Product Overview
The KEYENCE IL-1500 is a panel mount amplifier unit designed for precise measurement and control applications. It offers versatile display options, analog outputs, and digital control inputs.
An image shows the front panel of the KEYENCE IL-1500 amplifier unit. It features a digital display showing '12.395' and '12.345', along with several buttons and indicator lights. The unit is labeled 'IL-1500 Amplifier unit, Panel mount type' and displays 'CE' and 'C US' certification marks.
Specifications
Category | Detail | Value |
---|---|---|
Model | IL-1500 | |
Type | Panel mount | |
Main unit/Expansion unit | Main unit | |
Head compatibility | Compatible | |
Display | Minimum displayable unit | IL-030: 1 µm, IL-065/IL-100: 2 µm, IL-300: 10 µm, IL-600: 50 µm, IL-2000: 100 µm |
Display range | IL-030/IL-065/IL-100: ±99.999 mm to ±99 mm (4 levels selectable), IL-300/IL-600: ±999.99 mm to ±999 mm (3 levels selectable), IL-2000: ±9999.9 mm to ±9999 mm (2 levels selectable) | |
Analog voltage output | Display rate | Approx. 10 times/sec. |
Analog current output | ±5 V, 1 to 5 V, 0 to 5 V Output impedance 100 Ω*1 | |
Control input | Bank switch input | 4 to 20 mA Maximum load resistance of 350 Ω*1 |
Zero-shift input | Non-voltage input*2 | |
Stop emission input | ||
Timing input | ||
Reset input | ||
Control output | Judgement output | Open collector output (NPN, PNP changeover possible/N.O., N.C. changeover possible)*3 |
Alarm output | Open collector output (NPN, PNP changeover possible/N.C.)*3 | |
Current | Power voltage | 10 to 30 VDC ripple (P-P) 10% included, Class 2*4 |
Power consumption | 2500 mW or less (at 30 V: 84 mA or less) | |
Environmental resistance | Pollution degree | 2 |
Ambient temperature | -10 to +50 °C (14 to 122 °F) (No condensation or freezing) | |
Relative humidity | 35 to 85% RH (No condensation) | |
Vibration resistance | 10 to 55 Hz, Double amplitude 1.5 mm (0.06"), 2 hours in each of the X, Y, and Z directions | |
Material | Case / Front sheet | Polycarbonate; Key tops: Polyacetel; Cable: PVC |
Weight | Approx. 170 g (including attachments) |
Notes:
*1 Select and use one of ±5 V, 1 to 5 V, 0 to 5 V or 4 to 20 mA.
*2 Assign an input of your choice to the 4 external input lines before using.
*3 The NPN open collector rated output is: 50 mA max./ch (20 mA when adding an expansion unit) less than 30 V, residual voltage less than 1 V (less than 1.5 V when adding over 6 units including the main unit). The PNP open collector rated output is: 50 mA max./ch (20 mA/ch when adding expansion units), less than power voltage, and less than 2 V residual voltage (less than 2.5 V when adding over 6 units including the main unit).
*4 If there are over 6 additional expansion units, please use a power voltage of 20 to 30 V.
Dimensions
A technical drawing illustrates the dimensions of the IL-1500/IL-1550 unit. Key dimensions include a height of 37.1 mm, a width of 44.7 mm, and a depth of approximately 76 mm for the main body. It also shows panel mounting details, indicating a panel thickness range and cable diameters of ø4.7 mm / ø4.8 mm for cables with a 2 m length.
Download CAD file or product manual for larger image/text and more detail.
I/O Circuit Connection Diagram
A wiring diagram shows the connections for the IL-1500 unit. The diagram connects colored wires to terminal labels on the unit.
- Brown wire connects to '10 to 30 VDC' (Power input)*4.
- Blue wire connects to '0 V'*1.
- Black wire connects to 'HIGH judgment output'.
- White wire connects to 'LOW judgment output'.
- Grey wire connects to 'GO judgment output'.
- Green wire connects to 'Alarm output'.
- Light blue wire connects to 'Analogue output +'*2.
- Orange wire connects to 'Analogue output -'*2.
- Shield wire connects to 'Analogue output GND'*2.
- Pink wire connects to 'External input 1 (zero shift input)'*3.
- Yellow wire connects to 'External input 2 (reset input)'*3.
- Pink/purple wire connects to 'External input 3 (timing input)'*3.
- Purple wire connects to 'External input 4 (not used)'*3.
Notes for Wiring Diagram:
- The brown, blue, and light blue cables are not provided in an IL-1050/IL-1550 unit (expansion unit).
- The power is supplied to the expansion unit from the IL-1000/IL-1500 unit (main unit).
- For an analogue output, OFF (not used), 0 to 5 V, ±5 V, 1 to 5 V, or 4 to 20 mA can be selected.
- For an external input, bank A input, bank B input, laser emission stop input, or OFF (not used) can also be selected.
- If there are over 6 additional expansion units, please use a power voltage of 20 to 30 V.
For details, refer to the User's Manual.