WOODPECKER HW-3H Ultrasonic Scaler Handpiece Instruction Manual
1 Introduction of product
1.1 Foreword
Guilin Woodpecker Medical Instrument Co., Ltd. is a professional manufacturer in researching, developing and producing dental equipment which has a wholesome quality assurance system. There are two brands of WOODPECKER company: WOODPECKER and DTE. Products include ultrasonic scaler, curing light, apex locator and ultrasurgery, etc.
1.2 Description of device
Handpiece of Ultrasonic Scaler HW-3H is the accessory of WOODPECKER series scalers, it is matched the WOODPECKER brand Ultrasonic scaler of our company. It is the handheld energy transformation part of ultrasonic scaler, by using piezoceramics to transfer electric energy to mechanical energy for eliminate the dental calculus.
1.3 Model and dimensions
Model: HW-3H
Dimensions: 117 mm (length), 18 mm (diameter), 47 g (weight)
1.4 Product performance and structure
Handpiece of Ultrasonic Scaler HW-3H is composed of cover and ultrasonic transducer.
1.5 Scope of application
Handpiece of Ultrasonic Scaler HW-3H is used for the dental calculus elimination when compatible with the scalers UDS-K, UDS-N2 which belong to WOODPECKER.
Handpiece of Ultrasonic Scaler HW-3H is used for the dental calculus elimination and root canal cleaning when compatible with the scalers UDS-L, UDS-M, UDS-N3, UDS-A, UDS-P, UDS-E which belong to WOODPECKER.
1.6 Contraindication
The hemophilia disease patient is not allowed to use this equipment.
The patients with heart pacemaker are forbidden to use this equipment.
The doctors with heart pacemaker are forbidden to use this equipment.
The heart disease patient, pregnant woman and children should be cautious to use the equipment.
1.7 The classification of the device
Operating mode: Continuous operation
Protection degree against water (used on the pedal): IPX1 [Drip-proof]
Degree of safety of application in the presence of a Flammable Anaesthetic Mixture with air or with Oxygen or Nitrous Oxide: Equipment not suitable for being used in the presence of a flammable anaesthetic mixture with air or with oxygen or nitrous oxide.
1.8 The main technical specifications
Power source Input: ~190V-230V
Output primary tip Vibration excursion: 1µm~100µm
Output tip Vibration frequency: 28kHz±3kHz
Output half-excursion force: 0.1N~2N
Tip's output power: 3W~20W
Water pressure: 0.01MPa~0.5MPa
1.9 Working condition
Environmental temperature: +5°C ~ +40°C
Relative humidity: 30%~75%
Atmospheric pressure: 70kPa ~106kPa
2 Installation of the product
2.1 Installation of handpiece
Open the packing box, make sure that all the parts and accessories are complete according to the packing list. Contact with dealer if necessary.
Keep dry when the detachable handpiece connects to the connector of the cable before installing the handpiece.
Make sure that the bottom jack point at the plug, make the handpiece fix to the cable correctly. (See Figure 1)
Figure 1: Illustration of handpiece components showing: Nipple, Handpiece decorative ring, Shaft, Handpiece, O-ring, CE mark, Brand, Autoclavable symbol, Connector of cable.
2.2 Installation of tips
Fix the tip to handpiece by torque wrench.
Please use "WOODPECKER" brand scaling tips.
The installing of the tip and endochuck toward direction as showed in Figure 2.
Figure 2: Illustration of tip and endochuck installation using a torque wrench and endo wrench. Shows screwing on the tip, endochuck, and file.
Notes: The detachable handpiece is a high-tech device which should be used carefully. Don't use too much torque force to screw the tip. Screw the tips by torque wrench. In order to prevent vibration by touch the foot switch by accident, unscrew the scaling tips after each operation.
3 Product function and operation
3.1 Scaling function
Connect the handpiece and the connector of cable of the main unit of the scaler correctly, and adjust the water and vibration intensity to a suitable level.
The handpiece can be handled in the same gesture as a pen in hand.
Select a suitable scaling tip as you need, screw it on the handpiece tightly by the torque wrench.
During the clinical treatment, be sure not to make the end of tip touch the teeth vertically.
Not to make the tip overexert on the surface of the teeth in case of hurting the teeth and damaging the tip.
After finishing operation, keep the machine working for 30 seconds on the water supply condition in order to clean the handpiece and the scaling tip.
3.2 Endo function
Usage process:
- Fix endochuck to handpiece by endochuck.
- Unscrew the screw cap on the endochuck.
- Put the ultrasonic file into the hole in the front of endochuck.
- Screw down the screw cap with endo wrench to tight up the ultrasonic file.
- Press option key, turn to endo root canal cleaning function.
- Press option key, turn the water and vibration intensity to a suitable level.
- Put the ultrasonic file into the patient's root canal slowly, step on the foot pedal, then make endo root canal cleaning. During the treatment, turn up the power gradually according to the needs.
Notes:
- When fixing endochuck, it must be screwed down.
- The screw on the endochuck must be screwed down.
- Don't press it too much when the ultrasonic file in root canal.
- Don't start up the foot pedal until the ultrasonic file is in root canal.
- The power range is supposed from 1st to 5th grades.
- Please sterilize the endo chuch and handpiece after operation.
3.3 Instruction for main components of detachable handpiece
Nipple: The nipple can be removed, for easy cleaning.
Handpiece: The main part of the whole handpiece, can be autoclaved under high temperature and pressure.
O ring: Please lubricate the waterproof O ring with dental lube frequently, as sterilization and repeated pulling and inserting will reduce their using life. Change a new one once it is damaged or worn excessively.
4 Notes
Don't knock or rub the handpiece.
