EASYPULSE POC

Service Manual

Model PM4150

Precision Medical, Inc.
300 Held Drive
Northampton, PA 18067 USA

Document ID: 507719 Rev 2, 101718

Preface

Precision Medical, Inc.
300 Held Drive
Northampton, PA 18067 USA

Customer Service / Tech Support
Phone: 1-610-262-6090
Phone: 1-800-272-7285
Fax: 1-610-262-6080
Web: www.precisionmedical.com

This manual is intended to guide and help a qualified service technician in the safe handling, service, and repair and performance verification of the PM4100 Series “EasyPulse POC" Portable Oxygen Concentrator. A qualified service technician should be trained in the safe handling of oxygen equipment and understand its inherent dangers.

DO NOT attempt to use or perform any service function on the PM4100 series POC unless you have read and understand this manual as well as the User Manual.

Definition of terms

Safety information - Warnings and Cautions

DANGER: Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.

WARNING: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

CAUTION: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.

Theory of Operation

The EasyPulse POC is intended to provide supplemental oxygen to the ambulatory oxygen patient in the smallest lightest package possible. Utilizing Precision Medical, Inc.'s patented Oxygen Conserving Technology, the EasyPulsePOC is a state of the art device, designed to be versatile and reliable.

This oxygen conserving technology delivers a true "pulse" of oxygen within the first 1/3 of the patient's inspiration and is unique in several additional ways. A true "pulse” is defined by a high peak flow (4 to 10 LPM) for a very short duration of time (0.2 to 0.4 seconds). It also controls the delivered minute volume which remains consistent between 15 and 35 BPM. The total oxygen volume (delivered minute volume) the patient receives is always the same, as the patient's breath rate decreases the pulse volume increases and conversely when breath rate increases.

The EasyPulse POC senses the patient's Inspiratory effort by means of a micro sensor. The sensor signals a micro valve to open for a specific duration based on setting, patient breathe rate and altitude delivering the bolus of oxygen to the patient. This all happens in less than 0.5 seconds. The delivered minute volume is controlled by the setting, the micro valve timing and the patient's breath rate.

Warnings and Cautions

WARNING: Oxygen supplied from this device is for supplemental use and is not intended to be life supporting or life sustaining. This device is not intended for use by patients who would suffer immediate, permanent, or serious health consequences as a result of an interruption in the oxygen supply.

TO AVOID INCREASED RISK OF FIRE

CAUTION

Specifications

Height10.1 in (25.7 cm)
Width6.5 in (16.5 cm)
Depth4.5 in (11.4 cm)
Weight6.8 lbs (3.08 kg)
Shipping Weight13.1 lbs (5.94 kg)
AltitudeSea level to 9000 ft (2743 m)
Operating Temperature41°F to 104°F (5°C to 40°C)
Storage / Transport Temperature-4°F to 140°F (-20°C to 60°C)
Operating / Storage / Transport Relative HumidityUp to 95% Non-Condensing
Oxygen Concentration87% to 95%
Start Up Time87% O2 Concentration ≤ 10 min
Oxygen Minute Volume*
Setting 1240 +/-15% mL/min
Setting 2380 +/-15% mL/min
Setting 3520 +/-15% mL/min
Setting 4660 +/-15% mL/min
Setting 5780 +/-15% mL/min
Mode of DeliveryPulsed Dose only
Trigger Sensitivity0.50 cmH2O maximum
Trigger Delay0.11 seconds typical
PowerAC Power Adapter: 100-240 VAC (1.0 A @ 120 VAC)
DC Power Adapter: 11-16 VDC (5.0 A @ 12 VDC)
Battery Duration (Approximate)Setting 1: 4.5 hrs
Setting 2: 3.2 hrs
Setting 3: 2.5 hrs
Setting 4: 1.8 hrs
Setting 5: 1.5 hrs
Sound (@ Setting 2)44 dBA
CannulaMaximum 7 ft (2.1 m) long standard adult single lumen nasal cannula.

Equipment Classification

Volatile Organic Compound (VOC) and Particulate Requirements

The oxygen delivered from the EasyPulse POC meets the following requirements for particulate levels, VOC levels, carbon monoxide levels, carbon dioxide levels and ozone levels.

* At Sea level

Specifications are subject to change without prior notice.

