TN420 Portable Turbidity Meter Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de

TABLE OF CONTENTS
1. INTRODUCTION...................................................................................................................... - 1 2. TECHNICAL SPECIFICATIONS ............................................................................................... - 2 3. INSTRUMENTATION ILLUSTRATION ..................................................................................... - 3 -
3.1 Overview............................................................................................................................ - 3 3.2 Configuration ..................................................................................................................... - 4 3.3 Display Mode ..................................................................................................................... - 4 3.4 Keypad .............................................................................................................................. - 5 3.5 Power Supply..................................................................................................................... - 6 4. METER CALIBRATION ............................................................................................................ - 6 4.1 Preparation for Calibration ................................................................................................. - 6 4.2 Calibration Procedure (Take 0 NTU and 20 NTU as an example) ...................................... - 8 4.3 Notes for Calibration .......................................................................................................... - 9 5. Turbidity Measurement ........................................................................................................... - 10 5.1 Sample Vial Handling....................................................................................................... - 10 5.2 Measurement Preparation................................................................................................ - 10 5.3 Measurement Mode ......................................................................................................... - 10 5.4 Notes for Measurement.................................................................................................... - 11 6. Lamp replacement.................................................................................................................. - 12 6.1 Light Source Check.......................................................................................................... - 12 6.2 Lamp Replacement .......................................................................................................... - 12 7. WARRANTY ........................................................................................................................... - 13 8. REPLACEMENT PARTS ........................................................................................................ - 13 -
*Please fully charge the battery before first-time use.
V1.1

1. INTRODUCTION
Thank you for purchasing Apera Instruments TN420 Turbidity Meter (hereafter referred to as the instrument). The instrument uses tungsten filament lamp (400 ­ 600nm) as the light source and 90° scattering method, which is compliant with U.S EPA 180.1 method for the determination of of turbidity in drinking, ground, surface, and saline waters, domestic and industrial wastes. It is suitable for onsite and laboratory use.
The instrument provides accurate results with simple measurement and calibration, giving you unparalled confidence in turbidity test results and easy of use. The main functions and features include:  Intelligent functions such as automatic calibration, operation navigation, parameter setup,
automatic power off, low voltage indication, and a replaceable tungsten filament lamp.  Large TFT color screen with blue background for measurement mode, green background for
calibration mode.  Operation guidance and reminders in the process of calibration, measurement and parameter
setting.  Average measurement mode automatically takes multiple consecutive readings, and calculate
the average from 10 times of measurement, which is ideal for sample solutions with rapid settling and continuous measurement changes.  3.7V rechargeable lithium battery can supply power for over 20 hours of continuous measurement. The battery life is 5 times longer compared with tungsten lamp turbidimeter in the market using AA alkaline batteries.  Meets IP67 rating, ideal for use under harsh environment and field.  Comes with a carrying case, which includes the meter, calibration solutions, and other accessories, convenient for use and carry.
- 1 -

2. TECHNICAL SPECIFICATIONS

Measurement Method Light Source

90° scattering measurement
Tungsten (400 ­ 600nm) filament lamp, compliant with US EPA 180.1 Method

Measuring Range

0 to 1000 NTU (FNU), automatic range switch 0.01 to 19.99 NTU (FNU) 20.0 to 99.9 NTU (FNU) 100 to 1000 NTU (FNU)

Accuracy Repeatability
Resolution Stray Light

 ±2% of reading+ stray light  ±1% of reading or 0.02 NTU (FNU) (whichever is greater)
0.01/0.1/1 NTU (FNU) 0.02 NTU (FNU)

Calibration Standard

AMCO Polymer solution or Formazin Solution: 0, 20, 100, 400 and 800 NTU (FNU)

Detector Measuring Mode
Display Sample Vial Sample Vial Volume Power supply

Silicone photovoltaic Normal measurement; Average Measurement
TFT color screen Ø25×60 mm, high borosilicate glass with lid
18 ml 3.7V Rechargeable lithium battery

Working Condition

Temperature: 0 to 50°C (32°F to 122°F) Relative humidity: 0 to 90% at 30°C; 0 to 80% at 40°C;
0 to 70% at 50°C; no condensation

Storage Condition
Instrument sealing grade
Certificates Limited Warranty

Instrument: -40 to 60°C (-40 to 140°F) Calibration Solution: 5 to 30°C (41 to 86°F)
IP67
CE and RoHS 2 years

Dimension and Weight

Meter: (90×203×80) mm/ 385 g Test Kit: (310×295×110) mm/ 1.5 kg

- 2 -

3. INSTRUMENTATION ILLUSTRATION

3.1 Overview























Diagram-1



Flip cover of the sample vial holder

 Dust-proof plug (Take off the plug when

(Close the cover when measuring)

measuring)



