Intelligent Digital Clamp Multimeter
Users Manual
1. Safety Information
▲ Warning Pay special attention to use of this instrument, for improper use may cause electric shock or damage to instrument. During use, observe usual safety regulations and observe safety measures regulated in use manual. In order to make full use of instrument functions and guarantee safe operation, please carefully read and observe use methods in this manual.
Instrument complies with safety requirements on electronic measuring instrument of EN-61010-1, EN-61010-2-030 and EN-61010-2-032, level II pollution, and over-voltage standard is CAT II 600V. Please observe safety operation guide, and guarantee to use instrument in a safe manner.
1.1 Preparation
When using this instrument, users must observe standard safety rules:
- General electric shock prevention.
- Prevention of misuse of instrument.
After receiving the instrument, check whether it is damaged during transportation.
After storing and shipping under adverse conditions, check whether the instrument has been damaged.
Pens of the instrument must be in good condition. Before use, check whether insulation of pens is damaged, and whether metal wire is exposed.
1.2 Symbol
- ! Note (refer to use manual for important safety information)
- ⚡ Able to be used on dangerous electrified conductors.
- [ ] Dual-insulation protection (Category II)
- CAT II Over-voltage according to IEC-61010-1 standard (installation), level II, pollution degree 2 refers to protection level of pulse withstand voltage provided.
- CE Comply with EU standard.
- ⏚ Grounded
1.3 Maintenance
- Do not try to open bottom case to adjust or repair instrument, for such operation could only be conducted by technicians fully understanding the instrument and electric shock danger.
- Before opening instrument bottom case or battery cover, remove the pens from the wire being measured.
- In order to avoid electric shock possibly caused by error reading, when symbol [low battery symbol] displays on instrument, replace battery immediately.
- Use wet cloth and gentle detergent to clean the instrument, and do no use any abrasive or solvent.
- Power off when the instrument is not used, and rotate range switch to OFF position.
- If the instrument is not used for a long time, take out the battery to avoid any damage to the instrument.
2. Description
2.1 Part name
1 Non-contact voltage detecting & inducing area
2 Function key
3 electric torch
4 Liquid crystal display (LCD)
5 Trigger
6 On/off key
7 Trigger
8 alarm light
9 Current clamp: Used in current measurement
2.2 LCD display
The LCD display shows various symbols and readings:
Symbol | Meaning |
~ / ⎓ | AC & DC |
))) | Connection/Disconnection indication |
AUTO | Automatic range mode |
↻ | Automatic shutdown indication |
[battery symbol] | Low battery |
HOLD | Reading hold state |
V, A | Volt (voltage), ampere (current) |
Ω, kΩ, MΩ | Ohm kilohm and megohm (resistance) |
Hz | hertz |
⏓ | DIODE |
nF, uF, mF | Microfarad |
°C / °F | Centigrade and Fahrenheit |
3. Specification
The instrument specifies one year as a cycle, and shall be re-calibrated under 18°C ~ 28°C, with relative humidity less than 75%.
3.1 Overview
- Select measurement function and range automatically.
- Overload protection throughout the range.
- Max. voltage between measurement terminal and ground: 600V DC or 600V AC
- Operating height: Max. 2000m
- Display: LCD
- Max. display value: 6000 digit.
- Polarity indication: Automatic indication, and '-' indicates negative.
- Over range display: 'OL' or '-OL'.
- Sampling time: About 3 times/second., bargraph 10 times/second
- Unit display: Function and electricity quantity unit display.
- Automatic shutdown time: 15 minutes
- Power supply: 1.5V AAA battery × 2
- Battery under-voltage indication: LCD display [battery symbol] symbol
- Temperature coefficient: Less than 0.1×accuracy/°C.
- Operating temperature: 18°C ~ 28°C.
- Storage temperature: -10°C ~ 50°C.
3.2 Technical indexes
3.2.1 AC current
Range | Resolution | Accuracy |
---|---|---|
60A | 0.01A | ± (2.5% reading + 8 digits) |
200A | 0.1A |
- Min. input current: 0.2A AC current.
- Max. input current: 200A AC current.
