Honeywell BW™ Ultra Portable Five-gas Detector User Manual

Gas Detection

Limited Warranty and Limitation of Liability

BW Technologies by Honeywell LP (Honeywell) warrants the product to be free from defects in material and workmanship under normal use and service for a period of three years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer. Honeywell's warranty obligation is limited, at Honeywell's option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a Honeywell authorized service center within the warranty period. In no event shall Honeywell's liability here under exceed the purchase price actually paid by the buyer for the product.

This warranty does not include:

The obligations set forth in this warranty are conditional on:

THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. HONEYWELL SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, OR RELIANCE OR ANY OTHER THEORY.

Since some countries and states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.

Warranty Registration: www.honeywellanalytics.com/support/product-registration

Before You Begin

About this Publication

While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in its written agreement with and for its customers. In no event is Honeywell liable to anyone for any indirect, special, or consequential damages. The information and specifications in this document are subject to change without notice.

Trademarks

Brand or product names are trademarks of their respective owners. The following brand or product names are trademarks of Honeywell:

Introduction

The Honeywell BW™ Ultra gas detector warns of hazardous gas at levels above user-defined alarm setpoints. The detector is a personal safety device. It is your responsibility to respond properly to the alarm. This publication is intended for people who understand how to configure, maintain and use personal gas detectors, docking systems, and accessories.

Monitored Gases

The detector can monitor up to five gases at a time. Four gases detected by default, and one optional gas selected from the following list.

Monitored Gas Unit of Measure Default detected gases
Hydrogen Sulfide (H₂S) parts per million (ppm)
Carbon Monoxide (CO) parts per million (ppm)
Oxygen (O₂) % volume
Combustible gases (LEL) Lower explosive limit a) percent of lower explosive limit (%LEL) b) percent by volume methane 0-5.0% v/v
Monitored Gas Unit of Measure Optional gases
IR Flammable (IR LEL) Hydrogen (H₂) % volume parts per million (ppm)
Sulfur Dioxide (SO₂) parts per million (ppm)
IR-Carbon Dioxide (CO₂) parts per million (ppm)
Ammonia (NH₃) parts per million (ppm)
Volatile Organic Compounds (VOC) parts per million (ppm)
Chlorine (Cl₂) parts per million (ppm)
Nitrogen Dioxide (NO₂) parts per million (ppm)
Hydrogen Cyanide (HCN) parts per million (ppm)
Nitric Oxide (NO) parts per million (ppm)
CO sensor with a Hydrogen filter (CO-H) parts per million (ppm)

What's in the Box

Safety Information

CAUTION READ THIS FIRST

Use the detector only as specified in this manual, otherwise the protection provided by the detector may be impaired.

WARNINGS

CAUTION

French Warnings (Translated):

Getting Started

Appearance

The detector's physical components are labeled:

Main Screen

The start-up main screen displays gas readings, typically showing Oxygen (O₂), LEL (Lower Explosive Limit), CO (Carbon Monoxide), H₂S (Hydrogen Sulfide), and SO₂ (Sulfur Dioxide) concentrations.

Screen Icons

The detector's screen displays various icons to indicate its status:

Alarms

When the detector enters an alarm state, it provides audible, visual, and vibrator alerts. The type and intensity of these alerts vary depending on the alarm condition. In Stealth mode, only vibration is active.

Alarm Type Description Screen Example
Low Alarm Slow siren (upward tone), Slow flash, Black box around gas flashes, Vibrator alarm activates [Example screen showing low alarm]
High Alarm Fast siren (downward tone), Fast flash, Black box around gas flashes, Vibrator alarm flashes [Example screen showing high alarm]
Time Weighted Average (TWA) Alarm Fast siren (downward tone), Fast flash, Black box around gas flashes, Vibrator alarm activates [Example screen showing TWA alarm]
Short Term Exposure Limit (STEL) Alarm Fast siren (downward tone), Fast flash, Black box around gas flashes, Vibrator alarm activates [Example screen showing STEL alarm]
Multi Alarm Alternating low and high alarm, Black box around gas flashes, Type of alarm alternates, Vibrator alarm alternates [Example screen showing multi-alarm]
Sensor Failure Alarm 'X' is displayed [Example screen showing sensor failure]
Over Limit (OL) Alarm Fast siren (downward tone), Fast flash, Black box around gas flashes, Vibrator alarm activates [Example screen showing over limit alarm]
Normal Deactivation Sequence of alternating beeps and alternating flashes, Vibrator alarm activates [Example screen showing normal deactivation]
Low Battery Alarm Countdown initiates, "OFF" is displayed, Sequence of 10 rapid sirens and alternating flashes followed by 7 seconds of silence (continues for 15 minutes), flashes, Vibrator alarm pulses [Example screen showing low battery alarm]
Critical Battery Alarm After 15 minutes of the low battery alarm sequence, the detector will enter critical alarm, Fifteen minutes after low battery alarm activates, sequence of 10 rapid sirens and alternating flashes with 1 second of silence in between (sequence reactivates seven times), Vibrator alarm pulses, "Low Battery Powering Off" is displayed and the detector deactivates [Example screen showing critical battery alarm]
Pump Alarm Detector is in pump alarm when gas is turned off during calibration. [Example screen showing pump alarm]

