Bullard Clean Air Boxes (CAB) Series User Manual

Breathing Air Requirements

Respirable, at least Grade D breathable air must be supplied at all times to the Clean Air Box (CAB) Series units. This is the responsibility of the user or employer. Supplied High Pressure Compressed Air must AT A MINIMUM meet the requirements for Grade D breathable air as described in the Compressed Gas Association Commodity Specification G-7.1, and as specified by Federal Law 42 CFR Part 84, Subpart J, 84.141(b), and 29 CFR Part 1910, Subpart I, 1910.134(i)(1).

The requirements for Grade D breathable air include:

*Specific measurement of odor in gaseous air is impractical. Air may normally have a slight odor. The presence of a pronounced odor should render the air unsatisfactory.

Contact the Compressed Gas Association at www.cganet.com for complete details on Commodity Specification G-7.1.

Air Source for Clean Air Boxes (CAB) Units

Locate the source of supplied air in a clean air environment containing at least Grade D breathable air at all times. Ensure the air source is far enough from your work site or hazardous exhaust emissions to guarantee the air remains contaminant-free and at least Grade D at all times.

Always use an inlet filter on your air source. Utilize suitable after-cooler/dryers with filters, and carbon monoxide monitors and alarms, as necessary to assure clean, at least Grade D breathable air at all times.

Industrial high pressure compressors that send compressed air into a storage air tank are typically used as the air source for Clean Air Boxes (CAB) units. These compressors usually include after-coolers, dryers, filters, CO monitors, and alarms. As compressed air is distributed, certain areas may require additional filtering and CO monitoring to meet Grade D breathable air requirements.

WARNING: Bullard Clean Air Boxes (CAB) MUST be supplied with at least Grade D breathable air. Clean Air Boxes (CAB) do NOT create Grade D breathable air. The air source should be regularly sampled to ensure it meets at least Grade D requirements at all times. Failure to regularly sample the air can result in death, serious injury, or life-threatening delayed lung disease, including silicosis, pneumoconiosis, or asbestosis.

Filtration Efficiency

The CAB units feature a multi-stage filtration system:

WARNING: These units will not remove Carbon Monoxide (CO), Carbon Dioxide (CO2), Nitrogen, or other toxic gases or fumes. They do not increase the oxygen content of an air supply and should not be used when air entering the system is oxygen deficient. Air filtration systems must be operated in the upright position to allow auto drains to function correctly. Standard Clean Air Box (CAB) filtration units are not explosion proof and should not be located in an explosive environment. Failure to read, understand, and follow these instructions could cause death or serious injury.

California Proposition 65 Warnings

California Proposition 65 WARNING: Cancer and Reproductive Harm - www.P65Warnings.ca.gov.

Propuesta de California 65 ADVERTENCIA: Cáncer y Daño Reproductivo - www.P65Warnings.ca.gov.

Proposition de la Californie 65 AVERTISSMENT: Cancer et Troubles de l'appareil reproducteur - www.P65Warnings.ca.gov.

CAB Specifications

The Bullard Clean Air Box is supplied with up to 8 quick-disconnect fittings. The actual number of respirators that may be connected to the CAB depends on the inlet pressure to the Bullard air filtration system and the air flow and pressure requirements of the respirator being worn. Air supply pressure ranges and approved hose lengths for a respirator can be found in the manufacturer's User Manual for the respirator. Only NIOSH approved respirators should be used with this product.

Air Monitoring

The Grade D breathable air flows through the CO monitor. A small amount (~0.5 lpm) is diverted through a preset internal regulator and flow restrictor, delivering a continuous flow of sample air to the sensor chamber. The monitor analyzes the air and displays the CO concentration in parts-per-million (ppm).

A Green LED on the outside of the CAB illuminates during normal operation when the CO level is below the CO alarm point (10 ppm US and 5 ppm International). If the CO level rises above the alarm set point, the Red LED on the outside of the CAB illuminates, and an audible alarm sounds.

The red CO alarm and green LED lights indicating normal operation are located on the face of the monitor inside the CAB15 Clean Air Box.

Function Modes

The knob on the front of the CO monitor sets its functions and can be switched to RUN, TEST, and CAL.

F1 (fault): Indicated on the display when the cam valve is not securely positioned in one of the three modes.