Component Description

External

Description of the external view of the EasyPulse POC device:

  • Handle: Located at the top for carrying.
  • Setting Knob: Used to select the oxygen flow setting.
  • Oxygen Outlet Fitting: Where the oxygen tubing connects.
  • Internal Battery Level Indicator: Shows the remaining battery charge.
  • Alarm Indicator: Alerts the user to potential issues.
  • Exhaust Vents: Located at the bottom, allowing air to exit.
  • Power Input Jack: For connecting the AC power adapter.
  • Setting Indicator: Displays the currently selected flow setting.
  • Inlet Filter: Located on the back, filters incoming air.
  • Product Serial # Label: Contains the device's serial number and other product information.

Internal

Description of the internal components as shown in an exploded view:

  • Control Knob only (505846)
  • Case Handle (505829)
  • Reservoir Plate Assy (505848)
  • Sieve Bed & Reservoir Assy (505849)
  • Case Filter (505847)
  • Case Screws (506317)
  • Back Case (505828)
  • Compartment Plate (505832)
  • Feet (506881)
  • Compressor Plate Assy (506320)
  • Front Case Assy (506653)
  • PC Board (506729)

Maintenance / Cleaning

WARNINGS

Cabinet

  1. The patient should connect to an alternate Oxygen source.
  2. Turn the POC to the “OFF” position.
  3. Unplug the external power source before cleaning.
  4. Clean exterior surfaces of the POC with a cloth dampened with disinfectant solution. Example: Sporicidin ® Brand Disinfectant Solution.
  5. Allow POC to air dry.
  6. DO NOT place inside carry bag or run the POC until it is thoroughly dry.
  7. When not in use, store the POC in a clean area free from grease, oil and other sources of contamination.

Inlet Filter

  1. Remove the inlet filter by pulling on the protruding tab.
  2. Wash filter with mild detergent. Rinse thoroughly with water and allow to dry completely.
  3. Once filter is dry, push the filter back into the case so only the small tab is protruding.

Returns and Disposal

Returns

Returned products require a Returned Goods Authorization (RGA) number; contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com. Manuals available on our Website; www.precisionmedical.com

Disposal Instructions

This device may contain electrical components that are hazardous to the environment. DO NOT dispose of device into standard trash. Contact your local waste management for disposal of Electronic Equipment.

Troubleshooting

Visual Indicator/ProblemAudible AlarmProbable CauseRemedy
The last LED on the internal Battery Level Indicator turns redRepetitive beepInternal battery has dropped below 15% charge remainingConnect the POC to the AC Power Adapter, DC Power Adapter, or external battery pack
The last LED on the internal Battery Level Indicator flashes redRepetitive beepInternal battery has dropped below 5% charge remaining1. Connect the POC to the AC Power Adapter, DC Power Adapter, or external battery pack
2. Change to alternate oxygen source if power source is not available
Amber alarm indicator is illuminated and setting number is not flashing during inhalation to signal a detected breathRepetitive beepNo breath detected for 1 minute1. Verify the cannula is not kinked
2. Verify the cannula is properly connected to oxygen outlet
3. Verify the cannula is properly positioned in the nose
Red alarm indicator is illuminated and setting number is not flashing during inhalation to signal a detected breathRepetitive beepNo breath detected for 2 minutes1. Verify the cannula is not kinked
2. Verify the cannula is properly connected to oxygen outlet
3. Verify the cannula is properly positioned in the nose
4. If alarm persists change to alternate oxygen source and contact your Equipment Provider
Amber alarm indicator is illuminatedRepetitive beepBreath rate exceeding 35 breaths / minuteReduce level of activity to allow breath rate to slow
Red alarm indicator is illuminatedRepetitive beepBreath rate exceeding 35 breaths / minute for longer than 1 minute1. Reduce level of activity to allow breath rate to slow
2. If alarm persists, change to alternate oxygen source and contact your Equipment Provider
Red alarm indicator is illuminated, device is not running and setting numbers 1 & 2 are flashingRepetitive beepAbnormal compressor operation detected1. Turn the POC to the "Off" position.
2. Turn the POC back "ON" to prescribed setting
3. If alarm persists, change to alternate oxygen source and contact your Equipment Provider
Red alarm indicator is illuminated, device is not running and setting numbers 1& 3 are flashingRepetitive beepInternal battery is outside its operating temperature range1. Turn off the POC and disconnect any power supplies
2. Check that the inlet filter is clean
3. Ensure inlet and exhaust vents are not blocked in any way
4. Verify you are operating the POC within the specified operating temperature range.
5. It may be necessary to allow the POC to warm up or cool before turning back "ON"
Red alarm indicator is illuminated, device is not running and setting numbers 1& 4 are flashingRepetitive beepMotor is above its maximum operating temperature1. Turn off the POC
2. Check that the inlet filter is clean
3. Ensure inlet and exhaust vents are not blocked in any way
4. Verify you are operating the POC within the specified operating temperature range
5. It may be necessary to allow the POC to cool before turning back "ON"
Red alarm indicator is illuminated, device is not running and setting numbers 1& 5 are flashingRepetitive beepPressure outside of normal operation1. Turn the POC to the "Off" position
2. Turn the POC back "ON" to prescribed setting
3. If alarm persists, change to alternate oxygen source and contact your Equipment Provider
Red alarm indicator is illuminated, device is not running and setting numbers 2 & 3 are flashingRepetitive beepfan malfunction1. Turn the POC to the "Off" position
2. Turn the POC back "ON" to prescribed setting
3. If alarm persists, change to alternate oxygen source and contact your Equipment Provider
Red alarm indicator is illuminated, device is not running and setting numbers 2& 4 are flashingRepetitive beepfailed electrical connection1. Turn the POC to the "Off" position
2. Turn the POC back "ON" to prescribed setting3. If alarm persists, change to alternate oxygen source and contact your Equipment Provider
Amber alarm indicator is flashing and device is runningNAService Needed1. Continue to use POC
2. Contact Equipment Provider
Red alarm indicator is flashing and device is runningNAService Needed1. Switch to another source of oxygen
2. Contact Equipment Provider

