Detailed Specifications
High-End DT4281/DT4282 Specifications
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year.
DC Voltage
Range | Accuracy | Input Impedance |
60.000 mV | ±0.2 %rdg. ±25 dgt. | 1 GΩ or more //100 pF or less |
600.00 mV | ±0.025% rdg. ±5 dgt. |
6.0000 V | ±0.025% rdg. ±2 dgt. |
60.000 V | ±0.03 %rdg. ±2 dgt. | 10.3 MΩ ±2% //100 pF or less |
600.00 V | | 10.2 MΩ ± 2% //100 pF or less |
1000.0 V | | |
AC Voltage
Range | Accuracy | Input Impedance |
| 20 to 45 Hz | 45 to 65 Hz | 65 to 1 kHz | 1 k to 10 kHz | 10 k to 20 kHz | 20 k to 100 kHz | |
60.000 mV | ±1.3 %rdg. ±60 dgt. | ±0.4%rdg. ±40 dgt. | ±0.6 %rdg. ±40 dgt. | ±0.9 %rdg. ±40 dgt. | ±1.5 %rdg. ±40 dgt. | ±3.5 %rdg. ±40 dgt. | 1 MΩ ± 4%//100pF or less |
600.00 mV | ±1 %rdg. ±60 dgt. | ±0.2 %rdg. ±25 dgt. | ±0.3 %rdg. ±25 dgt. | ±0.4 %rdg. ±25 dgt. | ±0.7 %rdg. ±40 dgt. | Undefined |
6.0000 V | Undefined | Undefined | Undefined | Undefined | Undefined | Undefined |
60.000 V | | ±0.3 %rdg. ±30 dgt. | ±0.4 %rdg. ±30 dgt. | ±0.4 %rdg. ±30 dgt. | ±1.5 %rdg. ±45 dgt. | ±3.5 %rdg. ±125 dgt. | 10.3 MΩ ± 2.0% //100 pF or less |
600.00 V | | | | | | |
1000.0 V | | | | | | | 10.2 MΩ ± 2% //100 pF or less |
DCV + ACV Measurement
Range | Accuracy | Input Impedance |
| 20 to 45 Hz | 45 to 65 Hz | 65 to 1 kHz | 1 k to 10 kHz | 10 k to 20 kHz | 20 k to 100 kHz | |
6.0000 V | ±1.2 %rdg. ±65 dgt. | ±0.3 %rdg. ±30 dgt. | ±0.4 %rdg. ±30 dgt. | ±0.4 %rdg. ±30 dgt. | ±1.5 %rdg. ±45 dgt. | ±3.5 %rdg. ±125 dgt. | 1 MΩ ± 4%//100pF or less |
60.000 V | Undefined | Undefined | Undefined | Undefined | Undefined | Undefined |
600.00 V | | | | | | | 10.3 MΩ ± 2.0% //100 pF or less |
1000.0 V | | | | | | |
AC Clamp (AC Current)
Range | Accuracy | DT4281 only |
| 40 to 65 Hz | 65 to 1 kHz | 1 k to 10 kHz | 10 k to 20 kHz | 20 k to 100 kHz | |
10.00 A | ±0.6 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. | ±2 %rdg. +2 dgt. | ±4 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. |
20.00 A | ±0.6 %rdg. +4 dgt. | ±0.9 %rdg. +4 dgt. | ±2 %rdg. +4 dgt. | ±4 %rdg. +4 dgt. | ±0.9 %rdg. +4 dgt. | ±0.9 %rdg. +4 dgt. |
50.00 A | ±0.6 %rdg. +10 dgt. | ±0.9 %rdg. +10 dgt. | ±2 %rdg. +10 dgt. | ±4 %rdg. +10 dgt. | ±0.9 %rdg. +10 dgt. | ±0.9 %rdg. +10 dgt. |
100.0 A | ±0.6 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. | ±2 %rdg. +2 dgt. | ±4 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. |
200.0 A | ±0.6 %rdg. +4 dgt. | ±0.9 %rdg. +4 dgt. | ±2 %rdg. +4 dgt. | ±4 %rdg. +4 dgt. | ±0.9 %rdg. +4 dgt. | ±0.9 %rdg. +4 dgt. |
500.0 A | ±0.6 %rdg. +10 dgt. | ±0.9 %rdg. +10 dgt. | ±2 %rdg. +10 dgt. | ±4 %rdg. +10 dgt. | ±0.9 %rdg. +10 dgt. | ±0.9 %rdg. +10 dgt. |
1000 A | ±0.6 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. | ±2 %rdg. +2 dgt. | ±4 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. | ±0.9 %rdg. +2 dgt. |
The optional 9010-50, 9018-50, or 9132-50 CLAMP ON PROBE is used. Accuracy does not include the error of the clamp-on probe. Crest factor: 3 or less. Accuracy specification range: Minimum 1% of range; add ±5 dgt. when measuring at or below 5% of range.
