FSP500M-K47 Series: 450-500W Medical Open Frame Power Supply
Technical Datasheet
Product Image
A photograph of the FSP500M-K47 Series power supply unit, showing its open-frame construction and internal components.
Features
- Compact size 4 x 7 x 1.7 inches
- Certified medical safety IEC 60601-1
- High altitude 5000 meters operation
- Power Fail Detect (PFD) signal
- Inhibit - TTL high to disable output
- BF Class insulation
- Meet EN55011
- Over voltage protection
- Over current protection
- Over temperature protection
- Compliant with RoHS requirement
Safety Standard Approval
Approvals include TÜV Rheinland, CE, with certification number E178020.
Description
The FSP500M-K47 series is a Class-I design, 4 x 7 inch AC/DC switching power supply. It delivers 450-500 watts of continuous output power with 30 CFM forced air cooling, or 350-400 watts with convection cooling. The unit is constructed on a printed circuit board with a U-bracket or enclosed form for mechanical support. All models meet EN55011 and FCC class B emission limits and are designed for medical applications.
Input Specifications
- Input voltage:
- 90-264 VAC
- Input frequency:
- 47-63 Hz
- Input current:
- < 5.2 A (rms) for 115 VAC, < 2.6 A (rms) for 230 VAC
- Earth leakage current:
- < 300 μA@264 VAC, 63 Hz
- Touch current:
- < 100 μA@264 VAC, 63 Hz
Output Specifications
- Output voltage/current:
- See rating chart
- Maximum output power:
- See rating chart
- Remote sense:
- Compensation for cable losses up to 0.5 V
- Protection:
- Over voltage: Provided on output. Set at 112% to 140% of its nominal output voltage.
- Over current: The power supply will shut down without damage and enter auto-recovery mode.
- Over temperature: The power supply will enter into shut down while the abnormal thermal rise occurs.
- Temperature coefficient:
- All outputs ±0.04% /°C maximum.
- Transient response:
- Maximum excursion of 4%, recovering to 1% of final value within 500s after a 25% step load change.
- Fan power:
- 12 V at 300 mA maximum
- Standby power:
- 5 V at 500 mA maximum
General Specifications
- Switching frequency:
- 85 KHz (typical)
- Power factor:
- 0.98 typical at 115 VAC
- Efficiency:
- See rating chart
- Hold-up time:
- 20 ms minimum at 110 VAC
- Line regulation:
- ±0.5% maximum at full load
- Inrush current:
- 30 A @ 115 VAC, or 60 A @ 230 VAC, at 25°C cold start
- Operating altitude:
- 5000 meters
- Withstand voltage:
- 4000 VAC from input to output (2 MOPP), 1500 VAC from input to ground (1 MOPP), 1500 VAC from output to ground
- MTBF:
- 100,000 hours at full load at 25°C ambient, calculated per MIL-HDBK-217F
EMC Performance (IEC60601-1-2)
- EN55011:
- Class B conducted, class B radiated
- EN61000-3-2:
- Harmonic distortion, Class D
- EN61000-3-3:
- Line flicker
- EN61000-4-2:
- ESD, ±15 KV air and ±8 KV contact
- EN61000-4-3:
- Radiated immunity, 10 V/m
- EN61000-4-4:
- Fast transient/burst, ±2 KV
- EN61000-4-5:
- Surge, ±1 KV diff., ±2 KV com.
- EN61000-4-6:
- Conducted immunity, 10 Vrms
- EN61000-4-8:
- Magnetic field immunity, 30 A/m
- EN61000-4-11:
- Voltage dip immunity, 30% reduction for 500 ms, 60% reduction for 100 ms, and >95% reduction for 10 ms
Environmental Specifications
- Operating temperature:
- -10°C to +70°C
- Storage temperature:
- -40°C to +85°C
- Operating humidity:
- 10% to 90% RH non-condensing
- Storage humidity:
- 5% to 95% RH non-condensing
- Temperature derating:
- Derate from 100% at +50°C linearly to 50% at +70°C, applicable to convection and forced-air cooling conditions.