Alarm Indicators

Indicator ColorStatusDescription
GreenReady for useReady for use
AmberCautionCaution - Prompt response by the OPERATOR is required
RedWarningWarning - Immediate response by the OPERATOR is required

The device features indicator lights on the control panel to signal its status. The amber indicator light signifies a caution, requiring prompt operator attention, while the red indicator light signifies a warning, demanding immediate operator response.

Service Procedures

Sieve Bed Assembly Replacement (505849)

Tools and equipment required:

  • #1 Phillips screwdriver
  • Small diagonal wire cutters
  • Needle nose pliers
  • 3/16" (4.7625 mm) ID silicone tubing x 6 inch (15 cm) long (minimum)
  • Cable Ties 4" x 3/32"

Disassembly

  1. Disconnect the POC from any external power source.
  2. Ensure the device is set to the off position.
  3. Lay the POC on to its front with the handle away from you.
  4. Using a #1 Phillips screwdriver, remove the four screws from the back cover and set aside.
  5. Remove the Back Cover by lifting up and off the Sieve Bed Assembly, set it aside as shown.
  6. Stand the POC up. Remove the tubing clip from the right large Sieve Bed Tube as shown below.
  7. Using diagonal cutters carefully cut the Cable Ties securing the two (2) large Sieve Bed Tubes to the Sieve Bed Assembly as shown below and discard the old cable ties.
  8. Carefully remove the two (2) large diameter Tubes from the Sieve Bed assembly.
  9. Using the diagonal cutters cut the Cable Tie on the Tube connecting the Sieve Bed Assembly to the 'Y' Connector (as shown above) and remove the Tube from the 'Y' Connector.

Image Description: The image shows cable ties securing the sieve bed tubes to the assembly. Arrows indicate the cable ties to be cut.

STOP! From this point you must follow Electrostatic Discharge procedures to avoid damage to the electronic components.

  1. Disconnect the small black 6 pin connector from the PC board and remove the sieve bed assembly from the POC as shown below.
  2. Attach a piece of 3/16" (4.7625 mm) tubing to each of the large tubing fittings on the sieve bed assembly as shown and set aside.

Image Description: The image shows the sieve bed assembly disconnected from the PC board. Another image shows new tubing attached to the sieve bed assembly fittings.