- Frequency range: 40 ~ 1000 Hz;
3.2.2 DC current (Optional features)
Range | Resolution | Accuracy |
---|---|---|
60A | 0.01A | ± (2.5% reading + 8 digits) |
200A | 0.1A |
- Min. input current: 0.2A DC current
- Max. input current: 200A DC current
3.2.3 DC voltage
Range | Resolution | Accuracy |
---|---|---|
6V | 0.001V | ± (0.5% reading + 3 digits) |
60V | 0.01V | |
600V | 0.1V |
- Min. input voltage 0.5 V DC
- Max. input voltage: 600V DC
3.2.4 AC voltage
Range | Resolution | Accuracy |
---|---|---|
6V | 0.001V | ±(0.8% reading + 5 digits) |
60V | 0.01V | |
600V | 0.1V |
- Min. input voltage: 0.5 V AC
- Max. input voltage: 600V AC (valid value)
- Frequency range: 40 ~ 1000Hz
3.2.5 Line on-off test
Range | Resolution | Function |
---|---|---|
))) | 1Ω | If resistance of the line being measured is less than 50Ω, buzzer in instrument will make continuous alarming sounds. |
- Overload protection: 600V DC or AC (valid value)
3.2.6 Resistance
Range | Resolution | Accuracy |
---|---|---|
600Ω | 0.1Ω | ± (0.8% reading + 3 digits) |
6kΩ | 0.001kΩ | |
60kΩ | 0.01kΩ | |
600kΩ | 0.1kΩ | |
6MΩ | 0.001MΩ | |
60MΩ | 0.01MΩ |
- Overload protection: 600V DC or AC (valid value)
3.2.7 DIODE
Range | Resolution | Accuracy | Testing environment |
---|---|---|---|
Diode test ⏓ | 1 V | 0.001 V | Testing current: about 1mA; Open circuit voltage: about 2.8V. The display shows an approximation of the diode forward pressure drop |
- Overload protection: 250V DC or AC (valid value)
3.2.8 Capacitor
Range | Resolution | Accuracy |
---|---|---|
60.00nF | 0.01nF | ±(2.5%reading+ 20 digits) |
600.0nF | 0.1nF | |
6.000uF | 1nF | |
60.00uF | 10nF | |
600.0uF | 100nF | |
6.000mF | 1μF | |
60.00mF | 10μF |
- Overload protection: 250V DC or AC (valid value)
3.2.9 Temperature
Range | Resolution | Accuracy |
---|---|---|
-50~1300 °C (-58~2372 °F) |
1°C / 1°F | ± (1% reading +3°C) / ± (1% reading +3°F) |
The precision does not include the error of the thermocouple probe
- Overload protection: 250V DC or AC (valid value)
3.2.10 Frequency
3.2.10.1 Clamp frequency measurement (via gear A):
Range | Resolution | Accuracy |
---|---|---|
60.0Hz | 0.1Hz | ± (1.0%reading+ 5digits) |
1000Hz | 1Hz |
- Measurement range: 40Hz ~ 1000Hz
- Input signal range: ≥ 1/4 Full- scale Value
3.2.10.2 Via gear V:
Range | Resolution | Accuracy |
---|---|---|
60.0Hz | 0.1Hz | ± (1.0%reading+ 5digits) |
1000Hz | 1Hz |
- Measurement range: 40Hz ~ 1000Hz
- Input signal range: ≥ 0.8 AC voltage (valid value)
4. Operation guide
4.1 Reading hold
During measurement, if it is required to hold reading, touch [HOLD symbol] button, value on display will be locked, touch [HOLD symbol] button again, to cancel reading hold.
4.2 Backlight/Light
- During measurement, if environment light is too dark, causing it difficult to read, you could press [HOLD symbol] key for more than 2 seconds, to open backlight or light, which will turn off in about 15 seconds automatically.
- Pressing [HOLD symbol] the button for more than 2 seconds will turn off the backlight.
- Press "NCV/⚡" the button for 2 seconds, turn on the flashlight, and turn it off automatically after about 15 minutes.
4.3 Automatic shutdown
- If there's no operation within 15 minutes after start, instrument will enter standby state, and shut down to save energy.
- After automatic shutdown, Press the "[SELECT symbol]" button, to wake the instrument enter work state.
4.4 To measure
- Touch the power button to start the machine. If the battery voltage is insufficient, the display will display a low voltage symbol [battery symbol], at which point the battery should be replaced. Press the power button again to shut down the machine.
- When the instrument is not measured, it will enter the automatic scanning state. The instrument is shown as follows:
AUTO
4.5 NCV (Non-contact voltage detection)
Switch the rotation switch to any gear, hold down the "NCV" key to display EF, and bring the NCV sensor close to the conductor under test. Instrument could detect whether the lead being measured is >90V AC voltage. When instrument detects AC voltage, instrument buzzer will alarm and NCV alarm light will flash.
Note:
- Even if there's no alarm indication, voltage may still exist. Do not depend on non-contact voltage detector to judge whether there's voltage in the lead. Detection operation may be impacted by factors like different socket designs and insulation thickness types etc.
4.6 AC/DC current measurement (DC current optional)
- Hold the trigger open clamp head and clamp one lead of the cable under test into the clamp head.
- When the measured signal is greater than 0.2A, the meter displays the measured current value.
- Before measuring DC current, press "ZERO" to set the reading to "0" before measuring.
4.7 DC voltage measurements
Connect the gauge pen to the measured signal. When the measured signal is greater than 0.5V, the meter will display the current DC voltage value. When the measured signal is less than 0.5V, the meter will default to the resistance value and display the internal resistance value of the measured signal. When the measured signal voltage is greater than 42V, the red backlight will turn on the alarm.
