OMRON HEM-7124 Automatic Blood Pressure Monitor

Instruction Manual

Introduction

Thank you for purchasing the OMRON HEM-7124 Automatic Blood Pressure Monitor. The OMRON HEM-7124 is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation, the monitor uses its advanced “IntelliSense” technology.

Intended Use: This product is designed to measure the blood pressure and pulse rate of people within the range of the designated arm cuff, following the instructions in this instruction manual. It is mainly designed for general household use.

Please follow this instruction manual thoroughly for your safety. Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR PHYSICIAN.

Contents: Monitor, arm cuff, instruction manual, battery set.

Diagram showing the monitor components: A. Display, B. START/STOP button, C. Battery compartment, D. Air jack, E. Arm cuff (Arm circumference 22 - 32 cm), F. Air plug, G. Air tube, H. Marker, I. Memory symbol, J. Systolic blood pressure, K. Diastolic blood pressure, L. Battery symbol (low/depleted), M. Heartbeat symbol (Flashes during measurement.), N. Pulse display, O. Deflation symbol.

Important Safety Information

General Usage Precautions:

Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.

General Usage (Caution):

Battery Usage:

General Precautions:

1. Know Your Monitor

The monitor is a device that measures blood pressure and pulse rate. It is designed for general household use.

2. Preparation

2.1 Battery Installation

  1. Remove the battery cover.
  2. Insert 4 “AA” batteries as indicated in the battery compartment.
  3. Replace the battery cover.

Notes:

3. Using the Monitor

Before Taking a Measurement:

  1. Avoid bathing, drinking alcohol or caffeine, smoking, exercising and eating for 30 minutes before taking a measurement.
  2. Rest for at least 5 minutes before taking the measurement.
  3. Stress raises blood pressure. Avoid taking measurements during stressful times.
  4. Measurements should be taken in a quiet place.
  5. Remove tight-fitting clothing from your arm.

3.1 Applying the Arm Cuff

Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm. Do not place the arm cuff over thick clothes.

  1. Insert the air plug into the air jack securely.
  2. Wrap the arm cuff firmly in place around your left upper arm. The bottom edge of the arm cuff should be 1 to 2 cm above the elbow. The air tube is on the inside of your arm and aligned with your middle finger.
  3. Secure closed with the fabric fastener.

Notes:

3.2 How to Sit Correctly

To take a measurement, you need to be relaxed and comfortably seated at a comfortable room temperature.

3.3 Taking a Measurement

If your systolic pressure is more than 210 mmHg: After the arm cuff starts to inflate, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.

  1. Press the START/STOP button. The arm cuff will start to inflate automatically.
  2. Remove the arm cuff.
  3. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes.

Notes:

Note: If your systolic or diastolic pressure is outside the standard range (above 134/84 mmHg) the Heartbeat symbol ( ) will blink. Please refer to Section 3.3.

3.4 Using the Memory Function

The monitor automatically stores the last measurement values (blood pressure and pulse rate).

To View the Measurement Values Stored in Memory:

  1. Press and hold the START/STOP button for more than 5 seconds. The last measurement value is displayed along with the memory symbol.
  2. Press the START/STOP button to turn the monitor off. It will automatically turn off after 2 minutes.

To Delete the Values Stored in Memory: Press and hold the START/STOP button for more than 15 seconds.

Notes:

4. Error Messages and Troubleshooting

In case of any of the below problems occur during measurement, first check that no other electrical device is within 30cm. If the problem persists, please refer to the table below.

Problem Cause Solution
Nothing happens when you press the buttons. The batteries are depleted. Replace 4 batteries with new ones at once. Refer to section 2.1.
The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity. Refer to section 2.1.
Arm cuff pressure does not rise. The air connector is not securely connected into the air jack. Insert the plug securely. Refer to section 3.1.
Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to section 3.1.
Air is leaking from the arm cuff. Replace the cuff with a new one. Refer to section 5.3.
Arm cuff deflates too soon. The arm cuff has not been inflated sufficiently. Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result. Refer to section 3.3.
Cannot measure or the results are too low or too high. Movement or talking during measurement. Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3.
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to section 3.1.
The air connector is not securely connected into the air jack. Make sure that the air tube is connected securely. Refer to section 3.1.
Air is leaking from the arm cuff. Replace the arm cuff with a new one. Refer to section 5.3.
The arm cuff is loose. Apply the cuff correctly so that it is firmly wrapped around the arm. Refer to section 3.1.
The arm cuff has not been inflated sufficiently. Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result. Refer to section 3.3.
The batteries are depleted. Replace the batteries with new ones. Refer to section 2.1.
The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity. Refer to section 2.1.
Device error. Contact your local OMRON representative. Contact your local OMRON representative.
Other problems.
  • Press the START/STOP button and repeat measurement.
  • Replace the batteries with new ones.
  • If the problem continues, contact your local OMRON representative.

