OMRON HEM-7600T Upper Arm Blood Pressure Monitor
Instruction Manual
Introduction
Thank you for purchasing the OMRON Upper Arm Blood Pressure Monitor. Your new upper arm blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope, making it simple to use.
Intended Use
This device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patients. The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. It is mainly designed for general household use.
Important Safety Information
Please read the Important Safety Information in this instruction manual before using the unit. Please follow this instruction manual thoroughly for your safety. Please keep for future reference. For specific information about your own blood pressure, consult your doctor.
Warning
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your doctor. Only a doctor is qualified to diagnose and treat high blood pressure.
DO NOT use this monitor on an injured arm or an arm under medical treatment.
DO NOT wear the arm cuff on your arm while on an intravenous drip or blood transfusion.
DO NOT apply or inflate the arm cuff around any body parts other than your upper arm.
DO NOT use this monitor in areas containing high frequency (HF) surgical equipment, magnetic resonance imaging (MRI) equipment, computerized tomography (CT) scanners or in oxygen-rich environments.
DO NOT use this monitor on infants, toddlers, children or persons who cannot express themselves.
Consult your doctor before using this monitor if you have common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation; arterial sclerosis; poor perfusion; diabetes; pregnancy; pre-eclampsia; or renal disease.
Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
This product contains small parts that may cause a choking hazard if swallowed by children, toddlers or infants.
Keep the BATTERY out of the reach of children, toddlers and infants.
This product emits radio frequencies (RF) in the 2.4 GHz band. DO NOT use this product in locations where RF is restricted, such as on an aircraft or in hospitals.
Caution
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
Stop using this monitor and consult your doctor if you experience skin irritation or discomfort.
Consult your doctor before using this monitor on an arm with an arterio-venous (A-V) shunt.
Note that PATIENT motion, trembling, shivering may affect the measurement reading.
Consult your doctor before using this monitor if you have had a mastectomy.
Consult your doctor before using the monitor if you have severe blood flow problems or blood disorders as cuff inflation can cause bruising.
Do not take measurements more often than necessary because bruising, due to blood flow interference, may occur.
Remove the arm cuff if it does not start deflating during the measurement.
Do not use this monitor for any purpose other than measuring blood pressure.
Do not use a mobile phone or other electrical devices that emit electromagnetic fields near this monitor except when transferring your readings to your smart device. This may result in incorrect operation of this monitor.
Do not disassemble or attempt to repair the monitor or components. This may cause an inaccurate reading.
Do not use in a location where there is moisture or a risk of water splashing the monitor. This may damage the monitor.
Do not use this monitor in a moving vehicle such as in a car or on an aircraft. Do not drop, or subject this monitor to strong shocks or vibrations.
Do not use this monitor in places with high or low humidity or high or low temperatures. See the "6. Specifications" section of the manual.
Please check, by observation of the arm being used, that the monitor is not causing any impairment of patient blood circulation.
Do not use this monitor in high-use environments such as medical clinics or doctor offices.
Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 30 minutes before taking a measurement.
Rest for at least 5 minutes before taking the measurement.
Remove tight-fitting or thick clothing from your arm while taking a measurement.
Remain still and do not talk while taking a measurement.
Read and follow the "Important information regarding Electro Magnetic Compatibility (EMC)" in the EMC information provided with this device.
Read and follow the "Correct Disposal of This Product" in "6. Specifications" when disposing of the device and any used accessories or optional parts.
Data Transmission
Do not replace the battery while your reading is being transferred to your smart device. This may result in the incorrect operation of your monitor and failure to transfer your blood pressure data.
Do not place integrated circuit cards, magnets, metal objects, or other devices that emit electromagnetic fields near this monitor while your reading is being transferred to your smart device. This may result in the incorrect operation of your monitor and failure to transfer your blood pressure data.
Battery Handling and Usage
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 "AAA" alkaline batteries with this monitor. Do not use other types of batteries. Do not use new and used batteries together.
Remove the batteries if the monitor will not be used for a long period of time.
If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.
If battery fluid should get on your skin, wash your skin immediately with plenty of clean, lukewarm water. If irritation, injury or pain persists, consult a doctor.
Do not use batteries after their expiration date.
Periodically check your batteries to ensure they are in good working condition.
Only use batteries specified for this monitor. Use of unsupported batteries may damage and/or may be hazardous to the monitor.
