Apera PH700 Benchtop pH Meter Instruction Manual
1 Brief Introduction
Thank you for purchasing the Apera Instruments PH700 Benchtop pH meter. This instrument combines advanced electronics design and sensor technology, offering an economical and reliable lab-grade benchtop pH meter. Please read this manual carefully for proper use and maintenance. Apera Instruments reserves the right to update the manual's contents without prior notice.
1.1 Main Features
- The meter features a built-in microprocessor chip enabling intelligent functions such as automatic calibration, automatic temperature compensation, data storage, parameter setup, and max/min reading display.
- The advanced automatic calibration mode includes a calibration guide and self-diagnosis functions, automatically recognizing standard buffer solutions (USA or NIST series).
- Advanced digital processing technology enhances response time and measurement accuracy with a stable reading display mode.
- The kit includes a 3-in-1 combination pH Electrode that measures pH and temperature simultaneously.
- The instrument is IP54 dustproof and spillproof, with connectors protected by silicone seal caps for reliability in harsh environments.
- A detachable electrode holder and three bottles of pH buffer solutions are included for convenient use.
2 What's in the Box?
Content | Quantity |
---|---|
PH700 Benchtop pH Meter | 1 |
201T-F Plastic 3-in-1 Combination pH Electrode | 1 |
pH standard buffer solutions (4.00/7.00/10.01 pH) | 1 for each 50ml |
9V Power Adapter | 1 |
Electrode Holder | 1 set |
Instruction Manual | 1 |
3 Technical Specifications
Parameter | Specification | |
---|---|---|
pH | Range | 0 to 14.00 pH |
Resolution | 0.01/0.1 pH | |
Accuracy | ±0.01 pH ±1 digit | |
Temperature Compensation | 0 to 100 °C (Auto or Manual) | |
mV | Range | ±1999 mV |
Resolution | 1 mV | |
Accuracy | ±0.1% F.S ±1 digit | |
Temp. | Range | 0 to 100°C; 32 to 212°F |
Resolution | 0.1°C; 0.1/1°F | |
Accuracy | ±0.5°C ±1 digit | |
Others | Data Storage | 50 groups |
Storage Content | Numberings, Measurement, Unit, Temperature, Temperature Compensation Status | |
Power | DC9V/300mA | |
IP Ratings | IP54 dustproof and spillproof | |
Size and Dimension | (240*235*103) mm/1kg |
4 Instrument Description
4.1 LCD Display
The LCD display shows various measurement values, icons, and status indicators. Key elements include:
- (1) Measurement value: Displays the current pH or mV reading.
- (2) Calibration icons and numberings: Indicates calibration status and step.
- (3) Measurement unit: Shows the unit (pH or mV).
- (4) Temperature and reminder icons: Displays temperature and other status reminders.
- (5) pH unit during calibration: Shows the pH value being calibrated.
- (6) pH value during calibration: Displays the calibration buffer value and storage/recall numberings.
- (7) Data storage and recall icons: M+ for storing data, RM for recalling data.
- (8) Self-Diagnosis icons and numberings: Indicates error codes or diagnostic information.
- (9) Temperature compensation mode icons: ATC for Auto Temperature Compensation, MTC for Manual Temperature Compensation.
- (10) Completed calibration icons: Indicates successful calibration steps.
- (11) Stable reading icon: A smiley face ? indicates a stable measurement.
- (12) Icons for maximum and minimum values: Indicates max/min reading mode.
Diagram Description: Figure-1 illustrates the Apera PH700 pH meter's LCD display with numbered indicators pointing to different functional areas and icons.
4.2 Keypad
The keypad provides control for all meter functions. Operations are categorized by short press (<1.5 seconds) and long press (>1.5 seconds).