Monitor Specs

Specification Details
Operating Temperature -4°F to 120°F (-20°C to 48.89°C)
Humidity Range 15% to 90% Non-Condensing
Sample Flow Rate 0.85 cfh
Display 2 digit LCD
Sensor Type Electrochemical; Carbon Monoxide
Range 0 - 99 ppm
Calibration Manual CO and Zero - Auto Cal®; No manual adjustments
Alarm Setting 10 ppm (5 ppm International)

Error Codes

Warning Signals

Warranty

Two years on sensor.

Shielding

Internal RFI/EMI coating.

CAB Series Specifications Table

The following table details the specifications for the different CAB models:

Specification CAB15 Series CAB30 Series CAB50 Series CAB100 Series
Unit Dimensions 19" x 13" x 7" 22" x 14" x 7" 24" x 18" x 9" 27" x 20" x 10"
Weight 17 lbs 18 lbs 36 lbs 41 lbs
Inlet Fitting Size - QD Nipple 1/2" 1/2" 1/2" 1/2"
Outlets (1/4") - QD Coupling 1 2 4 8
Maximum Air Flow (lpm/bar) 15 scfm at 110 psi
425 lpm at 7.5 bar
30 scfm at 110 psi
850 lpm at 7.5 bar
50 scfm at 110 psi
1415 lpm at 7.5 bar
100 scfm at 110 psi
2830 lpm at 7.5 bar
Remote Alarm Jack N Y Y Y
Maximum Inlet Pressure 150 psi 150 psi 150 psi 150 psi
Relief Valve 125 psi 125 psi 125 psi 125 psi
Monitoring Inline Continuous CO Monitoring Inline Continuous CO Monitoring Inline Continuous CO Monitoring Inline Continuous CO Monitoring
Power 12V DC or 110-220V AC 50/60Hz 12V DC or 110-220V AC 50/60Hz 12V DC or 110-220V AC 50/60Hz 12V DC or 110-220V AC 50/60Hz
Shipping Dimensions 24.5" L x 17" W x 8" H 24.5" L x 17" W x 8" H 29" L x 22.5" W x 10.75" H 29" L x 22.5" W x 10.75" H

CAB Clean Air Box Set Up and Operation

WARNING: Always operate the CAB Clean Air Box in the upright position. Failure to comply may result in improper auto drain function, filter contamination, or moisture passing into the respirator.

Set Up and Operation

  1. Prior to connecting incoming air supply lines, purge all moisture by running air through the airlines before connecting the CAB.
  2. Attach to an incoming air source containing at least Grade D breathable air; do not exceed 150 psi.
  3. Connect the AC cord to the power inlet on the side of the CAB and plug it into an AC power source.
  4. NOTE: This unit may also be operated using DC power or eight AA batteries if no AC power is available.
  5. Flip the switch on the Control Box to the ON position. (See Figure 1)
  6. Ensure the knob on the CO Monitor is rotated to the TEST position. This prevents the Low Flow alarm from sounding during start-up if no air is flowing. (See Figure 2)
  7. Push the ON button located behind the front cover of the CO Monitor. (See Figure 2)
  8. Wait for the 60-second countdown shown on the CO Monitor display screen.
  9. Calibrate the CO Monitor if necessary. Refer to Calibration Instructions.
  10. Connect the remote alarm assembly (optional) to the remote alarm jack.
  11. NOTE: The knob must be in the RUN position for active monitoring of CO in the air supply.
  12. Attach respirators and adjust regulators to the appropriate pressure per the respirator manufacturer's recommendation. For units with independent regulators, adjust each regulator as appropriate for each outlet being used.

Calibration and Zero-Point Process

Zero-Point Adjustment

NOTE: It is recommended that the Clean Air box be under pressure with no air going through the outlet fitting when performing zero point adjustment or calibration.

NOTE: If the monitor has been relocated to a different working area and ambient temperatures change, allow the unit to stabilize for 15-20 minutes prior to Zero Point adjustment and calibration. Zero calibration should be performed before calibrating the monitor with CO gas. Bullard recommends using impurity-free test gas to re-zero whenever the zero point has drifted.

