METHOD

Calibration of Gas Detectors

Document ID: SOP-CAL-001

Revision: 1.0

Effective Date: May 14, 2025

1. Purpose and Scope

1.1 Purpose

This Standard Operating Procedure (SOP) describes the method for calibrating gas detectors to ensure accurate measurement of specific gas concentrations in accordance with ISO/IEC 17025:2017 requirements.

1.2 Scope

This procedure applies to the calibration of portable gas detectors used for the detection and measurement of gases in the below table. The calibration process involves both zero calibration using fresh air or nitrogen, and span calibration using NIST-traceable reference gases as listed in the below table.

Target Gas NameChemical SymbolGas Range
AcetyleneC2H20 - 50 ppm
OxygenO20% - 21%
EthyleneC2H40 - 80 ppm
ChlorineCl20 - 10 ppm
Carbon MonoxideCO0 - 5000 ppm
Carbon DioxideCO20 - 100 %vol
Ethylene OxideETO0 - 10 ppm
MethaneCH40 - 50 %LEL
HydrogenH20 - 5,000 ppm
Hydrogen SulfideH2S0 - 100 ppm
FormaldehydeCH2O0 - 10 ppm
Hydrogen ChlorideHCl0 - 10 ppm
Hydrogen CyanideHCN0 - 20 ppm
NitrogenN20 - 100 %vol
AmmoniaNH30 - 50 ppm
Nitrous OxideN2O0 - 200 ppm
Nitrogen DioxideNO20 - 200 ppm
PhosphinePH30 - 10 ppm
Sulfur DioxideSO20 - 10 ppm
IsobutyleneC4H80 - 50 ppm
ArgonAr0 - 100 %vol
HeliumHe0 - 100 %vol
Nitric OxideNO0 - 1000 ppm
EthanolC2H6O0 - 500 ppm

2. References

3. Definitions

4. Responsibilities

4.1 Calibration Technician

4.2 Quality Manager

5. Equipment and Materials

5.1 Equipment

5.2 Materials

6. Environmental Conditions

6.1 Required Conditions

6.2 Documentation

Record the environmental conditions at the beginning and end of each calibration procedure. If conditions fall outside the specified ranges, document this as a deviation and assess the impact on calibration results.

7. Calibration Procedure

7.1 Pre-Calibration Checks

  1. Verify that the gas detector is in good working condition with no physical damage.
  2. Check battery level to ensure adequate power for calibration.
  3. Inspect all calibration equipment, including tubing and all connections.
  4. Verify that reference gas cylinders are within their validity period.
  5. Record the serial number of the gas detector and the certificate numbers of reference gases.
  6. Record environmental conditions.

7.2 Zero Calibration Procedure

  1. Power on the gas detector and allow 2 minutes for stabilization in fresh air.
  2. If using fresh air for zero calibration: a. Ensure the calibration area is free from target gas contamination. b. Allow the detector to stabilize in fresh air for at least 2 minutes.
  3. If using nitrogen for zero calibration: a. Connect the nitrogen cylinder to the gas detector using the appropriate calibration cap. b. Set the flow rate to 0.5 LPM. c. Allow nitrogen to flow for at least 2 minutes.
  4. The gas detector reading may not read precisely 0 ppm. Depending on the model, it may display 0.0 or 0.00 ppm. Depending on the scale it may also read in %vol.
  5. Save the zero calibration point to the detector per the gas detector model instructions.
  6. Verify that the real-time display reading shows 0 ppm (or 0.0/0.00 ppm according to the detector's display precision or scale %vol).
  7. Record the pre- and post-calibration zero readings.

7.3 Span Calibration Procedure

  1. Ensure that zero calibration has been completed successfully before proceeding.
  2. Connect the reference gas cylinder to the gas detector using a calibration cap or equivalent.
  3. Set the flow rate to 0.5 LPM.
  4. Expose the gas analyzer to the reference gas for 1 minute or until the reading has stabilized. Document any deviation from the standard exposure time.
  5. Verify that the displayed gas value has stabilized.
  6. Save this span calibration point to the gas detector per instructions of the gas detector model.
  7. Record the following:
    • Pre-calibration reading (before adjusting)
    • Span gas concentration from the Certificate of Analysis
    • Post-calibration reading
  8. Allow the detector to return to fresh air and verify that the reading returns to zero.