Keypad | Operation | Functions |
---|---|---|
POWER | Short Press | Power on/off |
MODE SETUP | Short Press | Select measuring mode: pH → mV. |
Long Press | Enter parameter setting: P1 → P2 → ... → P4. | |
CAL MEAS | Short Press | Enter Calibration mode |
Short Press | Cancel any operation and go back to measurement mode | |
ENTER | Short Press | In calibration mode, press to conduct calibration. In parameter setting, press to confirm selection. |
M+ RM | Short Press | Store measuring data |
Long Press | Recall stored measuring data | |
▲ / ▼ | Short Press | In manual temperature compensation mode, short press to change temperature; in parameter setting mode, short press to change parameter and settings; in recall mode, short press to change numberings. |
Long Press | In manual temperature compensation mode, long press to change temperature quickly; in parameter setting mode, long press to change settings quickly; in recall mode, long press to change numberings quickly. | |
MAX MIN | Short Press | Buttons for max/min values (refer to section 4.6). |
4.3 Connectors
The meter features various connectors for electrodes and power:
Symbol | Connector Name | Connector Type |
---|---|---|
REF | Socket for reference electrode | Φ2 banana type |
pH/mV | Socket for pH and ORP electrode | BNC |
TEMP | Socket for temperature sensor | RCA |
DC9V | DC9V power socket | Φ2.5 direct type |
Diagram Description: An illustration shows the rear panel of the pH meter with labeled sockets for REF, pH/mV, TEMP, and DC9V power input.
4.4 Stable reading display mode
When a measurement is stable, a smiley face icon ? appears on the LCD. Calibration or recording should only be performed when this icon is visible and remains steady.
Figure-3 shows the LCD displaying a stable pH reading with the smiley face icon ?.
Figure-4 shows the LCD after data has been stored, displaying the M+ icon and storage numberings.
4.5 Data storage, recall, and deletion
Storage: When the measurement is stable, press the M+ RM key. The meter displays the M+ icon and storage numberings, saving the current measurement. Pressing M+ RM again continues storing data.
Recall stored data: In measurement mode, long press the M+ RM key to recall the last stored measurement. The meter displays the RM icon and storage numberings. Use the ▲ and ▼ keys to scroll through stored data. Press and hold ▲ or ▼ to scroll quickly. Short press CAL MEAS to return to measurement mode.
Clear stored data: The meter can store up to 50 groups of data. If storage is full, pressing M+ RM will display a FUL icon. Data can be cleared via parameter setting P4.
Figure-5 shows the LCD displaying recalled data with the RM icon and a specific data number (e.g., 28).
Figure-6 shows the LCD indicating that the storage is full, displaying the FUL icon.
The numberings for M+ indicate the total stored data across all modes. The numberings for RM indicate data stored within the current mode (pH or mV). For example, if 20 groups are stored in pH mode and 10 in mV mode, M+ will show "M+30", RM in pH mode will be "RM20", and in mV mode "RM10".
4.6 Max/Min Function
This function displays the maximum or minimum recorded values from a series of measurements. In measurement mode, press the MAX MIN key; the LCD will show flashing MAX/MIN icons. After testing, pressing MAX MIN again will alternately display the maximum and minimum values. Press CAL MEAS to return to regular measurement mode.
5 pH Measurement
5.1 Preparation
Install the electrode holder
The electrode holder consists of a metal stand base, a supporting pole, and an electrode clip. Attach the metal stand base to the meter (screw it on), ensuring it can be installed on either side. Then, attach the supporting bar to the base and the electrode clip to the supporting bar.
Connect the power adapter
Plug the power adapter into the DC9V socket. Ensure the adapter's voltage meets the meter's requirements.
Things needed in addition to what's included in the box
A clean cup, distilled water (8-16oz), and tissue papers for rinsing and drying the probe are required.
5.2 pH Electrode Information
pH Electrode
The meter is supplied with a 201T-F 3-in-1 combination electrode featuring a built-in temperature sensor for automatic temperature compensation. This electrode is suitable for regular water solutions. For specific applications, refer to Section 7. The electrode housing is made of polycarbonate, offering resistance to shocks and corrosion. A 3M KCL storage bottle is provided to maintain electrode sensitivity when not in use.
