Precision Medical EasyGauge Reg 1600 Series Oxygen Regulator
Model: 168715G (Shown)
Precision Medical
300 Held Drive • Northampton, PA 18067 USA
T: 800-272-7285 • F: 800-352-1240
www.precisionmedical.com
ISO 13485 Certified
Receiving / Inspection
Remove the Precision Medical, Inc. Gauge Regulator from packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.
Intended Use
To regulate flow of oxygen from a high pressure oxygen cylinder for medical use only. To be used by or after instruction from a trained, healthcare professional.
Read All Instructions Before Using
This manual instructs a Professional to install and operate the Gauge Regulator. This is provided for your safety and to prevent damage to the Gauge Regulator. If you do not understand this manual, DO NOT USE the Gauge Regulator and contact your Provider.
Explanation of Abbreviations
Abbreviation
Meaning
psi
Pounds Per Square Inch
kPa
Kilopascal
l/min
Liters Per Minute
Safety Information - Warnings and Cautions
AWARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
ACAUTION Indicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury.
CAUTION Used without the safety alert symbol, indicates a potentially hazardous situation which, if not avoided, could result in property damage.
i OR ! CONSULT ACCOMPANYING DOCUMENTS
[No Oil Symbol] Symbol for "USE NO OIL"
WARNING
To prevent risk of burns, fire or injury to person(s):
Handle the Gauge Regulator carefully at all times and use the Regulator according to the instructions in this manual.
Oxygen cylinder should be in an upright position when in use.
ALWAYS confirm prescribed flow before administering to patient and monitor flow on a frequent basis.
Always follow ANSI and CGA standards for Medical Gas Products and Flowmeters and Oxygen Handling.
DO NOT use tape, oils, greases, petroleum jellies, organic lubricants, paint or any other hydrocarbon based or combustible materials on or near this Gauge Regulator.
DO NOT use if oil, grease or contaminants are present on or around Regulator, cylinder valve or connecting devices.
DO NOT use near any type of flame or flammable/explosive substances, vapors or atmosphere.
DO NOT smoke in an area where oxygen is being administered.
CAUTION
Only personnel instructed and trained in its use should operate this Gauge Regulator.
This Gauge Regulator contains magnetic, ferrous material that may affect the results of an MRI.
Ensure all connections are tight and leak free.
Only use oxygen safe leak detector to test for leaks.
Store Regulator in a clean area when not in use.
DO NOT direct flow of oxygen at any person, or flammable material when "cracking" cylinder.
DO NOT autoclave.
DO NOT gas sterilize with EtO (Ethylene Oxide).
DO NOT clean with aromatic hydrocarbons.
DO NOT attempt to repair Regulator.
DO NOT immerse the Regulator in any kind of liquid. This will void the warranty.
Specifications
Accuracy: ±10% of indicated setting @ 70°F (21°C)
Transport / Storage Temperature Range: -4°F (-20°C) to 140°F (60°C)
Specifications are subject to change without prior notice.
Product Diagram
CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc.
Flow Gauge
Gauge Guard
T-Handle
Outlet Connection
Flow Control Knob
Contents Gauge
Operating Instructions
WARNING Read this User Manual before installing or operating the Gauge Regulator.
Check cylinder valve to be sure it is not bent or damaged. A damaged valve could leak or fail, or it may not make a tight connection when cylinder is placed into use.
Make sure the Regulator and cylinder valve are free of oil, grease or other contaminants.
Before installing Regulator, "crack" cylinder valve by opening the valve slightly and closing to remove any foreign particles.
Turn Flow Control Knob fully clockwise to the "OFF" position.
a) For CGA 870 connection (Yoke):
Be sure the high pressure inlet port has a Yoke Seal.
Attach Regulator to the oxygen cylinder post valve and align index pins with mating holes.
Hand tighten T-handle.
b) For CGA 540 connection (Nut & Nipple):
Attach the Regulator to the oxygen cylinder valve and tighten with wrench.
Position the Regulator so the cylinder valve opening is pointing away from the user and any other person(s).
S-L-O-W-L-Y open cylinder valve counterclockwise, until completely open.
Adjust Flow:
To INCREASE flow - Turn Flow Control Knob counterclockwise
To DECREASE flow - Turn Flow Control Knob clockwise
Set flow by aligning indicator needle with marks on flow gauge face.