Detector Operations

Activate the detector

Turn on the detector in a safe area with an atmosphere of 20.9% oxygen and free of hazardous gas.

  1. For first time use, charge the battery for up to 8 hours or until LED light turns green using the charging adapter provided. Refer to Charge the battery for more information.
  2. Press and hold the button for three seconds.
  3. For first time use, a Warming sensors message is displayed and a 30 minutes countdown is displayed. In most cases, this countdown only lasts a couple of minutes.
  4. When the detector displays Pump test Block inlet, block the pump inlet with a finger, and then after a couple of seconds unblock the pump inlet. The detector performs a quick pump test. A Pump Test passed message is displayed. If you do not block the pump inlet, the detector will turn off after two minutes. The detector will then perform a self-test, including testing the sensors. This process could take several minutes. If necessary, you will be instructed by screen prompts to calibrate newly-installed sensors.
  5. When the self-test is complete, press and hold the button to zero sensors. After the autozero is complete, the detector then checks the sensors for calibration and bump test. If the detector identifies sensors requiring calibration or bump testing, press the button and follow the on-screen instructions.

Activate the backlight

To activate the backlight on the screen, press the button.

Deactivate the detector

  1. Press and hold the button during the powering-off countdown.
  2. Release the button when OFF is displayed.

Navigate the menu

There are four main menu items:

  1. Double pressing the button displays all four options on the screen. See Information is selected and highlighted by default.
  2. Press the button to transition the selection to the next choice.
  3. Press and hold for three seconds to enter the selected option.
  4. Follow on-screen instructions for your selected operation. Most of the detector's procedures are described in this guide.

View detector's general information

  1. Double-press the button to enter the main menu.
  2. Select See information and press the button to scroll through the following information:
    • Peak readings
    • STEL readings
    • TWA readings
    • Bump test intervals
    • Calibration
    • BLE information (BLE should be enabled)
    • LEL Correction Factor
    • Low Alam setpoint
    • Hi Alam setpoints
    • STEL setpoints
    • TWA setpoints

Reset TWA or STEL readings

Before you begin: You need to enable TWA/STEL Reset in Fleet Manager II to reset readings in the detector.

  1. Go to the main menu and select > See Information > TWA readings or STEL readings.
  2. Press and hold for 3 seconds to reset readings. A reset message is displayed.

Reset Peak readings

Before you begin: You need to enable Peak Reset in Fleet Manager II to reset readings in the detector.

  1. Go to the main menu and select > See Information > Peak readings.
  2. Select Hold to reset peak readings. Press and hold for 3 seconds to reset readings.

Reset TWA/STEL and Peak readings

Before you begin: You need to enable TWA/STEL Reset and Peak Reset in Fleet Manager II to reset readings in the detector.

  1. Go to the main menu and select > See Information > Peak readings.
  2. Select Hold to reset all readings. Press and hold for 3 seconds to reset readings.

Zero sensors

Before you begin: Connect the Nitrogen if this is a CO₂ unit.