Technical Specifications of the 201T-F pH Electrode
- Measurement Range: 0 - 14 pH, 0 - 80°C (32 – 176°F)
- Junction: Single Ceramic
- Reference Electrode: Ag/AgCl
- Connector: BNC/RCA
- Size: Ø12*160 mm
- Temperature unit: 30K Thermistor
Electrode Connectors
The pH electrode has two connectors: the BNC connector for the pH probe and the RCA connector for the temperature sensor. Connect these to the 'pH/mV' and 'TEMP' sockets respectively. Avoid pulling the cables to prevent poor contact. Keep connectors clean and dry. Refer to section 4.7 for maintenance.
Use the Electrode
Remove the electrode from its KCL storage bottle (do not spill the solution). Rinse the electrode in distilled water for a few seconds. Gently shake off excess water and pat dry with clean tissue paper (do not rub). Stir the electrode gently in the sample solution for a few seconds, then let it stand still until a stable reading (indicated by ? icon) is achieved. Record the reading. After testing, place the electrode back into the storage bottle and tighten the cap to preserve its sensitivity.
5.3 Information regarding Calibration
pH buffer solutions
The meter supports two series of standard buffer solutions: USA and NIST, selectable via parameter P1 (see 4.6.1). Calibration can be performed using 1, 2, or 3 points.
Calibration icon | pH standard buffer series (USA) | pH standard buffer series (NIST) |
---|---|---|
[L] | 1.68 and 4.00 pH | 1.68 and 4.01 pH |
[M] | 7.00 pH | 6.86 pH |
[H] | 10.01 pH | 9.18 pH |
Table-2 lists the standard buffer solutions for USA and NIST series, associated with calibration icons [L], [M], and [H].
3-Point Calibration
Calibration can be performed using 1, 2, or 3 points. The first point must use the 7.00 pH (or 6.86 pH) buffer. Subsequent points use other buffers. The calibration process displays the slope of acidity and alkalinity ranges, indicating probe health; a slope closer to 100% signifies better probe condition.
Calibration | USA | NIS | Calibration icon | When to adopt |
---|---|---|---|---|
1-Point Calibration | 7.00 pH | 6.86 pH | [M] | accuracy ≤ ±0.1 pH |
2-Point Calibration | 7.00 pH and 4.00/1.68 pH | 6.86 pH and 4.01/1.68 pH | [L] [M] | Measuring range: 0 to 7.00 pH |
7.00 pH and 10.01 pH | 6.86 pH and 9.18 pH | [M] [H] | Measuring range: >7.00 pH | |
3-Point Calibration | 7.00pH, 4.00/ 1.68 pH and 10.01 pH | 6.86pH, 4.01/ 1.68pH, 9.18 pH | [L] [M] [H] | Wide measuring range |
Table-3 details the calibration modes, corresponding buffer solutions, icons, and recommended usage scenarios.
How often to calibrate
Calibration frequency depends on the samples, electrode condition, and accuracy requirements. For high-accuracy measurements (≤ ±0.02pH), calibrate before each test. For ordinary-accuracy measurements (≥ ±0.1pH), calibration is typically valid for about a week. Recalibration is necessary if:
- The electrode has not been used for a long time or a new electrode is connected.
- After measuring strong acid (pH<2) or strong base (pH>12) solutions.
- After measuring fluoride-containing or strong organic solutions.
- There is a significant temperature difference between the test sample and the buffer solution used in the last calibration.
5.4 pH Meter Calibration
Follow these steps for pH meter calibration:
- Press CAL MEAS to enter calibration mode. The CAL1 icon and '7.00 pH' will flash, indicating the first calibration point.