Housing



Sample vial holder



Display



Calibration vials or sample vials

Positioning mark (Align the mark with



Keypad



the mark on the calibration vial or

sample vial)



USB Port

 Built-in lithium battery (Rechargeable)



Lamp cover

- 3 -

3.2 Configuration

    
Diagram-2
3.3 Display Mode
Measuring Mode



Calibration Solutions 0.0, 20.0, 100, 400, 800 NTU





Carrying case



TN420 Turbidimeter



Microfiber cloth







Power adaptor (5V 1A) Silicone oil (10 ml)



Sample vials×6





USB Cable (under the meter)





User Manual

 Measurement Mode  Measurements and unit  Operation Guide  Battery icon

Calibration Mode

 Calibration solution  Measurements and unit  Progress bar  Operation guide

- 4 -

Calibration Menu
3.4 Keypad

 Calibration mode  Standard values  Operation guide  Finished calibration indicator

Diagram 3

Key

Functions

 Power on/off
 In measurement mode: press to enter calibration mode  In calibration mode: press to exit calibration mode  In measurement mode: long press to switch measurement units NTU-FTU
 In measuring mode: short press the key to perform normal measurement; long press the key to perform average measurement
 In calibration mode: press the key to confirm the calibration.

 In calibration mode, press the key to select standard solution.

- 5 -

3.5 Power Supply

The instrument adopts 3.7V rechargeable lithium battery. Fully charge the battery before first-
time use. a) Charging mode

· Charge via Power adaptor: connect instrument and power adaptor with a USB cable.

Adaptor specification: AC100 to 240V, 50/60Hz, output: 5V/1A.

*Note: Please use the power adapter we provide in the kit to charge your turbidimeter.

· Charge via Computer: connect instrument and a computer with a USB cable.

Under normal circumstances, it is recommended to use the power adapter to charge the

lithium battery to ensure the performance. When the lithium battery voltage is lower than

3V, the instrument will shut down, and it's time to recharge it. b) Battery capacity indication

· Battery capacity icon:

, please recharge the battery when

icon displays to ensure measuring accuracy; when icon displays, the battery must be

recharged, otherwise the instrument cannot work properly.

· If charging the instrument when it is turned on,

icon will be displayed. Users can

continue using the instrument while it is being charged.

· If charging the instrument when it is turned off, "Charging......" will be displayed, and

"Charging is completed" will be displayed after a full charge.

4. METER CALIBRATION
4.1 Preparation for Calibration a) Calibration Point
The instrument has 5 calibration points: 0 NTU, 20 NTU, 100 NTU, 400 NTU, and 800 NTU. Among them, 0 NTU point uses AMCO 0.0 NTU calibration solution or laboratory distilled water, and the remaining 4 calibration points use AMCO polymer solutions. Note that the cap of the 0.0 NTU solution vial can be unscrewed. After the solution is invalidated, users can replace the 0.0 NTU calibration solution or use fresh laboratory distilled water. The AMCO 0.0 NTU calibration solution refill can be purchased from the supplier; For the remaining 4 calibration solutions, their vial caps cannot be opened. Simply dispose the solutions after they are expired or invalidated and buy new ones from your supplier to replace. See the detail in Section 8 Replacement Parts. b) Replace zero turbidity solution
- 6 -

· Open the 0.0 NTU vial cap, pour out the original solution, add 1/2 distilled water, screw on the cap and shake the vial to rinse it off and pour out the water. Repeat it 3 times. Shake off the distilled water in the vial. Pour in new AMCO 0.0 NTU calibration solution or fresh laboratory distilled water. Then close the vial cap.
· AMCO 0.0 NTU calibration solution has 6 to 12 months of shelf-life, distilled water is only valid for a couple of days.
AMCO 0.0 NTU solution is provided in the default kit. For purchase details, please refer to Section 8 Replacement Parts. c) Clean vial surface Apply a small drop of silicone oil on the surface of the vial and wipe it off with a lint-free cloth to evenly distribute the silicone oil on the surface in order to cover smudges and scratches, which helps light scattering. But please pay attention to the following points:  The silicone oil applied should not be too much. After wiping with lint-free cloth, please wipe
with filter paper or high-quality tissue paper to clean off. Excessive residual silicone oil on the vial surface will affect the measurement accuracy.  It's not necessary to use silicone oil for each calibration and measurement. Apply silicone oil every several days or once a week. In between, just clean the surface with filter paper or high-quality tissue paper.  Clean the calibration vial and sample vial together and keep the steps and actions consistent to achieve same degree of cleanliness.  Stability of calibration solutions The U.S EPA approved AMCO polymer standard calibration solution is very uniform and stable. It does not precipitate, drift or condense. Generally, it can be used directly without shaking or flipping the vial (to make the solution even). For AMCO polymer calibration solutions that have not been used for a long time, slowly flip the vial twice and let it stand for 2 minutes. Be careful not to shake the solution vigorously, as this will create air bubbles which will destabilize the measurement; For 0.0 NTU calibration solution, do NOT shake or flip. If using Formazin calibration solution, as it tends to precipitate easily, each time users must flip and shake the vial to make the solution even. But sediment can still occur during tests and would make the measurement unstable. Users need to have rich experience handling Formazin calibration solutions.
- 7 -