Reassembly

(Reference photos from the "Disassembly" section above)

  1. Stand the POC up.
  2. Remove the temporary Tube from the Sieve Bed Assembly.
  3. Ensure the Elbow Fittings on the Sieve Bed Assembly are pointing away from one another.
  4. Set the Sieve Bed Assembly into position.
  5. Attach the large Tubes to the Sieve Bed Assembly, left Tube to the left Elbow, right Tube to the right Elbow, secure both with Cable Ties.
  6. Attach the small Tube from the Pressure Tank to the 'Y' Connector and secure with a Cable Tie.
  7. Snap the Tubing Clamp on to the right side large Tube.
  8. Connect the black 6 pin Connector to the PC board.
  9. Lay the POC on to its front.
  10. Position the Rear Cover over the Sieve Bed Assembly and slide into place.
  11. Secure the Rear Cover with 4 screws.
  12. Perform the Oxygen Purity Test to verify performance.

Control Panel Replacement (507646)

Tools and equipment required:

  • #1 Phillips screwdriver
  • Small flat blade screwdriver
  • Small diagonal wire cutters
  • Needle nose pliers
  • 3/16" (4.7625 mm) ID silicone tubing x 6 inch (15 cm) long (minimum)
  • Cable Ties 4" x 3/32"
  • ESD mat

Disassembly

  1. Follow steps 1 thru 11 of the "Disassembly" section of the "Sieve Bed Replacement" in this manual.
  2. Lay the POC on its front cover.
  3. Models with O2 Sensor
    • Using diagonal cutters, carefully cut and discard the Cable Tie securing the O2 Sensor Board to the 4 pin Battery Connector wires.
    • Using diagonal cutters, carefully cut and discard the Cable Tie that holds the tubing bundle together located to the right of the welded filter.
  4. Carefully remove the large white 4 pin Battery Connector from the PC Board.
  5. Carefully remove the large white 3 pin Motor Connector from the PC Board.
  6. Carefully remove the small white 9 pin Motor Connector from the PC Board.
  7. Models with O2 Sensor
    • Carefully remove the small white 4 pin Oxygen Sensor Connector from the Oxygen Sensor PC Board.
  8. Carefully remove the two (2) small 3 pin Fan Connectors from the PC Board.
  9. Fold the Wire Harness down and out of the way.
  10. Using diagonal cutters carefully cut the Cable Tie securing the small Tube to the PC Board as shown below and discard the used Cable Tie.

Image Description: The image shows a cable tie securing a small tube to the PC board, with an arrow pointing to the tie.

  1. Disconnect the small Tube from the PC Board.
  2. Using needle nose pliers carefully disconnect the small blue Tube from the cannula sensor on the PC Board as shown above.
  3. Carefully disconnect the small 2 pin connector from the Conserver Valve from the PC Board as shown above.
  4. Using a #1 Phillips screwdriver, remove the four (4) Screws securing the PC Board in place.

Image Description: The image shows the PC Board with connectors and screws, indicating the location of ribbon cables.

  1. Carefully rotate the PC Board up to expose the ribbon cables as shown.
  2. Release the Ribbon Cable connectors by sliding the clip away from the connector, disconnect both ribbon cables and remove the PC Board from the POC and place in a static free bag.
  3. Using a small flat blade screwdriver, from the top of the control panel, carefully pry off the Control Knob and set aside.
  4. Using a small flat blade screwdriver, from the top of the control panel, carefully pry it away from the front POC housing until it releases. The control panel is attached to the housing with an adhesive backing.
  5. When the adhesive has released remove the Control Panel and discard.

Reassembly

STOP! Ensure to follow Electrostatic Discharge procedures to avoid damage to electronic components

(Reference photos from the "Disassembly" section above)