Output Voltage/Current Rating Chart
Model(1) | V1 | Min. Current at convection | Max. Current at 30 CFM | Tolerance | Ripple & Noise(2) | Max. Power(3) | Average Active Efficiency (typical) @115 / 230 VAC |
---|---|---|---|---|---|---|---|
FSP500M-K47-12B | 12 V | 0 A | 29.17 A | ±2% | 120 mV | 350 W / 450 W | 88% / 90% |
FSP500M-K47-15B | 15 V | 0 A | 23.34 A | ±2% | 150 mV | 350 W / 450 W | 88% / 90% |
FSP500M-K47-18B | 18 V | 0 A | 22.23 A | ±2% | 180 mV | 400 W / 500 W | 88% / 90% |
FSP500M-K47-24B | 24 V | 0 A | 16.67 A | ±2% | 240 mV | 400 W / 500 W | 89% / 91% |
FSP500M-K47-28B | 28 V | 0 A | 14.29 A | ±2% | 280 mV | 400 W / 500 W | 89% / 91% |
FSP500M-K47-36B | 36 V | 0 A | 11.12 A | ±2% | 360 mV | 400 W / 500 W | 89% / 91% |
FSP500M-K47-48B | 48 V | 0 A | 8.34 A | ±2% | 480 mV | 400 W / 500 W | 89% / 91% |
FSP500M-K47-57B | 57 V | 0 A | 7.02 A | ±2% | 570 mV | 400 W / 500 W | 89% / 91% |
Notes:
- Change suffix "B" for U-bracket form to "C" for enclosed form with cover and fan assembly, e.g. FSP500M-K47-12C.
- Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output.
- 350 W or 400 W of max. output power is at convection cooling, 450 W or 500 W is with 30 CFM forced air provided by user for "B" version. 450 W or 500 W for "C" version with cover and fan assembly.
Interface Signals
- PFD:
- TTL logic high for normal operation and TTL logic low upon loss of input power. Turn-on delay time 100-1000 ms, turn-off delay time 1 ms minimum.
- Inhibit:
- TTL high level signal to inhibit output.
Output Power Derating Curve
A line graph illustrating the output power derating based on ambient temperature. The graph shows output power percentage on the y-axis and ambient temperature in degrees Celsius on the x-axis. The curve starts at 100% power at 50°C and decreases linearly to 50% power at 70°C.
Mechanical Specifications
Diagrams illustrating the mechanical dimensions and mounting hole locations for both the U-bracket form and the enclosed form of the FSP500M-K47 Series power supply.
Notes:
- 1. Dimensions:
- Shown in inches [mm].
- 2. Tolerance:
- 0.02 [0.5] maximum.
- 3. Input connector P1:
- Dinkle terminal P/N DT-35C-B01W-03, with nickel plated M3 screws.
- 4. Output connectors P2 and P3:
- For M4x0.7 screw connections.
- 5. Output connector P4:
- Molex header 87833-08 or equivalent, mating with Molex housing 51110-0850 or equivalent.
- 6. Fan connector P5:
- JST header S2B-ZR-3.4 or equivalent, mating with JST housing ZHR-2 or equivalent.
- 7. Screw penetration depth:
- Maximum 4 mm from the outer surface of chassis.
- 8. Weight:
- 1.0 Kgs (2.23 lbs.) approx. for U-bracket form, 1.14 Kgs (2.52 lbs.) approx. for enclosed form.
Pin Chart
Connector P1
Pin No. | 1 | 2 | 3 |
---|---|---|---|
Polarity | Ground | Live | Neutral |
Connector P5
Pin No. | 1 | 2 |
---|---|---|
Polarity | +V1 | Common Return |
Connector P4
Pin No. | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 |
---|---|---|---|---|---|---|---|---|
Polarity | Common Return | +V1 Sense | -V1 Sense | PFD | Inhibit | +5V Standby | NC | NC |