ResMed ClimateLineAir™ and ClimateLineAir Oxy™
User Guide
About your ClimateLineAir
The ClimateLineAir™ is heated air tubing that delivers the desired temperature at your mask. ClimateLineAir Oxy™ is a variant of the ClimateLineAir that has a built-in oxygen connector to attach a supplemental oxygen supply.
When used with the HumidAir™ humidifier, ClimateLineAir heated air tubing allows you to use Climate Control. Climate Control is designed to make therapy more comfortable by enabling constant temperature and maintaining humidity.
Intended use
The ClimateLineAir heated air tubing is designed for use with AirSense 10 / AirCurve 10 and Lumis™ devices.
Read this entire user guide together with your device user guide before using your ClimateLineAir heated air tubing.
Notes:
- Not all types of air tubing are available in all regions.
- Lumis does not include S9™ Lumis Tx.
Setup
- Make sure the device is connected and turned on.
- Hold the cuff of the air tubing and line up the air tubing connector with the connector port.
- Push the cuff until the connector clicks into place. ClimateLineAir connected will briefly display on the device screen.
- Connect the assembled mask to the free end of the air tubing.
WARNING: Make sure that you arrange the air tubing so that it will not twist around the head or neck.
Setup with supplemental oxygen
To set up the ClimateLineAir Oxy with supplemental oxygen, perform the steps outlined in the setup section and the following additional steps.
- Remove the cap from the oxygen connector.
- Connect the oxygen supply tubing to the oxygen connector. Make sure it is firmly attached.
WARNING: Supplemental oxygen must not be used while smoking or in the presence of an open flame.
Notes:
- Before adding oxygen, familiarize yourself with the specific warnings relating to the use of supplemental oxygen. These can be found at the end of your device User Guide.
- If you use the ClimateLineAir Oxy without supplemental oxygen, make sure the cap of the oxygen connector is firmly attached.
Climate Control
Climate Control is an intelligent system that controls the humidifier and the ClimateLineAir heated air tubing to deliver constant, comfortable temperature and humidity levels during therapy.
Designed to prevent dryness of the nose and mouth, it maintains the set temperature and relative humidity while you sleep. Climate Control can be set to either Auto or Manual and is only available when both the ClimateLineAir and the HumidAir humidifier are attached.
Climate Control Auto
Climate Control Auto is the recommended and default setting. Climate Control Auto is designed to make therapy as easy as possible, so there is no need to change the temperature or humidity settings.
The Tube Temperature is set to Auto (27°C) and Climate Control adjusts the humidifier output to maintain a constant, comfortable humidity level of 85% relative humidity while protecting against rainout (water droplets in the air tubing and mask).
Tube Temperature
In Climate Control Auto there is no need to change any settings, but if the air in the mask feels too warm or cold you can adjust the tube temperature to find what is most comfortable for you. You can set the Tube Temperature to anywhere between 16–30°C, or turn it off completely.
To adjust the Tube Temperature:
- In My Options, turn the dial to highlight Tube Temp. and then press the dial.
- Turn the dial to adjust the temperature and press the dial to save the change.
Climate Control Manual
Designed to offer you more flexibility and control over your settings, Climate Control Manual lets you adjust the temperature and humidity to the setting which is most comfortable for you. In Climate Control Manual, the tube temperature and the humidity level can be set independently however, rainout protection is not guaranteed. If rainout does occur, first try increasing the tube temperature. If the air temperature becomes too warm and rainout continues, try decreasing the humidity.
Humidity Level
The humidifier moistens the air and is designed to make therapy more comfortable. If you are getting a dry nose or mouth, turn up the humidity. If you are getting any moisture in your mask, turn down the humidity.
You can set the Humidity Level to Off or between 1 and 8, where 1 is the lowest humidity setting and 8 is the highest humidity setting.
To adjust the Humidity Level:
- In My Options, turn the dial to highlight Humidity Level and then press the dial.
- Turn the dial to adjust the humidity level and press the dial to save the change.
To change to Climate Control Auto:
- In My Options, turn the dial to highlight Climate Ctrl and then press the dial.
- Turn the dial to select Auto and then press the dial to save the change.
Caring for your ClimateLineAir
It is important that you regularly clean your air tubing to make sure you receive optimal therapy. The following sections will help you with disconnecting, cleaning, checking and reconnecting your air tubing.
Disconnecting
- Hold the cuff of the air tubing, apply gentle pressure to the release buttons and pull it away from the device.
- Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart.
Note: Only hold and pull the cuff of the air tubing. Do not hold or pull the tubing itself as it may cause damage.
Cleaning
You should clean the air tubing weekly as described.
- Wash the air tubing in warm water using mild detergent. Do not wash in a dishwasher or washing machine.
- Rinse the air tubing thoroughly and allow to dry, out of direct sunlight and/or heat.
Checking
You should regularly check the air tubing for any damage. If there are any holes, tears or cracks you should replace it.
Reconnecting the air tubing
When the air tubing is dry, you can reconnect it to the device.