- Rinse the electrode with distilled water and pat dry. Dip it into the pH 7.00 buffer solution, stir gently, and wait for a stable reading (indicated by ?). The meter will show the buffer recognition process. Pressing ENTER before recognition may result in an 'Er2' error.
- When the meter recognizes 7.00 pH and the ? icon appears, press ENTER to calibrate. An 'End' icon confirms completion. The CAL2 icon will flash, and '4.00 pH'/'10.01 pH' will alternate, prompting for the second calibration point.
- Rinse and dry the electrode. Dip it into the pH 4.00 buffer solution, stir gently, and wait for a stable reading. When 4.00 pH is recognized (? icon appears), press ENTER to calibrate. The 'End' icon and the electrode slope will be displayed. The CL3 icon will flash, and '10.01 pH' will appear, prompting for the third calibration point.
- Rinse and dry the electrode. Dip it into the pH 10.01 buffer solution, stir gently, and wait for a stable reading. When 10.01 pH is recognized (? icon appears), press ENTER to calibrate. The 'End' icon and the electrode slope of alkalinity range will be displayed. The meter returns to measurement mode with calibration icons L M H shown.
- Press CAL MEAS at any time to exit calibration mode. The meter supports one-point, two-point, and three-point calibrations.
Figure-7 illustrates the LCD display sequence during a 3-point calibration process.
5.5 Measuring pH
To measure pH:
- Remove the electrode from its storage bottle. Rinse with distilled water and pat dry.
- Gently stir the electrode in the sample solution for a few seconds, then let it stand still.
- Wait for the stable measurement indicator (? icon) to appear and remain steady.
- Record the reading.
Figure-8 provides a flowchart outlining the calibration and measurement process.
Notes:
- The default 201T-F electrode may not provide accurate readings for purified water (distilled or deionized) due to low ion concentration. For such samples, a specialized electrode like the LabSen™ 803 Pure Water pH Electrode is recommended. Contact Apera Instruments for details. Testing purified water like spring or drinking water may require longer stabilization times (3-5 minutes) due to fewer ions.
- Distilled or deionized water is recommended only for rinsing the probe.
- The 201T-F electrode is not suitable for high-temperature samples (>176°F), as it may cause permanent damage. Use specialized electrodes like LabSen™ 213 (up to 225°F) for such applications.
- The electrode housing is polycarbonate. Avoid cleaning agents like carbon tetrachloride, trichlorethylene, tetrahydrofuran, or acetone, as they can dissolve the housing.
For best results, the test sample's temperature should be close to the calibration solution's temperature. For samples at 150°F, warm calibration solutions to the same temperature before calibrating.
The instrument has a factory default reset function (parameter P5, section 5.6). This restores theoretical values (pH 7.00, slope 100%) and default settings. Use this function if calibration or measurement is abnormal. This reset is irreversible.
5.5.3 Self-Diagnosis
The meter performs self-diagnosis during calibration and measurement. Error codes and their solutions are listed below:
Display Icons | Contents | How to fix |
---|---|---|
Er1 | Wrong pH buffer solution or the buffer solution is out of range. | 1. Check pH buffer solution correctness. 2. Ensure proper electrode connection. 3. Check electrode for damage. 4. Ensure no air bubble in the electrode's bulb. |
Er2 | pH electrode zero electric potential out of range (<-60mV or >60mV) | 1. Ensure no air bubble in the electrode's bulb. 2. Check electrode for damage. 3. Check pH buffer solutions. 4. The pH Electrode is aged; replace it. |
Er3 | pH electrode slope is out of range (<85% or >110%) | 1. Ensure no air bubble in the electrode's bulb. 2. Check electrode for damage. 3. The pH Electrode is aged; replace it. |
Er4 | Press ENTER key when reading is not stable during calibration. | Press ENTER key only when the ? icon appears and stays on the screen. |
Er5 | During calibration, the measuring value is not stable for ≥3min. | 1. Ensure no air bubble in the electrode's bulb. 2. Check electrode for damage. 3. The pH Electrode is aged; replace it. |
Note: An air bubble in the pH sensor's glass bulb can significantly decrease measurement stability. Gently shake the probe to remove it.
