USER MANUAL

EASYPULSE 5

Oxygen Conserving Regulator

1900 SERIES
(CGA 870 Connection)

(198705 shown)

SAVE THESE INSTRUCTIONS

Federal (USA) law restricts this device to sale by or on the order of a physician.

PRECISION MEDICAL®
300 Held Drive
Northampton, PA 18067 USA
ISO 13485 Certified

Tel: (+001) 610-262-6090
Fax: (+001) 610-262-6080
www.precisionmedical.com

RECEIVING / INSPECTION

Remove the Precision Medical, Inc. EasyPulse5 Oxygen Conserving Regulator (Conserver) from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider.

INTENDED USE

To regulate high pressure cylinders that provide supplemental oxygen to patients who may have difficulty extracting oxygen from the air they breathe. It is for patients who would normally receive the oxygen via a nasal cannula. The device delivers 100% oxygen at flow settings. It is intended to be used as an oxygen saving device that reduces the drying of the airways.

READ ALL INSTRUCTIONS BEFORE USING

This manual instructs a user to install and operate the Oxygen Conserving Regulator (Conserver). This is provided for your safety and to prevent damage to the Conserver. If you do not understand this manual, DO NOT USE the Conserver and contact your Provider.

[DANGER]

This product is not intended as a life-sustaining or life-supporting device.

EXPLANATION OF ABBREVIATIONS

kPaKilopascal
psiPounds Per Square Inch
l/minLiters Per Minute
b/minBreaths Per Minute

SAFETY INFORMATION - WARNINGS AND CAUTIONS

[DANGER] Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.

[WARNING] Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

[CAUTION] Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury.

[CAUTION] Used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage.

WARNING

CAUTION

PRINCIPLES OF OPERATION

The Oxygen Conserver is designed to be used with high-pressure oxygen systems. It consists of a cylinder connection, cylinder contents gauge (if equipped), high-to-low pressure regulator, orifice plate and a conserving demand module. The regulator reduces the high pressure of the cylinder to the working pressure of the orifice plate. The orifice plate uses calibrated orifices to deliver a selected flow to the conserving demand module. The conserving demand module controls the pulse size and timing to the patient. It supplies a pulse of oxygen at the beginning of each breath. This reduces the oxygen demand on the system and limits the drying of the airways. The flow is determined by setting the flow control knob to the prescribed flow. The oxygen is supplied to the patient through the cannula.

SPECIFICATIONS

Inlet Pressure Range: 300 - 3000 psi (2068 - 20684 kPa)

Pressure Gauge Accuracy: 3 - 2 - 3% of full scale

Dimensions: (Are approximate and may vary by model)

Model: 198705

Pulse Settings: 1, 2, 3, 4, & 5 l/min Equivalents

Flow Settings:
2 l/min Continuous (198705)
1 through 6 l/min Continuous (198712)

Accuracy: Continuous Settings: within ± 10%

Accuracy: Pulse Settings: Within +/- 15% of the nominal bolus value (at each breath rate)

Savings Ratio: Up to 5.7:1

Trigger Method: Inspiratory effort (Negative pressure from patient inhalation)

Breathing Frequency: Up to 35 b/min

Cannula Requirement: Maximum 7 foot (2.1 m) long standard adult single lumen nasal cannula.

Operating Conditions:
Temperature: 35°F to 105°F (1.7°C to 40.6°C)
Altitude: Sea level to 10,000 ft (0 to 3,048 m)

Storage Conditions:
Temperature: -40°F to 140°F (-40°C to 60°C)
Maximum Humidity: 95% Noncondensing

Oxygen Cylinder Connection: CGA 870 Valve

Ignition and Fault Tolerance: Meets ASTM G175-03

Specifications are subject to change without prior notice. Manuals available on our Website; www.precisionmedical.com

COMPONENT DESCRIPTION

[CAUTION]

Missing or illegible labels must be replaced, contact Precision Medical, Inc.

INSTALLATION

[WARNING]

[CAUTION]

NOTE: For operational instructions on how to operate the cylinder consult your Provider.

  1. Position the cylinder so the oxygen cylinder valve outlet is pointing away from the user and any other person(s).
  2. Before installing the Conserver, "crack" cylinder valve by opening the valve slightly to remove any foreign particles and then close valve.
  3. Be sure the Conserver is in the "OFF" position. Align "OFF" with the indicating pointer ▼.
  4. Be sure the CGA 870 connection high pressure inlet has a yoke seal.
  5. Align Index Pins with Index Pin Alignment Holes in Post Valve.
  6. Hand tighten T-handle until snug.

