0.5 ... 450.0 mg/l / 0.5 mg/l
Cl-: 0.1 ... 1,000 mg/l / 0.1 mg/l
Measurement Accuracy (Lab Standards)±5% of reading ± 0.2 mg/l in standard solutions
Calibration TypesMatrix adjustment against any reference value, 2-point calibration with combined standards possible
Service Life (typical)Reference electrode: 18 months, Measuring and compensation electrode: 18 months in typical application (municipal wastewater treatment plants)
Temperature Measurement and CompensationIntegrated NTC, Range 0 °C ... +40 °C, Accuracy ±0.5 K, Resolution 0.1 K, t95 < 20 s
pH RangepH 4 ... pH 8.5pH 4 ... pH 11
Pressure ResistanceMaximum 0.2 bar (incl. sensor connection cable, with integrated measuring chains)
Environmental ConditionsOperating temperature: 0 °C ... +40 °C, Storage temperature: 0 °C ... +40 °C
Electrical Connection2-wire shielded cable with quick-connect at the sensor
Electromagnetic CompatibilityEN 61326, Emission: Class B, FCC Class A. Immunity for essential operation.
Test MarksCE, UKCA
MechanicsHousing shaft: V4A stainless steel 1.4571
Temperature sensor: V4A stainless steel 1.4571
Measuring chain holder: POM
Protective cage: POM, Protection class: IP 68 (0.2 bar, with integrated measuring chains)
WeightApprox. 670 g (without measuring chains, without sensor connection cable)
WarrantyVARION®Plus 700 IQ: 2 years; Electrodes: 1 year for material defects

Product Variants and Ordering Information

ModelDescriptionOrder No.
VARION®Plus 700 IQDigital armature for ion-selective measurement of Ammonium and Nitrate, without electrodes (Cable please order separately)107040
VARION®Plus AN/A comp SET NH4 & NO3VARION®Plus 700 IQ, VARION®Ref, VARION®Plus NH4 and VARION®Plus NO3, VARION®Plus K (Potassium) (Cable please order separately)107066
VARION®Plus AN/N comp SET NH4 & NO3VARION®Plus 700 IQ, VARION®Ref, VARION®Plus NH4 and VARION®Plus NO3, VARION®Plus Cl (Chloride) (Cable please order separately)107068

PDF preview unavailable. Download the PDF instead.

v1 , WTW, Prozess-Segment Adobe PDF Library 17.0

Related Documents

Preview WTW VARION Plus, AmmoLyt Plus, NitraLyt Plus: Ion-Selective Electrodes for Ammonium and Nitrate Measurement
Detailed information on WTW's VARION Plus, AmmoLyt Plus, and NitraLyt Plus ion-selective electrode sensors for accurate ammonium and nitrate measurement in water analysis, featuring easy operation, cost savings, and advanced capabilities.
Preview WTW VARION® Plus, AmmoLyt® Plus, NitraLyt® Plus: Soluzioni per Misurazione Ammonio e Nitrato
Scopri le soluzioni WTW VARION® Plus, AmmoLyt® Plus e NitraLyt® Plus per la misurazione accurata e semplice di ammonio e nitrato in acqua. Caratteristiche avanzate, calibrazione facile e robustezza per applicazioni industriali.
Preview WTW VARION® Plus, AmmoLyt® Plus, NitraLyt® Plus Sensoren für Ammonium- und Nitratmessung
Entdecken Sie die WTW VARION® Plus, AmmoLyt® Plus und NitraLyt® Plus Sensoren von Xylem für präzise und einfache In-Situ-Messungen von Ammonium und Nitrat in Wasser. Erfahren Sie mehr über Kalibrierfreiheit, Robustheit und Integration ins IQ Sensor Net.
Preview WTW VARION Plus, AmmoLyt Plus, NitraLyt Plus: Ammonium and Nitrate Measurement Sensors
Discover WTW's advanced sensors for precise ammonium and nitrate measurement in water quality monitoring. Featuring VARION Plus, AmmoLyt Plus, and NitraLyt Plus, these solutions offer easy operation, cost savings, and high performance in challenging media, integrating seamlessly with the IQ Sensor Net system.
Preview Capteurs WTW VARION Plus, AmmoLyt Plus, NitraLyt Plus : Mesure d'Ammonium et de Nitrate
Découvrez les capteurs ion-sélectifs avancés de WTW pour la mesure précise de l'Ammonium et du Nitrate dans l'eau. Apprenez les caractéristiques, spécifications techniques et informations de commande des VARION Plus, AmmoLyt Plus et NitraLyt Plus.
Preview WTW Continuous Color Measurement Sensors for Water Analysis
Discover WTW's advanced sensors for continuous, compliant color measurement in water. Featuring ColorVis, CarboVis, and NiCaVis models, these instruments ensure reliable analysis of wastewater and surface water, adhering to ISO standards and offering benefits like turbidity compensation and ultrasonic cleaning.
Preview Alyza IQ PO4 Analyzer for Orthophosphate Measurement by WTW
Discover the Alyza IQ PO4, the most economical analyzer for direct orthophosphate measurement at the edge of basins. Learn about its revolutionary Multiport Valve, low reagent consumption, and precise results for wastewater treatment applications.
Preview WTW FCML 412 N/M12 pH-Independent Free Chlorine Sensor Operating Manual
This manual provides comprehensive instructions for the WTW FCML 412 N and FCML 412 M12 pH-independent amperometric sensors used for free chlorine measurement in water. It covers setup, operation, maintenance, and technical specifications for reliable water quality monitoring.