HI991300 / HI991301 Waterproof pH, EC, TDS & Temperature Meter

Instruction Manual

Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using this meter. This manual will provide you with the necessary information for correct use of this meter, as well as a precise idea of its versatility. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com or view our worldwide contact list at www.hannainst.com.

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.

Table of Contents

Preliminary Examination

Remove the instrument and accessories from the packaging and examine it carefully to make sure that no damage has occurred during shipping. Notify your nearest Hanna Instruments Customer Service Center if damage is observed.

HI991300 Package Contents:

HI991301 Package Contents:

Note: Save all packing material until you are sure that the instrument works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.

General Description and Intended Use

The HI991300 and HI991301 are lightweight, portable pH, conductivity (or total dissolved solids) and temperature meters for portable applications requiring both pH and conductivity (or TDS) measurement. To increase conductivity accuracy, two models are available, with different conductivity ranges, for applications from purified to brackish waters.

The HI991300 and HI991301 feature a two-button operation system and are simple to use. All operations and settings are made through these two buttons. They have a waterproof and compact casing rated for IP67 conditions and a large three-line display. The meters have automatic pH calibration at one or two points and a single conductivity calibration. Other user-selectable features include different TDS factors from 0.45 to 1.00, and a range of temperature coefficients (β) from 0.0 to 2.4% for better conductivity or TDS solution temperature compensation.

The supplied HI12883 multi-parameter probe incorporates a dome-shaped pH bulb rated from 0-13 pH, a single junction Ag/AgCl reference electrode with gelled electrolyte and a retractable cloth wick junction, a graphite EC/TDS cell, and a temperature sensor in one convenient, rugged polypropylene body. In addition, to ensure against interference from transient electrical noise to pH, a solid-state preamplifier is integrated into the probe. The probe is rated from 0 to 50°C.

Main Features

Product Image Description: The image shows the Hanna HI991300 meter, a handheld device with a black casing. The LCD screen displays multiple readings: "CAL" indicator, "EC" with a value of "7.00", "pH" with a value of "9.54", "ATC" (Automatic Temperature Compensation), "µS" (microsiemens), "mV" (millivolts), and "23.2°C". Below the screen are two buttons labeled "SET" and "MODE/HOLD". The meter is connected to a probe via a coiled cable. The probe, labeled HI12883, is black and has a sensor tip with a pH bulb, temperature sensor, reference junction, and EC/TDS cell.

Specifications

Feature HI991300 HI991301
Range* -2.00 to 16.00 pH / -2.0 to 16.0 pH
±825 mV (pH-mV)
0 to 3999 µS/cm**
0 to 2000 ppm
-5.0 to 105.0 °C / 23.0 to 221.0 °F
-2.00 to 16.00 pH / -2.0 to 16.0 pH
±825 mV (pH-mV)
0.00 to 20.00 mS/cm**
0.00 to 10.00 ppt
-5.0 to 105.0 °C / 23.0 to 221.0 °F
Resolution 0.01 pH / 0.1 pH
1 mV
1 µS/cm
1 ppm (mg/L)
0.1 °C / 0.1 °F
0.01 pH / 0.1 pH
1 mV
0.01 mS/cm
0.01 ppt (g/L)
0.1 °C / 0.1 °F
Accuracy @ 25°C/77°F ±0.02 pH / ±0.1 pH
±1 mV (pH-mV)
±2% F.S. (EC/TDS)
±0.5 °C / ±1.0 °F Temperature
±0.02 pH / ±0.1 pH
±1 mV (pH-mV)
±2% F.S. (EC/TDS)
±0.5 °C / ±1.0 °F Temperature
Temperature compensation pH - Automatic
EC/TDS - Automatic, with β selectable from 0.0 to 2.4 %/°C (0.1 increments)
pH calibration Automatic, 1 or 2 points choice between 2 sets of buffers (standard: 4.01, 7.01, 10.01 pH or NIST: 4.01, 6.86, 9.18 pH)
EC/TDS calibration Automatic, one-point at 1413 µS/cm or 1382 ppm (CONV=0.5) or 1500 ppm (CONV=0.7) Automatic, one-point at 12.88 mS/cm or 6.44 ppt (CONV=0.5) or 9.02 ppt (CONV=0.7)
TDS conversion factor Selectable from 0.45 to 1.00 (0.01 increments)
Probe (included) HI12883 pH/EC/TDS/temperature sensor, DIN connector and 1 m (3.3') cable
Battery type/life 1.5V AAA (3 pcs.) approx. 600 hours of continuous use
Auto-off User selectable: after 8 min, 60 min or disabled
Environment 0 to 50 °C (32 to 122 °F) RH max. 100%
Meter dimensions 154 x 63 x 30 mm (6.1 x 2.5 x 1.2")
Weight (with batteries) 196 g (6.91 oz.)
Casing ingress protection rating IP67