OPERATING INSTRUCTIONS

  1. Attach a standard adult single lumen oxygen nasal cannula, no longer than 7 feet (2.1 m) to the Conserver's outlet according to the cannula manufacturer's instructions.

    [CAUTION]

    • DO NOT use pediatric, low flow nasal cannulas or oxygen masks with the Conserver.
    • DO NOT block the cannula connection or kink cannula tubing when the Conserver is in use, this may damage the Conserver.
  2. Place the cannula over your ears and position the prongs in your nose as instructed by your Provider or cannula manufacturer.
  3. Slowly open cylinder valve counterclockwise, until completely open.
  4. Turn dial on the Conserver until indicating pointer is aligned with the prescribed setting.

    NOTE: When changing from a Pulse setting to a continuous flow setting, the device must be placed in the off position for at least 5 seconds to release back pressure. Not following this step will cause the device to rapidly pulse in the continuous flow position.

  5. Breathe through your nose.

    NOTE: When on a pulse setting, there is flow or a pulse only at the beginning of each breath. If you do not feel the pulse at the beginning of each breath, check the setting. If there is still no pulse, turn dial to an equivalent "Continuous" setting.

    Because each patient's breathing pattern is different and the environment varies, it may be difficult to feel some low setting pulses.

  6. To remove the Conserver from cylinder:
    • Completely close oxygen cylinder valve.
    • Turn the Conserver dial to any "Continuous" position.
    • Wait for oxygen to stop flowing from Conserver.
    • Remove the Conserver from cylinder valve.
    • DO NOT try to loosen T-handle beyond stop. Damage to the Conserver will occur.

[DANGER]

NEVER attempt to remove the Conserver from a cylinder unless the cylinder valve is closed.

[WARNING]

To avoid injury to patient:

[WARNING]

MAINTENANCE / CLEANING

  1. Disconnect all connections before cleaning.
  2. Clean exterior surfaces of the Conserver with a cloth dampened with mild detergent and water.
  3. Wipe dry with a clean cloth.
  4. Store the Conserver in a clean area free of grease, oil, and other sources of contamination.

[CAUTION]

RETURNS

Returned products require a Returned Goods Authorization (RGA) number, contact Precision Medical, Inc. All returns must be packaged in sealed containers to prevent damage. Precision Medical, Inc. will not be responsible for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy available on the Internet, www.precisionmedical.com.

DISPOSAL INSTRUCTIONS

This device and its packaging contain no hazardous materials. No special precautions need to be taken when disposing the device and/ / or its packaging.

[Recycle Symbol] Please Recycle

TROUBLESHOOTING

If the Conserver fails to function, consult the Troubleshooting Guide below. If problem cannot be solved, consult your Provider.

ProblemProbable CauseRemedy
No flow1. Cylinder valve closed
2. Conserver in "OFF" position
3. Cylinder empty
4. Dial not set at proper position
5. Conserver not sensing breath
1. Open the valve on cylinder
2. Set to prescribed setting
3. Replace cylinder
4. Set dial so indicator points to a setting
5. Check position of cannula in nose, DO NOT breathe through mouth
Leaking at cylinder connection1. Missing or defective yoke seal
2. Defective cylinder valve
1. Replace with Precision Medical yoke seal
2. Contact your Provider

REPLACEMENT PARTS

DescriptionPart #
Cannula504833
Yoke Seal1221

ACCESSORIES

DescriptionPart #
Carry Bag M6503920
Carry Bag M4/M6504184
Carry Bag ML6/ML9504185

Carry Bag with shoulder strap

Conserver properly installed in Carry Bag

LIMITED WARRANTY AND LIMITATION OF LIABILITY

Precision Medical, Inc. warrants that the Easy Pulse Oxygen Conserving Regulator (the Product) will be free of defects in workmanship and/or material for the following periods:

Two (2) years from date of shipment.

Should any failure to conform to this warranty appear within the applicable period, Precision Medical, Inc. shall, upon written notification thereof and substantiation that the goods have been stored, installed, maintained and operated in accordance with Precision Medical, Inc.'s instructions and standard industry practice, and that no modifications, substitutions, or alterations have been made to the goods, correct such defect by suitable repair or replacement at its own expense.

ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES.

The representative of Precision Medical, Inc. or any retailers are not authorized to make oral warranties about the merchandise described in this contract, and any such statements shall not be relied upon and are not part of the contract for sale. Thus, this writing is a final, complete and exclusive statement of the terms of that contract.

THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ANY WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESS OR IMPLIED.