- Connect the air tubing firmly to the air outlet located on the rear of the device.
- Connect the free end of the air tubing firmly onto the assembled mask.
Troubleshooting
If you have any problems, refer to the troubleshooting section of your device user guide or www.resmed.com. For any serious incidents that occur in relation to this device, these should be reported to ResMed and the competent authority in your country.
Technical specifications
Environmental conditions
- Operating temperature: +5°C to +35°C. Note: The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme ambient temperature conditions (40°C) the device remains safe.
- Operating humidity: 10 to 95% relative humidity, non-condensing
- Operating altitude: Sea level to 2,591 m; air pressure range 1013 hPa to 738 hPa
- Storage and transport temperature: -20°C to +60°C
- Storage and transport humidity: 5 to 95% relative humidity, non-condensing
Electromagnetic compatibility / Air tubing resistance to flow and compliance values
Refer to the device User Guide or ResMed.com.
Temperature
- Temperature range: +16 to +30°C
- Cut-out: +41°C
Operating pressure range
- Maximum recommended pressure: 30 cm H2O (30 hPa)
Physical
Air tubing | Material | Length | Inner diameter |
---|---|---|---|
ClimateLineAir | Flexible plastic and electrical components | 2 m | 15 mm |
ClimateLineAir Oxy | Flexible plastic and electrical components | 1.9 m | 19 mm |
Design Life
- ClimateLineAir, ClimateLineAir Oxy: 6 months
Symbols
The following symbols may appear on the product or packaging.
- Manufacturer: ?
- EC REP: European Authorized Representative.
- REF: Catalog number.
- LOT: Batch code.
- Rx Only: Prescription only (In the US, Federal law restricts these devices to sale by or on the order of a physician).
- Humidity: ?
- Temperature limitation: ?️
- Atmospheric pressure limitation: ?️
- China pollution control logo 1: ♻️
- China pollution control logo 2: ♻️
- Importer: ?
- Medical device: MD
See symbols glossary at ResMed.com/symbols.
Environmental information
This device should be disposed of separately, not as unsorted municipal waste. To dispose of your device, you should use appropriate collection, reuse and recycling systems available in your region. The use of these collection, reuse and recycling systems is designed to reduce pressure on natural resources and prevent hazardous substances from damaging the environment.
If you need information on these disposal systems, please contact your local waste administration. The crossed-bin symbol ?️ invites you to use these disposal systems. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to ResMed.com/environment.
Limited warranty
Refer to your device user guide.
File Info : application/pdf, 8 Pages, 946.09KB
DocumentDocumentReferences
Resmed | The World Leader in Sleep Health
Environmental policy | Resmed
ResMed.com/symbols
Resmed | The World Leader in Sleep Health
Related Documents
![]() |
ResMed ClimateLineAir 11 User Guide This user guide provides instructions for setting up, using, and caring for the ResMed ClimateLineAir 11 heated tubing system, designed to enhance CPAP therapy comfort by maintaining consistent temperature and humidity. |
![]() |
ResMed AirSense 11 CPAP Machine: Advanced Sleep Therapy Discover the ResMed AirSense 11, a state-of-the-art CPAP device designed for successful sleep therapy. Featuring personalized support, integrated coaching, and advanced therapy modes like AutoSet and AutoSet for Her, the AirSense 11 enhances patient comfort and adherence. Learn about its features, comparison with other ResMed models, and how it integrates with the myAir app and AirView system for comprehensive sleep apnea management. |
![]() |
ResMed ClimateLineAir 11 Heated Tubing User Guide This user guide provides instructions for setting up, using, cleaning, and troubleshooting the ResMed ClimateLineAir 11 heated tubing for the Air11 system. It includes technical specifications and warranty information. |
![]() |
ResMed myAir App: Setup Guide for AirSense and AirCurve Devices A comprehensive guide to creating and activating your ResMed myAir account and entering device information for AirSense 10 and AirSense 11 PAP devices to track sleep data and receive coaching. |
![]() |
ResMed AirCurve 11 User Guide Comprehensive user guide for the ResMed AirCurve 11 series of positive airway pressure devices, including VAuto, S, and ASV models. Learn about setup, operation, maintenance, troubleshooting, and technical specifications. |
![]() |
ResMed Oximeter Adapter User Guide User guide for the ResMed Oximeter Adapter, detailing its intended use, setup, troubleshooting, and technical specifications for compatibility with AirSense, AirCurve, Lumis, and S9 devices. |
![]() |
ResMed Corp Product List A comprehensive product list from ResMed Corp, detailing various respiratory support devices, masks, tubing, and accessories with item numbers, descriptions, units of measure, and HCPCS codes. |
![]() |
AirSense 11 AutoSet Elite CPAP Machine User Guide Comprehensive user guide for the ResMed AirSense 11 AutoSet Elite CPAP machine, covering setup, operation, maintenance, troubleshooting, and features for treating obstructive sleep apnea. |