Don't screw or unscrew the scaling tip and endochuck when step on the foot pedal.
The handpiece, scaling tip, torque wrench, endo wrench and endochuck must be sterilized before each treatment.
The internal screw threat of the scaling tips produced by some manufactures is coarse, rusty and collapsed. This will damage the external screw threat of the handpiece irretrievably. Please use "WOODPCKER" brand scaling tips.
We are only responsible for the safety on the following conditions: The maintenance, repair and modification are made by the manufacturer or the authorized dealer.
5 Troubleshooting
Fault | Possible causes | Solutions |
---|---|---|
Handpiece doesn't work | The tip is loosen | Tighten it |
Something wrong with the Handpiece | Take out the handpiece and send it back for repair | |
Something wrong with the Cable | Contact our dealers or us | |
The connect plug between the cable and circuit board is loosen | Contact our dealers or us | |
There is no spray when getting through the electricity | Turn on the water control switch is vibrates not on | Turn on the water control switch |
There is no water when getting through the electricity | No pressure or low pressure | Check the presure |
The water control switch is vibrates not on | Turn on the water control switch | |
Low vibration | There is impurity in the electric-magnetic valve | Contact our dealers or us |
The handpiece generates heat | The tip hasn't been screwed on the handpiece tightly | Screw the tips tightly by torque wrench |
The tip is shaken loose by vibration | Screw on the tip tightly | |
There is water seeping from the coupling between the handpiece and the cable | The coupling between the handpiece and the cable isn't dry | Dry it, especially the water between handpiece and connector |
The tip is worn out or deformation | Change another tip | |
The water control switch is in a low setting | Turn the water control switch to a higher grade | |
The waterproof "O ring is damaged | Change a new "O" ring |
If the problem still can't be solved, please contact with local dealer or manufacturer.
6 Cleaning and sterilization
Clean and sterilize the handpiece after using it.
Clean the handpiece by pure water before autoclaved, dry it by three-way syringe.
Autoclaved under high temperature, pressure, time: 134°C, 2.0bar~2.3bar (0.20MPa~0.23MPa), 4min;
Pack the handpiece with sterile gauze or sterile bag before sterilizing. Reuse handpiece after it cools naturally in case of burning hand.
6.5 Notice
Clear the cleaning liquid in the handpiece with three-way syringe before sterilization.
Be sure that the scaling tip has been unscrewed from the handpiece and it cannot be sterilized with others.
Please notice whether the outer of the handpiece is damaged during the treatment or sterilization.
Please don't smear any protective oil on the surface of handpiece.
There are two waterproof "o" rings at the end of handpiece. Please lubricate them with dental lube frequently, as sterilization and repeated pulling and inserting will reduce their using life. Change a new one once it is damaged or worn excessively.
6.6 The following sterilizing methods are forbidden:
- Put handpiece into any liquid for boiling.
- Dip handpiece in disinfectors such as iodine, alcohol and glutaraldehyde.
- Put handpiece into oven or microwave oven for baking.
- Put handpiece into ultrasonic cleaner to clean.
7 Storage, maintenance and transportation
7.1 Storage, maintenance
The equipment should be handled carefully and lightly. Be sure that it is far from the vibration, and is installed or kept in a cool, dry and ventilated place.
Don't store the machine together with the articles that are combustible, poisonous, caustic or explosive.
This equipment should be stored in a room where the relative humidity is 10%-93%, atmospheric pressure is 70kPa to 106kPa, and the temperature is -20°C ~ +55°C.
7.2 Transportation
Excessive impact and shake should be prevented in transportation. Lay it carefully and lightly and don't invert it.
Don't put it together with dangerous goods during transportation.
Avoid solarization and getting wet in rain or snow during transportation.
8 Environmental Protection
There is not any harm factor in our product. You can deal with it based on the local law.
9 After Service
We offer one year's free repair to the equipment according to the warranty card. The repair of the equipment should be carried out by our professional technician. We are not responsible for any irretrievable damage caused by the non-professional person.
10 Symbol instruction
- WOODPECKER: Trademark
- IPX1: [Drip-proof]
- 134°C: [Autoclavable 134°C]
- SN: Serial number
- CE 0197: CE marked product
- EC REP: Authorised Representative in the EUROPEAN COMMUNITY
- [Consult the accompanying documents]
- [Used indoor only]
- [Appliance compliance WEEE directive]
- Atmospheric pressure for storage: 70kPa - 106kPa
- Humidity limitation for storage: 10% - 93%
- Temperature limitation for storage: -20°C ~ +55°C
11. Statement
All rights of modifying the product are reserved to the manufacturer without further notice. The pictures are only for reference. The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO., LTD. The industrial design, inner structure, etc, have claimed for several patents by WOODPECKER, any copy or fake product must take legal responsibilities.
Be sure that the connecting side of handpiece and seal socket is completely dried before installing handpiece each time.
Illustration showing the connecting side of handpiece and seal socket.
Contact Information
Guilin Woodpecker Medical Instrument Co., Ltd.
Information Industrial Park, Guilin National High-Tech Zone, Guilin, Guangxi, 541004 P. R. China
Europe Sales Dept.: +86-773-5873196, +86-773-2125222
North America, South America & Oceania Sales Dept.: +86-773-5873198, +86-773-2125123
Asia & Africa Sales Dept.: +86-773-5855350, +86-773-2125896
Fax: +86-773-5822450
E-mail: woodpecker@glwoodpecker.com, sales@glwoodpecker.com
Website: http://www.glwoodpecker.com
ECREP MedNet GmbH, Borkstrasse 10, 48163 Muenster Germany
Document Version: ZMN/WI-09-163 V1.8-20180927