  1. Remove the front cover of the CO Monitor.
  2. Push the ON button. The unit will beep and begin a 60-second countdown.
  3. Rotate the knob on the front of the CO Monitor to the CAL position. 'AC' will appear on the display.
  4. Press and hold the On/Off button until it beeps. 'AO' will now appear on the display.
  5. Attach the regulator to the zero gas cylinder.
  6. Insert the calibration connector fitting into the center of the knob on the front of the monitor.
  7. Open the cylinder valve fully. The flow rate is preset.
  8. The Red LED will continue to blink for approximately 90 seconds during the zero adjustment process. Watch for the Green LED to begin blinking, indicating a successful zero.
  9. With the zero air flowing, rotate the knob to the TEST position. The reading should alternate between CO and 1 or 0. If any other reading is displayed, repeat Zero Point Adjustment.
  10. Close the regulator on the zero gas cylinder completely and remove the calibration connector fitting.
  11. With the incoming air flowing, turn the knob to RUN.

NOTE: Knob must be in RUN position for active monitoring of CO in the air supply.

Calibration Instructions

The COM5 and COM10 are Auto-Cal instruments; all calibration adjustments are made automatically. CO Monitor calibration should be performed monthly or whenever the reading may be questionable. A calibration schedule should be maintained for future reference. Bullard recommends the use of Bullard CABCK calibration kits for accurate calibration.

Calibration Procedure:

  1. Remove the front cover of the CO monitor and push the ON button if necessary.
  2. Rotate the knob on the front of the monitor to the CAL position. 'AC' will appear on the display.
  3. Attach the regulator to the calibration gas cylinder (10 ppm CO required).
  4. Insert the calibration connector fitting into the center of the knob.
  5. Open the cylinder valve fully. The flow rate is preset.
  6. The Red LED will blink for approximately 90 seconds during calibration. Watch for the Green LED to blink, indicating successful calibration.
  7. With the 10ppm gas still flowing, turn the knob to TEST. The reading should alternate between CO and 10 (+/- 1). If any other reading is displayed, repeat both the Zero Point adjustment and CO calibration.
  8. Close the regulator on the CO gas cylinder completely and remove the calibration connector fitting.
  9. With the incoming air flowing, rotate the knob to the RUN position.

NOTE: Typical calibration takes approximately 90 seconds. If the unit will not calibrate within 3 minutes, repeat calibration steps with compressed air turned off. If the problem persists, contact Bullard Customer Service at 1-877-BULLARD (285-5273).

NOTE: Knob must be in RUN position for active monitoring of CO in the air supply.

Shut Down

  1. Ensure all personnel have left the work area.
  2. Shut off the incoming air to the Clean Air Box.
  3. Release air pressure from the box by pulling the relief valve ring out.
  4. Turn the CO Monitor OFF by pressing and holding the ON/OFF button for three seconds. Three short beeps will indicate the monitor is OFF.
  5. Turn the control box switch to the OFF position.
  6. Disconnect the airline hoses.
  7. Install dust caps to prevent entry of debris into the outlet fittings.

System Maintenance

CAUTION: Always depressurize the system before performing service.

Filter Housing/Bowls

Periodic cleaning of the polycarbonate bowls may be necessary. Remove the auto drains. Clean the bowls with a mild soapy solution. Re-install into the filter housing.

Auto Drains

The automatic drains are designed to remove bulk liquid contaminants. They will automatically drain liquids after the level reaches 1/3 of the bowl capacity. For periodic cleaning, use mild soap and water.

Filter Change

The filtration system includes a filter change indicator that gradually changes from green to orange/red when the filter life is spent.

NOTE: Air must be flowing through the Clean Air Box for the filter change indicators to function.

Drain Lines

Ensure the auto drain tubes are placed in the holes at the bottom of the box to allow liquids to drain outside.

Battery Replacement

CO Monitor Battery

A 9-volt battery is installed under the front panel cover to power the CO monitor. The monitor may also run solely on DC power if the clean air box is not attached to an AC power supply. (See Figure 3)

To replace the 9-volt battery, remove the panel cover from the front of the CO monitor. Lift it off from either the left or right side. If a 9-volt battery is used for power, the monitor will indicate low battery ('LB') on the display when it is time to change the battery.