  1. Stand the POC up.
  2. Remove the adhesive backing's paper cover from the replacement Control Panel and feed the Ribbon Cable through the slots in the front POC housing and secure with the adhesive backing. Make sure the Control Panel is centered in its recess in the Housing.
  3. Place the Control Knob that was removed from the old Control Panel onto the stem on the new Control Panel.
  4. Ensure the Ribbon Cable connectors on the PC Board are in their open position.
  5. Position the PC Board into the POC so the Ribbon Cables can be connected.
  6. Using needle nose pliers gently grasp the narrow Ribbon Cable and slide into its corresponding Connector on the PC Board. Secure the Ribbon Cable in place by pushing the Connector latch down and onto the Ribbon Cable. You should hear a faint click.
  7. Repeat steps 4 for the wide Ribbon Cable.
  8. Carefully position the PC Board into place and secure with 4 Screws.
  9. Connect the Conserver valve 2 pin Connector to the PC Board.
  10. Using needle nose pliers, connect the small blue Tube to the Cannula Sensors outer Fitting on the PC Board.
  11. Connect the small Tube from the Pressure Tank to the Pressure Sensor on the PC Board and secure with a Cable Tie.
  12. Connect the two (2) small 3 pin Fan connectors to the PC board, orientation does not matter.
  13. Connect the small white 9 pin Motor connector to the PC board.
  14. Connect the large white 3 pin Motor connector to the PC board.
  15. Models with O2 Sensor
    • Connect the small 4 pin Oxygen Sensor Cable to the Oxygen Sensor PC Board.
  16. Connect the large 4 pin Battery connector to the PC Board.
  17. Models with O2 Sensor
    • Install a Cable Tie through the right side mounting hole of the Oxygen Sensor PC Board and around the Battery Connector wires, closest to the Battery Plug. Snug and trim Cable Tie.
    • Install a Cable Tie around the bundle of tubing to the right side of the welded filter. Snug and trim Cable Tie. NOTE: Do NOT over tighten Cable Tie. Will crimp tubing if too tight.
  18. Be sure the wires from the Pump are pushed down and into their original position.
  19. Stand the POC up.
  20. Follow steps 2 thru 12 of the "Reassembly" section of the "Sieve Bed Replacement" in this manual.

PC Board Replacement (506729/506729S O2 Sensor Models)

Tools and equipment required:

  • #1 Phillips screwdriver
  • Small diagonal wire cutters
  • Needle nose pliers
  • 3/16" (4.7625 mm) ID silicone tubing x 6 inch (15 cm) long (minimum)
  • Cable Ties 4" x 3/32"
  • ESD Mat

Disassembly

  1. Follow steps 1 thru 11 of the "Disassembly" section of the "Sieve Bed Replacement" in this manual.
  2. Follow steps 2 thru 14 of the "Disassembly" section of the "Control Panel Replacement" in this manual.

Reassembly

  1. Follow steps 4 thru 16 of the "Reassembly" section of the "Control Panel Replacement" in this manual.
  2. Follow steps 2 thru 12 of the "Reassembly" section of the "Sieve Bed Replacement" in this manual.

NOTICE!

The correct Firmware must be installed for the device to operate correctly.

Compressor Motor 505852 has been replaced with new Compressor Motor 507640.

If Motor PN# 505852 is installed, firmware should be _PXXX.

If Motor PN# 507640 is installed, firmware should be _AXXX.

Models with O2 Sensor, the firmware should be _LXXX.

To determine the installed firmware revision:

  • Remove Inlet Filter from back of device and look thru window. (or remove back cover)
  • Locate LCD Display on circuit board in the upper left hand corner.
  • With the device Control Knob in the off position, plug an external power source into the DC jack.
  • Upon power-up the LCD Display will display “HELLO”; followed by the Firmware revision (_XXXX); followed by run Hours.

When ordering a replacement Circuit Board, be sure to indicate the required Firmware version.

Oxygen Sensor Replacement (508419) Models with O2 Sensor

Tools and equipment required:

  • #1 Phillips screwdriver
  • Small diagonal wire cutters
  • Needle nose pliers
  • Cable Ties 4"x 3/32"
  • ESD Mat

STOP! Ensure to follow Electrostatic Discharge procedures to avoid damage to electronic components

Disassembly

  1. Follow steps 1 thru 11 of the "Disassembly" section of the "Sieve Bed Replacement" in this manual.
  2. Lay the POC on its front cover.
  3. Using diagonal cutters, carefully cut and discard the Cable Tie securing the O2 Sensor PC Board to the 4 pin Battery Connector wires.
  4. Using diagonal cutters, carefully cut and discard the Cable Tie that holds the tubing bundle together located to the right of the welded filter.
  5. Carefully remove the large white 4 pin Battery Connector from the PC Board.
  6. Carefully remove the small 4 pin Oxygen Sensor Connector from the Oxygen Sensor PC Board.
  7. Carefully remove both the top and bottom tubing from the Oxygen Sensor