Resistance Measurement
Range | Accuracy | Measurement Current | Open-terminal Voltage |
600.0 Ω | ±0.3 %rdg. ±20 dgt. | 640 μA ±10% | DC2.5 V or less |
6.000 kΩ | ±0.03 %rdg. ±10 dgt. | 96 μA ±10% |
60.00 kΩ | ±0.03 %rdg. ±2 dgt. | 9.3 μA ±10% |
600.00 kΩ | ±0.15 %rdg. ±4 dgt. | 0.96 μA ±10% | |
6.0000 MΩ | ±1.5 %rdg. ±10 dgt. | 96 nA ±10% | |
60.00 MΩ | ±3.0 %rdg. ±20 dgt. | | |
600.0 MΩ | ±8.0 %rdg. ±20 dgt. | | |
Accuracy guarantee condition: After zero adjustment has been performed.
Conductance (nS)
Range | Accuracy | Measurement Current | Open-circuit Voltage |
600.00 nS | ±1.5%rdg. ±10 dgt. | 96 nA ±10% | DC2.5 V or less |
Accuracy is defined for humidity 60% RH or less. Accuracy is defined for the range 20nS or more. In the case of 300 nS or more, ±20 dgt. is added.
Capacitance Measurement
Range | Accuracy | Measurement Current | Open-circuit Voltage |
1.000 nF | ±1 %rdg. ±20 dgt. | 32 μA ±10% | DC2.5 V or less |
10.00 nF |
100.0 nF | ±1 %rdg. ±5 dgt. | 680 μA ±20% | DC3.1 V or less |
1.000 µF |
10.00 µF | ±2 %rdg. ±5 dgt. | Approx. 10 μ/100 μ/200 μA | DC2.1 V or less |
100.0 µF |
1.000 mF |
10.00 mF | ±2 %rdg. ±20 dgt. | Approx. 100 μ/200 μA | |
100.0 mF | | | |
Temperature
Thermocouple Type | Range | Accuracy |
K | -40.0 to 800.0 °C (-40.0 to 1472.0°F) | ±0.5%rdg. ±3 °C (5.4°F) |
The optional K Thermocouple DT4910 is used. Accuracy does not include the error of the K thermocouple.
Frequency (For AC V, DC+AC V, AC µA, AC mA, AC A)
Range | Accuracy |
99.999 Hz | ±0.005 %rdg. +3 dgt. |
999.99 Hz | ±0.005 %rdg. +3 dgt. |
9.9999 kHz | |
99.999 kHz | |
500.00 kHz | |
Measurement range: 0.5Hz or more ([----] is displayed when frequency is less than 0.5Hz). Pulse width: 1µs or more (DUTY ratio is 50%). With the filter ON, accuracy is defined only for frequencies 100Hz or less. (For ACV, DC+ACV).
