Agilent 1260 Infinity II Vialsampler

Data Sheet

Product Description

The Agilent 1260 Infinity II Vialsampler is an autosampler designed for the reliability and ease-of-use needed for routine pharmaceutical tasks and quality control, as well as for environmental and food analyses. It can house optionally the integrated column compartment for two LC columns with temperature control up to 80 °C as well as a sample thermostat for stable temperatures from 4 °C to 40 °C, all within one module.

Features

Specifications

Table 1 Physical Specifications

Type Specification Comments
Weight 19 kg (41.9 lbs) w/o sample thermostat
Dimensions (height x width x depth) 320 x 396 x 468 mm (12.8 x 15.6 x 18.4 inches)
Line voltage 100 – 240 V~, ± 10 % Wide-ranging capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 350 VA / 350 W / 1195 BTU/h
Ambient operating temperature 4 – 40 °C (39 – 104 °F), without sample thermostat up to 55 °C (131 °F)
Ambient non-operating temperature -40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F)¹ Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15092 ft) For storing the module
Safety standards: IEC, EN, CSA, UL Installation category II, Pollution degree 2 For indoor use only.
ISM Classification ISM Group 1 Class B According to CISPR 11
Permitted solvents Auto-ignition temperature ≥200 °C Boiling point ≥56 °C

¹ If a sample thermostat is included the upper value for humidity can be reduced. Please check your lab conditions to stay beyond dew point values for non-condensing operation.

Table 2 Performance Specifications 1260 Infinity II Vialsampler (G7129A)

Type Specification
Injection range 0.1 – 100 µL in 0.1 µL increments with 100 µL up to 60 MPa
0.1 – 900 µL in 0.1 µL increments with 900 µL up to 40 MPa
Injection Precision <0.25% RSD of peak areas from 5 µL to 100 µL
Pressure range 0 – 60 MPa (0 – 600 bar, 0 – 8702 psi)
0 – 40 MPa (0 – 400 bar, 0 – 5801 psi)
Sample viscosity range 0.2 – 5 cp
Sample capacity 132 x 2 mL vial (two trays default)
100 x 2 mL vial (two classic trays optional)
36 x 6 mL vials (two trays optional)
Carry over <0.004 % (40 ppm) with needle wash
Injection cycle time 18 s for draw speed 200 µL/min
Ejection speed: 200 µL/min
Injection volume: 1 µL
Minimum sample volume 1 µL from 5 µL sample in 100 µL microvial, or
1 µL from 10 µL sample in 300 µL microvial.
Instrument Control LC & CE Drivers A.02.12 or above
Instrument Control Framework (ICF) A.02.03 or above
Instant Pilot (G4208A) with firmware B.02.19 or above
Lab Advisor B.02.07 or above
Communications Controller-area network (CAN), Local Area Network (LAN)
ERI: ready, start, stop and shut-down signals
Maintenance and safety-related features Leak detection, safe leak handling, leak output signal for shutdown of pumping system, and low voltages in major maintenance areas
GLP features Extensive diagnostics, error detection and display with Agilent Lab Advisor software
Housing All materials recyclable.

Table 3 Performance Specifications 1260 Infinity II Vialsampler (G7129C)

Type Specification
Injection range 0.1 – 100 µL in 0.1 µL increments with 100 µL up to 80 MPa
Injection precision <0.25% RSD of peak areas from 5 µL to 100 µL
Pressure range 0 – 80 MPa (0 – 800 bar, 0 – 11603 psi)
Sample viscosity range 0.2 – 5 cp
Sample capacity 132 x 2 mL vial (two trays default)
100 x 2 mL vial (two classic trays optional)
36 x 6 mL vials (two trays optional)
Carry over <0.004% (40 ppm) with needle wash
Injection cycle time 18 s for draw speed 200 µL/min
Ejection speed: 200 µL/min
Injection volume: 1 µL
Minimum sample volume 1 µL from 5 µL sample in 100 µL microvial, or
1 µL from 10 µL sample in 300 µL microvial.
Instrument Control LC & CE Drivers A.02.17 or above
Instrument Control Framework (ICF) A.02.05 or above
Instant Pilot (G4208A) with firmware B.02.22 or above
InfinityLab LC Companion (G7108A)
Lab Advisor B.02.10 or above
Communications Controller-area network (CAN), Local Area Network (LAN)
ERI: ready, start, stop and shut-down signals
Maintenance and safety-related features Extensive diagnostics, error detection and display with Agilent Lab Advisor software
GLP features Leak detection, safe leak handling, leak output signal for shutdown of pumping system, and low voltages in major maintenance areas
Housing All materials recyclable.