NOTE: Adjustment of flow beyond the last calibrated indicator line will result in an undetermined oxygen flow.
Continue procedure as prescribed by your healthcare professional.
NOTE: Precision Medical, Inc. strongly recommends the use of kink proof patient tubing.
To remove Regulator from cylinder:
Completely close oxygen cylinder valve.
Turn Flow Control Knob fully clockwise to the "OFF" position.
S-L-O-W-L-Y loosen T-handle or Nut and remove Regulator.
WARNING
To prevent risk of burns, fire or injury to person(s):
Prior to handling regulator and cylinder valve, be sure hands are free of oil, grease and other contaminates.
DO NOT use Regulator if oil, grease or dirt is found. Particles propelled in a high-velocity oxygen stream could create a spark, causing a fire. Oil and grease can react with oxygen and other oxidizers, causing an explosion.
CAUTION
DO NOT use tools to tighten T-handle. This can lead to over tightening and will cause damage to the Regulator.
CAUTION
Inspect the Regulator for visual damage before use, DO NOT USE if damaged.
For optimum safety, use only a Precision Medical, Inc. Yoke Seal part number 1221. The Yoke Seal supplied with the Regulator is reusable.
DO NOT use plastic yoke seals.
Tell us how we are doing! Visit us at www.precisionmedical.com
Maintenance / Cleaning
Disconnect all connections before cleaning.
Clean exterior surfaces of the Gauge Regulator with a cloth dampened with disinfectant SporicidinR or equivalent.
Wipe dry with a clean cloth.
Store the Regulator in a clean area free from grease oil and other sources of contamination.
CAUTION
DO NOT use cleaning solutions.
DO NOT attempt to repair the Regulator. All repairs must be performed by Precision Medical, Inc.
Returns
Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the internet, www.precisionmedical.com.
Manuals available on our website: www.precisionmedical.com
Troubleshooting
If the Gauge Regulator fails to function, consult the Troubleshooting Guide. If problem cannot be solved, consult your Provider.
Problem
Probable Cause
Remedy
No flow
Cylinder valve closed
Regulator in "OFF" position
Cylinder empty
Turn "ON" cylinder
Set to prescribed flow
Replace cylinder
Limited Warranty and Limitation of Liability
Precision Medical, Inc. warrants that the Medical Gas Gauge Regulator (the Product) will be free of defects in workmanship and/or material for the following period: Five (5) years from date of shipment.
Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.'s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.
ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.
The representative of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.
Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.
Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.
503369 Rev7 Printed in USA
Models: 1600 Series Oxygen Regulator, 1600 Series, Oxygen Regulator, Regulator
Precision Medical EasyPulse 5 Oxygen Conserving Regulator User Manual User manual for the Precision Medical EasyPulse 5 Oxygen Conserving Regulator (1900 Series, CGA 870 Connection). Provides instructions on installation, operation, maintenance, safety, specifications, and troubleshooting.
Precision Medical PM4150 EasyPulse POC Service Manual Comprehensive service manual for the Precision Medical PM4150 EasyPulse POC portable oxygen concentrator. Covers theory of operation, specifications, component descriptions, maintenance, troubleshooting, and detailed replacement procedures for various parts.
Precision Medical PM4150 POC Carry Bag Instructions & Guide Official instructions for placing and using the Precision Medical PM4150 Portable Oxygen Concentrator (POC) in its carry bag. Learn correct placement, strap configurations, and troubleshooting tips.
Precision Medical PM4155 Live Active Five Portable Oxygen Concentrator Service Manual This service manual provides detailed instructions for qualified technicians on the maintenance, repair, and troubleshooting of the Precision Medical PM4155 Live Active Five Portable Oxygen Concentrator. It includes component descriptions, disassembly and reassembly procedures, preventative maintenance guidelines, and diagnostic information.
Precision Medical Fluximeter User Manual User manual for Precision Medical fluximeters, series 1MFA, 4MFA, 6MFA, and 8MFA. Provides instructions for use, safety precautions, specifications, troubleshooting, and warranty information.
Precision Medical EasyPulse POC PM4150 Service Manual This service manual provides detailed instructions for the Precision Medical EasyPulse POC Model PM4150, including theory of operation, specifications, maintenance, troubleshooting, and component replacement procedures for qualified service technicians.