1. Know Your Device Contents
The package typically includes:
Monitor
Battery set
Instruction manual
Setup instructions
Monitor Components
The monitor features:
A. START/STOP button: [▶️ Start/Stop]
B. Connection button: [? Connect]
C. Display: [?️ Display Screen]
D. Arm cuff: [? Arm Cuff]
E. Battery compartment: [? Battery Compartment]
Display and Symbols
The display shows measurement values and status indicators:
F. Systolic blood pressure: [⬆️ Systolic]
G. Diastolic blood pressure: [⬇️ Diastolic]
H. Pulse display: [? Pulse]
Status Symbols:
Connection symbol: [? Bluetooth Connected]
Battery symbol (low): [? Low Battery]
Connection OFF symbol: [?? Bluetooth Off]
Battery symbol (depleted): [? Depleted Battery]
Connection ON symbol: [✅? Bluetooth On]
Movement error symbol: [? Movement Error]
SYNC symbol: [? Sync]
Deflation symbol: [⬇️ Deflating]
Heartbeat symbol: [❤️ Heartbeat] (Flashes during measurement.)
Irregular Heartbeat Symbol (??): When the monitor detects an irregular rhythm two or more times during the measurement, this symbol appears on the display with the measurement values. An irregular heartbeat rhythm is defined as a rhythm that is 25% less or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure. If this symbol displays with your measurement results, consult your doctor.
Movement Error Symbol (?): The movement error symbol is displayed if you move your body during the measurement. Remove the arm cuff, wait 2-3 minutes, and take another measurement, remaining still during the process.
Cuff Wrap Guide Symbol (??): If the cuff was wrapped too loosely, it may cause unreliable results. If the wrapping of the cuff is too loose, the cuff wrap guide symbol "?" appears. Otherwise, "?" appears. This function helps determine if the cuff is wrapped snugly enough.
2013 ESH/ESC Guidelines for Hypertension
Definitions of hypertension by office and home blood pressure levels:
Blood Pressure Type
Office
Home
Systolic Blood Pressure
140 mmHg
135 mmHg
Diastolic Blood Pressure
90 mmHg
85 mmHg
These values are based on statistical data for blood pressure.
1.2 Before Taking a Measurement
To help ensure an accurate reading, follow these directions:
Avoid bathing, drinking alcohol or caffeine, smoking, exercising, and eating for 30 minutes before taking a measurement.
Rest for at least 5 minutes before taking the measurement.
Stress raises blood pressure. Avoid taking measurements during stressful times.
Measurements should be taken in a quiet place.
Remove tight-fitting clothing from your arm.
Keep a record of your blood pressure and pulse readings for your doctor. A single measurement may not provide an accurate indication of your true blood pressure. Take and record several readings over a period of time. Try to measure your blood pressure at the same time each day for consistency.
2. Preparation
2.1 Battery Installation
Follow these steps to install batteries:
Remove the battery cover by gently slipping it into the gap of the hook and pulling downward.
Insert 4 "AAA" batteries as indicated into the battery compartment, ensuring correct polarity.
Replace the battery cover.
When batteries are securely placed, the message and SYNC symbol [?] will appear on the screen. Ensure to press the connection button [?] to pair with the smart device or transfer data when the SYNC symbol [?] appears.
Notes:
When the depleted battery symbol [?] appears, turn the monitor off and replace all 4 batteries with new ones at the same time. Long-life alkaline batteries are recommended.
Measurement values are stored in memory even after battery replacement.
Supplied batteries may have a shorter life than new ones.
Dispose of used batteries according to national/local regulations.
2.2 Pairing Your Monitor with a Smart Device
The date and time on your monitor will automatically be set when paired with your smart device. For first-time use of the "OMRON connect" app, visit omronconnect.com/setup for initial setup instructions.
Turn on the Bluetooth feature of your smart device.
Download and install the free "OMRON connect" app on your smart device. (If you already have the app, go to: Menu > Devices > Add Device).
Open the app and follow the setup and pairing instructions.
Confirm the monitor is connected successfully. "OK" will appear on the display if connected. If "Err" appears, refer to section 4.2.
Notes:
Pairing with the "OMRON connect" app sets the date and time on the monitor. All future measurements will include date and time upon transfer. If not paired, measurements will not have date/time.
Using an app other than "OMRON connect" may prevent automatic date/time setting.
OMRON is not responsible for data loss.
Refer to "Using your Monitor without Pairing" in section 3.3 if not pairing.
3. Using the Device
3.1 Applying the Arm Cuff
Remove tight-fitting clothing from your left upper arm. Do not place the arm cuff over thick clothes.
