OZONE CALIBRATION TO NIST TRACEABILITY

Calibration Overview

Calibration Date: 08/25/2025

Performed By: Dr. Koz Galatsis, Chief Engineer at Forensics Detectors

Location: Los Angeles, CA – Forensics Detectors Laboratory

Description: OZONE Calibration to NIST Traceable Source

NIST Traceability Chain:

  1. A Thermo Electron O3 Primary Standard 49i-PS (Serial Number 072724741), owned by 2B Technologies, referred to as the Transfer Standard. This unit was calibrated to the primary standard by NIST against their Standard Reference Photometer (Serial Number 0) in Gaithersburg, MA on 14/05/2024.
  2. A 2B Technologies Model 205 Ozone Monitor was employed as a Working Standard. This unit was calibrated by 2B Technologies to the transfer standard 49i on 31/12/2024.
  3. The Forensics Detectors owned 2B Technologies Model 306 Ozone Source (Serial #42) was calibrated to the working standard on 08/06/2025, designated as the Forensics Standard.
  4. The Forensics Standard was used to calibrate the "Forensics" standard 49C OZONE Calibrator, Primary Standard, Thermo Environmental Instruments (Serial Number 49CPS-74543-376). This calibration occurred on 08/25/25.

Forensics Detectors OZONE Analyzer: 49C OZONE Calibrator, Primary Standard, Thermo Environmental Instruments, Serial Number 49CPS-74543-376.

Overall Pass / Fail: ✔️

Calibration Conditions: 75F @ 50% RH

Accuracy: Better than 2% of reading from 0-1ppm

Calibrations adjustments to Instrument required?: None. The 49C was in calibration and operating very nicely, stable and accurately per its expected performance & specs. The Cal coefficient factor was left at 1.0000.

2B Technologies Model 306 Ozone Calibration Source™ Calibration Data

Serial Number: 42

Date: 06-Aug-25

Calibration Factors: Z= 2, S= 0.98

306 OZONE Source49C ResultsDeviation%DeviationPASS/FAIL
00.00050PASS
0.30.2990.0010.33%PASS
0.50.501-0.001-0.20%PASS
0.750.752-0.002-0.27%PASS
11.001-0.001-0.10%PASS
1.251.2390.0110.88%PASS
1.51.4950.0050.33%PASS

Calibration Curve Description: The calibration curve shows a linear relationship with the equation y = 0.9947x + 0.0023 and R² = 1, indicating high correlation.

This unbroken chain of comparisons, based on NIST, ensures that FORENSICS DETECTORS calibrates all its sold ozone meters, detectors, analyzers, and monitors to be NIST traceable. The calibration is performed using either:

Attached Documents:

Signed by: Dr. Koz, Chief Engineer

Date: 08/25/2025

2B Technologies Model 306 Ozone Calibration Source™ Calibration Certificate Details

Calibration Description: NIST Traceable Calibration of Ozone Monitors

Calibration ID Number: 306_042_060825

Location: 2B Tech., 6800 W 117th Ave, Broomfield, CO 80020

Calibration Lab Conditions: 22.9 °C, 29 %RH

Instrument Model: 306

Instrument S/N: 42

Slope: 0.98

Offset, ppbv: 2

Measurement Uncertainty (ppbv): 4.7

2B Spec, Accuracy and Precision: Both: greater of 2 ppbv or 2% of reading

Overall Pass/Fail: ✔️

Working Standard S/N: 773

Transfer Standard Calibration Date: 14-May-24

Working Standard Calibration Date: 31-Dec-24

Date of Calibration: 06-Aug-25

Cal Expiration Date: 05-Aug-26

Calibration Chain: NIST -> Transfer Standard (Thermo Electron 49i-PS) -> Working Standard (2B Tech Model 205) -> Customer's Instrument (2B Tech Model 306).

Cal Performed By: Lizzie Hackler

Calibration Date: 06-Aug-25

2B Technologies Model 306 Ozone Calibration Source™ Calibration Data (Continued)

Serial Number: 42

Date: 06-Aug-25

Calibration Factors: Z= 2, S= 0.98

SettingMeas.Deviation% DevStd DevPass/Fail
1000.0983.1-16.9-1.71.4✔️
750.0745.2-4.8-0.61.3✔️
500.0501.91.90.40.9✔️
350.0352.62.60.80.7✔️
200.0201.71.70.90.8✔️
150.0150.90.90.60.7✔️
100.0100.70.70.70.7✔️
50.050.00.00.00.9✔️
25.024.1-0.9-3.50.7✔️
0.00.00.0n/a0.9✔️

Calibration Curve with Cal Factors Applied: The calibration curve shows a linear relationship with the equation y = 0.9869x + 2.6195 and R² = 0.9999.