Peak Measurement (For AC V, DC V, DC+AC V, Clamp, DC µA, DC mA, AC A, AC mA, AC A)
Main measurement | Signal width | Accuracy |
DCV | 4ms or more (single) | ±2.0%rdg. ±40 dgt. |
| 1ms or more (repeated) | ±2.0%rdg. ±100 dgt. |
Other than DCV | 1ms or more (single) | ±2.0%rdg. ±40 dgt. |
| 250µs or more (repeated) | ±2.0%rdg. ±100 dgt. |
Decibel Conversion Measurement : Standard impedance (dBm)
4/8/16/32/50/75/93/110/125/135/150/200/250/300/500/600/800/900/1000/1200 Ω (default : 600 Ω)
General Specifications
Specification | Details |
Durability | Drop proof: YES |
Operating temperature and humidity*1 | -15°C to 55°C. Up to 40°C: 80%RH or less (non-condensating); 40°C to 45°C: 60%RH or less; 45°C to 55°C: 50%RH or less. |
Storage temperature and humidity*2 | -30°C to 60°C. 80%RH or less (non-condensating). |
Applicable standards | Safety: EN61010, EMC: EN61326, Waterproof and dustproof: IP40 |
Safety | Maximum rated voltage between input terminals and ground: CATIII1000 V/ CATIV600 V. Maximum rated voltage between terminals: Between V and COM: 1000 V DC/AC. Maximum rated current between terminals: mA and COM: 600mA DC/AC; A and COM: 10A DC/AC. |
Dimensions/Mass | 93mm(W)×197mm(H)×53mm(D) / 650g (including batteries) |
Accessories | TEST LEAD L9207-10, Instruction Manual, LR6 alkaline batteryx4 |
Standard Models DT4252/DT4253/DT4254/DT4255/DT4256 Specifications
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year.
DC Voltage
Range | Accuracy | Input Impedance |
High precision 600mV range*1 | ±0.2 %rdg. ±5 dgt. | 10.2 MΩ ± 1.5 % |
600.0 mV | ±0.5%rdg. ±5 dgt. |
6.000 V | | 11.2 MΩ ± 2.0 % |
60.00 V | | 10.3 MΩ ± 2.0 % |
600.0 V | | 10.2 MΩ ± 1.5 % |
1000 V | ±0.3 %rdg. ±3 dgt. | |
1500 V*2 | ±0.3 %rdg. ±3 dgt. | |
*1: DT4252 only *2: DT4254 only
AC Voltage
Range | Accuracy | Input Impedance |
| 40 to 500 Hz | 500 or more to 1kHz | |
6.000 V | ±2.5 %rdg. ±3 dgt. | ±1.0 %rdg. ±3 dgt. | 11.2 MΩ ± 2.0%//100 pF or less |
60.00 V | ±1.0 %rdg. ±3 dgt. | ±2.0%rdg. ±3 dgt. | 10.3 MΩ ± 2.0%//100 pF or less |
600.0 V | ±2.0 %rdg. ±3 dgt. | ±4.0 %rdg. ±3 dgt. | 10.2 MΩ ± 1.5%//100 pF or less |
1000 V | | | |
AUTO V (Identification)
Range | Accuracy | Input Impedance |
600.0 V | DC, 40 to 500 Hz ±2.0%rdg. ±3 dgt. | 500 or more to 1kHz ±4.0 %rdg. ±3 dgt. | 900 kΩ ± 20% |
DT4253, DT4254, DT4255, DT4256 only. Crest factor: 3 up to 4000 counts and reduces linearly to 2 at 6000 counts. For ACV, minimum 1% of range; add ±5 dgt. when measuring at or below 5% of range. With the filter ON, the accuracy is not specified in 100Hz/500Hz or more.
DCA Measurement
Range | Accuracy | Input Impedance |
60.00 μA | ±0.8 %rdg. ±5 dgt. | 1 kΩ±5% |
600.0 μA | ±0.8 %rdg. ±5 dgt. |
6.000 mA | ±0.8 %rdg. ±5 dgt. |
60.00 mA | ±0.8 %rdg. ±5 dgt.*1 |
600.0 mA | ±0.9 %rdg. ±5 dgt. | 35 mΩ±30% |
6.000 A | ±0.9 %rdg. ±3 dgt.*2 |
10.00 A | ±0.9 %rdg. ±3 dgt.*2 | 35 mΩ±30% |
*1: DT4256: ±1.8 %rdg. ±15 dgt. Input Impedance: 35 mΩ±30%
*2: DT4252: ±0.9 %rdg. ±5 dgt.
