Product Overview

The InTemp CX405 RTD dry ice logger is designed for Cold Chain monitoring and features a built-in external probe capable of measuring temperatures as low as -95°C (-139°F). This Bluetooth® Low Energy-enabled logger facilitates wireless communication with a mobile device. The InTemp app allows users to easily configure the logger, perform daily checks, download reports, and monitor alarms. Alternatively, InTempConnect can be used with the CX5000 Gateway for logger configuration and data download. The logger's LCD screen provides real-time temperature, logging status, and battery usage information. Data uploaded to InTempConnect can be used for tracking logger configurations and building custom reports.

Specifications

External Temperature Probe

Range -95° to 50°C (-139° to 122°F)
Accuracy ±1.5°C from -95° to -40°C (±2.7°F from -139° to -40°F)
±1.0°C from -40° to -10°C (±1.8°F from -40° to 14°F)
±0.5°C from -10° to 50°C (±0.9°F from -14° to 122°F)
Resolution 0.05°C at 25°C (0.09°F at 77°F)
Drift <0.1°C (0.18°F) per year
NIST Calibration Two-point NIST calibration, probe only
Cable Length 1.5 m (4.92 ft)
Probe Dimensions Food-grade stainless steel probe with blunt tip, 31.75 mm (1.25 inches) long, 4.78 mm (0.188 inches) diameter

Ambient Temperature Sensor

Range -30° to 70°C (-22° to 158°F)
Accuracy ±0.5°C from -15° to 70°C (±0.9°F from 5° to 158°F)
±1.0°C from -30° to -15°C (±1.8°F from -22° to 5°F)
Resolution 0.024°C at 25°C (0.04°F at 77°F)
Drift <0.1°C (0.18°F) per year

Logger

Radio Power 1 mW (0 dBm)
Transmission Range Approximately 30.5 m (100 ft) line-of-sight
Wireless Data Standard Bluetooth Low Energy (Bluetooth Smart)
Logger Operating Range -30° to 70°C (-22° to 158°F), 0 to 95% RH (non-condensing)
Logging Rate 1 second to 18 hours
Time Accuracy ±1 minute per month at 25°C (77°F)
Battery Type Two AAA 1.5 V alkaline or lithium batteries, user replaceable
Battery Life 1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the app, excessive report generation, numerous audible alarms, and paging all impact battery life.
Memory 128 KB (75,245 total measurements or 37,622 probe & internal paired measurements)
Full Memory Download Time Approximately 60 seconds; may take longer the farther the device is from the logger
LCD LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Dimensions 9.4 x 4.5 x 2.59 cm (3.7 x 1.77 x 1.02 inches)
Weight 90.2 g (3.18 oz)
Environmental Rating IP54
Certifications CE, FC

Included Items

Required Items

Logger Components and Operation

The logger features an LCD screen, Start button, Mute or Next button, audible alarm speaker, alarm LED, temperature sensor, and four magnets on the back for mounting.

LCD Symbols and Descriptions

LCD Symbol Description
CHECK An alarm has tripped because the temperature reading is outside the specified range.
MEM USED Indicates memory usage. Approximately 40% used in the example.
Battery Indicator Shows the approximate battery power remaining.
LOGGING The logger is currently logging.
Bluetooth Signal Bars Indicates connection strength to a phone or tablet via Bluetooth.
START The logger is waiting to be started. Press and hold the Start button for 3 seconds.
PROBE Temperature reading from the external probe.
AMBIENT Temperature reading from the internal sensor.
MIN The minimum temperature reading within the current 24-hour period or entire logging period.
MAX The maximum temperature reading within the current 24-hour period or entire logging period.
MUTE / MUTED Indicates an alarm is beeping (MUTE) or has been muted (MUTED).
Delayed Start (e.g., 05:38) The logger is configured to start logging after a delay. The display shows the remaining time.
LoAd Profile settings are being loaded onto the logger.
Err An error occurred during profile loading. Reconfigure the logger.
HELLO The logger was paged from the InTemp app.
STOP The logger has been downloaded and stopped, or memory is full.
book The logger is being updated with new firmware.

Logger Operations

Getting Started

InTempConnect is a web-based software for monitoring logger configurations and downloaded data. The InTemp app allows configuration and data download via mobile devices, with reports saved in the app and uploaded to InTempConnect. The CX5000 Gateway can also be used for automatic configuration and data upload.

For users who do not require cloud-based data access, the logger can be used solely with the InTemp app.