Optional DC Power Supply

The CAB Clean Air Box may be operated using 8 AA batteries to provide power to the external lights, CO alarm, and the CO monitor. Lights located on the top of the box indicate whether the box is operating on AC or DC power. If powered by DC (8 AA batteries), the amber light on top will burn steady. If battery power falls below the required voltage, the amber light will flash, indicating it is time to change the batteries. An audible alert of short intermittent beeps will also notify that batteries are low. The DC battery compartment is located inside the Clean Air Box. (See Figure 4)

WARNING: Removing the front of the housing by prying at the top will result in damage to the housing and render the product unsafe for use. This damage is not covered under warranty.

Sensor Replacement

NOTE: Before installing a replacement sensor, verify you have the correct sensor. The CABRS is the replacement sensor for COM10 or COM5 monitors with serial numbers ending in 'A' or no suffix. The CABRS2 is the replacement sensor for COM10 or COM5 monitors with serial numbers ending in 'B'. Monitors requiring CABRS2 will have an extra board for sensor mounting. (See Figure 5)

The sensor is accessible by removing the front of the housing from the COM5 or COM10 monitor.

  1. To gain access to the sensor, remove the four screws on the front of the CO monitor.
  2. After removing the screws, gently pry the front of the housing off from the bottom. (See Figure 6)
  3. Gently pull out the old sensor.
  4. Remove the shorting clip cover from the new sensor. Failure to remove this clip will result in an inability to calibrate.
  5. Align the pins and gently push the new sensor into the sockets. (See Figure 7)
  6. Replace the front cover of the monitor, ensuring LED lights align with the housing before securing with screws.
  7. After replacing the sensor, allow the instrument to warm up for at least 30 minutes. The replacement sensor must be zeroed and calibrated before first use. Refer to Zero Point Adjustment and Calibration Instructions.

Replacement Parts Breakdown

The following tables list replacement parts for the CAB units.

CAB15 (15 cfm)

Description Part Number Description Part Number
CO Monitor COM5 or COM10 Snap-Tite Coupling CABQDSN
Filter Change Indicator CAB15FCI Hansen Stainless Coupler CABQDSS
Pressure Regulator with Gauge CABREG1 Dust Cap, Universal CABQDCAP
Relief Valve, 125 psi 41PRV A Filter Element - First Stage CAB30FEA
Hansen Coupling CABQDHA B Filter Element - Second Stage CAB30FEB
Schrader Coupling CABQDSC C Filter Element - Third Stage CAB30FEC
CEJN Coupling CABQDCE CO Sensor CABRS2

CAB30 (30 cfm)

Description Part Number Description Part Number
CO Monitor COM5 or COM10 CEJN Coupling CABQDCE
Filter Change Indicator CAB30FCI Snap-Tite Coupling CABQDSN
Pressure Regulator with Gauge CABREG1 Hansen Stainless Coupler CABQDSS
Independent Regulator CABREG4 Dust Cap, Universal CABQDCAP
Pressure Gauge for Independent Regulator S19685 A Filter Element - First Stage CAB30FEA
Relief Valve, 125 psi 41PRV B Filter Element - Second Stage CAB30FEB
Hansen Coupling CABQDHA C Filter Element - Third Stage CAB30FEC
Schrader Coupling CABQDSC CO Sensor CABRS2

CAB50 (50 cfm)

Description Part Number Description Part Number
CO Monitor COM5 or COM10 CEJN Coupling CABQDCE
Filter Change Indicator CAB50FCI Snap-Tite Coupling CABQDSN
Pressure Gauge S19686 Hansen Stainless Coupler CABQDSS
Pressure Regulator CABREG2 Dust Cap, Universal CABQDCAP
Independent Regulator CABREG4 A Filter Element - First Stage CAB50FEA
Pressure Gauge for Independent Regulator S19685 B Filter Element - Second Stage CAB50FEB
Relief Valve, 125 psi 41PRV C Filter Element - Third Stage CAB50FEC
Hansen Coupling CABQDHA CO Sensor CABRS2
Schrader Coupling CABQDSC

CAB100 (100 cfm)

Description Part Number Description Part Number
CO Monitor COM5 or COM10 CEJN Coupling CABQDCE
Filter Change Indicator CAB100FCI Snap-Tite Coupling CABQDSN
Pressure Gauge S19686 Hansen Stainless Coupler CABQDSS
Pressure Regulator CABREG3 Dust Cap, Universal CABQDCAP
Independent Regulator CABREG4 A Filter Element - First Stage CAB100FEA
Pressure Gauge for Independent Regulator S19685 B Filter Element - Second Stage CAB100FEB
Relief Valve, 125 psi 41PRV C Filter Element - Third Stage CAB100FEC
Hansen Coupling CABQDHA CO Sensor CABRS2
Schrader Coupling CABQDSC