4.2 Calibration Procedure (Take 0 NTU and 20 NTU as an example)

1) Instrument warm up (only required for low-range or high-accuracy measurement): Power on

the instrument. Long press

to start continuous measurement (don't insert the vial), then

wait for 3 to 5 minutes. Then short press

to exit continuous measurement.

2) Open the flip cover and place the 0.0 NTU calibration vial. Align the arrow on the vial lid with the arrow on the sample vial holder and close the flip cover.

3) Press

to enter the calibration menu, the cursor

is at 0 NTU. Press

to start calibration.

4) After calibration is done, press

to confirm.

The instrument will return to the calibration menu (see Diagram on the right side). The OK sign indicates that 0 NTU has been calibrated and prompt to calibrate 20 NTU.

5) Place the 20.0 NTU calibration vial in the sample

vial holder, press

to start 20 NTU calibration.

After calibration is done, press

to confirm. The

instrument will return to the calibration menu (see diagram on the right side). The OK sign indicates that 20 NTU has been calibrated.

6) If you want to continue to calibrate 100, 400 and 800 NTU, place the 100 NTU calibration

vial in the sample vial holder and press

to calibrate. After calibration is done, press

to confirm; repeat the steps to calibrate 400 NTU and 800 NTU. After calibration is

done, press

to return to measurement mode.

- 8 -

7) Press

to exit calibration mode, the instrument

will return to measurement mode as shown in Diagram on the right side.

4.3 Notes for Calibration a) Calibration point verification: The calibration point can be verified after the calibration is
completed. If the calibration point has a large error, enter the calibration mode and repeat the calibration. For calibration point accuracy, users can refer to the following standards:

Calibration point 0 NTU 20 NTU
100 NTU 400 NTU and 800 NTU

Accuracy for reference  0.05 NTU  ±0.2 NTU  ±2 NTU  ±5 NTU

b) Calibration point selection: The instrument has been calibrated at full range before leaving the factory. For subsequent use, you can select 2 or more points as needed, as long as the estimated
measurement range is between the two calibration points. In calibration setup mode, press
or to select the calibration point. c) Low turbidity calibration requirement
a) For low turbidity measurements (measurement less than 2 NTU), please test 0.0 NTU calibration solution. If the accuracy is not meeting requirements, calibrate the instrument at 0.0 NTU and 20.0 NTU before test; then use 1# or 2# sample vial for measurement.
b) Using the same sample vial to calibrate and measure can eliminate the error caused by different vials, thus achieving higher accuracy. For example, add laboratory distilled water to 1# vial for calibration and then add sample solution to 1# vial for measurement. Note that the solution vial should be rinsed thoroughly when changing solutions.
d) High turbidity calibration requirement: For turbidity measurement greater than 2 NTU, it is recommended to calibrate once a week, or to test a calibration solution close to the sample solution. If the error is large, the instrument needs to be recalibrated.
e) The instrument does not automatically recognize the calibration solution. If the wrong solution is selected for calibration, the measurement will be completely wrong. In this case, it can be restored by recalibrating with the correct calibration solutions.

- 9 -

f) Place the instrument on a flat and level surface. Do not hold the instrument in hand while operating.
g) If using Formazin standards for calibration, please make sure to use the freshly made Formazin standard to ensure calibration accuracy.

5. Turbidity Measurement

5.1 Sample Vial Handling a) 6 sample vials are included in the test kit. The cap is marked with 1# to 6#, and the bottom of the
vial also has the same number. The number of the vial and the cap should always be the same. *Pay attention that 1# and 2# vials are only for low turbidity solution measurement. (< 2 NTU) b) The vial has been rigorously cleaned and sterilized. They can be used directly for the first time. For subsequent uses, follow the steps below to perform a thorough cleaning.  Clean the sample vial inside and outside with detergent  rinse with distilled water or
deionized water multiple times  Rinse the vial twice with the sample solution  Pour the sample solution into the vial  Close the cap.

5.2 Measurement Preparation

a) Collect the sample solution with a clean container and add the

solution to the 4/5 of the vial (about 18ml), see Diagram 4. Then

close the lid.

b) Before the measurement, users can slowly flip the sample vial a

few times and let it stand for 2 to 5 minutes to eliminate potential

air bubbles (see Diagram 5).