4.8 AC voltage measurement
Connect the meter pen to the measured signal. When the measured signal is greater than 0.5V, the meter will display the current AC voltage value and the frequency value of the voltage. When the measured signal is less than 0.5V, the meter will default to the resistance value and display the internal resistance value of the measured signal. When the measured signal voltage is greater than 42V, the red backlight will turn on the alarm.
4.9 Resistance, on off measurement
Pen connection resistance measurement, the instrument shows resistance, when measuring the resistance is greater than 60 M Ω instrument will show a "----".When measuring the resistance is less than 50 Ω, buzzer alarm lamp, instrument lights flashing.
4.10 Single watch pen wire test
- Press "SEL" button to switch to LIVE mode, and the meter shows LIVE.
- Insert the meter stylus into the power socket. If the instrument detects ac voltage, it will display the voltage signal strength based on the detected signal strength. The instrument induction LCD segment code will display the voltage signal strength. The red backlight is lit up and the buzzer emits an alarm of varying intensity.
4.11 Capacitance measure
- Press "SEL" key to switch to capacitance.
- Measure the capacitance with both ends of the test pen and read the measurement from the LCD.
Notice:
- When measuring large capacitors, it takes time to stabilize the reading.
- When measuring the capacitance with polarity, pay attention to the corresponding polarity to avoid damaging the instrument.
4.12 Temperature measurement
Press "SEL" key to switch TEMP, and the red plug of the thermocouple is inserted into the "°C/°F" end. The black plug can be inserted into the COM jack. When the reading is stable, the temperature value can be read directly from the display screen.
Notice:
- The maximum temperature of the randomly packaged K-type thermocouple is 250°C, and the instantaneous measurement can be up to 300°C.
5. Maintenance
5.1 Change battery
▲ Warning Before opening battery cover of the instrument, remove the pens from the circuit being measured, to avoid electric shock.
- If symbol [low battery symbol] appears, it indicates to change the battery.
- Unfasten bolts on battery cover of instrument and remove the cover.
- Change the old battery.
- Place the battery cover.
Note: Do not reverse battery polarity.
5.2 Change pens
▲ Warning When changing the pen, it is required to replace with an identical pen or a pen of the same level. The pen must be in good condition, and level of the pens is: 1000V 10A.
If insulation layer of the pen is damaged, or metal wire of the lead exposes, it is required to change the pen.
6. Accessories
- Pens: Level: 1000V 10A (One pair)
- Use Manual (One copy)
- Battery: 1.5V AAA battery (2PCS)
- Cloth bag (1PCS)
- K-Type: Optional function accessories (1PCS)
Related Documents
![]() |
ANENG ST180 4000 Counts Digital AC Current Clamp Multimeter Specifications and Features Detailed specifications and features of the ANENG ST180, a 4000 counts digital AC current clamp multimeter. Includes measurements for AC/DC voltage, AC current, resistance, capacitance, frequency, and temperature. |
![]() |
ANENG ST212 Clamp Meter: Specifications and Features Detailed specifications and features of the ANENG ST212 clamp meter, a versatile digital multimeter for AC/DC current, voltage, resistance, capacitance, frequency, and temperature measurements. Includes NCV detection, data hold, and auto power off. |
![]() |
ANENG ST181 Clamp Multimeter Specifications and Features Detailed specifications and features of the ANENG ST181 clamp multimeter, a digital AC/DC current meter suitable for electrical equipment testing and maintenance. |
![]() |
Mini Digital Clamp-On Multimeter Operating Instructions This document provides operating instructions, safety information, and specifications for the Mini Digital Clamp-On Multimeter. It covers measurements for AC/DC voltage, DC/AC current, resistance, audible continuity, diode, temperature, and Non-Contact ACV detection. |
![]() |
Aneng AN9002 Pro Digital Multimeter: Features, Specs, and Package Contents Comprehensive overview of the Aneng AN9002 Pro Digital Multimeter, detailing its advanced features like Bluetooth connectivity and True RMS, extensive measurement ranges for voltage, current, resistance, capacitance, frequency, and temperature, along with its package contents. |
![]() |
Intelligent Digital Multimeter User Manual User manual for the intelligent digital multimeter, covering safety information, product description, specifications, operation guide, maintenance, and accessories. Features include AC/DC voltage and current measurement, resistance, capacitance, temperature, and non-contact voltage detection. |
![]() |
Aneng 683 Pro Smart Digital Multimeter with Touchscreen Detailed specifications and features of the Aneng 683 Pro advanced digital multimeter, including its touchscreen VA display, wide measurement ranges for voltage, current, resistance, capacitance, frequency, and temperature, along with its automatic AC/DC detection and safety features. |
![]() |
ANENG PN200 Multifunctional Clamp Ammeter: Features, Specifications, and Usage Explore the ANENG PN200, a versatile AC/DC clamp ammeter for electrical testing. Learn about its functions including current, voltage, resistance, capacitance, frequency, NCV detection, and its key specifications and accessories. |