Honeywell IntelliDoX Docking Module Quick Reference Guide

About this Publication

This publication is a quick-start reference guide for assembling the IntelliDoX Docking Module and preparing it for first use. Ensure familiarity with personal gas detection devices and accessories, and take appropriate action in the event of an alarm condition. For additional information regarding IntelliDoX installation, configuration, operation, and maintenance, refer to the IntelliDoX User Manual available at Honeywell Analytics website.

About the IntelliDoX Docking Module

The IntelliDoX Docking Module (referred to as 'IntelliDoX' or 'IntelliDoX module') is an automatic bump test and calibration docking station for Honeywell portable gas detectors. It automatically performs essential procedures including sensor identification, bump tests, calibrations, alarm tests, and data transfers. The module stores a cumulative record of detector data logs transferred to its onboard memory. The IntelliDoX can be used on a tabletop or other flat surface. A built-in retractable stand allows it to be held upright at an angle suitable for routine use. IntelliDoX modules can also be mounted on a wall or other flat surface. Refer to the User Manual for more details.

Intended Use

Unless otherwise specified at the time of purchase, IntelliDoX modules shipped from Honeywell are:

  • Factory configured for use with portable gas detectors.
  • Operate as stand-alone bump test and calibration stations.
  • Can be connected to a network via Ethernet cable for enhanced access to and control of administrative and maintenance tasks.
  • Compatible with PC Configuration software.

If the IntelliDoX module or any of its parts are damaged or missing, contact Honeywell or an authorized distributor immediately.

Normal Operating Conditions

The IntelliDoX is designed to be safe under the following conditions:

  • Indoor use only.
  • Normal atmosphere (20.9% O2) that is free of hazardous gas.
  • Temperature range of +10°C to +35°C (50°F to 95°F).
  • Relative humidity of 0% to 50%.

If the intended operating environment does not match these criteria, Honeywell recommends consulting a qualified professional specialist before installing and using any IntelliDoX modules. This equipment uses potentially harmful gas for calibrations and must be attached to a venting system or used in a well-ventilated area.

What's in the Box

IntelliDoX Docking Module Package

Each IntelliDoX Docking Module package contains one IntelliDoX module with a factory-installed nest. The power adapter, power cord, calibration gas tubing, and exhaust tubing are packaged separately as the IntelliDoX Enabler Kit. One Enabler Kit is required for each stand-alone IntelliDoX module. If the IntelliDoX module is damaged or parts are missing, contact Honeywell or an authorized distributor immediately.

IntelliDoX Enabler Kit

One IntelliDoX Enabler Kit is required for each stand-alone IntelliDoX module. Each Enabler Kit contains:

  • Power supply and AC power cord appropriate to shipping destination.
  • Ethernet cable.
  • Calibration gas and purge gas tubing, cut to 1 meter (3.3 feet).
  • Quick connect fittings.
  • Exhaust tubing, cut to 4.57 meters (15 feet).
  • Inlet (purge) filter assembly.
  • IntelliDoX end plate.
  • Download PC Configuration software that supports your product from the Honeywell Analytics website: www.honeywellanalytics.com/en/products/IntelliDox.

If Enabler Kit parts are damaged or missing, or if additional Enabler Kits are required, contact Honeywell or an authorized distributor immediately.

Diagrams and Components

At a Glance: Module Views

The IntelliDoX Docking Module is depicted with various views and numbered components:

Top View:

  • 1. Lid: The main cover of the module.
  • 2. Lid button: Button to operate the lid.
  • 3. View screen (LCD): Displays information and menus.
  • 4. Navigation buttons: Buttons used to move through menus and options.
  • 5. Gas module: Port for gas connections.
  • 6. Arm: Part of the module's structure.

Left View:

  • 7. Arm unlock: Mechanism to unlock the arm.
  • 8. Hinge: For opening or moving parts.
  • 9. Panel latch: Secures a panel.
  • 10. Detachable back panel: Removable rear panel.
  • 11. Expansion module: Port for expansion modules.
  • 12. Network connection: Ethernet port.
  • 13. Power connection: Port for the power supply.
  • 14. Exhaust outlet: Port for exhaust tubing.
  • 15. Purge inlet: Port for purge gas or ambient air inlet.
  • 16. Gas inlet 1: Connection for calibration gas cylinder 1.
  • 17. Gas inlet 2: Connection for calibration gas cylinder 2.
  • 18. Gas inlet 3: Connection for calibration gas cylinder 3.
  • 19. Gas inlet 4: Connection for calibration gas cylinder 4.
  • 20. End plate: Final component for sealing or connection.

