Welch Allyn® KleenSpec® 78900 and 78950 Corded Illumination System
Quick Start Guide
Introduction
The KleenSpec Corded Illumination System is designed to work with the following Welch Allyn specula: 58000S, 58001S, 58004S, 590XS, 59000, 59001, 59004, 59005, and 59006 for gynecological and obstetric procedures.
The power supply provides power from the hospital-grade wall outlet to the illuminator assembly, which houses the lamp. This provides illumination to the specula. This device is intended for use only with Welch Allyn-appropriate accessories.
This product is for use with the Welch Allyn KleenSpec Vaginal Specula only. Do not use the illuminator for any other use. The lamp is hot and can cause a burn if used outside of the vaginal specula.
The illuminator slides into the appropriate Welch Allyn KleenSpec Vaginal Specula listed above. Verify all connections are made prior to applying power with the on/off switch.
Diagram Description: The Welch Allyn KleenSpec Corded Illumination System consists of an illuminator (a pen-like device with a tip), a connecting cord with an in-line on/off switch, and a power supply adapter that plugs into a wall outlet.
For a complete copy of the directions for use, visit www.welchallyn.com.
For a printed copy of the directions for use, contact Welch Allyn Technical Support: (www.welchallyn.com/support), or visit www.welchallyn.com/about/company/locations.htm.
Inserting the Illuminator
Diagram Description: Two illustrations demonstrate the insertion of the illuminator into specula. The left illustration shows the illuminator being inserted into a Model 580 series speculum, with a 'Sheath' label pointing to a part of the speculum. The right illustration shows the illuminator being inserted into a Model 590 series speculum.
Classification
- Class II, Type BF Applied Part
- Ordinary Equipment
- Clean and disinfect as directed in Cleaning Section
- Non-continuous Operation
Symbol Descriptions
Symbol/Description | Textual Representation |
---|---|
Warning | ⚠️ Warning |
Caution | ⚠️ Caution |
Consult Directions for Use | ¡️⃣ Consult Directions for Use |
Lamp On | ? Lamp On |
Lamp Off | ? Lamp Off |
Caution, high temperatures | ?️ Caution, high temperatures |
Caution, high intensity light | ? Caution, high intensity light |
Type BF Applied Part | [Type BF Applied Part] |
Double Insulation | [Double Insulation] |
IPX7 (Illuminator) | Illuminator is protected against the effects of temporary immersion in water (and the solutions described in this manual) |
Direct Current | [DC Symbol] |
IPX0 (Power supply) | Power supply - Not protected from the ingress of water. |
Warnings and Cautions
⚠️ WARNING: To avoid eye injury, do not look into the beam.
⚠️ WARNING: Lamp is harmful to skin if touched. Allow lamp to cool for 5 minutes before replacing. Lamp generates heat and should remain off until just prior to and immediately after use. Do not place illuminator onto a flammable surface if lamp is illuminated.
⚠️ WARNING: Disconnect power supply from outlet prior to any maintenance.
⚠️ WARNING: This product is for use with the Welch Allyn KleenSpec Vaginal Specula only. Do not use the illuminator for any other use. Lamp is hot and can cause a burn if used outside of the vaginal specula.
⚠️ WARNING: KleenSpec Corded Illumination System is not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.
⚠️ WARNING: The lamp is hot and can cause burns if used outside of the vaginal specula.
⚠️ CAUTION: Properly secure all cords and cables to minimize trip hazards.
⚠️ CAUTION: Use only the cleaning methods and disinfection solutions/methods described in this manual. Clean the device prior to disinfection.
⚠️ CAUTION: The Illuminator is a biohazard after use. Follow good hospital practices and the instructions for cleaning and disinfection of the illuminator within this manual prior to the next usage.
⚠️ CAUTION: Ensure lamp is installed and fully seated in illuminator throughout cleaning and disinfection process. The illuminator assembly is immersible (IPX7) provided the lamp is properly installed.
⚠️ CAUTION: This device complies with current required standards for electromagnetic interference, and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using this device in close proximity to other equipment.
⚠️ CAUTION: The illuminator cable is capable of temporary immersion to a depth of 0.15 to 1 meter, at ambient pressure, at room and fluid temperatures of 22° C to 43° C. The power supply is not. Do not immerse the power supply.
U.S. Federal Law restricts this device to sale by or on the order of a physician.
To remove all power from the Vaginal Illumination System, unplug the external power supply from the wall outlet.
There are no user-serviceable parts inside the device except the lamp.
Contact Welch Allyn Technical Support: www.welchallyn.com/support.
Cleaning and Disinfecting
Cleaning and Disinfecting Table
Procedure | KleenSpec Corded Illuminator | Power Supply |
---|---|---|
Cleaning | ||
CaviWipe (Wipe) | Yes | Yes |
Enzol (Soak) | Yes | No* |
Metrizyme (Soak) | Yes | No* |
Disinfecting | ||
Caviwipe | Intermediate Level | Intermediate Level |
* Do not immerse the power supply plug. Follow instructions per this manual.
Cleaning (Wipe)
KleenSpec Corded Illumination System
Refer to Cleaning Table to identify the appropriate cleaning product.
- Disconnect the power supply from the wall outlet.
- ⚠️ CAUTION: Lamp must remain installed during all cleaning and disinfecting processes.
- Follow the disinfecting wipe manufacturer's instruction and wipe cord or illuminator completely to remove any residue.
- This process is a rough clean only. Refer to disinfection for further processing.
Soak
Refer to Cleaning Table to identify the appropriate cleaning product.
KleenSpec Corded Illuminator
- Disconnect the power supply from the wall outlet.
- ⚠️ CAUTION: Lamp must remain installed during all cleaning and disinfecting processes.
- Follow the cleaning solution manufacturer's instructions and soak the illuminator as instructed. Solution temperature is 22° C to 43° C. Make sure to keep the power supply out of the soak solution. Do not immerse the power supply.
- Scrub any area of visible debris with a soft brush and thoroughly rinse the illuminator with 22° C to 43° C tap water to remove the cleaning solution. Do not rinse the power supply; wipe only.
- Rough dry the illuminator.
- The illuminator is now ready for disinfection.
Disinfecting
Refer to the Cleaning Table to identify the appropriate disinfection method/product for the illuminator and power supply.
Illuminator
- Clean or disinfect as described in the table above.
- Follow the disinfecting wipe manufacturer's disinfection instructions.
- Thoroughly rinse illuminator with 22° C to 43° C tap water to remove disinfectant.
- Allow to dry.
Power supply
DO NOT IMMERSE THE POWER SUPPLY.
- Clean or disinfect as described in the table above.
- Follow the disinfecting wipe manufacturer's disinfection instructions on the cord only.
- Thoroughly rinse the cord only with 22° C to 43° C tap water to remove disinfectant. Do not immerse the power supply.
- Allow to dry.
Disposal
There are no harmful materials associated with the Illumination System. Device is considered a biohazard after use. Follow good hospital practices and local ordinances in the disposal of the illuminator.
There are no risks associated with the disposal of this device after its useful life if it is cleaned and/or disinfected per the instructions in this manual.
Approvals
CERTIFIED TO: CAN/CSA STD C22.2 NO. 601.1 CONFORMS TO: IEC 60601-1, IEC60601-1-2
Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153 USA
Tel: (800) 535-6663 or (315) 685-4560
Fax: (315) 685-3361
Website: www.welchallyn.com
Printed in the U.S.A.
Material No. 728645