ACA Measurement
Range | Accuracy | Input Impedance |
| 40 to 500 Hz | 500 or more to 1kHz | |
600.0 mA*1 | ±1.4 %rdg. ±5 dgt. | ±1.8 %rdg. ±5 dgt. | 35 mΩ±30% |
6.000 A | ±1.4 %rdg. ±3 dgt. | ±1.8 %rdg. ±3 dgt. | 35 mΩ±30% |
10.00 A | ±1.4 %rdg. ±3 dgt. | ±1.8 %rdg. ±3 dgt. | 35 mΩ±30% |
*1: DT4256 only. Crest factor: 3 up to 4000 counts and reduces linearly to 2 at 6000 counts. Accuracy specification range: Minimum 1% of range; add ±5 dgt. when measuring 300 counts or less.
Continuity Check
Range | Accuracy | Measurement Current | Open-terminal Voltage |
600.0 Ω | ±0.7 %rdg. ±5 dgt. | Approx. 200 μA | DC1.8 V or less |
Continuity ON threshold: Approx. 25 Ω or less (continuous buzzer sound, red LED lights). Continuity OFF threshold: Approx. 245 Ω or more.
Diode Check
Range | Accuracy | Measurement Current | Open-terminal Voltage |
1.500 V | ±0.5 %rdg. ±5 dgt.*1 | Approx. 0.5 mA (DT4222) Approx. 0.2 mA (DT4224) | DC2.5 V or less |
*1: DT4255: ±0.5 %rdg. ±8 dgt. Forward threshold: Buzzer sounds intermittently at 0.15V to 1.5V, the red LED flashes.
Resistance Measurement
Range | Accuracy | Measurement Current | Open-terminal Voltage |
600.0 Ω | ±0.7 %rdg. ±5 dgt. | Approx. 200 μA | DC1.8 V or less |
6.000 kΩ | ±0.7 %rdg. ±3 dgt.*1 | Approx. 100 μA |
60.00 kΩ | ±0.9 %rdg. ±3 dgt.*1 | Approx. 10 μA |
600.00 kΩ | ±1.5 %rdg. ±3 dgt.*1 | Approx. 1 μA |
6.000 MΩ | | Approx. 100 nA | |
60.00 MΩ | | Approx. 10 nA | |
Accuracy guarantee condition: After zero adjustment has been performed. *1: DT4252/4253: ±5dgt.
Capacitance Measurement
Range | Accuracy | Measurement Current | Open-terminal Voltage |
1.000 µF | ±1.9 %rdg. ±5 dgt. | Approx. 10 n/100 n/1 μA | DC1.8 V or less |
10.00 µF |
100.0 µF | ±5.0 %rdg. ±20 dgt. | Approx. 1 μ/10 μ/100 μA | |
1.000 mF |
10.00 mF | | Approx. 10 μ/100 μ/200 μA | |
100.0 mF | | Approx. 100 μ/200 μA | |
Temperature
Thermocouple Type | Range | Accuracy |
K | -40.0 to 400.0 °C | ±0.5%rdg. ±2 °C |
The optional K Thermocouple DT4910 is used. Accuracy does not include the error of the K thermocouple.
General Specifications
Specification | Details |
Durability | Drop proof: YES |
Operating temperature and humidity*1 | -10°C to 50°C (DT4221, DT4222); -10°C to 65°C (DT4223, DT4224). Up to 40°C: 80%RH or less (non-condensating); 40°C to 45°C: 60%RH or less; 45°C to 65°C: 50%RH or less. |
Storage temperature and humidity*2 | -30°C to 60°C (DT4221, DT4222); -30°C to 70°C (DT4223, DT4224). 80%RH or less (non-condensating). |
Applicable standards | Safety: EN61010, EMC: EN61326, Waterproof and dustproof: IP42 |
Safety | Maximum rated voltage between input terminals and ground: CAT III 600V/ CAT IV 300V. Maximum rated voltage between terminals: Between V and COM: 600 V DC/AC. |
Dimensions/Mass | 72mm(W)×149mm(H)×38mm(D) / 190g (including batteries and holster) |
Accessories | TEST LEAD DT4911, Instruction Manual, LR03 Alkaline batteryx1, Holster |
Pocket Models DT4221/DT4222/DT4223/DT4224 Specifications
Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year.
