Optimized GC/MS Analysis for PAHs in Challenging Matrices

Using the Agilent 5977 Series GC/MSD with JetClean and midcolumn backflush

Authors: Anastasia A. Andrianova and Bruce D. Quimby, Agilent Technologies, Inc.

Abstract

The Agilent 8890 GC combined with an Agilent 5977 Series MSD system was used for the analysis of polycyclic aromatic hydrocarbons (PAHs). By proper selection of instrument configuration and operating conditions, the system provides a robust means of analyzing PAHs in difficult matrices. Midcolumn backflushing, continuous hydrogen source cleaning (JetClean), and use of an alternative drawout lens result in excellent linearity across a calibration range of 1 to 1,000 pg. System precision and robustness are demonstrated with replicate injections of an extract from high organic content soil.

Introduction

PAHs are toxic to aquatic life and are suspected human carcinogens. Because they originate from multiple sources, they are widely distributed as contaminants throughout the world.

PAH Sources:

Given their ubiquitous nature, they are monitored as trace contaminants in many different food products ranging from seafood to edible oils to smoked meats. They are also monitored in the environment including air, water, and soil. PAHs have been analyzed by multiple techniques including HPLC/UV, GC/FID, GC/MS, or GC/MS/MS.

This Application Note focuses on GC/MS in SIM mode. A common calibration range is from 1 to 1,000 pg with an acceptable linearity of R² >0.99. Internal standard (ISTD) area reproducibility is typically specified at ±20 % with calibration standards, and ±30 % with samples.

A number of issues arise with the analysis due to the properties of PAHs. They span wide molecular weight and boiling temperature ranges. Although not considered active or subject to degradation, they are sticky and readily adhere to surfaces. PAHs are subject to desublimation (deposition) and are difficult to vaporize. High temperatures and minimizing surface contact are important. Peak tailing is often seen on the later eluters, resulting in manual integration and extending data review. In some cases, the ISTD response is inconsistent across the calibration range and can lead to problems with linearity of the method.

In addition to the PAH-related challenges, there are often matrix-related problems with the analysis. For example, in food and soil analyses, high boiling matrix contaminants that elute after the analytes can require extended bakeout times to prevent ghost peaks in subsequent runs. The highest boiling contaminants can deposit in the head of the column, requiring more frequent column trimming and adjustment of SIM and data analysis time windows from the resulting retention time shift.

Experimental

This system was configured to minimize the potential problems with the analysis of PAHs in high-matrix samples. The important techniques used were:

System Configuration

Figure 1 shows the system configuration used. Tables 1 and 2 list the instrument operating parameters. Instrument temperatures must be kept high enough to prevent deposition of the highest boiling PAHs. The inlet and MSD transfer line are maintained at 320 °C. The MS source should be a minimum of 320 °C. Pulsed splitless injections are used to maximize transfer of the PAHs, especially the heavy ones, into the column. The straight bore 4 mm liner with glass wool is a must. The wool transfers heat to the PAHs and blocks the line of sight to the inlet base. If the PAHs condense on the inlet base, they are difficult to vaporize, and sweep back into the column.

Figure 1. System configuration. [Diagram description: Shows the Agilent 8890 GC connected to a 5977 Series GC/MSD. The 8890 GC includes a Liquid Injector and a PSD (Helium) module. The GC column setup involves a 9-mm Extractor lens, an EI Source, and two 15 m DB-EUPAH columns connected via a Purged Ultimate Union (PUU) with a makeup flow from the PSD module. The 5977 Series GC/MSD has JetClean (Hydrogen) connected to its source.]

Table 1. GC and MS conditions for the PAH analysis.