Co Monitor Troubleshooting Guide

Issue Troubleshooting Steps
Monitor will not calibrate at 10 ppm. 1. Allow monitor temperature to stabilize to the environment.
2. Ensure CAB is under pressure with air flowing, but no air through outlet couplers.
3. Perform zero point adjustment using impurity-free air.
4. If monitor zeroes, perform calibration using 10 ppm CO.
5. Verify calibration gas is 10 ppm and cylinder is not empty.
6. If calibration fails, remove air supply and repeat zero point adjustment and 10 ppm CO calibration.
7. If calibration still fails, replace the sensor and repeat zero point adjustment and 10 ppm CO calibration.
Monitor will not successfully complete a zero point adjustment. Wait 5 minutes and repeat steps 1-4 above.
Monitor displays a (-0) reading or rC. Wait 5 minutes and repeat steps 1-6 above.
Monitor zeroes during the zero point adjustment and the reading elevates after returning to the RUN mode. A. Allow unit to stabilize in RUN mode for 5 minutes.
B. Repeat steps 1-6 above.
C. If calibration fails, supply air may be contaminated.
D. Check the supply air for CO level.
Monitor will not turn ON. Verify the ON/OFF button under the front cover has been pressed.
Check AC power supply: Confirm ON/OFF switch is ON, AC power cord is connected, and unit is wired correctly.
Using Battery backup: Confirm AA batteries are good; replace if needed.
Using 9V battery: Verify the 9V battery is good; replace if needed.
NOTE: Monitor requires a 9V battery installed correctly to turn ON.
Monitor does not display flashing green light after calibration or a partial display appears on the monitor screen. If the power supply is good and the monitor will not power up, return it to the factory for repair or replacement.
With the monitor ON and on battery power only, remove and reinstall the 9 volt battery.
NOTE: 9V battery must be good for this step to work.
Repeat steps 1-6 above.
Monitor does not turn OFF. Monitor must be switched to the RUN or TEST mode to turn OFF.
Ensure the rotary cam switch is not locked in CAL, TEST, or RUN position. Turn the switch knob slightly until it is in the correct position.
Monitor displays an (F1) reading. This can occur when the ON/OFF button is pushed while the monitor is in the fault mode with F1 showing.
Continue to push the ON/OFF button until F1 is displayed.
Turn the switch knob slightly to lock the monitor into the CAL, TEST, or RUN mode.
Monitor displays (8.88, 2F or 9E) with flashing lights. Check power supply.
Verify that the 9V battery and AA batteries are good.
If the power supply is good and the monitor will not power up, return it to the factory for repair or replacement.
Monitor shuts down after 60 second countdown. This indicates a low battery condition. The 9V battery in the CO monitor should be replaced.
Monitor displays "lb". This is normal for TEST mode and no action is necessary.
If the unit is not in TEST mode, it may indicate a low battery condition. Replace the 9V battery in the CO monitor.
Yellow caution LED flashes. This is normal for TEST mode and no action is necessary.
If the unit is not in TEST mode, then it may indicate a low battery condition. Replace the 9V battery in the CO monitor.

Warranty Information

One Year LIMITED Warranty

Bullard warrants to the original purchaser that the Clean Air Box will be free of defects in material and workmanship under normal use and service for a period of one (1) year from the date of purchase. Bullard's obligation under this warranty is limited to repairing or replacing, at its option, parts that are returned within the warranty period and that are, after examination, shown to Bullard's satisfaction to be defective, subject to the following limitations:

Bullard shall not be responsible for damages for loss of use or other indirect, incidental, consequential, or special costs, expenses, or damages incurred by the purchaser, even if Bullard has been advised of the possibility of such damages.

ANY IMPLIED WARRANTIES, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE SPECIFICALLY EXCLUDED FROM THIS WARRANTY.

Some states do not allow the exclusion or limitation of incidental or consequential damages, or limitations on how long an implied warranty lasts, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights, and you may have other rights which vary from state to state.