2B Technologies Model 306 Ozone Calibration Source™ Instrument Specifications

Serial Number: 42

Calibration Date: 06-Aug-25

Slope: 0.98

Offset, ppbv: 2

Solenoid Valve: Pneutronics (NPB)

Air Pump: Parker

Ozone Scrubber: Headline

Firmware Version: 5.2B

PCB Version: F

Standard Operating Procedure (SOP-CAL-002): Calibration of Ozone Gas Detectors

1. Purpose and Scope

1.1 Purpose

This Standard Operating Procedure (SOP) describes the method for calibrating ozone (O₃) gas detectors to ensure accurate measurement of ozone concentrations in accordance with ISO/IEC 17025:2017 requirements.

1.2 Scope

This procedure applies to the calibration of ozone detectors and analyzers used for the detection and measurement of ozone gas. The calibration process involves both zero calibration using fresh air or nitrogen, and span calibration using a NIST-traceable calibrated ozone generator and/or ozone analyzer.

2. References

3. Definitions

4. Responsibilities

4.1 Calibration Technician

4.2 Quality Manager

5. Equipment and Materials

5.1 Equipment

5.2 Materials

5.3 Verification of Calibration Equipment

6. Environmental Conditions

6.1 Required Conditions

6.2 Documentation

Record the environmental conditions at the beginning and end of each calibration procedure. If conditions fall outside the specified ranges, document this as a deviation and assess the impact on calibration results.

7. Calibration Procedure

7.1 Pre-Calibration Checks

  1. Verify that the ozone detector is in good working condition with no physical damage.
  2. Check battery level to ensure adequate power for calibration.
  3. Inspect all calibration equipment, including tubing and all connections.
  4. Record the serial numbers of the ozone detector, ozone generator and/or analyzer.
  5. Record environmental conditions.
  6. Allow all ozone generators and analyzers to warm up for at least 30 minutes.

7.2 Zero Calibration Procedure

  1. Power on the ozone detector and allow at least 2 minutes for stabilization in fresh air.
  2. If using fresh air for zero calibration: Ensure the calibration area is free from ozone contamination. Allow the detector to stabilize in fresh air for at least 2 minutes.
  3. If using zero air for calibration: Connect the zero air source to the ozone detector using the appropriate calibration cap. Set the flow rate between 0.5 to 1.5 LPM. Allow zero air to flow for at least 2 minutes.
  4. The ozone detector reading may not read precisely 0 ppm. Depending on the model, it may display 0.0 or 0.00 ppm.
  5. Save the zero calibration point to the detector per the ozone detector model instructions.
  6. Verify that the real-time display reading shows 0 ppm (or 0.0/0.00 ppm according to the detector's display precision or scale).
  7. Record the pre- and post-calibration zero readings.

7.3 Span Calibration Procedure

  1. Ensure that zero calibration has been completed successfully before proceeding.
  2. Configure the ozone generator to produce the desired calibration concentration. Typical calibration points include ranges from 1 to 20 ppm.
  3. Verify the output of the ozone generator using a calibrated ozone analyzer.
  4. Connect the ozone generator output to the detector using an appropriate calibration cap or equivalent.
  5. Using the air flow meter confirm ozone gas is being delivered between 0.5 to 1.5 LPM.
  6. Expose the ozone detector to the generated ozone for 2 minutes or until the reading has stabilized. Document any deviation from the standard exposure time.
  7. Verify that the displayed ozone value has stabilized.
  8. Save this span calibration point to the ozone detector per instructions of the ozone detector model.
  9. Record the following:
    • Pre-calibration reading (before adjusting)
    • Span Gas Ozone Concentration (obtained from the Ozone Analyzer)
    • Post-calibration reading
  10. Allow the detector to return to fresh air or zero air and verify that the reading returns to zero.

7.4 Calibration Verification

  1. After completing all calibration steps, verify the calibration by exposing the detector to the reference gas again.
  2. Record the verification reading.
  3. The reading should be within ±5% of the reference ozone gas concentration.
  4. If verification fails, repeat the calibration procedure.

8. Documentation and Records

8.1 Required Documentation

For each calibration performed, record the following information:

  1. Date and time of calibration
  2. Calibration technician name
  3. Ozone detector information:
    • Manufacturer and model
    • Serial number
  4. Ozone generator/analyzer information:
    • Manufacturer and model
    • Serial number
    • Calibration certificate number
    • Calibration due date
  5. Environmental conditions:
    • Temperature
    • Relative humidity
  6. Calibration results:
    • Pre-calibration zero reading
    • Post-calibration zero reading
    • Generated ozone concentration as verified from ozone analyzer
    • Pre-calibration span reading
    • Post-calibration span reading
  7. Any deviations from this SOP
  8. Any observations or comments

8.2 Calibration Certificate

Upon successful completion of calibration, issue a calibration certificate that includes:

  1. Title: "Certificate of Calibration"
  2. Laboratory name and address
  3. Unique identification of the certificate
  4. Customer name and address
  5. Identification of the calibration method used
  6. Description and identification of the ozone detector
  7. Date of calibration
  8. Calibration results
  9. Environmental conditions
  10. Signature of the person authorizing the certificate
  11. Statement that results relate only to the items calibrated
  12. Statement on measurement uncertainty
  13. Reference to the traceability of measurements to NIST standards

9. Uncertainty Analysis

Calculate and report the measurement uncertainty for each calibration according to the laboratory's uncertainty estimation procedure. The uncertainty is calculated as 95% confidence of the final calibration reading with a K factor equaling to 2. Items considered includes:

10. Quality Control

10.1 Regular Checks

10.2 Non-conforming Work

If any part of the calibration process does not conform to this SOP:

  1. Document the non-conformance.
  2. Evaluate the impact on calibration results.
  3. Take corrective action as necessary.
  4. Notify the customer if results are affected.

