Protek Power PM100 Series 100W Medical Power Supplies
FSP Power Solution GmbH
Description
The PM100 series consists of compact, open PCB constructed, AC-DC switching power supplies capable of delivering 100 watts of continuous output power at convection cooling. They are designed for medical applications. Approval to EN60601-1 safety standards streamlines design-in time and reduces end equipment compliance costs.
Features
- Medical approval
- Compact size: 2" x 4" x 1.26"
- High power density: 10 W/cubic inch
- 100 W output with convection cooling up to +50°C
- Low earth leakage current
- EN55011 class B emissions
- RoHS compliant
Safety Standard Approvals
The power supplies meet the following safety standards:
- UL ES 60601-1, CSA C22.2 No. 60601-1 (File No. E178020) [UL Certified]
- TUV EN 60601-1, IEC 60601-1 [TUV Certified]
Input Specifications
- Input voltage:
- 90-132 / 180-264 VAC (Universal mains supply operation)
- Input frequency:
- 47-63 Hz
- Input current:
- 1.9 A (rms) for 100-120 VAC; 1.1 A (rms) for 200-240 VAC
- Earth leakage current:
- 150 μA max. @ 264 VAC, 63 Hz
- Touch current:
- 100 μA max. @ 264 VAC, 63 Hz
Output Specifications
- Output voltage/current:
- See rating chart.
- Total output power:
- 100 watts maximum
- Ripple and noise:
- 150 mV peak to peak on 5.0 V model, 1% peak to peak on other models
- Overvoltage protection:
- Provided on output; set at 110-140% of its nominal output voltage, latching by recycle input to reset.
- Short circuit protection:
- Automatic recovery
- Temperature coefficient:
- All outputs ±0.04% /°C maximum
- Transient response:
- Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 μs after a 25% step load change.
Environmental Specifications
- Operating temperature:
- -10°C to +70°C
- Storage temperature:
- -40°C to +85°C
- Relative humidity:
- 5% to 95% non-condensing
- Temperature derating:
- Derate from 100% at +50°C linearly to 50% at +70°C
- Cooling:
- Convection
General Specifications
- Switching frequency:
- 100 KHz (typical)
- Efficiency:
- 88-90% @ 230 VAC full load
- Hold-up time:
- 12 ms minimum at 110 VAC
- Line regulation:
- ±0.2% maximum at full load
- Inrush current:
- 40 A @ 115 VAC or 80 A @ 230 VAC, at 25°C cold start
- Withstand voltage:
- 5600 VDC from input to output (2 MOPP); 2100 VDC from input to ground (1 MOPP); 700 VDC from output to ground. (Note: Verify AC strength with correct test method to avoid power supply damage.)
- MTBF:
- 140,000 hours at full load at 25°C ambient temperature, calculated per MIL-HDBK-217F
EMC Performance
- EN55011:
- Class B conducted, class B radiated
- EN61000-3-2:
- Harmonic distortion, class A
- EN61000-3-3:
- Line flicker
- EN60601-1-2:
- Standard for medical electrical equipment EMC requirements.
- EN61000-4-2 (ESD):
- ±15 KV air and ±8 KV contact
- EN61000-4-3 (Radiated Immunity):
- 9-28 V/m
- EN61000-4-4 (Fast Transient/Burst):
- ±2 KV
- EN61000-4-5 (Surge):
- ±1 KV differential, ±2 KV common mode
- EN61000-4-6 (Conducted Immunity):
- 10 Vrms
- EN61000-4-8 (Magnetic Field Immunity):
- 30 A/m
- EN61000-4-11 (Voltage Dip Immunity):
- 30% reduction for 500 ms (criteria A @ 230 VAC), criteria B @ 100 VAC; 100% reduction for 10 ms
Output Voltage/Current Rating Chart
Model | V1 | Min. Current | Max. Current | Tol. | Ripple & Noise (2) | Max. Power | Efficiency (typical) @115/230 Vac |
---|---|---|---|---|---|---|---|
PM100-10A | 5 V | 0 A | 20.00 A | ±2% | 150 mV | 100 W | 87/88% |
PM100-12A | 12 V | 0 A | 8.34 A | ±2% | 120 mV | 100 W | 88/89% |
PM100-13A | 15 V | 0 A | 6.70 A | ±2% | 150 mV | 100 W | 88/89% |
PM100-13-1A | 18 V | 0 A | 5.56 A | ±2% | 180 mV | 100 W | 88/89% |
PM100-14A | 24 V | 0 A | 4.20 A | ±2% | 240 mV | 100 W | 87/90% |
PM100-15A | 28 V | 0 A | 3.58 A | ±2% | 280 mV | 100 W | 87/90% |
PM100-17A | 36 V | 0 A | 2.78 A | ±2% | 360 mV | 100 W | 88/89% |
PM100-18A | 48 V | 0 A | 2.10 A | ±2% | 480 mV | 100 W | 88/89% |
Notes:
- Safety approvals are for PCB form only. To order models with metallic L-bracket or box, change suffix "A" to "B" for L-bracket form, to "C" for enclosed form (see Outline Drawing of Cased Internal Switchers), e.g. PM100-14C.
- Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.
Mechanical Specifications
The power supply unit has the following dimensions: Length 4.00 inches (101.60 mm), Width 2.00 inches (50.80 mm), Height 1.26 inches (32.0 mm). Mounting holes are 0.157 inches (4.00 mm) in diameter, located at 4 places.
Notes:
- Dimensions shown in inches [mm].
- Tolerance ±0.02 [0.5] maximum.
- Connector P1: Molex header 09-65-2038 or equivalent, mating with Molex housing 09-50-1031 or equivalent.
- Connector P2: Molex header 09-65-2068 or equivalent, mating with Molex housing 09-50-1061 or equivalent.
- To ensure compliance with level B emissions, connect the three [asterisk] marked mounting holes with metallic standoffs to chassis.
- Weight: 190 grams (0.44 lbs.) approx.
Output Power Derating Curve
The output power derating curve shows the relationship between ambient temperature and maximum output power. The power supply can deliver 100% of its rated output power up to an ambient temperature of +50°C. Beyond +50°C, the output power derates linearly, reaching 50% of its rated power at +70°C.
Pin Chart
MODEL | PIN 1 | PIN 2 | PIN 3 | PIN 4 | PIN 5 | PIN 6 |
---|---|---|---|---|---|---|
PM100-10A | +V1 | +V1 | +V1 | |||
PM100-12A | +V1 | +V1 | +V1 | |||
PM100-13A | +V1 | +V1 | +V1 | |||
PM100-13-1A | V1 Return | V1 Return | V1 Return | |||
PM100-14A | V1 Return | V1 Return | V1 Return | |||
PM100-15A | V1 Return | V1 Return | V1 Return | |||
PM100-17A | +V1 | +V1 | +V1 | |||
PM100-18A | +V1 | +V1 | +V1 |