* The HI12883 is limited to be used from 0 to 13 pH and from 0 to 50 °C (32 to 122 °F)

** LCD displays µS for µS/cm (HI991300) or mS for mS/cm (HI991301)

Display Description

Display Diagram Description: The diagram shows a representation of the meter's LCD screen with various segments and indicators labeled numerically from 1 to 14. Key elements include: CAL (calibration tag), EC, pH, ATC, µS, mV, ppm, %, °C, °F, OFF, CAL, MODE. The primary display shows large numerical values, while secondary and tertiary displays show units and status indicators.

Operational Guide

Turning the Meter On

Each meter is supplied with batteries. Before using the meter for the first time, open the battery compartment and insert batteries, observing the polarity (see "Battery Replacement"). To turn the meter ON, press the [power-symbol] button on the front of the meter. If it does not turn on, make sure that the batteries are properly installed in their place. The meter is provided with an active acoustic signal when a key is pressed.

At start-up, the meter displays all LCD segments for a few seconds, followed by the percentage indication of the remaining battery life, displaying "WAIT" until electrode check is in process, then the meter enters the normal measurement mode.

Startup Sequence Description: The meter performs an LCD check-up, then displays battery life percentage, followed by an "Electrode check" (showing "WAIT"), and finally enters the normal "Measurement mode".

Note: The meter detects the presence and the type of the probe at its input. If the probe is not connected, the message "NO PROBE" appears alternatively on the tertiary LCD. If the probe is not compatible, "WRONG PROBE" message appears alternatively on the tertiary LCD line with "---" blinking on the first LCD line.

Connecting the Electrode

With the meter turned off, connect the HI12883 probe to the DIN socket on the bottom of the meter by aligning the pins and pushing in the plug firmly. Remove the protective cap from the probe before taking any measurements.

Selecting Measurement Range

While in measurement mode, press the SET button to select EC, TDS or mV of pH measurement on the secondary LCD.

Freezing Measurement Values

While in measurement mode, press and hold the SET button until "HOLD" appears on the tertiary LCD line. The "HOLD" remains for 1 second and readings will be frozen on the LCD with "H" blinking. Press any button to resume active measurements.

Hold Function Description: The display shows a transition from normal readings (e.g., 7.00 pH, 9.54 µS, 23.2°C) to a "HOLD" state where the values are frozen and an "H" indicator appears.

Entering Calibration Mode

Press and hold the [power-symbol] button until "POWER" and the [power-symbol] tag is replaced by "PH STD", [CAL-symbol] tag or "EC STD", [CAL-symbol] tag if standard calibration is selected from the setup menu. Release the button.

Entering Setup Mode

Press and hold the [mode-symbol] button until "EC STD" and the [CAL-symbol] tag is replaced by "SETUP" and the [mode-symbol] tag. Release the button.