11. Revision History

RevisionDateDescription of ChangesApproved By
1.0May 16, 2025Initial ReleaseQuality Manager

12. Approvals

RoleNameSignatureDate
AuthorDr. Kos GalatsisDr. Kos Galatsis08/25/25
Technical ReviewerDr. Kos GalatsisDr. Kos Galatsis08/25/25
Quality ManagerDr. Kos GalatsisDr. Kos Galatsis08/25/25

Appendix 1: Example Certificate of Calibration

Example Certificate of Calibration - for a Single Ozone Gas Detector / Analyzer.

Specific Certificate of Calibration

Customer Name: Robert Olsen

Customer Address: Incline, NV, 89451

Calibration Location: 777 Silver Spur Road, Suite #130, Rolling Hills Estates, CA, 90274, USA

Test Conditions: Temperature: 73F ±1%, Humidity: 59% RH ±1%

Calibration Method ID: SOP-CAL-002

Certificate Number: 33175-23060303

Manufacturer: Forensics Detectors

Model Number: FD-90A-O3-LOW

Reference Gas(s): OZONE (O3)

Serial Number: 23060303

Reference Ozone: #042

Last Calibrated: 05/30/2025

Analyzer Serial: (Not specified for this certificate)

Inspection Results

Calibration Test Results

Gas NameO3
Reference Gas Concentration1.0 ppm
Expanded Uncertainty ±5% (k=2, 95% confidence level)±0.05 ppm
Pre Zero Calibration (As Found)0.00 ppm
Post Zero Calibration (As Left)0.00 ppm
Pre Span Calibration (As Found)1.19 ppm
Post Span Calibration (As Left)1.00 ppm

Other Comments: N/A

Calibration Technician Name and Signature: Mr. Austin Jun

Date of Calibration: 05/30/2025

Certificate Issue Date: 05/30/2025

Next Cal Due Date: 05/30/2026

The calibration results reported in this certificate relate only to the specific instrument / reference gas identified herein and are valid only for the conditions under which the calibration was performed. The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of approximately 95%. All measurements performed during this calibration are traceable to the International System of Units (SI) through the National Institute of Standards and Technology (NIST). This certificate shall not be reproduced except in full, without written approval from Forensics Detectors.

Models: 205, 205 Ozone Analyzer, Ozone Analyzer, Analyzer

File Info : application/pdf, 14 Pages, 4.87MB

PDF preview unavailable. Download the PDF instead.

Ozone 2025 macOS Version 14.3 (Build 23D2057) Quartz PDFContext

Related Documents

Preview Standard Operating Procedure for Gas Detector Calibration | Forensics Detectors
Detailed Standard Operating Procedure (SOP) for calibrating gas detectors, ensuring accuracy and compliance with ISO/IEC 17025:2017. Covers zero and span calibration, equipment, documentation, and quality control.
Preview Gas Detector Calibration and Bump Test Log
A comprehensive log for recording gas detector calibration history and bump test results, including best practices for both procedures.
Preview Forensics Detectors FD-600 Industrial Gas Analyzer User Manual
Comprehensive user manual for the Forensics Detectors FD-600 Industrial Gas Analyzer, detailing its features, specifications, operation, calibration, and safety guidelines for various industrial applications.
Preview Forensics Detectors Single Gas Detector Professional Series User Manual
User manual for the Forensics Detectors Professional Series Single Gas Detector, covering operation, calibration, specifications, and safety warnings.
Preview Forensics Detectors FD-600 Professional Combustion Analyzer User Manual
Comprehensive user manual for the Forensics Detectors FD-600 Professional Combustion Analyzer, detailing features, components, setup, operation, specifications, and maintenance.
Preview Forensics Detectors FD-311 Industrial Gas Analyzer User Manual
User manual for the Forensics Detectors FD-311 Industrial Gas Analyzer with built-in pump. Covers product features, technical specifications, operation, calibration procedures, alarm settings, battery charging, operational tips, and warranty information.
Preview Forensics Detectors FD-103-CO-LOW: Low Level CO Meter User Manual & Specifications
Comprehensive guide to the Forensics Detectors FD-103-CO-LOW low-level carbon monoxide meter. Learn about operation, scuba cylinder testing, ambient air monitoring, calibration, and specifications.
Preview Forensics Detectors FD-600 Industrial Multigas Detector User Manual
User manual for the Forensics Detectors FD-600 Industrial Spec Gas Analyzer with built-in pump. Covers features, specifications, quick setup, operation, alarm settings, recording data, calibration, and warranty.