Parameter 8890 GC with fast oven, autoinjector, and tray 5977 Series GC/MSD
Inlet ModeEPC Split/splitless
ModePulsed SplitlessSIM
Injection pulse pressure50 psi until 0.7 minutes
Purge flow to split vent50 mL/min at 0.75 minutes
Septum purge flow modeStandard
Injection volume1.0 µL
Inlet temperature320 °C
Carrier gasHelium
Inlet linerAgilent 4 mm single taper, with glass wool (p/n 5190-2293)
Oven80 °C for 1 minute, 25 °C/min to 200 °C, 8 °C/min to 335 °C, hold 6.325 minutes
Total run time: 29 minutes
Post run time: 1.5 minutes
Equilibration time: 0.5 minutes
Column 1 Control modeConstant flow, 0.9272 mL/min
Inlet connectionSplit/Splitless
Outlet connectionPSD (PUU)
Post run flow (backflushing)-12.027 mL/min
Column 2 Control modeConstant flow, 1.1272 mL/min
Inlet connectionPUU
Outlet connectionMSD
Post run flow (backflushing)12.518 mL/min
Source Drawout lensInert Extractor, 9 mm
Vacuum pumpPerformance turbo
Tune fileAtune.U
ModeSIM
Solvent delay4 minutes
EM voltage gain mode1.0
TIDon
Quadrupole temperature150 °C
Source temperature320 °C
Transfer line temperature320 °C
JetClean modeAcquire and Clean
JetClean hydrogen flow0.33 mL/min

Table 2. SIM ions used for quantifier and qualifiers.

Compound RT (min) Quantifier Qualifier 1 Qualifier 2 Qualifier 3
Naphthalene-d85.126136134108
Naphthalene5.149128127129102
1-Methylnaphthalene5.758142141115139
2-Methylnaphthalene5.926142141115143
Biphenyl6.30415415376155
2,6-Dimethylnaphthalene6.346156141155115
Acenaphthylene7.04215215115376
Acenaphthene-d107.15016480
Acenaphthene7.204153154151155
2,3,5-Trimethylnaphthalene7.416170155169153
Fluorene7.912166165163167
Dibenzothiophene9.675184185139152
Phenanthrene-d109.881188189
Phenanthrene9.935178179177152
Anthracene10.002178179177152
1-Methylphenanthrene11.282192191193190
Fluoranthene12.952202203201101
Pyrene13.764202203201101
Benz[a]anthracene17.215228226229114
Chrysene-d1217.381240236
Chrysene17.474228226229114
Benzo[b]fluoranthene20.461252126
Benzo[k]fluoranthene20.528252126
Benzo[j]fluoranthene20.624252126
Benzo[e]pyrene21.494252253126250
Benzo[a]pyrene21.631252253250126
Perylene-d1221.889264260
Perylene21.966252253126250
Dibenz[a,c]anthracene24.460278279139138
Dibenz[a,h]anthracene24.588278279139138
Indeno[1,2,3-cd]pyrene24.622276138277137
Benzo[ghi]perylene25.778276138277137

Results and discussion

Initial calibration

Figure 2 shows the SIM TIC of the 100 pg/µL calibration standard. With the parameters chosen, the peak shapes for all PAHs, especially the latest ones, are very good.

The use of the 9 mm lens and continuous hydrogen cleaning often results in a reduced signal-to-noise ratio (S/N), so it is important to check the lowest desired calibration level. As an example, Figure 3 shows the response at the quantifier ion for several of the compounds at the 1 pg level. All analytes at the 1 pg level had sufficient signal for calibration.

Figure 2. SIM TIC of the 100 pg/µL calibration standard. [Description: A chromatogram showing the Total Ion Chromatogram (TIC) in Selected Ion Monitoring (SIM) mode for a 100 pg/µL calibration standard, with numbered peaks corresponding to identified PAHs and ISTDs. The x-axis represents acquisition time in minutes, ranging from approximately 5 to 28 minutes.]

Figure 3. Response at quantifier ion for select compounds in the lowest calibration standard (1 pg). [Description: Three separate, zoomed-in chromatogram plots. The first shows the response for Fluorene (SIM 166) between 7.75 and 8.65 minutes. The second shows responses for Fluoranthene and Pyrene (SIM 202) between 12.3 and 14.7 minutes. The third shows the response for Benzo[ghi]perylene (SIM 276) between 24.8 and 27.5 minutes.]

Stability of response

Table 3 shows the R² values for three ISTD calibrations of the system with seven levels from 1 to 1,000 pg. All analytes show excellent linearity across the entire range.

Figure 4 shows the precision of ISTD peak responses for 60 sequential replicate injections of the 100 pg standard. The RSDs of the ISTD areas were: Naphthalene-d8 (3.3 %), Acenaphthene-d10 (3.2 %), Phenanthrene-d10 (3.4 %), Chrysene-d12 (2.7 %), Perylene-d12 (2.0 %).

