Apera Instruments WS100 Fluoride/pH Portable Meter Kit

Instruction Manual

1 Overview

Thank you for purchasing the Apera Instruments WS100 Fluoride/pH Portable Meter Kit (Hereon after referred as the meter). This fluoride/pH combo meter is widely used for drinking water, surface water and other water treatment applications, ideal for both in-field and laboratory testing. Before using the product, please read this manual carefully before use to help you properly use and maintain the instrument.

For technical support, please contact us at info@aperainst.de or +49 202 51988998

1.1 Basic Functions

1.2 Features of Fluoride Measurement

1.3 Features of pH measurement

2 Technical Parameters

2.1 Fluoride

Measuring range0.02 ppm to 1900 ppm
Display unitsmg/L, ppm, pF
Accuracy±0.02 ppm or ±5% of reading (which ever is greater)
Temperature Compensation Range(0-80) °C (Automatic)

2.2 pH

Measuring range(-2.00 - 19.99) pH
Resolution0.1/0.01 pH
Accuracy±0.01 pH ± 1 digit
Temperature Compensation Range(0 – 100) °C (Automatic)

2.3 mV

Measuring range-1999mV ~ 0 ~ 1999mV
Resolution1mV
Accuracy±0.1% FS

2.4 Other technical parameters:

Data Storage300 Groups
Storage ContentNumberings, measurements, units of measurement and a temperature value
Power SupplyAA Battery (1.5VX2)
Size and WeightInstrument: (65X120X31) mm/180g; Case: (360X270X76) mm/1.5Kg
Quality and Safety CertificationISO9001:2015, CE
IP ClassIP57 Dustproof and waterproof

3 Instrument Description

3.1 LCD Display:

1) Parameter mode; 2) Measuring value; 3) Measured value storage and display of numbers, symbols indicate: M+ -- Store Measured value; RM - Recall measurement values Icon; 4) Measurement unit; 5) Temperature measurement and unit; 6) Smiley face ? icon indicates stable readings; 7) Instrument calibration indicator; 8) Low battery reminder.

3.2 Operation keys

3.3 Data Storage, Recall Data, and Clear Data:

3.3.1 Data Storage

While in the measurement mode, when the measuring data is stable and the smiley face icon ? appears, short press [M+] / [RM] key (<1.5s), LCD will display “M+” icon and a storage serial, and meanwhile storing all of the measuring information. It can store up to 300 groups.

3.3.2 Display measuring information:

(a) While in the measurement mode, long press [M+] / [RM] (>2s), meter will display the last stored information, and the storage number and “RM” icon will appear in the lower right corner of the LCD, and the complete measuring information. Again short press [CAL] or [M+] / [RM] meter will in turn display all the measuring data, Hold [CAL] or [RM] to quickly check the measuring data. (b) In the data displaying mode (there are “RM” and storage number in the upper right corner of the LCD), press [ENTER] key to return to the measuring mode.

(c) Clear the stored measuring information: In the data displaying mode, press the [ENTER] key for 5 seconds, LCD will appear CLr for 2 seconds. It means that all the storage has been eliminated and returns to measuring mode.

4 Fluoride concentration measurement

4.1 Preparation

4.2 Calibration

4.3 Sample Measurement

4.4 Parameter Settings

SymbolParameter sets the projectCodeParameters
P1Temperature unit selection°C / °F
P2Backlight display time setting0-1-3-6 min
P3Auto power off time setting0-10-20 min
P4Restore to factory default settingsOFF-ON

4.5 Notes

5 pH Measurement

5.1 Preparing the Meter

Connect the 201T-S pH Electrode to the meter; (pay attention to the inserting direction); loosen the cap of the storage bottle of electrode; screw off the storage bottle; Press [ON/OFF] to turn on meter. The meter automatically switches to pH measurement mode.

5.2 Meter Calibration

5.3 Sample Measurement

Soak the pH electrode into the sample solution after rinsing and drying, hold the electrode still. Take the readings when the smiley face icon ? shows up and stays on screen (indicates a stable reading).

pH isothermal measurement principle: According to the theory of Isothermal measurements, the smaller the difference of the temperatures between test solution and the calibration solution, the higher the measuring accuracy is. Typically, temperature difference between test solution and calibration solution should ≤±15°C; for high accuracy measurement, the temperatures should be the same.