Diagram 4

Diagram 5

c) Clean off the surface of the vial to ensure it is dry, clean and free of stains. Apply a small drop

of silicone oil on the surface of the vial and wipe it off with a micro-fiber cloth. Then wipe again

with filter paper or high-quality tissue paper. See section 4.1(c) for details.

5.3 Measurement Mode a) Normal Measurement Mode

Press

, the screen will display the progress bar, and the measured value will be

displayed in 20 seconds. To take the next measurement, press b) Average Measurement Mode

again.

Long press

button, release it until you hear a beep, then the instrument enters average

reading mode. In this mode, the instrument will perform 10 times continuous measurement, measure and display 1 data every 20 seconds, and displays the average value at the end, see diagram 6. The average measurement mode can be used for observing the stabilizing process of turbidity and is ideal for testing rapid-settling solutions.

- 10 -

Diagram 6
5.4 Notes for Measurement a) Keep the sample stable: After the vial is placed into the sample cell, it is recommended to wait
for 1 to 2 minutes before calibration, as the solution will experience some shaking when the vial moves, which may result in inaccurate measurements. b) Sample Vial cleaning requirement: Sample vial must be rigorously cleaned and free from smudges and scratches. When wiping, user should grip the cap and bottom to avoid leaving fingerprints on the surface of the vial. Its surface should be applied with a drop of silicone oil be wiped with a micro-fiber cloth. After that, please clean with filter paper or high-quality tissue paper. See section 4.1(c) for details. c) Mixing and Degassing: Samples should not be vigorously shaken or vibrated. It is recommended that users gently shake the sample vial to make solution evenly distributed. Air bubbles in solution will cause big error to turbidity measurement. So, the vial should be left stand still for 2 to 5 minutes to eliminate potential air bubbles before measuring. But mixing and degassing simultaneously is a difficult process to handle, especially for solution with precipitates, which requires some operating experience or making some limits in test conditions, for example, limiting the mixing condition and waiting time for degassing to be the same before comparing measurements. d) Other Requirements  On the premise of ensuring evenly distributed samples, sample solution should be
measured immediately to prevent temperature changing and precipitates from affecting measurements.  Avoid sample dilution for measurement as much as possible.  Avoid operating under direct sunlight.  Do not pour solution into the vial holder. Sample vials must be used for measurement.  Please do not wash the vial holder as this may damage its optical structure.
- 11 -

6. Lamp replacement

6.1 Light Source Check

Observe the light source on the left side of the sample vial holder as in Diagram 11. Turn on the

instrument, and press

. The light source will light up for 5 seconds. If it's not lighting up or it's

flashing, please replace the lamp.

6.2 Lamp Replacement

The tungsten filament lamp module (sold separately) is shown as in Diagram 12. The lamp hole and

socket are as in Diagram 13. Diagram 14 shows you how to replace the lamp.

socket

Light source

tungsten filament lamp

gasket lamp cover

lamp hole

Diagram 11 Check light source

Socket Diagram 12 Tungsten filament lamp module

Diagram 13 lamp hole and socket

Unscrew the lamp cover with a coin

Use a tweezer to unplug the socket

Use a tweezer to unscrew the lamp cover.

Take out the old tungsten filament lamp module

Install the new tungsten filament lamp module.

Close the lamp cover and screw tight with a tweezer.

Use a tweezer to plug wire into the gap and plug in the socket

- 12 -

Screw on the lamp cover tightly

7. WARRANTY
We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of TWO YEARS from the delivery.
This limited warranty does not cover any damages due to: Transportation, storage, improper use, failure to follow the product instructions or to perform any preventive maintenance, modifications, combination or use with any products, materials, processes, systems or other matter not provided or authorized in writing by us, unauthorized repair, normal wear and tear, or external causes such as accidents, abuse, or other actions or events beyond our reasonable control.

8. REPLACEMENT PARTS

Model T500-2 T500-1 T500-3 T500-5 T500-4 TN400-S3

Name
AMCO 0.0 NTU calibration solution 20/100/400/800 NTU AMCO polymer Sample vials with lid
Tungsten filament lamp Rechargeable lithium battery
Silicone oil

Description
0.0 NTU/100mL Compatible with TN500&TN420 25×60mm1# to 6#
/
3.7 V 10mL

Quantity 1 bottle 4 bottles 6 pcs
1 pc 1 pc 1 bottle

APERA INSTRUMENTS (Europe) GmbH Address: Wilhelm-Muthmann-Str.18 42329 Wuppertal, Germany Tel: +49 202 51988998 Email: info@aperainst.de Website: aperainst.de
- 13 -