Figure 5 shows the calculated concentration for several analytes in the 60 sequential replicate runs of the 100 pg standard. The system exhibits excellent stability of response. The average RSD of the calculated concentrations for all 27 analytes is 1.1 %.

Table 3. R² values of three seven-level ISTD calibrations, 1 to 1,000 pg SIM. [Table content listing compounds, RT, and R² values for three calibration sets.]

Figure 4. ISTD response stability over 60 injections for a 100 pg calibration standard. Areas are normalized to that of the first injection. [Description: A line graph showing the normalized ISTD response (percentage) over 60 injection numbers for five deuterated PAH standards: Naphthalene-d8, Acenaphthene-d10, Phenanthrene-d10, Chrysene-d12, and Perylene-d12. The y-axis ranges from 50% to 120%.]

Figure 5. Stability of calculated concentrations over 60 sequential injections for a 100 pg calibration standard. [Description: A line graph showing the calculated concentration (normalized to 100%) over 60 injection numbers for Naphthalene, Acenaphthylene, Benzo[k]fluoranthene, and Benzo[ghi]perylene. The y-axis ranges from 60% to 120%.]

Stability of response with soil extracts

The soil extract used for the robustness test was deliberately chosen to have a high-matrix content to challenge the system. Figure 6 compares the scan TIC of the extract to that of the 100 pg PAH standard. The soil extract has a very high level of matrix. Note that, for soils with this level of organic content, further sample cleanup should be considered for routine analysis. The sample preparation used was for test purposes only.

To test the robustness of the system, the soil extract was spiked with 100 pg each of the 27 analytes and 500 pg each of the ISTDs. The spiked extract was then injected 60 times. The PAHs were quantitated against the solvent-based calibration curve for each run, and the resulting calculated concentrations were plotted. Figure 7 shows the calculated concentrations for several of the analytes. Naphthalene and benzo[ghi]perylene both show measured concentrations higher than the spiked 100 pg level. These compounds were found to be present in the soil at levels roughly corresponding to the offset in Figure 7. Perylene (not shown) was found at almost 200 pg in the soil.

The average RSD for the calculated concentrations of all 27 analytes was 4.4 %. For 22 of the 27 analytes, the calculated concentration was within 20 % after 60 soil shots, compared to the first injection in the soil. As expected, the heaviest analytes, such as benzo[ghi]perylene, lost response quickest.

Figure 6. Scan TIC of soil extract and PAH 100 pg standard with 500 pg ISTDs, both drawn in the same scale, showing a large amount of material in the extract. [Description: A chromatogram comparing the Total Ion Chromatogram (TIC) of a heavy soil matrix extract against a PAH standard, illustrating the significant matrix interference. The x-axis represents acquisition time in minutes, ranging from approximately 5 to 28 minutes.]

Figure 7. Stability of calculated concentrations over 60 injections of a soil matrix spiked with 100 pg PAH standards and 500 pg ISTDs. [Description: A line graph showing the calculated concentration (normalized to 100% or higher) over 60 injection numbers for Naphthalene, Acenaphthylene, Benzo[k]fluoranthene, and Benzo[ghi]perylene when analyzing a spiked soil matrix. The y-axis ranges from 40% to 140%.]

Post-Maintenance Calibration Check

After the 60 injections of soil extract, inlet maintenance was performed. This consisted of changing the septum, inlet liner, and gold seal, and removing 30 cm from the head of column 1. While the liner and gold seal were out, the inlet was cleaned with cotton swabs saturated with methanol. After maintenance, the 100 ppb calibration standard was run and quantitated using the original calibration curve generated before both of the replicate studies. Table 4 shows the measured concentrations. All analytes were within 7 % of the expected concentration. Table 4 presents the R² values for a full calibration after inlet maintenance. The data in Table 4 demonstrate that the degradation in system performance with the soil is limited to the inlet and column head, as expected.

The source did not require cleaning, as is often the case with matrix levels such as those used here. The use of JetClean and the 9 mm drawout lens greatly reduce the deposits that normally degrade source performance.

Table 4. Calibration check and R² values of 7 level ISTD-calibration 1 to 1,000 pg SIM after the system maintenance. [Table content listing compounds, RT, calculated concentration (%), and R² values after maintenance.]