Error Display:

Error Cause Solution
E1 Movement during measurement and the arm cuff has not been inflated sufficiently. Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3.
Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to section 3.1.
E2 The arm cuff was inflated exceeding the maximum allowable pressure, and then deflated automatically. If “E2” appears repeatedly, inflate the cuff manually until it is 30 to 40 mmHg above your previous measurement result. Refer to section 3.3. Do not touch the arm cuff and/or bend the air tube while taking a measurement. Do not inflate the arm cuff more than necessary. Refer to section 3.3.
E3 Movement during measurement. Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3.
E4 Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to section 3.1.
E5 The air connector is not securely connected into the air jack. Make sure that the air tube is connected securely. Refer to section 3.1.
Air is leaking from the arm cuff. Replace the arm cuff with a new one. Refer to section 5.3.
The arm cuff is loose. Apply the cuff correctly so that it is firmly wrapped around the arm. Refer to section 3.1.
The arm cuff has not been inflated sufficiently. Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result. Refer to section 3.3.
The batteries are depleted. Replace the batteries with new ones. Refer to section 2.1.
The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity. Refer to section 2.1.

5. Maintenance and Storage

5.1 Maintenance

To protect your monitor from damage, please observe the following:

Calibration and Service:

5.2 Storage

  1. Unplug the air plug from the air jack.
  2. Gently fold the air tube into the arm cuff.

Note: Do not bend or crease the air tube excessively.

Do not store your monitor in the following situations:

5.3 Optional Parts

Arm Cuff:

6. Specifications

Product description Automatic Blood Pressure Monitor
Model HEM-7124
Display LCD Digital Display
Measurement method Oscillometric method
Measurement range Blood pressure: 0 to 299 mmHg
Pulse measurement range: 40 to 180 beats/min.
Accuracy Pressure: ±3 mmHg
Pulse: ±5 % of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Memory Last Measurement
Rating DC6 V 4 W
Operation mode Continuous operation
Power source 4 “AA” batteries 1.5 V
Battery life Approx. 1000 measurements (using new alkaline batteries)
Durable Period Monitor: 5 years
Cuff: 1 year
Applied part Type BF (Cuff)
Protection against electric shock Internally powered ME equipment
Operating conditions +10 to +40 °C / 15 to 90% RH (non-condensing) / 700-1060 hPa
Storage/ Transport conditions -20 to +60 °C / 10 to 95 % RH (non-condensing) / 700 to 1060 hPa
IP classification IP20
Weight Monitor: Approx. 250 g without batteries
Arm cuff: Approx. 130 g
Outer dimensions Monitor: Approx. 103 (w) mm x 80 (h) mm x 129 (l) mm
Arm cuff: Approx. 145 mm x 466 mm
Cuff circumference 22 to 32 cm
Cuff/ Tube material Nylon, polyester, polyvinyl chloride
Package contents Monitor, arm cuff, instruction manual, battery set

Notes:

Important information regarding Electro Magnetic Compatibility (EMC):

HEM-7124 manufactured by OMRON HEALTHCARE Co., Ltd. conforms to EN60601-1-2:2015 Electro Magnetic Compatibility (EMC) standard. Further documentation in accordance with this EMC standard is available at http://www.omronhealthcare-ap.com/emc-information. Refer to the EMC information for HEM-7124 on the website.

Symbols description:

Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.

The batteries are low.

The batteries are depleted.