Bottom View:

Shows components 8 (Hinge), 9 (Panel latch), and 10 (Detachable back panel).

Right View:

Shows components 11 (Expansion module) through 20 (End plate).

Touchpad and Buttons

The following table lists the buttons and their corresponding actions and part numbers:

ButtonActionPart Num
Move rightLid1
Move leftLid button2
Move upView screen3
Move downNavigation buttons4
Select menu itemGas module5
Save changesArm6
Close menu itemArm unlock7
Return to previous screenHinge8
Cancel changesPanel latch9
Detachable back panel10
Expansion module11
Network connection12
Power connection13
Exhaust outlet14
Purge inlet15
Gas inlet 116
Gas inlet 217
Gas inlet 318
Gas inlet 419
End plate20

LCD Backlight Status

The LCD backlight indicates the status of the module:

BacklightColorStatus
BacklightGrayIdle
BluePrompt for user action
YellowActivity in progress
GreenActivity successful
RedActivity Failed

Setup and Connections

Connect the Purge Inlet Filter Assembly

Each IntelliDoX Enabler Kit includes one purge inlet filter assembly. The purge inlet is typically configured to use ambient air in a fresh air environment with a normal atmosphere of 20.9% O2, free of hazardous gas. Ensure the filter assembly is attached before using the IntelliDoX module. An extension tubing can be attached to the filter assembly to draw ambient air from an adjacent fresh air environment.

  1. Ensure the filter assembly is free of obstructions and defects.
  2. Connect the filter assembly to the purge inlet (Part 15).
  3. If necessary, attach an extension tubing to the filter assembly to draw ambient air from an adjacent fresh air environment.

Connect the Exhaust Tubing

Each IntelliDoX Enabler Kit includes one exhaust tubing, 4.57 meters (15 feet) long.

  1. Inspect the exhaust tubing to ensure it is free of obstructions and defects.
  2. Connect the exhaust tubing to the exhaust outlet (Part 14).
  3. Ensure the exhaust tubing is not connected to a negative pressure system or obstructed in any way.

Connect the Power

Each IntelliDoX Enabler Kit contains one power supply and AC power cord. Use only the provided power supply to connect the IntelliDoX Docking Module to an appropriate electrical power outlet. When power is connected, the IntelliDoX activates and performs a self-test.

  1. Connect the AC power cord to the power supply.
  2. Connect the power supply to the IntelliDoX power port (Part 13).
  3. Plug the AC power cord into a suitable wall outlet.
  4. When power is connected, the IntelliDoX LCD activates and a self-test is performed.

NOTICE: To prevent corruption or loss of data, software, or firmware, do not deactivate the equipment while performing datalog transfers, bump tests, calibrations, or other operations.

Charging a Detector

Use the IntelliDoX to charge detectors fitted with rechargeable batteries. For battery maintenance information, refer to the detector's manual.

  1. Charge only in a normal environment (20.9% O2, free of hazardous gas). Do not operate the docking module in a hazardous area, as this could result in personal injury and/or property damage.
  2. Deactivate the detector.
  3. Insert the detector into the IntelliDoX module.
  4. Battery charging begins immediately. Battery charging is disabled during bump test and calibration procedures.

Once tests and other routines are completed, the detector can remain in the module for charging. If the detector is activated, the module will deactivate it after 10 minutes of inactivity. When charging is complete, remove the detector. Do not store the detector in the module.

Prepare for Bump Tests and Calibrations

The IntelliDoX module is factory-configured for portable gas detectors. Gas inlets are configured at the factory and cannot be altered. The IntelliDoX Enabler Kit includes quick connect fittings and calibration tubing cut to a minimum recommended length of 1 meter (39 inches). Use only tubing between 1 meter (39 inches) and 10 meters (33 feet) in length when connecting gas cylinders to an IntelliDoX module.

Connect Calibration Gas

  1. Connect a demand flow regulator to the calibration gas cylinder.
  2. Use the quick connect fittings and calibration gas tubing to connect the calibration gas cylinder to the gas inlet on the IntelliDoX module.