Table 4 Physical Specifications of the Sample Thermostat

Type Specification Comment
Weight <6 kg
Dimensions (height x width x depth) 205 mm x 340 mm x 370 mm
Refrigerant gas R600a (0.030 kg) Ozone depletion potential (ODP) =0
Global warming potential (GWP) =3
Supply voltage 24VDC
Current 10 A max.
Ambient operating temperature 4 – 40 °C (39.2 – 104 °F)
Ambient non-operating temperature -40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 3000 m (9842 ft)
Non-operating altitude Up to 4600 m (15091 ft) For indoor use only
Safety standards: IEC, EN, CSA, UL Installation category II, Pollution degree 2
ISM Classification ISM Group 1 Class B According to CISPR 11

Table 5 Performance Specifications for the Sample Thermostat

Type Specifications
Operating principle High performance, low-energy consumption micro-compressor based cooler with natural R600a coolant (Butane 30 g), user-upgradable
Temperature range from 4 – 40 °C
Temperature settable from 4 – 40 °C in 1° increments
Temperature accuracy (<25 °C, <50 % r.H.) 2 – 6 °C at a setpoint of 4 °C

Table 6 Ordering Details 1260 Infinity II Vialsampler

Description Part Number
Agilent 1260 Infinity II Vialsampler
UHPLC design autosampler up to 800 bar for 0.1 – 100 µL injections. Standard needle flush port and peristaltic pump. Default setup with 100 µL loop and 100 µL analytical head.
G7129C
Drawer choice 6 x 11 vials (2 mL), 2 drawers included G7129C #010
Drawer choice 3x 6 vials (6 mL), 2 drawers included G7129C #011
Drawer choice 5x 10 vials (2 mL), 2 drawers included G7129C #012
Extended injection range option
900 µL analytical head and sample loop for large single-stroke injections (400 bar pressure limit)
G7129C #020
InfinityLab Sample Thermostat
Thermostat unit to fit G7167A/B samplers as well as G7129 A/B/C samplers. Slide-in unit, customer installable.
G7129C #101
InfinityLab Sample Thermostat Upgrade
Slide-in thermostat for existing Multisampler instruments. Customer installable.
Requires FW 7.22 or higher.
G4761A
Integrated column compartment
Integrated column compartment for up to two columns with 3 µL heater volume, for standard analytical flow rates up to 5 mL/min.
Recommended for lowest dispersion and filtered samples
G7129C #063
Integrated column compartment
Integrated column compartment for up to two columns with 6 µL heater volume, for standard analytical flow rates up to 5 mL/min
Recommended for robust routine analysis
G7129C #066
2 mL Vial Drawer (6x11 positions)
Each drawer can hold 66 x 2 mL vials. Samples are set up in an A1 to F11 format (microtiter-plate style). 2 drawers/pack.
G7129C #010
6 mL vial drawer (3x6 positions)
Each drawer can hold 18 x 6 mL vials. Samples are set up in a A1 to C6 format (microtiter-plate style). 1 drawer/pack.
G7129C #011
Classic drawer kit (10x10 positions split in 2 drawers)
Kit of 2 drawers to hold 100x 2 mL vials. Samples are set up in a 1 to 100 format (classic Agilent vial sampler style).
G7129C #012
Analytical head 40 µL
40 µL analytical head, 800 bar pressure limit.
G7129C #162
Sample loop 40 µL
Sample loop 40 µL, for maximum 40 µL injection, 800 bar pressure limit.
G7129C #150
Sample loop 20 µL
Sample loop 20 µL for maximum 20 µL injection, 800 bar pressure limit.
G7129C #148
Multidraw option adding 400 µL / 1400 µL
Includes two seat capillary extensions to inject up to 500 or up to 1500 µL volume.
G7129C #021
Walkup extension: external tray
External tray offering 5 vial positions (2 mL) and 1 vial disposal position.
G7129-60000
Walkup extension: vial disposal tube
Disposal tube to fit on vial disposal position.
G1313-27302