7.4 Multi-Sensor Gas Calibration

For gas detectors with more than one gas sensor that require calibration to more than one target gas:

  1. Repeat the zero and span calibration procedure (Section 7.2 and 7.3) for each target gas.
  2. A mixed reference gas can be used to expose a multi-sensor gas detector. Undertake zero and span calibration simultaneously for each gas sensor within the multi-gas detector.
  3. If a mixed reference gas is not available, treat each gas sensor separately and following Section 7.2 and 7.3.
  4. Document the calibration results for each sensor separately (but on the same gas detector calibration certificate).

7.5 Calibration Verification

  1. After completing all calibration steps, verify the calibration by exposing the detector to the reference gas again.
  2. Record the verification reading.
  3. The reading should be within ±4% of the reference gas concentration.
  4. If verification fails, repeat the calibration procedure.

8. Documentation and Records

8.1 Required Documentation

For each calibration performed, record the following information:

  1. Date and time of calibration
  2. Calibration technician name
  3. Gas detector information:
    • Manufacturer and model
    • Serial number
  4. Reference gas information:
    • Gas type and concentration
    • Cylinder number
    • Certificate number
    • Expiration date
  5. Environmental conditions:
    • Temperature
    • Relative humidity
  6. Calibration results:
    • Pre-calibration zero reading
    • Post-calibration zero reading
    • Pre-calibration span reading
    • Reference gas concentration used
    • Post-calibration span reading
  7. Any deviations from this SOP
  8. Any observations or comments

8.2 Calibration Certificate

Upon successful completion of calibration, issue a calibration certificate that includes:

  1. Title: "Certificate of Calibration"
  2. Laboratory name and address
  3. Unique identification of the certificate
  4. Identification of the calibration method used
  5. Description and identification of the gas detector
  6. Date of calibration
  7. Calibration results
  8. Environmental conditions
  9. Signature of the person authorizing the certificate
  10. Statement that results relate only to the items calibrated
  11. Statement on measurement uncertainty
  12. Reference to the traceability of measurements to NIST standards

9. Uncertainty Analysis

Calculate and report the measurement uncertainty for each calibration according to the laboratory's uncertainty estimation procedure. The uncertainty is calculated as 95% confidence of the final calibration reading with a K factor equaling to 2. Items considered includes:

10. Quality Control

10.1 Regular Checks

  1. Perform regular checks of reference gases using secondary standards.
  2. Perform regular quality control checks on calibration equipment (as required).

10.2 Non-conforming Work

If any part of the calibration process does not conform to this SOP:

  1. Document the non-conformance.
  2. Evaluate the impact on calibration results.
  3. Take corrective action as necessary.
  4. Notify the customer if results are affected.

11. Revision History

RevisionDateDescription of ChangesApproved By
1.0May 16, 2025Initial ReleaseQuality Manager

12. Approvals

RoleNameSignatureDate
AuthorDr. Koz GalatsisDr. Koz Galatsis08/25/2025
Technical ReviewerDr. Koz GalatsisDr. Koz Galatsis08/25/2025
Quality ManagerDr. Koz GalatsisDr. Koz Galatsis08/25/2025

APPENDIX 1

Example Certificate of Calibration - for a Single Gas Detector.

Certificate of Calibration

USA NIST Traceability Calibration

Calibration Location: 777 Silver Spur Road, Suite #130, Rolling Hills Estates, CA, 90274, USA

Test Conditions:Temperature: 70F ±1%, Humidity: 52% RH ±1%Certificate Number:4567
Calibration Method ID:SOP-CAL-001Manufacturer:Forensics Detectors
Detector Model Number:FD-90A-COReference Gas(s):CO
Detector Serial Number:245686867Cylinder Lot Number:304-403306989-1
Reference Gas Expiration:04/05/2029

INSPECTION RESULTS

Visual Inspection?Normal, no anomaliesPASS
Detector Functionality?Normal, no anomaliesPASS
Alarm Functionality?Normal, no anomaliesPASS

CALIBRATION TEST RESULTS

Gas NameReference Gas ConcentrationExpanded Uncertainty 5% (k=2, 95% confidence level)Pre Zero Calibration (As Found)Post Zero Calibration (As Left)Pre Span Calibration (As Found)Post Span Calibration (As Left)
CO200 ppm±10 ppm0 ppm0 ppm189 ppm200 ppm
Other Comments: N/A

Calibration Technician Name and Signature: Austin Jun

Date of Calibration: 05/23/2025

Certificate Issue Date: 05/23/2025

Next Cal Due Date: 05/23/2026

The calibration results reported in this certificate relate only to the specific instrument / reference gas identified herein and are valid only for the conditions under which the calibration was performed. The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of approximately 95%. All measurements performed during this calibration are traceable to the International System of Units (SI) through the National Institute of Standards and Technology (NIST). This certificate shall not be reproduced except in full, without written approval from Forensics Detectors.