Turning the Meter Off

While in measurement mode, press the [power-symbol] button. "POWER" and the [power-symbol] tag will appear. Release the button.

pH Measurement & Calibration

Make sure the meter has been calibrated before use. If the probe is dry, soak it in HI70300 storage solution for 30 minutes to reactivate it. If soiled, clean the electrode by soaking in cleaning solution for 20 minutes, then rinse the tip and soak in storage solution at least 30 minutes before use. Rinse the electrode well and shake off excess droplets. Recalibrate before using. Submerge the probe in the sample to be tested while stirring it gently. Wait until the [stability-indicator-symbol] tag on the LCD disappears. The LCD displays the pH value (automatically compensated for temperature) on the primary LCD, the EC, TDS or pH-mV value on the secondary LCD, while the tertiary LCD line displays the sample temperature. If measurements are taken in different samples successively, rinse the probe tip thoroughly in distilled or deionized water to eliminate cross-contamination. For better accuracy, frequent calibration of the pH sensor with the meter is recommended. In addition, the meter must be recalibrated:

pH Calibration

Select calibration type "STD" [CAL-symbol]. Place the sensor into the first calibration buffer. If performing a two-point calibration, use pH 7.01 (pH 6.86 for NIST) buffer first. The meter will enter the calibration mode, displaying "pH 7.01 USE" [CAL-symbol] tag blinking (or "pH 6.86 USE" for NIST). Follow directions for single and two-point calibration below:

Single-point calibration
  1. Place the probe in any buffer from the selected buffer set. The meter will automatically recognize the buffer value.
  2. If the buffer is not recognized or the calibration offset is out of the accepted range, "---- WRONG" is displayed.
  3. If the buffer is recognized, "REC" is displayed, then "WAIT" until the calibration is accepted. If using pH 7.01 (or pH 6.86 for NIST), after acceptance of the buffer, press any key to exit. The "SAVE" message is displayed and the meter returns to pH measurement mode. If using pH 4.01 or 10.01 (or pH 9.18 for NIST) buffer, the "SAVE" message is displayed and the meter returns to pH measurement mode.
Two-point calibration

Proceed with steps 1 through 3 under single-point calibration using 7.01 pH (pH 6.86 for NIST) buffer first. Then follow steps below: The "pH 4.01 USE" message is then displayed. Place the probe in the second calibration buffer (pH 4.01 or 10.01, or, if using NIST, pH 4.01 or 9.18). When the second buffer is accepted, the LCD will display "SAVE" for 1 second and the meter will return to the normal measurement mode. If the buffer is not recognized or the slope is out of accepted range "---- WRONG" is displayed. Change the buffer, clean the electrode or press any key to exit calibration. For better accuracy, it is always recommended to carry out a two-point calibration. After the calibration procedure has been completed, the "CAL" is turned on together with the calibrated points.

Exiting Calibration and Resetting Default Values

After entering the calibration mode and before the first point is accepted, it is possible to quit the procedure and return to the last calibration data by pressing the [esc-symbol] button. The LCD displays "ESC" for 1 second and the meter returns to normal mode. To reset the default values and clear a previous calibration, press the SET button after entering the calibration mode and before the first point is accepted. The LCD displays "CLEAR" for 1 second, the meter resets to the default calibration and the "CAL" with the calibrated points on the LCD disappears.

pH Electrode Condition

The display is provided with a probe icon (unless the feature is disabled from setup) which indicates the pH electrode status after calibration. The "condition" remains active for 12 hours (unless the batteries are removed). The electrode condition is evaluated only if the current pH calibration has two points.

With 1 bar, it is recommended to clean the pH electrode and recalibrate. If there is still only 1 bar or 1 bar blinking, replace the probe.

Sensor Check

Setting the meter to pH-mV range, the user can check the sensor status at any time. The offset value is the reading in pH 7.01 buffer (@ 25 °C/77 °F). If this reading is outside the range ±30 mV, the electrode is considered "very poor". The slope value of the sensor is the difference between readings in pH 7.01 and in pH 4.01 buffers. When the slope reaches the value of about 150 mV, the electrode is considered "very poor". When "poor" or "very poor", it is recommended to replace it with a new one.