Reassembly

  1. Connect the tubing, that is connected to the patient outlet port, to the top barb fitting of the replacement Oxygen PC Board.
  2. Connect the tubing, that is connected to the Bolus Valve, to the bottom barb fitting of the replacement Oxygen PC Board.
  3. Install Cable Ties around each tubing connected to the Oxygen Sensor PC Board, and trim.
  4. Connect the small 4 pin Oxygen Sensor Cable to the Oxygen Sensor PC Board.
  5. Connect the large 4 pin Battery connector to the PC Board.
  6. Install a Cable Tie through the right side mounting hole of the Oxygen Sensor PC Board and around the Battery Connector wires closest to the Battery Plug. Snug and trim Cable Tie.
  7. Install a Cable Tie around the bundle of tubing to the right side of the welded filter. Snug and trim Cable Tie. NOTE: Do NOT over tighten Cable Tie. Will crimp tubing if too tight.
  8. Be sure the wires from the Pump are pushed down and into their original position.
  9. Stand the POC up.
  10. Follow steps 2 thru 12 of the "Reassembly" section of the "Sieve Bed Replacement" in this manual.

Internal Battery Replacement (507547)

Tools and equipment required:

  • #1 Phillips screwdriver
  • Small diagonal wire cutters
  • Needle nose pliers
  • Cable Ties 4"x 3/32"
  • 3/16" (4.7625 mm) ID silicone tubing x 6 inch (15 cm) long (minimum)

Disassembly

  1. Follow steps 1 thru 11 of the "Disassembly" section of the "Sieve Bed Replacement" in this manual.
  2. Follow Steps 2 thru 7 of the "Disassembly" section of the "Control Panel Replacement" in this manual.
  3. Remove the 4 Screws securing Pump Compartment Cover and set aside
  4. Using diagonal cutters cut the Cable Tie that secures the large rear Tube connecting the Inlet Muffler to the Pump and remove the Tube from the Inlet Muffler.
  5. Slide the Pump/Fan Assembly out of the POC, set aside.
  6. Using diagonal cutters, cut the two (2) Cable Ties securing the Battery to the Battery Shelf as shown and discard.

Image Description: The image shows the internal battery secured by two cable ties, with arrows pointing to the cable ties.

  1. Remove the battery and dispose of in accordance to local recycling requirements.

Reassembly

(Reference photos from the "Disassembly" section above)

  1. From the bottom of the Battery Shelf, feed the Cable Tie up and through the rear right slot in the Shelf, and back through the rear left slot in the Shelf.
  2. Engage the cable tie into itself about 1 inch (25mm).
  3. Position replacement Battery so the wires are to the upper right corner.
  4. Slide the Battery down into position in the POC and through the Cable Ties.
  5. Tighten the rear Cable Tie so that the latch is positions on the left side of the Battery Shelf.
  6. Tighten the front Cable Tie so that the latch is positions on the left side of the Battery Shelf.
  7. Ensure the Battery is properly secured, trim off the excess Cable Tie.
  8. Start to secure a Cable Tie and slide it over the large rear Tube on the pump.
  9. Position the Pump/Fan assembly into the POC by sliding it into its corresponding slot.
  10. Using needle nose pliers connect the rear large Tube from the Pump to the Inlet Muffler.
  11. Ensure the large Tube is completely engaged on to the Inlet Muffler Fitting and the Pump is straight in the POC.
  12. Secure the Cable Tie on the large rear Tube from the Pump on to the Inlet Muffler Fitting, trim off the excess Cable Tie.
  13. Position the large Tubes and the Wire Harness into their corresponding Slots in the Battery Shelf.
  14. Position the Pump Compartment Cover into position and secure with the 4 screws.
  15. Follow steps 11 thru 15 of the "Reassembly" section of the "Control Panel Replacement" in this manual.
  16. Stand the POC up.
  17. Follow steps 2 thru 12 of the "Reassembly" section of the "Sieve Bed Replacement" in this manual.

Compressor Plate Assy Replacement (506320)

Tools and equipment required:

  • #1 Phillips screwdriver
  • Small diagonal wire cutters
  • Needle nose pliers
  • Cable Ties 4"x 3/32"
  • 3/16" (4.7625 mm) ID silicone tubing x 6 inch (15 cm) long (minimum)

Disassembly

  1. Follow steps 1 thru 11 of the "Disassembly" section of the "Sieve Bed Replacement" in this manual.
  2. Follow Steps 2 thru 7 of the "Disassembly" section of the "Control Panel Replacement" in this manual.
  3. Fold the large Tubes coming from the Pump down and out of the way.
  4. Using diagonal cutters cut the Cable Tie that secures the large rear Tube connecting the Inlet Muffler to the Pump and remove the Tube from the Inlet Muffler.
  5. Slide the Pump/Fan Assembly up and out of the POC.