Precision Medical, Inc. shall not under any circumstances be liable for special, incidental or consequential damages including but not limited to lost profits, lost sales, or injury to person or property. Correction of non-conformities as provided above shall constitute fulfillment of all liabilities of Precision Medical, Inc. whether based on contract, negligence, strict tort or otherwise. Precision Medical, Inc. reserves the right to discontinue manufacture of any product or change product materials, designs, or specifications without notice.

Precision Medical, Inc. reserves the right to correct clerical or typographical errors without penalty.

DECLARATION OF CONFORMITY

Precision Medical, Inc
300 Held Drive
Northampton PA 18067, USA

Emergo Europe (European Office)
Molenstraat 15
2513 BH, The Hague
The Netherlands
Phone: +31 (0) 70.345.8570
Fax: +31 (0) 70.346.7299

Product: EasyPulse 5 Oxygen Conserving Device

Model Numbers: 198705EN, 195405, 19DE05, 19ES05, 19ES05-EN, 19GB05, 19ES05, 19SE05-ES, 198705NG, 195405NG, 19ES05NG, 19ES05NG-EN, 19GB05NG, 19SE05NG, 19SE05NG-NO, 19WM0501

Classification: IIb

Classification criteria: Clause 3.2 Rule 11 of Annex IX of MDD

We hereby declare that an examination of the under mentioned production quality assurance system has been carried out following the requirements of the UK national legislation to which the undersigned is subjected, transposing Annex II, 3 of the Directive 93/42/EEC on medical devices.

We certify that the production quality system conforms with the relevant provisions of the aforementioned legislation, and the result entitles the organization to use the CE 0473 marking on those products listed above.

Applied Standards: BS 341-3, DIN 477-1, EN 837-1, EN1041, EN ISO 14971, EN ISO 15001, ISO 228-1, ISO 15223-1, ISO 10524-1, ISO 10524-3, ISO 18779, SS 367615

Notified Body: AMTAC Certification Services Limited

Address: Davy Avenue Knowlhill Milton Keynes MK5 8NL, UK

Certification Registration No's: 1126 CE

Date of Expiry: 03 August 2017

Devices already manufactured: S/N traceability Device History Records

Validity of DOC: 04 August 2012 to Date of Expiry

Manufacture Representative: Quality Manager

Position: Quality Systems/ISO Representative

Date of Issue: 04 August 2012

Tell us how we are doing! Visit us at www.precisionmedical.com

PDF preview unavailable. Download the PDF instead.

Precision Medical EasyPulse Five Oxygen Conserving Regulator User Manual Adobe InDesign CS6 (Macintosh) Adobe PDF Library 10.0.1

Related Documents

Preview Precision Medical EasyDialReg Oxygen Regulator User Manual and Specifications
User manual and technical specifications for the Precision Medical EasyDialReg Oxygen Regulator, including operating instructions, safety warnings, maintenance, troubleshooting, and warranty information.
Preview Precision Medical PM4150 EasyPulse POC Service Manual
Comprehensive service manual for the Precision Medical PM4150 EasyPulse POC portable oxygen concentrator. Covers theory of operation, specifications, component descriptions, maintenance, troubleshooting, and detailed replacement procedures for various parts.
Preview Precision Medical EasyGauge Reg 1600 Series Oxygen Regulator User Manual
User manual for the Precision Medical EasyGauge Reg 1600 Series Oxygen Regulator, providing instructions for use, safety information, specifications, maintenance, troubleshooting, and warranty details.
Preview Precision Medical Air-Oxygen Blender User Manual & Specifications
User manual for Precision Medical Air-Oxygen Blenders (PM5200, PM5300). Covers operation, safety, specifications, maintenance, and troubleshooting for medical professionals.
Preview Precision Medical EasyPulse POC PM4150 Service Manual
This service manual provides detailed instructions for the Precision Medical EasyPulse POC Model PM4150, including theory of operation, specifications, maintenance, troubleshooting, and component replacement procedures for qualified service technicians.
Preview Precision Medical PM4150 POC Carry Bag Instructions & Guide
Official instructions for placing and using the Precision Medical PM4150 Portable Oxygen Concentrator (POC) in its carry bag. Learn correct placement, strap configurations, and troubleshooting tips.
Preview Precision Medical Fluximeter User Manual
User manual for Precision Medical fluximeters, series 1MFA, 4MFA, 6MFA, and 8MFA. Provides instructions for use, safety precautions, specifications, troubleshooting, and warranty information.
Preview Precision Medical PM4155 Live Active Five Portable Oxygen Concentrator Service Manual
This service manual provides detailed instructions for qualified technicians on the maintenance, repair, and troubleshooting of the Precision Medical PM4155 Live Active Five Portable Oxygen Concentrator. It includes component descriptions, disassembly and reassembly procedures, preventative maintenance guidelines, and diagnostic information.