Note: To ensure reliable readings, the electrode must be cleaned with cleaning solution and then hydrated in storage solution for a minimum of 30 minutes before calibrating the probe.

EC Measurement and Calibration

Place the probe in the sample to be tested. Use plastic beakers or containers to minimize any electromagnetic interference. Tap the probe lightly on the bottom of the beaker or container to remove air bubbles that may be trapped inside the tip. Wait for a few minutes for the temperature sensor to reach thermal equilibrium, when the [stability-indicator-symbol] tag disappears. The LCD displays the EC or TDS value (automatically compensated for temperature) on the secondary LCD, while the third line LCD displays the sample temperature.

EC Calibration

Before calibration, rinse the sensor tip with a heavy stream of purified water then shake excess water from the probe. Select calibration type "EC STD" [CAL-symbol]. The meter enters the calibration mode and "µS 1.41 USE" (HI991300) or "mS 12.88 USE" (HI991301) is displayed with [CAL-symbol] tag blinking. Immerse the probe in calibration solution. If the standard value is recognized, "REC" is displayed, then "WAIT" until the calibration is accepted. The LCD will display "SAVE" for 1 second and return to normal measurement mode. If the standard is not recognized or the reading is out of accepted range "---- WRONG" is displayed. Change the calibration solution, clean the electrode or press any key to exit calibration. When the calibration procedure is completed, the "CAL" and [ec-calibration-tag-symbol] is displayed.

Note: β should be set to 1.9 during calibration. There is a known relationship between the EC and TDS readings; it is not necessary to calibrate the meter in TDS. The meter will allow a calibration in TDS using HI70032 or HI70038 TDS calibration solution.

Exiting Calibration and Resetting Default Values

After entering the calibration mode and before the point is accepted, it is possible to quit the procedure and return to the last calibration data by pressing the [esc-symbol] button. The LCD displays "ESC" for 1 second and the meter returns to normal mode. To reset the default values and clear a previous calibration, press the SET button after entering the calibration mode and before the first point is accepted. The LCD displays "CLEAR" for 1 second, the meter resets to the default calibration and the "CAL" and [ec-calibration-tag-symbol] on the LCD disappears.

Meter Setup

Setup mode allows the selection of the Temperature unit, Auto-off, Beep confirmation tone, the type of pH buffer set, the pH Resolution, enable/disable calibration information display, temperature compensation factor for EC and TDS conversion factor. To enter Setup mode, press and hold the [mode-symbol] button until "EC STD" and the [CAL-symbol] tag is replaced by "SETUP" and the [mode-symbol] tag. Release the button.

Battery Replacement

When the remaining battery life is less than 10%, the battery [battery-icon] tag blinks on the display to warn the user.

Battery Error Prevention System (BEPS)

If the battery is too weak ("0%"), the display shows "batt", "DEAD" for a few seconds, then the meter powers off. Immediately replace the batteries with new ones.

The batteries are accessed by opening the battery cover on the back of the instrument. Remove protective boot if present.

Battery Replacement Diagram Description: The diagrams illustrate the process of replacing the batteries. First, the battery cover on the back of the meter is shown being opened. Then, three 1.5V AAA alkaline batteries are shown being inserted into the battery compartment, with attention to correct polarity (positive and negative ends). Finally, the battery cover is replaced, ensuring the gasket is in place.

Accessories

Probe Diagram Description: The diagram shows the HI12883 probe. It highlights the "Quick connector with Rubber protection, and 1 meter cable". Key components of the probe tip are labeled: "Temperature sensor", "pH bulb", "Reference junction", and "EC/TDS cell". It also mentions "Code and serial number on the other side".

Electrode Maintenance

Preparation

Storage

Note: Never store the electrode in distilled water.

Periodic Maintenance

Certification

All Hanna Instruments conform to the CE European Directives.

[CE Mark] [RoHS Compliant]

Disposal of Electrical & Electronic Equipment: The product should not be treated as household waste. Instead, hand it over to the appropriate collection point for the recycling of electrical and electronic equipment which will conserve natural resources.