Apply the arm cuff to your left upper arm. The bottom edge should be 1 to 2 cm above the elbow. Positioning the display in clear view makes reading results easier.
Securely close with the fabric fastener.
Notes:
Blood pressure can differ between the right and left arm. OMRON recommends using the same arm for measurement. If values differ substantially, consult your doctor about which arm to use.
Positioning the monitor with the [▶️ Start/Stop] button towards the body makes it easier to use.
3.2 How to Sit Correctly
To take a measurement, be relaxed and comfortably seated at a comfortable room temperature.
Sit in a chair with legs uncrossed and feet flat on the floor.
Sit upright with your back straight.
Sit with your back and arm supported.
The arm cuff should be placed on your arm at the same level as your heart.
3.3 Taking a Measurement
Notes:
To stop measurement, press the [▶️ Start/Stop] button once to deflate the arm cuff.
Remain still and do not talk while taking a measurement.
This device stores 100 measurements. Stored measurements cannot be viewed on the monitor's display; use the "OMRON connect" app to view them.
Do not use the monitor with other electrical devices simultaneously, as this may cause incorrect operation.
Press the [▶️ Start/Stop] button. The arm cuff will start to inflate.
The monitor will automatically deflate the cuff and display readings.
During measurement, the display may show:
[? Cuff OK]: The arm cuff was wrapped snugly enough.
[? Cuff Loose]: The arm cuff was loose or not wrapped correctly.
The following symbols indicate incorrect measurement. If any appear, take another measurement:
[?? Irregular Heartbeat]
[? Movement Error]
[?? Cuff Wrap Guide] (if cuff is loose)
After measurement:
Remove the arm cuff.
Transfer your readings. Ensure your monitor is within 5 m (16 ft.) of your smart device and the "OMRON connect" app is active. Readings transfer automatically upon completion.
Transferring readings later:
Open the app on your smart device and follow instructions.
Press the connection button [?] to transfer readings.
If readings are successfully transferred, "OK" will flash on the display. If "Err" appears, refer to section 4.2.
Notes:
When the SYNC symbol [?] appears, transfer data immediately before readings are deleted.
When the Connection OFF symbol [??] appears, turn on the Bluetooth feature of your monitor (refer to section 3.4).
Once readings are sent to the app, they cannot be resent.
After readings are transferred, press the [▶️ Start/Stop] button to turn off the monitor. The monitor will automatically turn off after 2 minutes.
Wait 2-3 minutes before taking another measurement to allow arteries to recover.
If using an app other than "OMRON connect", data may not display correctly.
Important Reminder: DO NOT adjust medication based on measurement results. Consult your doctor. Self-diagnosis and self-treatment are dangerous.
Data Transfer With Multiple Smart Devices: Data transferred to one smart device cannot be transferred again to the same or another device.
About Multiple Users: This monitor is recommended for personal use. If used by multiple people, measurement data will be combined.
Using your Monitor without Pairing
You can use this blood pressure monitor without pairing to a smart device.
Apply the arm cuff. (Refer to section 3.1)
Press [▶️ Start/Stop] button. (Refer to step 1 to 2 in section 3.3)
Write down your SYS/DIA and Pulse readings, along with the date and time of your measurement on a piece of paper to keep a record of your history.
Note: When using this blood pressure monitor without a smart device, you will not be able to view your prior readings stored in the memory.
3.4 Turning "OFF" Bluetooth
Turn off the Bluetooth [?] in your monitor in areas where wireless equipment use is prohibited, such as on aircraft or in hospitals.
Your blood pressure monitor comes with Bluetooth [?] turned on by default.
Turn your monitor power off if it is on.
Press and hold the connection button [?] for more than 10 seconds.
Release the connection button [?] when "[?? Bluetooth Off]" appears on the display.
Press the [▶️ Start/Stop] button to turn the monitor off.
Note: The monitor will automatically turn off after 2 minutes.
How to turn "ON" Bluetooth
Operate the following step when the power has been off.
Press and hold the connection button [?] for more than 2 seconds and then release the connection button [?] when "[✅? Bluetooth On]" appears on the display.
3.5 Restoring Your Monitor to the Default Settings
To delete all information stored in your monitor, follow these instructions. Ensure the monitor is turned off (screen is not showing numbers or symbols).
While holding the connection button [?] down, the Connection symbol [?] and "P" appear on the display, then press the [▶️ Start/Stop] button for more than 5 seconds.