Bump Test Procedure

What is a Bump Test?

A bump test is a procedure that confirms a detector's ability to respond to target gases by exposing the detector to gas concentrations that exceed its alarm setpoints.

Automatic Bump Test

If 'AutoBump on Insertion' is set using PC configuration software, the bump test starts automatically when IntelliDoX recognizes the detector. If 'AutoCal on Overdue Sensors' is enabled and a calibration is also due, no bump test is performed; calibration starts automatically instead.

Performing a Bump Test

  1. Insert a detector into the IntelliDoX module.
  2. Use the navigation buttons (Part 4) to move to 'Bump test my detector' on the IntelliDoX user menu.
  3. Press the select button to choose 'Bump Test'. The LCD screen background changes to yellow, and the bump test progress screen is displayed.
  4. The bump test begins, showing progress screens during the test.
  5. If 'AutoDownload Datalog' is set using PC configuration software, detector datalogs are automatically transferred to the IntelliDoX.

Bump Test Results

Bump Test Pass

When the bump test is successful, the LCD screen background changes to green, displaying 'Bump test passed'. Checkmarks (✔️) indicate successful responses for Alarm and Sensor. Press the return button to return to the IntelliDoX user menu.

Bump Test Fail

When the bump test fails, the LCD screen background changes to red, displaying 'Bump test failed'. Crosses (❌) indicate failed alarm response and/or sensor response test items. If 'AutoCal on Failed Bump' is set using PC configuration software, calibration automatically begins.

Calibration Procedure

What is Calibration?

Calibration is a two-step procedure that determines the measurement scale for the detector's response to gas. The first step involves taking a baseline reading in a clean, uncontaminated environment. The second step exposes the sensors to known concentrations of gas. The detector uses the baseline and known gas concentrations to determine the measurement scale.

Calibration Guidelines

  1. Calibrate only in a normal environment (20.9% O2, free of hazardous gas). Do not operate the docking module in a hazardous area. Failure to adhere to this guideline can result in possible personal injury and/or property damage.
  2. Use only premium grade calibration gases and cylinders approved by Honeywell and supplied by Honeywell or an authorized distributor. Calibration gases must meet the detector's accuracy requirements. Refer to the User Manual for more information.
  3. Do not use a gas cylinder beyond its expiration date.
  4. All calibration cylinders must be used with a demand flow regulator and meet these maximum inlet pressure specifications: Disposable cylinders: 0-300 psig / 0-20.7 bar; Refillable cylinders: 0-3000 psig / 0-207 bar.
  5. Do not calibrate the detector during or immediately after charging.
  6. Calibrate the sensor if ambient gas readings vary during startup.
  7. Calibrate a new sensor before use. Allow the sensor to stabilize before starting calibration: wait 60 seconds for a used sensor, and 5 minutes for a new sensor.
  8. When calibrating the same gas detector multiple times, wait 10 minutes between calibrations to allow the sensor to stabilize.
  9. If a certified calibration is required, contact Honeywell or an authorized distributor.

Performing a Calibration

If 'AutoCal on Overdue Sensors' is set using PC configuration software and sensors are overdue, calibration starts automatically once IntelliDoX recognizes the detector.

  1. Insert a detector into the IntelliDoX module.
  2. Use the navigation buttons (Part 4) to move to 'Calibrate my detector' on the IntelliDoX user menu.
  3. Press the select button to choose 'Calibrate my detector'. The LCD screen background changes to yellow, and the calibration progress screen is displayed.
  4. Calibration begins, showing progress screens during the test.
  5. If 'AutoDownload Datalog' is set using PC configuration software, detector datalogs are automatically transferred to the IntelliDoX.

Calibration Results

Calibration Pass

When calibration is successful, the LCD screen background changes to green, displaying 'Calibration passed'. Checkmarks (✔️) indicate successful responses for Alarm and Sensor. Press the return button to return to the IntelliDoX user menu.

Calibration Fail

When calibration fails, the LCD screen background changes to red, displaying 'Calibration failed'. Crosses (❌) indicate failed alarm response and/or sensor response test items.

Symbol Definitions

This manual uses the following signal words, as defined by ANSI Z535.4-1998:

Signal WordDescription
DANGERHazardous situation which, if not avoided, will result in death or serious injury. This symbol identifies the most extreme hazardous situations.
WARNINGHazardous situation which, if not avoided, could result in death or serious injury.
CAUTIONHazardous situation which, if not avoided, may result in moderate or minor injury.
NOTICESituations which, if not avoided, may result in property damage.

