TraceableLIVE Wi-Fi Datalogging High Temperature Thermometer

Instruction Manual

CONTROLS

Note: "WiFi enabled" is indicated by the flashing WiFi symbol (?). It also indicates that WiFi network needs to be configured. If WiFi network has been configured, and the WiFi symbol flashes, it indicates an unsuccessful data transmission to the cloud server.

DEVICE SPECIFICATIONS

DISPLAY MODES

SINGLE CHANNEL MODE

DUAL CHANNEL MODE

SELECTING CHANNEL (PROBE)

Note: Plug probe sensor into the USB jacks on top of the unit. Updated temperatures will be displayed. With the probe sensor attached, the unit displays the current probe temperature and probe minimum/maximum temperatures. With the probe sensor removed, the unit simultaneously displays minimum/maximum temperatures since the memory was last cleared.

CLEAR CURRENT MINIMUM/MAXIMUM MEMORY

  1. Press CHANNEL SELECT to select the temperature probe channel to be cleared.
  2. CH1 clears Channel 1 (Probe 1); CH2 clears Channel 2 (Probe 2); in dual channel mode, CH12 clears Channels 1 and 2 (Probe 1 and 2).
  3. Press the ?️ CLEAR button to clear the current minimum and maximum temperature readings.
  4. Each clear of the Minimum/Maximum memory will also trigger a transmission of current reading(s) to TraceableLIVE service, if connected. This will display in EVENT HISTORY with the label "DEVICE CHECK".

DEVICE SETUP

SCENARIO 1: WiFi is disabled. All settings are configurable.

  1. Press and hold the ⚙️ SET button for 3 seconds to enter the setup menu.
  2. The first flashing number is the year date setting. Press the ⬆️ UP or ⬇️ DOWN arrow to set the current year. Press the ▶️/⏸️ PLAY/PAUSE button to save and proceed to the next setting.
  3. Continue to set the remaining parameters: Month -> Day -> Hour -> Minute -> Time Format (12H/24H) -> Channel 1 Minimum Alarm -> Channel 1 Maximum Alarm -> Channel 2 Minimum Alarm -> Channel 2 Maximum Alarm -> Alarm Repost Enable/Disable -> Alarm Repost Interval Setting (if Alarm Repost is enabled). Press ▶️/⏸️ PLAY/PAUSE to proceed to the next parameter. Pressing ▶️/⏸️ PLAY/PAUSE after the last parameter is set will exit setup mode.

SCENARIO 2: WiFi is enabled. Alarm settings are not configurable on the device and can only be set through the TraceableLIVE cloud service interface.

  1. Press and hold the ⚙️ SET button for 3 seconds to enter the setup menu.
  2. The first flashing number is the year date setting. Press the ⬆️ UP or ⬇️ DOWN arrow to set the current year. Press the ▶️/⏸️ PLAY/PAUSE button to save and proceed to the next setting.
  3. Continue to set the remaining parameters: Month -> Day -> Hour -> Minute -> Time Format (12H/24H) -> Alarm Repost Enable/Disable -> Alarm Repost Interval Setting (if Alarm Repost is enabled). Press ▶️/⏸️ PLAY/PAUSE to proceed to the next parameter. Pressing ▶️/⏸️ PLAY/PAUSE after the last parameter is set will exit setup mode.

Note: Setting the time while WiFi is enabled is only intended for initial device setup. Once connected to the TraceableLIVE service, the device time will be synchronized daily for the selected time zone in TraceableLIVE.

ALARM

  1. If an alarm triggers, the LCD will automatically display the alarming channel and the temperature reading. The ALM and MIN or MAX symbols will flash. If the temperature is below the low alarm setting, the MIN symbol flashes; if the temperature is above the high alarm setting, the MAX symbol flashes. An audible alarm will continue beeping for 30 seconds and will beep once every 15 seconds until the alarm is acknowledged by pressing the ?️ CLEAR button.
  2. If alarms trigger on both channels, the LCD will display Channel 1.
  3. Use CHANNEL SELECT to select which channel to display. If the displayed channel is not alarming, the LCD will not flash, but the buzzer will remain active.
  4. If an alarm is triggered, the second line of the LCD will no longer scroll. If the device is in single channel display mode, the alarming set point will display on the second line.
  5. To clear an alarm, press the ?️ CLEAR button. The LCD will stop flashing, the buzzer will stop beeping, and the LCD second line will resume scrolling.
  6. Once an alarm is triggered, the device will post the alert to the TraceableLIVE service immediately. If connectivity is currently lost, the device will store the alarm until it reconnects. Devices can store up to 100 alarm events in internal memory.

DISPLAYING °F OR °C

CONFIGURE WiFi NETWORK: AP PROVISIONING

NOTE: WiFi configuration will be cleared if aborted at this stage.

NOTE: The device date/time is automatically synchronized to the mobile phone or laptop once the setup webpage is shown.

NOTE: Ensure the Network ID and password are correct; otherwise, the device will wait to connect to the router until timeout, and then "Err" will be shown on the LCD.

CONFIGURE WiFi NETWORK: WPS PROVISIONING

NOTE: The router must support WPS, and the WPS function must be enabled. The device supports only the PUSH BUTTON Method. Pin Code Method is NOT supported.

NOTE: Using WPS provisioning will not update the device's date/time.

HOW TO CONFIGURE WiFi NETWORK: SMARTCONFIG PROVISIONING

NOTE: This method requires the user to install the TI WiFi Starter app for iOS or Android on mobile devices.

NOTE: Using SmartConfig provisioning will not update the device's date/time.