Table 7 Ordering Details 1260 Infinity II Vialsampler

Description Part Number
Agilent 1260 Infinity II Vialsampler
UHPLC design autosampler up to 600 bar for 0.1 – 100 µL injections. Includes two drawers, each with capacity for 66 sample vials (2 mL), 132 vials total capacity. Standard needle flush port and peristaltic pump. Default setup with 100 µL loop and 100 µL analytical head.
G7129A
Extended injection range option
900 µL analytical head and sample loop for large single-stroke injections (400 bar pressure limit)
G7129A #020
InfinityLab Sample Thermostat
Thermostat unit to fit G7167A/B samplers as well as G7129 A/B/C samplers. Slide-in unit, customer installable.
G7129A #101
InfinityLab Sample Thermostat Upgrade
Slide-in thermostat for existing Multisampler instruments. Customer installable.
Requires FW 7.22 or higher
G4761A
Integrated column compartment
Integrated column compartment for up to two columns with 3 µL heater volume, with 0.12 mm capillaries optimized for highest chromatographic performance.
Recommended for standard flow rates up to 3 mL/min and filtered samples.
G7129A #063
Integrated column compartment
Integrated column compartment for up to two columns with 6 µL heater volume, with 0.17 mm capillaries for standard analytical flow rates up to 5 mL/min and higher.
Recommended for robust routine analysis.
G7129A #066
6 mL vial drawer (3x6 positions)
Each drawer can hold 18 x 6 mL vials. Samples are set up in a A1 to C6 format (microtiter-plate style). 1 drawer/pack.
G7129A #011
Classic drawer kit (10x10 positions split in 2 drawers)
Kit of 2 drawers to hold 100x 2 mL vials. Samples are set up in a 1-100 format (classic Agilent vial sampler style).
G7129A #012
Analytical head 40 µL
40 µL analytical head, 600 bar pressure limit.
G7129A #162
Sample loop 40 µL
Sample loop 40 µL, for maximum 40 µL injection, 600 bar pressure limit.
G7129A #150
Sample loop 20 µL
Sample loop 20 µL for maximum 20 µL injection, 600 bar pressure limit.
G7129A #148
Multidraw option adding 400 µL / 1400 µL
Includes two seat capillary extensions to inject up to 500 or up to 1500 µL volume.
G7129A #021
Walkup extension: external tray
External tray offering 5 vial positions (2 mL) and 1 vial disposal position.
G7129-60000
Walkup extension: vial disposal tube
Disposal tube to fit on vial disposal position.
G1313-27302

PDF preview unavailable. Download the PDF instead.

5991-7093en 1260 ii vialsampler datasheet Acrobat Distiller 11.0 (Windows)

Related Documents

Preview Agilent 1260 Infinity II Binary LC System: Manual and Quick Guide
This manual and quick guide provides comprehensive information on the Agilent 1260 Infinity II Binary LC System, covering product description, features, system components, configuration settings, quick start procedures, parts, consumables, and safety information.
Preview Agilent InfinityLab LC Series Vialsamplers User Manual
User manual for the Agilent InfinityLab LC Series Vialsamplers, covering product descriptions, features, site requirements, usage, preparation, performance optimization, troubleshooting, maintenance, parts, cable identification, hardware information, and LAN configuration.
Preview Agilent LC System Best Practices: Technical Guide
Comprehensive guide to best practices for operating and maintaining Agilent Liquid Chromatography (LC) systems, covering housekeeping, daily tasks, power management, and component-specific recommendations for pumps, samplers, degassers, columns, and detectors.
Preview Agilent InfinityLab LC Supplies Catalog
Explore the comprehensive Agilent InfinityLab LC Supplies catalog, featuring a wide range of instruments, columns, and accessories designed to optimize your liquid chromatography workflows for maximum efficiency and performance.
Preview Agilent InfinityLab LC Series Installation Checklist Guide
A comprehensive installation checklist for Agilent InfinityLab LC Series instruments, detailing customer and service engineer responsibilities, preparation steps, module-specific installation procedures, and checkout methods for various detectors and configurations.
Preview Agilent InfinityLab LC Series Site Preparation Checklist for HPLC Systems
Comprehensive site preparation checklist for Agilent InfinityLab LC Series instruments, detailing laboratory space, environmental conditions, power requirements, solvent and gas needs, and stack configurations to ensure successful and timely installation of HPLC systems.
Preview Agilent 1260 Infinity バイナリポンプ ユーザーマニュアル
Agilent 1260 Infinity バイナリポンプの設置、操作、メンテナンス、トラブルシューティング、および仕様に関する詳細情報を提供するユーザーマニュアル。高性能液体クロマトグラフィーシステムに不可欠なコンポーネントについて解説します。
Preview Agilent ProSize Data Analysis Software: Features and Usage Guide
Comprehensive guide to Agilent's ProSize Data Analysis Software, detailing new features, main screen interface, and advanced functionalities such as Y-axis/X-axis adjustments, gel image annotation, inclusion regions, and logo integration for reports.