When the "[? Clear]" symbol appears on the display, release the connection button [?] and [▶️ Start/Stop] button. The "[? Clear]" symbol indicates your monitor settings have reverted to their original state.
Press the [▶️ Start/Stop] button to turn the monitor off.
Notes:
The monitor will automatically turn off after 2 minutes.
Reverting to initial settings does not delete information in the app.
When using this monitor again, delete the monitor's registration in the app first, then pair your monitor.
4. Error Messages and Troubleshooting
4.1 Error Messages
Error Display
Cause
Solution
[?? Irregular Heartbeat]
Irregular heartbeat is detected.
Remove the arm cuff. Wait 2-3 minutes and take another measurement. Repeat steps in section 3.1. If the error continues, contact your doctor.
[? Movement Error]
Movement during measurement.
Carefully read and repeat the steps in section 3.3.
[?? Cuff Wrap Guide] (Loose)
Arm cuff is applied too loosely.
Apply the arm cuff tighter. Refer to section 3.1.
[? Low Battery] / [? Depleted Battery]
The batteries are low. The batteries are depleted or battery polarities are not properly aligned.
Recommend replacement of all 4 batteries with new ones. Refer to section 2.1. Immediately replace 4 batteries with new ones. Refer to section 2.1. Confirm batteries are properly inserted with polarities correctly aligned. Refer to section 2.1.
[❌ Connection failure. Data is not being transmitted.]
Connection failure. Data is not being transmitted.
Refer to "Connection failure. Data is not being transmitted." in section 4.2.
[?? Cuff Wrap Guide] (Loose)
Arm cuff is applied too loosely.
Apply the arm cuff tighter. Refer to section 3.1.
[Error Code e.g., 1-5]
The arm cuff was inflated exceeding the maximum allowable pressure, and then deflated automatically. Movement during measurement. Clothing is interfering with the arm cuff.
Do not touch the arm cuff while taking a measurement. Refer to section 3.3. Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3. Remove any clothing interfering with the arm cuff. Refer to section 3.1.
[Error Code other than 1-5]
Device error.
Contact your local OMRON representative.
4.2 Troubleshooting
Problem
Cause and Solution
No power. No display appears on the monitor.
Replace all batteries with new ones. Check battery installation for proper placement of polarities. Refer to section 2.1.
Measurement values appear too high or too low.
Blood pressure varies constantly. Factors like stress, time of day, and cuff application affect blood pressure. Review sections 3.2 and 3.3.
Connection failure. / Data is not being transmitted.
The blood pressure monitor might not be properly placed within the smart device's transmission range. Move the monitor within 5 m (16 ft.) of the smart device and try again. The Bluetooth feature of the smart device is turned off. Turn on Bluetooth on both the smart device and your monitor, and try sending data again. The Bluetooth feature of the blood pressure monitor is turned off. Refer to "To turn "ON" the Bluetooth" in "3.4 Turning "OFF" the Bluetooth". The blood pressure monitor did not pair successfully. Try pairing again. Refer to "Pair Your Monitor with a Smart Device" in section 2.2. For more detail, go to "Help" in the "OMRON connect" app. The application on the smart device is not ready. Check the application then try sending data again. Refer to "Pair Your Monitor with a Smart Device" in section 2.2. If the "Err" symbol still displays, contact your local OMRON representative. Confirm your smart device compatibility with this monitor at omronconnect.com.
5. Maintenance and Storage
5.1 Maintenance
To protect your device from damage, please follow these directions:
Store the device and components in a clean, safe location.
Do not use abrasive or volatile cleaners.
Do not wash the device or components or immerse them in water.
Do not use gasoline, thinners, or similar solvents to clean the device.
Use a soft dry cloth, or a soft cloth moistened with neutral soap to clean the monitor and arm cuff, then wipe with a dry cloth.
Changes or modifications not approved by the manufacturer will void the user warranty. Do not disassemble or attempt to repair the device or components.
Calibration and Service
The accuracy of this blood pressure monitor has been carefully tested and is designed for a long service life. It is generally recommended to have the device inspected every 2 years to ensure correct functioning and accuracy. Please consult your local OMRON representatives.
5.2 Storage
Do not store the device in the following situations:
If the device is wet.
Locations exposed to extreme temperatures, humidity, direct sunlight, dust, or corrosive vapors such as bleach.
Locations exposed to vibrations, shocks, or where it will be at an angle.
6. Specifications
Category
Details
Product category
Electronic Sphygmomanometers
Product description
Upper Arm Blood Pressure Monitor
Model
HEM-7600T
Display
OLED display
Cuff pressure range
Pressure: 0 to 299 mmHg
Measurement range
Pressure: 40 to 260 mmHg; Pulse: 40 to 180 beats / min.