Reassembly

  1. Follow steps 9 thru 16 of the "Reassembly" section of the "Internal Battery Replacement" in this manual.
  2. Follow steps 2 thru 12 of the "Reassembly" section of the "Sieve Bed Replacement" in this manual

NOTICE!

The correct Firmware must be installed for the device to operate correctly.

Compressor Motor 505852 has been replaced with new Compressor Motor 507640.

If Motor PN# 505852 is installed, firmware should be _PXXX.

If Motor PN# 507640 is installed, firmware should be _AXXX.

Models with O2 Sensor, the firmware should be _LXXX.

To determine the installed firmware revision:

  • Remove Inlet Filter from back of device and look thru window. (or remove back cover)
  • Locate LCD Display on circuit board in the upper left hand corner.
  • With the device Control Knob in the off position, plug an external power source into the DC jack.
  • Upon power-up the LCD Display will display “HELLO”; followed by the Firmware revision (_XXXX); followed by run Hours.

When ordering a replacement Circuit Board, be sure to indicate the required Firmware version.

Checking the Oxygen Purity

To Put the EasyPulse POC in Test Mode

*It is very important to remove any external power source and allow the device to sit for at least 15 seconds before attempting to put in to the test mode.

*Do not block the outlet while in the test mode. Damage to the device may occur and is not covered under warranty.

  1. Attach the EasyPulse POC to the oxygen purity test device
  2. Remove all external power sources
  3. Turn device off for 15 seconds
  4. Then, like a combination lock, turn dial to: 5, then 1, then 3, then off.
  5. Turn device to any flow setting. Device will continuously pulse.
  6. The alarm LED on the front will blink yellow and red while in test mode. (Steps 3 & 4 must be completed within 8 seconds)
  7. Attach any external power source if needed.
  8. In this mode, you can go to any flow setting and test oxygen purity. The device will pulse as if it was attached to a simulator breathing 20 breaths per minute.
  9. To return device to normal function, turn dial to off position. Once the device is turned to the off position, you will need to go through the above steps to get the device back into the test mode.

PDF preview unavailable. Download the PDF instead.

507719 Rev2

Related Documents

Preview Precision Medical PM4150 EasyPulse POC Service Manual
Comprehensive service manual for the Precision Medical PM4150 EasyPulse POC portable oxygen concentrator. Covers theory of operation, specifications, component descriptions, maintenance, troubleshooting, and detailed replacement procedures for various parts.
Preview Precision Medical PM4150 POC Carry Bag Instructions & Guide
Official instructions for placing and using the Precision Medical PM4150 Portable Oxygen Concentrator (POC) in its carry bag. Learn correct placement, strap configurations, and troubleshooting tips.
Preview Precision Medical PM4155 Live Active Five Portable Oxygen Concentrator Service Manual
This service manual provides detailed instructions for qualified technicians on the maintenance, repair, and troubleshooting of the Precision Medical PM4155 Live Active Five Portable Oxygen Concentrator. It includes component descriptions, disassembly and reassembly procedures, preventative maintenance guidelines, and diagnostic information.
Preview Precision Medical EasyPulse 5 Oxygen Conserving Regulator User Manual
User manual for the Precision Medical EasyPulse 5 Oxygen Conserving Regulator (1900 Series, CGA 870 Connection). Provides instructions on installation, operation, maintenance, safety, specifications, and troubleshooting.
Preview Precision Medical Air-Oxygen Blender User Manual & Specifications
User manual for Precision Medical Air-Oxygen Blenders (PM5200, PM5300). Covers operation, safety, specifications, maintenance, and troubleshooting for medical professionals.
Preview Precision Medical Fluximeter User Manual
User manual for Precision Medical fluximeters, series 1MFA, 4MFA, 6MFA, and 8MFA. Provides instructions for use, safety precautions, specifications, troubleshooting, and warranty information.
Preview Precision Medical EasyDialReg Oxygen Regulator User Manual and Specifications
User manual and technical specifications for the Precision Medical EasyDialReg Oxygen Regulator, including operating instructions, safety warnings, maintenance, troubleshooting, and warranty information.
Preview Precision Medical EasyGauge Reg 1600 Series Oxygen Regulator User Manual
User manual for the Precision Medical EasyGauge Reg 1600 Series Oxygen Regulator, providing instructions for use, safety information, specifications, maintenance, troubleshooting, and warranty details.