APPENDIX 2

Example Certificate of Calibration - for a Multi-Gas Detector with 4 gas sensors.

Certificate of Calibration

USA NIST Traceability Calibration

Calibration Location: Forensics Detectors, 777 Silver Spur Road, Suite #130, Rolling Hills Estates, CA, 90274, USA

Test Conditions:Temperature: 70F ±1%, Humidity: 52% RH ±1%Certificate Number:4567
Calibration Method ID:SOP-CAL-001Manufacturer:Forensics Detectors
Model Number:FD-4SReference Gas(s):O2, CO, H2S, CH4
Serial Number:245686867Cylinder Lot Number:304-403306989-1
Reference Gas Expiration:04/05/2029

INSPECTION RESULTS

Visual Inspection?Normal, no anomaliesPASS
Detector Functionality?Normal, no anomaliesPASS
Alarm Functionality?Normal, no anomaliesPASS

CALIBRATION TEST RESULTS

Gas NameReference Gas ConcentrationExpanded Uncertainty 5% (k=2, 95% confidence level)Pre Zero Calibration (As Found)Post Zero Calibration (As Left)Pre Span Calibration (As Found)Post Span Calibration (As Found)Post Span Calibration (As Left)
CH450%LEL±3 %LEL0% LEL0% LEL40% LEL50% LEL50% LEL
O220.9%vol±1%vol0.0%0.0%20.9%20.9%20.9%
CO200 ppm±10 ppm0 ppm0 ppm200 ppm200 ppm200 ppm
H2S25 ppm±1 ppm0 ppm0 ppm25 ppm25 ppm25 ppm
Other Comments: N/A

Calibration Technician Name and Signature: Mr. Austin Jun

Date of Calibration: 05/23/2025

Certificate Issue Date: 05/23/2025

Next Cal Due Date: 05/23/2026

The calibration results reported in this certificate relate only to the specific instrument / reference gas identified herein and are valid only for the conditions under which the calibration was performed. The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by the coverage factor k=2, which for a normal distribution corresponds to a coverage probability of approximately 95%. All measurements performed during this calibration are traceable to the International System of Units (SI) through the National Institute of Standards and Technology (NIST). This certificate shall not be reproduced except in full, without written approval from Forensics Detectors.

Models: FD-91 Gas Detectors, FD-91, Gas Detectors, Detectors

File Info : application/pdf, 10 Pages, 263.52KB

PDF preview unavailable. Download the PDF instead.

00. Gas Calibration SOP Gas Cylinders

Related Documents

Preview Ozone Calibration to NIST Traceability: Forensics Detectors Analyzer and 2B Technologies Calibration Source SOP
Comprehensive documentation detailing ozone calibration procedures, NIST traceability, and calibration certificates for Forensics Detectors and 2B Technologies ozone measurement instruments. Includes Standard Operating Procedure SOP-CAL-002 for calibrating ozone gas detectors.
Preview Forensics Detectors FD-103-CO-LOW Calibration Certificate
Official calibration, test, and quality inspection certificate for the Forensics Detectors FD-103-CO-LOW Carbon Monoxide gas detector, confirming compliance with specifications and NIST traceability.
Preview Forensics Detectors Single Gas Detector Professional Series User Manual
User manual for the Forensics Detectors Professional Series Single Gas Detector, covering operation, calibration, specifications, and safety warnings.
Preview Gas Detector Calibration Log and Best Practices | Forensics Detectors
Comprehensive guide to gas detector calibration, including best practices, calibration history logging, and essential information for Forensics Detectors models.
Preview Gas Detector Calibration and Bump Test Log
A comprehensive log for recording gas detector calibration history and bump test results, including best practices for both procedures.
Preview Forensics Detectors FD-600 Industrial Gas Analyzer User Manual
Comprehensive user manual for the Forensics Detectors FD-600 Industrial Gas Analyzer, detailing its features, specifications, operation, calibration, and safety guidelines for various industrial applications.
Preview Forensics Detectors FD-311 Industrial Gas Analyzer User Manual
User manual for the Forensics Detectors FD-311 Industrial Gas Analyzer with built-in pump. Covers product features, technical specifications, operation, calibration procedures, alarm settings, battery charging, operational tips, and warranty information.
Preview Forensics Detectors FD-91 Gas Leak Detector: Features, Specs, and Operation
Comprehensive guide to the Forensics Detectors FD-91 Gas Leak Detector. Learn about its models (RED, YELLOW, BLACK), specifications, operation, safety tips, and applications for detecting combustible, sewer, and refrigerant gases.