Return Authorization

The following steps must be completed before Bullard will accept any returned goods:

  1. Contact Bullard Customer Service by telephone or in writing at:
    E.D. Bullard Company
    1898 Safety Way
    Cynthiana, KY 41031-9303
    Toll-Free: 877-BULLARD (285-5273)
    Phone: 859-234-6616
    Describe the problem as completely as possible. The representative will try to help correct the problem over the phone.
  2. Verify with your representative that the product should be returned to Bullard. Customer Service will provide written permission and a return authorization number.
  3. Before returning the product, decontaminate and clean it to remove any hazardous materials. Laws and/or regulations prohibit the shipment of hazardous or contaminated materials. Products suspected to be contaminated will be professionally discarded at the customer's expense.
  4. Ship returned products, including those under warranty, with all transportation charges pre-paid. Bullard cannot accept returned goods on a freight collect basis.
  5. Returned products will be inspected upon return. Bullard Customer Service will contact you with a quote for required repair work not covered by warranty. If the cost of repairs exceeds the quoted amount by more than 20%, your representative will call for authorization to complete repairs.

Contact Information

Bullard Center
2421 Fortune Drive
Lexington, KY 40509 • USA
877-BULLARD (285-5273)
Tel: +1-859-234-6616
Fax: +1-859-246-0243

Americas:
E.D. Bullard Company
1898 Safety Way
Cynthiana, KY 41031-9303
Toll free: 877-BULLARD (285-5273)
Tel: 859-234-6616
Fax: 859-234-8987

Europe:
Bullard GmbH
Dieselstrasse 8a
53424 Remagen • Germany
Tel: +49-2642 999980
Fax: +49-2642 9999829

Bullard Asia-Pacific Pte. Ltd.
51 Changi Business Park
Central 2
#03-04 The Signature
Singapore 486066
Tel: +65 6745 0556

©2024 Bullard. All rights reserved. Bullard is a registered trademark of Bullard.

Models: 6081257771F Clean Air Boxes, 6081257771F, Clean Air Boxes, Air Boxes

File Info : application/pdf, 12 Pages, 1.32MB

PDF preview unavailable. Download the PDF instead.

RP CAB UserManual 6081257771F 0324

References

Adobe PDF Library 17.0 Adobe InDesign 19.0 (Windows)

Related Documents

Preview Bullard AirGladiator Airline Filter: Seven Stage Filtration
Discover the Bullard AirGladiator, a seven-stage filtration system designed to provide cleaner, more comfortable breathable air. Learn about its performance benefits, design features, technical specifications, and ordering information.
Preview Bullard AC1000 Cool Climate Control Tube Instruction Sheet
Instruction sheet for the Bullard AC1000 Cool Climate Control Tube, detailing its function, preparation, use, and safety warnings for use with Bullard supplied air respirators. Includes assembly instructions for the heat shield.
Preview Bullard TrakLite User Instructions: Firefighting Helmet Light
Comprehensive user instructions for the Bullard TrakLite integrated helmet light. Learn about its features, operation, installation, maintenance, specifications, and warranty for American-style fire helmets.
Preview Bullard NXT Pro, QXT Pro, and DXT Thermal Imager Quick Setup Guide
A quick setup guide for Bullard NXT Pro, QXT Pro, and DXT Thermal Imagers, covering general operation, settings, and the MyBULLARD Imager Configurator.
Preview Bullard QXT Pro Thermal Imager User Manual
User manual for the Bullard QXT Pro Thermal Imager, providing instructions on operation, features, maintenance, and warranty.
Preview Bullard NXT Thermal Imager Bid Specifications
Detailed bid specifications for the Bullard NXT Thermal Imager, outlining requirements for warranty, service, quality standards, physical configuration, durability, technology, image colorization, housing, colors, display, lens, visual indicators, switches, power supply, operation, and truck mounting systems.
Preview Bullard NXT Thermal Imager User Manual
User manual for the Bullard NXT Thermal Imager, detailing its features, operation, maintenance, and safety guidelines. Covers Super Red Hot (SRH) colorization, TI BASIC PLUS mode, Electronic Thermal Throttle (ETT), digital zoom, and SceneCatcher DVR.
Preview Bullard TXS Thermal Imager Instruction Manual
Instruction manual for the Bullard TXS Thermal Imager, detailing its operation, features like Super Red Hot (SRH) colorization, battery charging, care instructions, troubleshooting, and important safety considerations for firefighters.