5.4 Parameter Settings

5.4.1 pH measuring parameter setting (Chart (5-1))

Prompt MarkParameter Setting ItemsCodeParameters
P1pH buffer solution series selectionSOLUSA(U.S.A series) NIS (NIST series) CH (China series)
P2Purified water pH temperature compensation settingPU1OFF-On
P3Ammonia added purified water pH temperature compensation settingPU2OFF-On
P4Temperature unit setting°C / °F
P5Back light display time settingBL0-1-3-6min
P6Auto power off settingAC0-10-20min
P7Restore to Factory DefaultOFF-On

5.4.2 pH buffer solution series selection (P1)

(a) Press [MODE] key, meter enters P1 mode. (b) Press [CAL] or [RM] key to choose buffer solution series: U.S.A series —1.68, 4.00, 7.00,10.01 and 12.45 pH; NIST series —1.68, 4.01, 6.86, 9.18 and 12.45 pH; China series —1.68, 4.00, 6.86, 9.18 and 12.46 pH. (c) Press [MODE] key to enter next parameter setting or press [ENTER] key to confirm and return to measuring mode.

5.4.3 Purified water pH temperature compensation setting (P2)

(a) Short press [MODE] key in the mode P1, the meter enters mode P2. (b) Press [CAL] or [RM] to choose ON or OFF. (c) Press [MODE] key to enter next parameter setting or press [ENTER] key to confirm and return to measuring mode. (d) The default setting is "OFF". Note: A PU-1 icon will appear in the upper right corner of the LCD if purified water temperature compensation function is set by the user.

5.4.4 Ammonia added purified water pH temperature compensation setting (P3)

(a) Short press [MODE] key in mode P2 to enter mode P3. (b) Press [CAL] or [RM] key to choose ON or OFF. (c) Press [MODE] to enter next parameter setting or press [ENTER] key to confirm and return to measuring mode. (d) The default setting is "OFF". Note: A PU-2 icon will appear in the upper right corner of the LCD if purified water temperature compensation function is set by the user.

5.4.5 Temperature unit °C/°F setting (P4)

(a) Short press [MODE] key in mode P3 to enter mode P4. (b) Press [CAL] or [RM] key to choose temperature unit: °C or °F. (c) Press [MODE] key to enter next parameter setting or press [ENTER] key to confirm and return to measuring mode.

5.4.6 Backlight display time setting (P5)

(a) Short press [MODE] key in mode P4 to enter mode P5. (b) Press [CAL] or [RM] key to choose the time of back light auto power off: 0 min, 1 min, 3 min or 6 min. The back light display function will be closed if choosing a 0min. (c) Press [MODE] key to enter next parameter setting or press [ENTER] key to confirm and return to measuring mode. (d) The producer setting for P5 is 1min.

5.4.7 Auto power off time setting (P6)

(a) Short press [MODE] key in mode P5 to enter mode P6. (b) Press [CAL] or [RM] key to choose the time: 0min, 10min or 20min. The auto power off function will be closed if choosing 0min. (c) Press [MODE] key to enter next parameter setting or press [ENTER] key to confirm and return to measuring mode. (d) The producer setting for P6 is 10min.

5.4.8 Restore to producer setting (P7)

(a) Short press [MODE] key in mode P6 to enter mode P7. (b) Press [CAL] or [RM] key to choose "ON", means that all the pH measuring parameters have been restored to the factory default setting.

5.5 Notes

For other special contaminants, refer to Table-8 Clean contaminated glass bulb and junction:

ContaminationCleaning Solutions
Inorganic metal oxideDilute acid less than 1mol/L
Organic lipidDilute detergent (weak alkaline)
Resin macromoleculeDilute alcohol, acetone, ether
Proteinic haematocyte sedimentAcidic enzymatic solution (saccharated yeast tablets)
PaintsDilute bleacher, peroxide

(a) Every pH electrode gradually loses its sensitivity and will eventually fail. The lifespan period for the 201T-S pH Electrode is about 1-2 years depending on how it's used and maintained. (b) When an abnormal reading appears during calibration or measurement, please set P7 to "ON" to restore the meter to Factory Default settings, then re-calibrate the meter before measurement. (c) Electrodes that have been used for a long period of time will become aged. Submerge the electrode in 0.1mol/L hydrochloric acid for 24 hours, then rinse the electrode in distilled water, then submerge it in 3M KCL soaking solution for 24 hours. The method to prepare 0.1mol/L hydrochloric acid: dilute 9mL hydrochloric acid in distilled or deionized water to 1000mL.