DATA MEMORY

  1. The device is capable of storing 7 days of data if a 15-minute logging interval is set.
  2. If data transmission fails, data will be stored in data memory. Stored data will be transmitted automatically on the next successful transmission.
  3. If the WiFi network has been configured and WiFi connection is lost, data will be stored in data memory at the user-defined logging interval.
  4. If the WiFi network has not been configured, data will not be stored in data memory.
  5. Stored data in data memory cannot be cleared by the user. It can only be cleared by a successful data transmission.

ALARM REPOST

DISPLAY MESSAGES

If no buttons are pressed and "--.--" appears on the display, this indicates that the temperature being measured is outside the unit's temperature range, or that the probe is disconnected or damaged.

BENCH STAND

The unit is supplied with a bench stand located on the back. To use the bench stand, locate the small opening at the bottom back of the unit. Place your fingernail into the opening and flip the stand out. To close the stand, simply snap it shut.

LOW BATTERY POWER INDICATOR

The unit is supplied with 4 AAA alkaline batteries. If the battery power drops to 20% or lower, a low battery symbol will appear on the device display, and an alert will be sent via TraceableLIVE.

ALL OPERATIONAL DIFFICULTIES

If this thermometer does not function properly for any reason, please replace the battery with a new, high-quality battery (see "Battery Replacement" section). Low battery power can occasionally cause any number of "apparent" operational difficulties. Replacing the battery with a new, fresh battery will solve most difficulties. If the battery voltage becomes low, the °C and °F symbols will flash.

BATTERY REPLACEMENT

Erratic readings, a faint display, or no display are all indications that the battery must be replaced. Slide the battery cover toward the end of the unit. Remove the exhausted battery and replace it with an AAA alkaline battery. Replace the battery cover.

REGULATORY INFORMATION

This equipment has been tested and found to comply with limits for a Class B digital device, pursuant to Part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with instructions, may cause harmful interference to radio communications.

However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Hereby, Control Company, declares that this digital thermometer is in compliance with the essential requirements and other relevant provisions of Directive 1999/5/EC.

This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.

Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.

NOTE: THE GRANTEE IS NOT RESPONSIBLE FOR ANY CHANGES OR MODIFICATIONS NOT EXPRESSLY APPROVED BY THE PARTY RESPONSIBLE FOR COMPLIANCE. SUCH MODIFICATIONS COULD VOID THE USER'S AUTHORITY TO OPERATE THE EQUIPMENT.

WARRANTY, SERVICE, OR RECALIBRATION

Traceable®

12554 Old Galveston Rd. Suite B230
Webster, Texas 77598 USA

Ph. 281-482-1714 Fax 281-482-9448
support@traceable.com
traceable.com

Traceable® Products are ISO 9001:2015 Quality-Certified by DNV and ISO/IEC 17025:2017 accredited as a Calibration Laboratory by A2LA.

Product Information

Item No. 18004-26, 18004-27

Model No. 6514, 6515

©2023 1065Y10_M_92-6514-00 Rev. 5 07292025

Image: Wi-Fi Datalogging High Temperature Thermometer with Remote Notifications Instruction Manual, showing the device itself.

Models: 6514, 68890d968c68b, 6514 High Temperature Wireless Datalogger, 6514, High Temperature Wireless Datalogger, Wireless Datalogger, Datalogger

File Info : application/pdf, 2 Pages, 204.14KB

PDF preview unavailable. Download the PDF instead.

68890d968c68b

References

Adobe InDesign 18.3 (Windows) Adobe PDF Library 17.0

Related Documents

Preview TraceableLIVE Wi-Fi Datalogging CO2 Meter: Instruction Manual
Comprehensive instruction manual for the TraceableLIVE Wi-Fi Datalogging CO2 Meter (Models 6525/6526), detailing setup, controls, device specifications, WiFi configuration, alarm functions, and regulatory information.
Preview TraceableLive Wi-Fi Datalogging Ultra-Low Temperature Thermometer Instruction Manual
Comprehensive instruction manual for the TraceableLive Wi-Fi Datalogging Ultra-Low Temperature Thermometer, covering controls, setup, WiFi configuration, alarms, and specifications for models 6510 and 6511.
Preview TraceableLIVE Ethernet Datalogging Thermometer with Remote Notification: User Instructions and Specifications
Comprehensive guide to the TraceableLIVE Ethernet Datalogging Thermometer (Models 6600 & 6601), covering controls, specifications, setup, data transmission, alarms, and regulatory compliance.
Preview Traceable LN2 Memory-Loc Datalogging Thermometer User Manual
Comprehensive user manual for the Traceable LN2 Memory-Loc Datalogging Thermometer, detailing specifications, setup, operation of logging, alarms, data download, battery replacement, and warranty information.
Preview TraceableLIVE Instructions: Setup, Usage, and Management Guide
A comprehensive guide to TraceableLIVE data loggers, covering setup, subscription plans, device management, user accounts, reporting, and Bluetooth integration. Learn to monitor temperature, humidity, and CO2 with TraceableLIVE.
Preview Traceable Data Logger Instructions
Comprehensive instructions for setting up and using Traceable Data Loggers with the TraceableGO™ and TraceableLIVE® mobile applications. Learn how to configure settings, download data, and understand device specifications.
Preview TraceableLIVE WiFi Datalogging Refrigerator/Freezer Thermometer Instructions
Comprehensive instructions for setting up and operating the TraceableLIVE WiFi Datalogging Refrigerator/Freezer Thermometer, covering controls, specifications, setup, WiFi configuration, alarms, and troubleshooting.
Preview TraceableLIVE WiFi Datalogging Refrigerator/Freezer Thermometer Instruction Manual
Instruction manual for the TraceableLIVE Wi-Fi Datalogging Refrigerator/Freezer Thermometer, covering controls, setup, alarms, Wi-Fi configuration, data memory, and troubleshooting.