Disposal of waste batteries: This product contains batteries; do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.

Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health. For more information, contact your city, your local household waste disposal service, the place of purchase or go to www.hannainst.com.

MAN991300 12/18

Recommendations for Users

Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any variation introduced by the user to the supplied equipment may degrade the meter's performance. For yours and the meter's safety, do not use or store the meter in hazardous environments.

Warranty

HI991300 and HI991301 are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Electrodes and probes are warranted for a period of six months. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered.

If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number (see engraved on the back of the meter), and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the meter is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Technical Service department and then send it with shipping costs prepaid. When shipping any meter, make sure it is properly packed for complete protection.

Hanna Instruments reserves the right to modify the design, construction or appearance of its products without advance notice.

World Headquarters

Hanna Instruments Inc.
Highland Industrial Park
584 Park East Drive
Woonsocket, RI 02895 USA
www.hannainst.com

Printed in ROMANIA

Models: HI991300 Waterproof TDS and Temperature Meter, HI991300, Waterproof TDS and Temperature Meter, Temperature Meter, Meter

File Info : application/pdf, 24 Pages, 584.63KB

PDF preview unavailable. Download the PDF instead.

HI991300

References

Adobe InDesign CC 14.0 (Macintosh) iLovePDF

Related Documents

Preview Hanna HI5522 Research Grade Meter: pH, ORP, ISE, EC, TDS, Salinity, Resistivity, and Temperature
Detailed overview of the Hanna HI5522 research-grade benchtop meter, highlighting its advanced features for measuring pH, ORP, ISE, EC, TDS, Salinity, Resistivity, and Temperature. Includes information on calibration, data logging, incremental methods, and USP compliance.
Preview Hanna Instruments HI98197 Waterproof Portable EC Meter for Ultrapure Water
The Hanna Instruments HI98197 is a professional, waterproof, portable EC (conductivity) meter designed for ultrapure water applications. It features a high-resolution display, multi-point calibration, automatic temperature compensation, and data logging capabilities. The meter is IP67 rated and includes a four-ring platinum probe for accurate measurements across a wide range.
Preview Hanna HI3512 Dual Input Benchtop Meter Quick Reference Guide
Quick reference guide for the Hanna HI3512 Dual Input Calibration Check pH, ORP, ISE, EC, Resistivity, TDS, NaCl Benchtop Meter, covering features, package contents, and operational overview.
Preview Hanna Instruments HI9812-51 Portable pH, EC, TDS, Temperature Meter Instruction Manual
User manual for the Hanna Instruments HI9812-51 portable meter, detailing its features, specifications, calibration, measurement, and maintenance for pH, EC, TDS, and temperature.
Preview Hanna Instruments HI5221 & HI5222 pH/mV/ISE/Temperature Bench Meters Instruction Manual
This instruction manual provides comprehensive details on the Hanna Instruments HI5221 and HI5222 pH/mV/ISE/Temperature Bench Meters, covering their features, operation, specifications, and maintenance for accurate and versatile measurements.
Preview Hanna Instruments HI6000 Modular Multiparameter Meters
Explore the Hanna Instruments HI6000 series of modular multiparameter meters, offering flexible and customizable solutions for pH/ORP, pH/ORP/ISE, EC, and DO measurements. Learn about their advanced features, connectivity options, and user-friendly interface.
Preview Hanna HI981520 Marine Monitor: pH, Salinity, and Temperature Meter
The Hanna HI981520 Marine Monitor provides accurate and continuous measurement of pH, salinity (ppt, PSU, S.G.), and temperature for saltwater aquariums and aquaculture. Features include automatic calibration, high/low alarms, and a waterproof design.
Preview Hanna Instruments HI6221 Research Grade pH/mV/Temperature Meter - Technical Overview
Detailed overview of the Hanna Instruments HI6221 research-grade meter, featuring advanced pH, mV, and temperature measurement capabilities, intuitive interface, extensive connectivity, and precise calibration. Includes specifications and ordering information.