5.6 Self-diagnose Information

When using the meter, the following icons might appear. This is the meter's self-diagnose information, which helps users identify problems occurring with the meter or the electrode:

a) A still display of -2.00 pH or 19.99 pH – this icon appears when the value has surpassed the measurement range. There will also appear such signs when the electrode is not well connected with the meter or when the electrode is not immerged into the solution. This is a normal phenomenon.

b) Err1 — Abnormal electrode zero potential (<-60mV or >60mV)

c) Err2 — Abnormal electrode slope (< 85% or >105%)

When “Err1” or “Err2” appears, the meter cannot work properly, please take the following steps to check and fix the problem:

a) Check if the electrode bulb has air bubble, if has, please shake the electrode with force for a few times to eliminate the air bubble inside the sensor.

b) Check the quality of buffer solution, if it is expired or contaminated. If so, please replace with new ones.

c) Set the meter to factory default mode then recalibrate it.

d) If the meter still cannot recover to the normal state after doing the above steps, please replace a new pH electrode.

6 ORP Measurement

6.1 Sample Test

7 Complete Kit

8 Warranty

Apera Instruments warrants this instrument to be free from defects in material and workmanship. Apera Instruments (Europe) GmbH agrees to repair or replace, free of charge, any malfunctioned or damaged product attributable to its responsibility for a period of two years from the delivery (a six-month limited warranty applies to probes). This warranty does not apply to defects resulting from actions such as misuse (violation of the instructions in this manual or operations in the manner not specified in this manual), accidental damage, improper maintenance, or unauthorized repairs. Warranty period is the time limit to provide free service for the products purchased by customers, not the service life of the tester or probe.

9 Appendix 1. Fluoride Concentration Conversion Table

mol/LpFmg/Lppm
1 ×10-33.0019.0019.00
1 ×10-44.001.901.90
1 ×10-55.000.190.19

APERA INSTRUMENTS (Europe) GmbH
Wilhelm-Muthmann-Straße 18
42329 Wuppertal, Germany
Tel. +49 202 51988998
Email: info@aperainst.de

Models: WS100 Fluoride pH Portable Meter Kit, WS100, WS100 Portable Meter Kit, Fluoride Portable Meter Kit, pH Portable Meter Kit, Portable Meter Kit, Meter Kit

File Info : application/pdf, 14 Pages, 364.04KB

PDF preview unavailable. Download the PDF instead.

WS100-EN

References

GPL Ghostscript 9.50 ����

Related Documents

Preview Apera PH700 Benchtop pH Meter Instruction Manual
Comprehensive instructions for operating, calibrating, and maintaining the Apera PH700 Benchtop pH Meter. It covers technical specifications, instrument description, measurement procedures for pH and ORP, parameter settings, electrode care, and troubleshooting.
Preview Apera SX811-SS Portable pH Meter for Food Testing: Instruction Manual & Specifications
Comprehensive instruction manual and technical specifications for the Apera SX811-SS portable pH meter, designed for accurate food testing. Learn about features, calibration, setup, USB communication, and troubleshooting.
Preview Apera Instruments PC60 Premium 5-in-1 Tester User Manual
Comprehensive user manual for the Apera Instruments PC60 Premium 5-in-1 Tester, covering pH, Conductivity, TDS, Salinity, and Temperature measurements. Includes setup, calibration, operation, maintenance, and troubleshooting.
Preview PH60Z-HT Smart pH Meter User Manual - Apera Instruments
User manual for the Apera Instruments PH60Z-HT Smart pH Meter, detailing its features, operation, calibration, maintenance, and troubleshooting for high temperature and caustic solutions.
Preview Apera PH20 Value pH Tester Instruction Manual
Comprehensive instruction manual for the Apera PH20 Value pH Tester, covering battery installation, calibration, measurement, parameter settings, technical specifications, and warranty information.
Preview Apera PH700 Benchtop pH Meter Instruction Manual
Comprehensive instruction manual for the Apera PH700 Benchtop pH Meter, covering features, specifications, operation, calibration, maintenance, and troubleshooting.