DC Voltage
Range | Accuracy | Input Impedance |
600.0 mV | ±0.5 %rdg. ±5 dgt. | 11.2 MΩ ± 2.0 % |
6.000 V | |
60.00 V | | 10.3 MΩ ± 2.0 % |
600.0 V | | 10.2 MΩ ± 1.5 % |
AC Voltage
Range | Accuracy | Input Impedance |
| 40 to 500Hz | 500 or more to 1kHz | |
6.000 V | ±2.5 %rdg. ±3 dgt. | ±1.0 %rdg. ±3 dgt. | 11.2 MΩ ± 2.0%//100 pF or less |
60.00 V | ±1.0 %rdg. ±3 dgt. | ±2.0%rdg. ±3 dgt. | 10.3 MΩ ± 2.0%//100 pF or less |
600.0 V | ±2.0 %rdg. ±3 dgt. | ±4.0 %rdg. ±3 dgt. | 10.2 MΩ ± 1.5%//100 pF or less |
AUTO V (Identification)
Range | Accuracy | Input Impedance |
600.0 V | DC, 40 to 500 Hz ±2.0%rdg. ±3 dgt. | 500 or more to 1kHz ±4.0 %rdg. ±3 dgt. | 900 kΩ ± 20% |
DT4221, DT4223 only. Crest factor: 3 up to 4000 counts and reduces linearly to 2 at 6000 counts. For ACV, minimum 1% of range; add ±5 dgt. when measuring at or below 5% of range. With the filter ON, the accuracy is not specified in 100Hz/500Hz or more.
Electric Charge
Detection Voltage Range | Detection Target Frequency |
AC80 V to AC600 V | 50 Hz / 60 Hz |
During voltage detection, a continuous buzzer sounds.
Continuity Check
Range | Accuracy | Measurement Current | Open-terminal Voltage |
600.0 Ω | ±1.0 %rdg. ±5 dgt. | Approx. 200 μA | DC1.8 V or less (DT4221/DT4222) DC2.0 V or less (DT4223/DT4224) |
Continuity ON threshold: Approx. 25 Ω or less (continuous buzzer sound). Continuity OFF threshold: Approx. 245 Ω or more.
Diode Check
Range | Accuracy | Measurement Current | Open-terminal Voltage |
1.500 V | ±0.9 %rdg. ±5 dgt. | Approx. 0.5 mA (DT4222) Approx. 0.2 mA (DT4224) | DC2.5 V or less |
Resistance Measurement
Range | Accuracy | Measurement Current | Open-terminal Voltage |
600.0 Ω | | Approx. 200 μA | DC1.8 V or less (DT4222) DC2.0 V or less (DT4223/DT4224) |
6.000 kΩ | ±0.9 %rdg. ±5 dgt. | Approx. 100 μA |
60.00 kΩ | | Approx. 10 μA |
600.00 kΩ | | Approx. 1 μA |
6.000 MΩ | | Approx. 100 nA | |
60.00 MΩ | ±1.5%rdg. ±5 dgt. | Approx. 10 nA | |
Accuracy guarantee condition: After zero adjustment has been performed.
Capacitance Measurement
Range | Accuracy | Measurement Current | Open-terminal Voltage |
1.000 µF | ±1.9 %rdg. ±5 dgt. | Approx. 10 n/100 n/1 μA | DC1.8 V or less (DT4222) DC2.0 V or less (DT4223/DT4224) |
10.00 µF |
100.0 µF | ±5.0 %rdg. ±20 dgt. | Approx. 1 μ/10 μ/100 μA | |
1.000 mF |
10.00 mF | | Approx. 10 μ/100 μ/200 μA | |
100.0 mF | | Approx. 100 μ/200 μA | |
Frequency
Range | Accuracy |
99.99 Hz | ±0.1 %rdg. +2 dgt. |
999.9 Hz | |
9.999 kHz | |