Conclusions

This system addresses many of the problems encountered with GC/MS PAH analysis. The use of JetClean, the 9 mm drawout lens, higher zone temperatures, and the appropriate liner result in substantial improvements in linearity, peak shape, and system robustness. The greatly reduced need for manual source cleaning provided by JetClean is a welcome productivity improvement for the lab.

For labs analyzing high volumes of samples containing significant matrix interferences, the Agilent 8890/7000D triple quadrupole GC/MS with JetClean and midcolumn backflush offers all the advantages demonstrated here plus the much higher specificity of MS/MS⁴. Use of GC/MS/MS simplifies the data review versus GC/MS by providing much higher selectivity over spectral interferences from the matrix.

References

  1. Szelewski, M.; Quimby, B. D. Optimized PAH Analysis Using the Agilent Self-Cleaning Ion Source and Enhanced PAH Analyzer, Agilent Technologies Application Note, publication number 5191-3003EN, 2013.
  2. Anderson, K. A.; et al. Modified ion source triple quadrupole mass spectrometer gas chromatograph for polycyclic aromatic hydrocarbons, Journal of Chromatography A 2015, 1419(6), 89-9US.
  3. Quimby, B. D.; Prest, H. F.; Szelewski, M. J.; Freed, M. K. In-situ conditioning in mass spectrometer systems, US Patent 8,378,293 Feb 19, 2013.
  4. Andrianova, A. A.; Quimby, B. D. Optimized GC/MS/MS Analysis for PAHs in Challenging Matrices, Agilent Technologies Application Note, publication number 5994-0498EN, 2019.

PDF preview unavailable. Download the PDF instead.

application-optimized-gc-ms-analysis-for-PAHs-in-challenging-matrices-8890-5977b-single-quadrupole-gc-ms-5994-0499en-agilent , Bruce D. Quimby | Agilent Technologies, Inc. Adobe PDF Library 15.0

Related Documents

Preview Agilent MassHunter GC/MS Hardware User Manuals and Tools
A comprehensive list of user manuals, quick start guides, and troubleshooting resources for Agilent's GC/MS systems, including the 5977B GC/MSD, 5975/77 Series MSD for OpenLAB, 7000D/7010B Series Triple Quad MS, and 7200 Accurate-Mass Q-TOF GC/MS systems, all designed to work with MassHunter software.
Preview 2021 Agilent University Customer Training Catalog
Explore the 2021 Agilent University Customer Training Catalog to advance your lab skills. Discover online learning, classroom training, and onsite options for chromatography and other laboratory techniques. Enhance your expertise with Agilent's comprehensive training programs.
Preview Agilent MassHunter Data Acquisition Software for 5975/5977 Series GC/MSD Familiarization Guide
This familiarization guide provides essential information for users of Agilent MassHunter Data Acquisition Software for 5975/5977 Series GC/MSD, covering system setup, method creation, and sample analysis.
Preview Agilent MassHunter Workstation Software Quantitative Analysis Familiarization Guide
A guide to help users learn the Agilent MassHunter Workstation Software Quantitative Analysis program through step-by-step exercises. Covers setting up batches, methods, target compounds, quantitation, and reviewing results for MRM and Q-TOF data.
Preview Agilent P200/P200H Micro Gas Chromatograph User's Manual
Comprehensive user's manual for the Agilent P200 and P200H Micro Gas Chromatographs, detailing operation, installation, theory, and troubleshooting for fast, accurate gas analysis.
Preview Agilent ICP-MS Solution for USP <232>/<233> and ICH Q3D Elemental Impurities Analysis
This white paper details Agilent's Inductively Coupled Plasma Mass Spectrometry (ICP-MS) solution for elemental impurities analysis in pharmaceutical products, aligning with USP <232>/<233> and ICH Q3D guidelines. It covers regulatory requirements, analytical methods, instrument capabilities, and data quality assurance.
Preview Agilent MassHunter Workstation Software Installation Guide: Offline Qualitative and Quantitative Analysis
Comprehensive installation guide for Agilent MassHunter Workstation Software, covering offline qualitative and quantitative analysis. Includes steps for preparing installation, installing programs, configuring Excel, and setting up reporting tools.
Preview Agilent 1260 Infinity II Vialsampler Data Sheet
Data sheet for the Agilent 1260 Infinity II Vialsampler, an autosampler designed for routine pharmaceutical tasks, quality control, environmental, and food analyses. Includes product description, features, and detailed physical and performance specifications.