Initial Setup Steps:

  1. Administrator Setup: Create an InTempConnect account or log in if an account already exists. Follow steps c and d for existing accounts.
  2. Logger Setup: Insert two AAA batteries, observing polarity. Secure the battery door with the included screw and screwdriver.
  3. Download InTemp App & Log In: Download the InTemp app from the App Store or Google Play™. Enable Bluetooth. Log in using InTempConnect credentials or create a new account for InTemp app only use. Ensure the correct login option is selected.
  4. Configure Logger: Access the logger via the app's Devices icon. Select a logger profile or set up a custom profile. Administrators can also set up daily logger checks and trip information fields.
  5. Deploy and Start: Deploy the logger to the monitoring location. Logging begins according to the selected profile settings. If daily checks are enabled, perform them daily via the app. The logger LCD will display current, minimum, and maximum readings.

Troubleshooting Connection: Ensure the logger is within range, move closer if needed, adjust device orientation for better signal, or restart the mobile device if connection fails.

Logger Alarms

Alarms can be configured in the logger profile to trigger when temperature readings fall outside specified values. When an alarm trips:

To mute a beeping alarm, press the Mute button. Download the logger to view alarm details and clear indicators.

Passkey Protection

The logger is protected by an encrypted passkey, automatically generated by the InTemp app for InTempConnect users, or optionally set by InTemp app only users. This passkey ensures that only authorized devices can connect to the logger.

Downloading the Logger

Download logger data via the InTemp app to generate reports including readings, events, and alarms. Reports can be shared immediately or accessed later.

Reports are available in PDF, XLSX, and VFC CSV formats. Access previously downloaded reports via the Reports icon in the app. For more details, visit www.intempconnect.com/help.

Performing Logger Checks

Daily or twice daily checks can be performed using the InTemp app. Enable this feature in the app's Settings under 'Record CX400 Logger Checks'.

Notifications can be set up in the app's Reminders option.

Logger Events

The logger records various events to track its operation and status:

Deploying the Logger

Use the four magnets on the logger's back to mount it to a magnetic surface. Protect the external probe cable's sheath during handling.

Protecting the Logger

The logger is designed for indoor use. Protect it from moisture and corrosion. If the logger gets wet, remove batteries immediately and dry the circuit board. Avoid static electricity; ground yourself when handling the logger.

Battery Information

The logger requires two user-replaceable AAA 1.5 V alkaline or lithium batteries. Battery life is approximately 1 year with typical usage, but can be affected by ambient temperature, connection frequency, alarm duration, and battery performance. Use batteries with flat negative terminals; avoid batteries with indentations.

To Install or Replace Batteries:

  1. Remove the battery door using a Phillips-head screwdriver.
  2. Remove old batteries.
  3. Insert new batteries, observing polarity.
  4. Screw the battery door back into place.

WARNING: Do not cut open, incinerate, heat above 85°C (185°F), recharge lithium batteries, or dispose of in fire. Avoid exposing batteries to water. Dispose of batteries according to local regulations.

Federal Communication Commission Interference Statement

This equipment complies with FCC Rules Part 15. Operation is subject to two conditions: (1) it may not cause harmful interference, and (2) it must accept any interference received, including interference that may cause undesired operation. Modifications not expressly approved by the responsible party may void the user's authority to operate the equipment.

Industry Canada Statements: This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to two conditions: (1) it may not cause interference, and (2) it must accept any interference, including interference that may cause undesired operation.

RF Radiation Exposure: To comply with FCC and Industry Canada RF radiation exposure limits, the logger must be installed to provide a separation distance of at least 20cm from all persons and must not be co-located or operating with any other antenna or transmitter.

Models: CX405, RTD Dry Ice Logger

PDF preview unavailable. Download the PDF instead.

cx405-rtd-manual

Related Documents

Preview InTemp CX600/CX700 Cryogenic Data Logger Manual and Specifications
Comprehensive manual for InTemp CX600 Dry Ice and CX700 Cryogenic Loggers. Covers specifications, setup, operation, data download, and reporting for cold chain temperature monitoring.
Preview InTemp CX450 Temperature and RH Logger User Manual
Comprehensive user manual for the InTemp CX450 Temperature and Relative Humidity (RH) Logger, detailing its features, specifications, setup, operation, and data management for pharmaceutical, life science, and medical monitoring.
Preview Getting Started with InTemp CX400 Series Data Loggers
A comprehensive guide for setting up and using InTemp CX400 Series data loggers, covering both cloud-based InTempConnect and standalone mobile app functionalities for temperature monitoring.
Preview InTemp CX5000 Gateway Manual: Setup, Configuration, and Deployment Guide
This manual provides comprehensive instructions for setting up, configuring, and deploying the InTemp CX5000 Gateway. Learn how to connect loggers, manage network settings, and monitor data via the InTempConnect website and InTemp app.
Preview InTemp CX400 Series Temperature Logger Quick Start Guide
A quick start guide for setting up and using the InTemp CX400 Series Temperature Logger, covering account setup, logger configuration, app download, deployment, and data download procedures.