Applied part - Type BF

Degree of protection against electric shock (leakage current)

Class II equipment. Protection against electric shock

CE CE Marking

SN Serial number

LOT LOT number

Temperature limitation

📈 Humidity limitation

🌐 Atmospheric pressure limitation

Indication of connector polarity

For indoor use only

IntelliSense OMRON's trademarked technology for blood pressure measurement

Identifier of cuffs compatible for the device

Cuff positioning indicator for the left arm

Marker on the cuff to be positioned above the artery

Range pointer and brachial artery alignment position

Not made with natural rubber latex

Alternating current

Direct current

Need for the user to follow this instruction manual thoroughly for your safety.

Need for the user to consult this instruction manual.

Range indicator of arm circumferences to help selection of the correct cuff size.

Date of manufacture

Product production date is integrated in the Serial number, which is placed on the product and/or sales package: the first 4 digits mean year of production, the next 2 digits mean month of production.

Depending on the product, some of the above symbols may not be applicable on the product.

Correct Disposal of This Product (Waste Electrical & Electronic Equipment):

This marking shown on the product or its literature, indicates that it should not be disposed of, with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this product from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources.

Household users should contact either the retailer where they purchased this product, or their local government office, for details of where and how they can return this item for environmentally safe recycling.

Business users should contact their supplier and check the terms and conditions of the purchase contract. This product should not be mixed with other commercial waste for disposal.

Manufacturer Information

OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, KYOTO, 617-0002 JAPAN

OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS
www.omron-healthcare.com

OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08 Alexandra Technopark, Singapore 119967
www.omronhealthcare-ap.com

OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.
Binh Duong Province, VIETNAM

This device fulfils the provisions of EC directive 93/42/EEC (Medical Device Directive).

This blood pressure monitor is designed according to the European Standard EN1060, Non-invasive sphygmomanometers Part 1: General Requirements and Part 3: Supplementary requirements for electromechanical blood pressure measuring systems.

This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan. The Core component for OMRON blood pressure monitors, which is the Pressure Sensor, is produced in Japan.

PDF preview unavailable. Download the PDF instead.

2e47740b0ec84674aa80d4b18c314ceb Adobe PDF library 9.90 Adobe Illustrator CS5

Related Documents

Preview OMRON JPN1 Automatic Blood Pressure Monitor Instruction Manual
Instruction manual for the OMRON JPN1 Automatic Blood Pressure Monitor, covering setup, usage, troubleshooting, and technical specifications.
Preview OMRON X2 Smart Blood Pressure Monitor: Accurate Home Health Monitoring
The OMRON X2 Smart Automatic Upper Arm Blood Pressure Monitor offers simple, accurate blood pressure and pulse rate measurement at home, featuring Bluetooth connectivity for easy health tracking.
Preview OMRON EVOLV Automatic Upper Arm Blood Pressure Monitor User Manual
User manual for the OMRON EVOLV Automatic Upper Arm Blood Pressure Monitor (HEM-7600T-E), detailing its features, operation, safety information, maintenance, and specifications.
Preview OMRON HEM-7142T1 Automatic Blood Pressure Monitor User Manual
Comprehensive guide for the OMRON HEM-7142T1 Automatic Blood Pressure Monitor, covering setup, usage with and without a smartphone, memory functions, troubleshooting, maintenance, and specifications. Learn how to accurately measure blood pressure and heart rate at home.
Preview OMRON MIT3 Automatic Blood Pressure Monitor Instruction Manual
Comprehensive guide for the OMRON MIT3 Automatic Blood Pressure Monitor, covering setup, usage, maintenance, troubleshooting, and specifications. Learn how to accurately measure blood pressure and pulse rate at home.
Preview OMRON HEM-7155T Automatic Blood Pressure Monitor Instruction Manual
Comprehensive instruction manual for the OMRON HEM-7155T Automatic Blood Pressure Monitor, covering setup, usage, safety precautions, troubleshooting, and specifications. Learn how to accurately measure blood pressure and pulse rate for general household use.
Preview OMRON HEM-7361T Automatic Blood Pressure Monitor Instruction Manual
Comprehensive instruction manual for the OMRON HEM-7361T Automatic Blood Pressure Monitor, covering safety instructions, intended use, troubleshooting, specifications, and maintenance.
Preview Návod na použitie OMRON M6 Comfort AFib (HEM-7380-E)
Kompletný návod na použitie pre automatický tlakomer OMRON M6 Comfort AFib (HEM-7380-E), vrátane pokynov na bezpečnú prevádzku, údržbu, riešenie problémov a špecifikácií produktu.