Important Safety Information

Read this section carefully before using the IntelliDoX:

  • To ensure personal safety, read Safety Information and Warnings before using the IntelliDoX.
  • Use the IntelliDoX only as specified by the manufacturer. Failure to do so may impair protection provided by the IntelliDoX.
  • The safety and security of any system or network incorporating the IntelliDoX and its accessory components is the responsibility of the assembler of the system.
  • Follow all required National Electric Codes (NEC) and safety standards.

Configuration Software

Please download PC configuration software on the IntelliDoX product page under the software tab on the Honeywell Analytics website: www.honeywellanalytics.com.

Available Software:

ProductConfiguration Software
IntelliDoX BWClipFleet Manager/SafetySuite DC
IntelliDoX BW MicroClipFleet Manager/SafetySuite DC
IntelliDoX BW Clip4Fleet Manager/SafetySuite DC
IntelliDoX BW UltraFleet Manager/SafetySuite DC
IntelliDoX BW SoloSafetySuite DC

Copyright, Notices, Trademarks

While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in its written agreement with and for its customers. In no event is Honeywell liable to anyone for any indirect, special, or consequential damages. The information and specifications in this document are subject to change without notice. IntelliDoX is a trademark of Honeywell. Other brand or product names are trademarks of their respective owners.

Product Disposal

This symbol indicates that the product must not be disposed of as general industrial or domestic waste. This product should be disposed of through suitable WEEE disposal facilities. For more information about disposal of this product, contact your local authority, distributor, or the manufacturer.

Contact Information

Honeywell Analytics
Suite 110
4411-6 Street SE
Calgary, Alberta Canada, T2G 4E8
Toll-free: 1-888-749-8878
www.honeywellanalytics.com

Document Number: 50104991-166 EN-B4 ©2018 Honeywell. All rights reserved.

Models: IntelliDoX Docking Module for BWC4

File Info : application/pdf, 2 Pages, 996.06KB

PDF preview unavailable. Download the PDF instead.

qrgenfmsu b2 intellidox quick reference guide

References

Adobe PDF Library 15.0 Adobe InDesign CC 13.1 (Windows)

Related Documents

Preview Honeywell BW Ultra Gas Detector: Quick Reference Guide & Safety
Get essential information on the Honeywell BW Ultra portable gas detector. This quick reference guide covers setup, safety precautions, operation, and alarms for reliable gas detection.
Preview Honeywell BW™ Ultra Portable Five-gas Detector User Manual
Comprehensive user manual for the Honeywell BW™ Ultra portable five-gas detector with internal pump, covering safety, operation, maintenance, and troubleshooting.
Preview Honeywell BWTM Ultra Portable Gas Detector User Manual - Operation, Maintenance, Safety
Detailed user manual for the Honeywell BWTM Ultra portable multi-gas detector. Covers setup, calibration, bump testing, maintenance, sensor replacement, troubleshooting, and safety guidelines for industrial gas monitoring.
Preview Honeywell BW Solo Portable Single-Gas Detector User Manual
Comprehensive user manual for the Honeywell BW Solo portable single-gas detector, covering operation, maintenance, safety, specifications, and troubleshooting.
Preview ToxiRAE Pro LEL QuickStart Guide | Honeywell Gas Detection
Concise guide for setting up, operating, and calibrating the ToxiRAE Pro LEL wireless gas detector from Honeywell. Learn about its user interface, charging, powering on/off, and essential testing procedures.
Preview Honeywell BW RigRat Quick Start Guide: Operation, Charging, and Calibration
A concise guide to operating, charging, performing bump tests, and calibrating the Honeywell BW RigRat portable gas detector. Includes basic navigation and app information.
Preview Honeywell BW™ MaxXT II Gas Detector Quick Reference Guide
Quick reference guide for the Honeywell BW™ MaxXT II multi-gas detector, covering operation, safety information, calibration, alarms, and specifications. Includes essential details for safe and effective use.
Preview Honeywell BW™ Flex Gas Detector Quick Reference Guide
A quick reference guide for the Honeywell BW™ Flex portable multiple gas detector, covering basic operations, bump testing, calibration, and safety precautions.