5.6 Parameter Setting
The meter allows adjustment of various parameters:
Symbol | Parameter setting content | Parameter | Factory default setting |
---|---|---|---|
P1 | Select pH buffer series | USA - NIST | USA |
P2 | Select resolution | 0.1/0.01 pH | 0.01 pH |
P3 | Select temperature unit | °C - °F | °C |
P4 | Clear stored data | No-Yes | No |
P5 | Return to factory default | No-Yes | No |
How to do the parameter setting:
From measurement mode, long press MODE SETUP to enter parameter setting P1. Use ▲ or ▼ to navigate through P1 to P5.
Parameter | Procedure |
---|---|
P1 -- Select pH buffer series (USA— NIS) | 1. Long press MODE SETUP to enter P1. 2. Press ENTER. 'USA' flashes. Press ▲ or ▼ to select USA or NIS, then press ENTER to confirm. 3. Press CAL MEAS to enter P2 or return to measurement mode. |
P2 -- Select resolution (0.01—0.1) | 1. Press ENTER. '0.01' flashes. Press ▲ or ▼ to select 0.01 or 0.1, then press ENTER to confirm. 2. Press CAL MEAS to enter P3 or return to measurement mode. |
P3 -- Select temperature unit (°C-°F) | 1. Press ENTER. '°C' flashes. Press ▲ or ▼ to select °C or °F, then press ENTER to confirm. 2. Press CAL MEAS to enter P4 or return to measurement mode. |
P4 -- Clear stored data (No-Yes) | 1. Press ENTER. 'No' flashes. Press ▲ or ▼ to select No or Yes, then press ENTER to confirm. 2. Press CAL MEAS to enter P5 or return to measurement mode. |
P5 -- Restore to factory default setting (No-Yes) | 1. Press ENTER. 'No' flashes. Press ▲ or ▼ to select No or Yes, then press ENTER to confirm. The meter returns to measurement mode. 2. If 'Yes' is not selected, press CAL MEAS to return to measurement mode. |
5.7 pH Electrode's Maintenance
5.7.1 Daily maintenance
The soaking solution in the protective bottle maintains electrode activation. Loosen the cap, remove the electrode, and rinse it in distilled water before measurement. Reinsert and tighten the cap after use. If the soaking solution becomes cloudy or moldy, replace it with fresh 3M KCL solution.
Note: Do not use other brands' soaking solutions, as incompatible chemicals may damage the electrode.
Note: The electrode should not be soaked or stored in purified/distilled water, protein solutions, acid fluoride solutions, or organic lipids. Distilled or purified water is only for rinsing.
5.7.2 Calibration buffer solution
For accurate calibration, use reliable buffer solutions. Refresh them frequently, especially after heavy use. It is recommended to replace buffers after 10-15 uses.
5.7.3 Protect glass bulb
The sensitive glass bulb should not contact hard surfaces, as scratches or cracks can cause inaccurate readings. Rinse the electrode with distilled or purified water before and after each measurement. If the electrode is contaminated, clean it with a soft brush and mild soap, rinse with distilled water, and soak in KCL solution for at least 6 hours.
5.7.4 Renew glass bulb
Aged electrodes can be renewed. Submerge the electrode in 0.1mol/L hydrochloric acid for 24 hours, rinse in distilled water, then soak in KCL solution for 24 hours. To prepare 0.1mol/L HCl: dilute 9mL HCl in distilled water to 1000mL. For serious passivation, briefly submerge the bulb in 4% HF for 3-5 seconds, rinse in distilled water, and soak in KCL solution for 24 hours.