These specifications are subject to change without notice.
In the clinical validation study, K5 was used on 85 subjects for determination of diastolic blood pressure.
This device is clinically investigated according to the requirements of ISO81060-2:2013.
This device has not been validated for use on pregnant patients.
IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This device is protected against solid foreign objects of 12 mm diameter and greater such as a finger, and against oblique falling water drops which may cause issues during a normal operation.
This device fulfills the provisions of EC directive 93/42/EEC (Medical Device Directive).
This blood pressure monitor is designed according to the European Standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.
This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure monitors, the Pressure Sensor, is produced in Japan.
Symbol descriptions
Please consult this instruction manual
Applied part - Type BF [⚕️ Type BF]
Degree of protection against electric shock (leakage current) [⚡ Electric Shock Protection]
To indicate generally elevated, potentially hazardous, levels of non-ionizing radiation, or to indicate equipment or systems e.g. in the medical electrical area that include RF transmitters or that intentionally apply RF electromagnetic energy for diagnosis or treatment. [? RF Transmitter]
Temperature limitation [?️ Temperature]
Humidity limitation [? Humidity]
Atmospheric pressure limitation [? Pressure]
Artery mark [? Artery]
Regulatory Compliance Mark that meets electrical safety requirements and EMC requirements in Australia [✅ Compliance]
About a wireless communication interference This Product operates in the unlicensed ISM band at 2.4GHz. In case this Product is used around other wireless devices including microwave and wireless LAN, which operate on the same frequency band as this Product, there is a possibility that interference occurs between this Product and such other devices. If such interference occurs, please stop the operation of other devices or relocate this Product before using this Product or do not use this Product around other wireless devices.
Correct Disposal of This Product (Waste Electrical & Electronic Equipment) This marking shown on the product or its literature indicates that it should not be disposed of with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources. Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling. Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial waste for disposal. [♻️ Recycle]
7. Trademarks
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co., Ltd. is under license. Other trademarks and trade names are those of their respective owners.
Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc. Android, the Google Play logo are trademarks of Google Inc.
8. Some Useful Information about Blood Pressure
What is Blood Pressure?
Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart's cycle. The highest pressure in the cycle is called the Systolic Blood Pressure; the lowest is the Diastolic Blood Pressure. Both pressures, the Systolic and Diastolic, are necessary for a doctor to evaluate the status of a patient's blood pressure.
What is Arrhythmia?
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in the bio-electrical system that drives the heartbeat. Typical symptoms are skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) or slow (bradycardia) pulse.
Why is it a Good Thing to Measure Blood Pressure at Home?
Many factors such as physical activity, anxiety, or the time of day, can influence your blood pressure. A single measurement may not be sufficient for an accurate diagnosis. Thus, it is best to try and measure your blood pressure at the same time each day to get an accurate indication of any changes in blood pressure. Blood pressure is typically low in the morning and increases from afternoon to evening. It is lower in the summer and higher in the winter.
Example: fluctuation within a day (male, 35 years old)
OMRON HEM-7142T1 Automatic Blood Pressure Monitor User Manual Comprehensive guide for the OMRON HEM-7142T1 Automatic Blood Pressure Monitor, covering setup, usage with and without a smartphone, memory functions, troubleshooting, maintenance, and specifications. Learn how to accurately measure blood pressure and heart rate at home.
OMRON BP5100 Bronze Upper Arm Blood Pressure Monitor Instruction Manual This instruction manual provides detailed information on the OMRON BP5100 Bronze Upper Arm Blood Pressure Monitor, including setup, operation, safety guidelines, troubleshooting, and specifications for accurate blood pressure measurement at home.
OMRON M4 Intelli IT & X4 Smart Automatic Upper Arm Blood Pressure Monitor Instruction Manual Comprehensive instruction manual for OMRON M4 Intelli IT and X4 Smart automatic upper arm blood pressure monitors, covering safe usage, features, troubleshooting, and maintenance for accurate home health monitoring. Learn how to use your OMRON blood pressure monitor effectively.
OMRON EVOLV Automatic Arm Blood Pressure Monitor User Manual User manual for the OMRON EVOLV Automatic Arm Blood Pressure Monitor (HEM-7600T-E), providing instructions on setup, measurement, troubleshooting, and maintenance. Learn how to accurately measure blood pressure at home and understand the results.