FORA COVID-19 Antigen Rapid Test
Quick Reference Guide
Key Features
- Easy to perform
- Fast results in 15 minutes
- Visual interpretation
- For In Vitro Diagnostic use
- No special equipment needed
- Technical cooperation with Taiwan's Academia Sinica
Performance and Features
Test Principle | Lateral Flow Chromatographic Immunoassay |
Target Antigen | SARS-CoV-2 Nucleocapsid Protein |
Sample Type | Fresh Nasopharyngeal Swab Specimen or Fresh Nasal Swab Specimen |
Limit of Detection (LoD) | 1.26 x10² TCID50 per mL |
Cross-Reactivity & Interferences | Viruses, Bacteria and Interferences tested do not cross-react or interfere |
Reaction time | 15 minutes |
Test content:
- Individually Foil Packaged Test Cassette
- Extraction Buffer Tube (0.5mL per bottle)
- User Manual
- Quick Reference Instruction
- Sterile Swab
Instruction for Use
SAMPLE COLLECTION AND HANDLING
Mid-turbinate or Anterior nasal Sample
- Take the test strip out from the sealed packaging and place it on a flat and clean surface. Once opened, start the test within 20 minutes.
- Gently blow your nose into a tissue and throw the tissue away in a closed bin. If you are testing a child, help them to blow their nose. This is to get rid of excess mucus.
- Wash your hands thoroughly, using soap and water, or hand sanitiser. Dry your hands before performing the test.
- Take the swab out. Never touch the soft, fabric tip of the swab with your hands.
- Tilt the head back about 70 degrees.
- Gently insert the entire absorbent tip of the swab into one nostril. The swab tip should be inserted up to 2.5 cm (1 inch) from the edge of the nostril and firmly sample the nasal wall by rotating the swab in a circular path against the nasal wall at least 5 times. Make sure the swab tip is touching the nasal walls as you rotate. Take approximately 15 seconds to collect the sample.
- Remove the swab from the nostril.
- Using the same swab, repeat Step 6 for the other nostril. It is important to collect samples from BOTH nostrils using the same swab.
- After sample collection is complete, proceed to the assay procedure as soon as possible. Do NOT place the swab back into the packaging after sample collection is complete.
Procedure Steps:
- Remove and discard the cap from the extraction buffer tube.
- Immerse the swab into the extraction buffer tube. Roll the swab at least ten-twenty (10-20) times while pressing the head against the bottom and side of the extraction buffer tube.
- Leave the swab in the extraction buffer tube for 30 seconds.
- When removing, roll the swab head toward the inside of the extraction buffer tube and squeeze the sides of the tube to extract the liquid from the swab. Dispose of the used swab in biohazardous waste.
- Press the attached cap tightly onto the extraction buffer tube containing the processed sample.
- Add 3-5 drops (100 µL) of the processed sample into the sample well.
- Visual interpretation at 15-20 minutes. Some positive results may appear sooner.
CAUTION: Do not read the results after 20 minutes. It may provide false results.
For Professional Use and for Individuals Self-collection
Interpretation of Results
Result | Visual Representation | Description |
---|---|---|
Positive | Two colored lines on the test cassette. |
Both colored test (T) and control (C) lines appear on the test cassette. Within the specified observation time, a weak colored T line should be judged as a positive result. |
Negative | Only the colored control line appears on the test cassette. |
Only the colored control line (C) appears on the test cassette. The absence of the test line (T) indicates a negative result. |
Invalid Assay | No colored control line appears. |
There should always be a colored control line (C) in the control region regardless of the test result. If the control line is not seen, repeat the assay with a new test cassette. |
Negative / High CT value | Very weak colored T line. |
Within a specified observation time, a very weak colored T line should be judged as negative or high CT value. Please refer to RT-PCR result. |
Frequently Asked Questions
- Q: Will this test hurt?