5.7.5 Clean contaminated glass bulb and junction
Contamination | Cleaning Solutions |
---|---|
Inorganic metal oxide | Dilute acid (less than 1mol/L) |
Organic lipid | Dilute detergent (weak alkaline) |
Resin macromolecule | Dilute alcohol, acetone, ether |
Proteinic haematocyte sediment | Acidic enzymatic solution (saccharated yeast tablets) |
Paints | Dilute bleach, peroxide |
6 mV measurement
6.1 ORP measurement
To switch to mV measurement mode, press the MODE SETUP key. Connect an ORP electrode (e.g., 301Pt-C, sold separately) and dip it into the sample solution. Stir briefly and allow it to stabilize until the ? icon appears. ORP stands for Oxidation Reduction Potential, measured in mV.
6.2 Notes on ORP measurement
6.2.1 ORP measurement
ORP measurement does not require calibration. To verify electrode or meter performance, use an ORP standard solution. Table-9 provides data for a standard ORP solution at 222 mV.
°C | 10 | 15 | 20 | 25 | 30 | 35 | 38 | 40 |
---|---|---|---|---|---|---|---|---|
mV | 242 | 235 | 227 | 222 | 215 | 209 | 205 | 201 |
Table-9 shows the mV values of a standard ORP solution at various temperatures.
6.2.2 Clean and activate ORP electrode
Polluted platinum surfaces on ORP electrodes can cause inaccurate measurements and slow response. Follow these cleaning methods:
- For inorganic pollutant: Submerge the electrode in 0.1mol/L dilute hydrochloric acid for 30 minutes, rinse in distilled water, then soak in the KCL solution for 6 hours.
- For organic or lipid pollutant: Clean the platinum surface with detergent, rinse in distilled water, then soak in the KCL solution for 6 hours.
- For heavily polluted platinum surface (oxidation film): Polish the platinum surface with toothpaste, rinse in distilled water, then soak in the KCL solution for 6 hours.
7 Recommended pH Electrodes for Specific Applications
The following table lists Apera pH electrodes recommended for use with the PH700 meter for various applications:
Application | Ideal Apera pH Electrodes to Use with PH700 Meter |
---|---|
Regular water solutions | 201T-F, LabSen 213 |
Beverage, beer, or wine analysis | LabSen 213 |
Cosmetics | LabSen 851-1, (MP500 temp. probe required) |
Dairy products (milk, cream, yogurt, mayo, etc.) | LabSen 823 |
High-Temperature liquid | LabSen 213 |
Low-temperature liquid | LabSen881 (MP500 temp. probe required) |
Meat | LabSen 763 |
Micro sample testing | LabSen 241-6, LabSen 241-3 (MP500 temp. probe required) |
Purified Water (Low ion concentration samples) | LabSen 803, LabSen 813 |
Soil | LabSen 553 |
Solid or semi-solid samples (cheese, rice, fruit, etc.) | LabSen 753 |
Strong acid samples | LabSen 831 (MP500 temp. probe required) |
Strong alkalined samples | LabSen 841 (MP500 temp. probe required) |
Surface test (skin, paper, carpet, etc.) | LabSen 371 (MP500 temp. probe required) |
Titration | LabSen 223 |
TRIS buffer solutions | LabSen 213, LabSen 223 |
Viscous liquid samples | LabSen 223, LabSen851-1 |
Wastewater or emulsion | LabSen 333 |
8 Warranty
APERA INSTRUMENTS (Europe) GmbH warrants this instrument to be free from defects in material and workmanship for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery date. This warranty covers malfunctions or damage attributable to APERA INSTRUMENTS (Europe) GmbH. It does not cover damages due to transportation, storage, improper use, failure to follow instructions, unauthorized modifications, combination with unauthorized products, unauthorized repairs, normal wear and tear, or external causes beyond reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Wilhelm-Muthmann-Straße 18, 42329 Wuppertal, Germany
Email: info@aperainst.de | www.aperainst.de | Tel. +49 202 51988998