- A: No, the nasal swab is not sharp and the part touching the nasal wall is soft. Some discomfort may be expected, but there should not be sharp pain. Do not insert the swab any deeper if you feel strong resistance or pain.
- Q: What do I have to do if the test is negative?
- A: This means that the virus was not found or the virus count was not high enough to be detected in your sample at that particular moment. However, it is only a snapshot. Hygiene rules should still be followed, and if you feel sick, please contact your doctor.
- Q: What do I have to do, if the result is unclear?
- A: If the result is unclear, you should consult your doctor to take further steps.
- Q: What do I have to do if the test is positive?
- A: If your test result is positive, you and your household must self-isolate according to the Government guidelines and call your doctor.
- Q: Can I use the test, if the test is expired?
- A: Do not use any accessory past the expiration date.
- Q: Can I read the results after 20 minutes?
- A: No. All results after 20 minutes are not valid.
- Q: What are the known and potential risks and benefits of this test?
- A: Potential risks include: Possible incorrect test result (see INTERPRETATION OF RESULTS section). You may feel uncomfortable during specimen collection. Potential benefits include: Slow the spread and help protect the most vulnerable in our families and communities. The results, along with other information, can help your doctor make informed recommendations about your care.
Copyright TaiDoc Technology Corporation. All rights reserved.
Study the User Manual thoroughly before using the Quick Reference Guide. This is not a complete User Manual.
EC REP MedNet EC-REP GmbH, Borkstraße 10, 48163 Münster, Germany
TaiDoc Technology Corp., B1-7F., No. 127, Wugong 2nd Rd., Wugu Dist., 24888 New Taipei City, Taiwan
Related Documents
![]() |
FORA COVID-19 Antigen Rapid Test: Instructions for Use and Performance Data Comprehensive guide to the FORA COVID-19 Antigen Rapid Test, covering intended use, assay procedure, interpretation of results, quality control, and analytical performance data for SARS-CoV-2 detection. |
![]() |
FORA 6 Blood Glucose Test Strips - User Guide and Specifications This document provides comprehensive information on the FORA 6 Blood Glucose Test Strips, including intended use, limitations, storage, handling, testing procedures, and accuracy results. It is designed for both healthcare professionals and individuals managing diabetes. |
![]() |
FORA 6 Blood Glucose Test Strip User Information and Accuracy Data Comprehensive guide for FORA 6 Blood Glucose Test Strips, covering warnings, intended use, limitations, handling, testing procedures, results interpretation, quality control, chemical composition, accuracy data, and user performance studies. Includes information for healthcare professionals and self-testing. |
![]() |
FORA COMFORT check G40 Blood Glucose Test Strip User Manual Comprehensive user manual for the FORA COMFORT check G40 Blood Glucose Test Strip, detailing warnings, intended use, limitations, testing procedures, accuracy data, and storage instructions. |
![]() |
FORA PO200 Fingertip Pulse Oximeter Operation Manual User manual for the FORA PO200 Fingertip Pulse Oximeter (Model TD-8255B), detailing operation, safety, troubleshooting, and specifications for monitoring blood oxygen saturation and pulse rate. |
![]() |
FORA P30 Plus BT Wireless Blood Pressure Monitor Quick Start Guide A concise guide to using the FORA P30 Plus BT wireless blood pressure monitor, covering single/average measurements, auscultatory mode, and smartphone app connectivity. |
![]() |
FORA GD20 Blood Glucose Monitoring System: Operations & Procedures Manual Comprehensive guide for the FORA GD20 Blood Glucose Monitoring System, covering operations, procedures, safety precautions, testing, maintenance, and troubleshooting for long-term care settings. |
![]() |
FORA TD-2555G TNG Scale 550 Mobile User Manual & Operating Instructions Comprehensive operating instructions and specifications for the FORA TD-2555G TNG Scale 550 Mobile. Learn how to set up, take measurements, understand results, and maintain your digital weight scale. |