  1. Go to the main menu and select Zero Sensors.
  2. Press and hold for 3 seconds. A Zeroing process starts automatically. The screen displays all current gas measurements, highlighting entries above zero. Ambient air is applied to zero all sensors that are not CO₂. The screen displays all current gas measurements, highlighting entries being reset to zero.
  3. If you are not zeroing O₂, Click NO in the prompted message: Is this a CO₂ unit?. Zero results are displayed. Press the button or wait for six seconds to end the Zeroing process.
  4. Click Yes if this is a CO₂ unit and you want to apply Nitrogen to zero CO₂. A two minutes gas measure process starts automatically. All current gas measurements and entries reset to zero are displayed.
  5. Turn off the gas following on-screen instructions. Zero results are displayed.
  6. Press the button or wait for six seconds to end the Zeroing process. Zero results are displayed as follows:
    • A check mark [✓] for sensors that passed Zero
    • A cross mark [✗] for sensors that failed Zero
    • An exclamation mark [!] for sensors that skipped Zero

Acknowledge alarms and messages

Press and release the button to perform any of the following:

Latching Alarms

If enabled, during an alarm condition the Latching Alarms option causes the low and high gas alarms (audible, visual, and vibrator) to persist until the alarm is acknowledged and the gas concentration is below the low alarm setpoint. The LCD displays the peak concentration until the alarm no longer exists. Local regulations in your region may require the Latching Alarms option be enabled. The detector is shipped with the Latching Alarms option disabled.

Calibrate the detector

Perform a calibration to adjust the sensitivity levels of sensors and ensure accurate responses to gases. The detector can be calibrated in two ways:

Before you begin: Move to a normal atmosphere (20.9% v/v O₂) that is free of hazardous gas.

  1. Go to the main menu and select > Start Calibration.
  2. Press and hold the button for three seconds to display the Powering Off countdown and continue to hold for the Starting Calibration countdown. The detector will enter the zero function. The Zeroing process starts automatically and lasts for five minutes. Current gas measurements are displayed, and entries above zero are highlighted.
  3. Connect the calibration hose to the pump inlet. Make sure to use a demand flow regulator.
  4. Confirm that you want to apply Nitrogen to zero CO₂.
  5. When Apply calibration gas now is displayed, apply the gas and wait for a maximum of five minutes. The detector first tests for a specific type of gas. When it detects enough of that gas for sensor calibration, a check box will be displayed next to that gas. Calibration then begins. Gas values will adjust on the screen during the calibration.
  6. When Turn gas off is displayed, disconnect the device from gas. Check marks will be displayed next to the calibrated sensors. These sensors reset to the number of days until the next calibration is due (for example, 180 days). The calibration cycle will take about two minutes after which the user will be prompted to Press button to continue.
  7. If the calibration was successful, Calibration Passed will be displayed. Press the button to exit calibration. If the calibration failed for some or all of the gases, either a Cal Error All gases applied mixed results message (if the detector was not successfully calibrated for all gases) or a Fail all gases message will be displayed. After the button is pressed, a Cal overdue message will be displayed.

Start a Bump Test

Perform a bump test regularly to test sensors and alarms. To bump test, expose the sensors to a gas concentration that exceeds alarm setpoints and confirm that the sensors and alarms work correctly. The detector can be bump tested in two ways:

Perform a manual bump test

Before you begin: Connect the calibration hose to a demand flow regulator on the gas cylinder.

  1. Double-press the button and select > Start Bump test.
  2. Press and Hold the button for three seconds. The detector displays Starting Bump test. Bump test started is displayed, and then the detector makes noise, flash, and vibrate.
  3. The detector will prompt you Did you see and hear the alarms?, select Pass, and Press and hold for three seconds to confirm that the visual, audible, and vibrator alarms work correctly. An Audio-Visual test passed message is displayed. Skip to Step 5.
  4. If the visual, audible, and vibrator alarms failed, select Fail, and press and hold the button. An Audio-Visual test failed message is displayed. Then you can:
    • a) Apply gas, Skip to Step 5
    • b) Press the button to skip gas application and follow on-screen instructions to end the Bump Test.
    Bump test results are displayed, and the test ends.
  5. If you want to apply gas, follow on-screen instructions. Wait for about 30 seconds; gas measurements are displayed for each pertinent gas sensor. A Bump Test pass confirmation is displayed.
  6. After the Turn gas off message is displayed, remove the hose from the pump inlet. The detector will remain in alarm until the gas clears from the sensors. Bump test results are displayed showing check marks next to the tested sensors. These sensors reset to the number of days until the next Bump Test is due.
  7. Press the button to finish the procedure.

CAUTION Honeywell recommends to bump test the sensors before each day's use to confirm their ability to respond to gas by exposing the sensors to a gas concentration that exceeds the alarm setpoints.

Set up IntelliFlash

The IntelliFlash® feature causes the detector, if it is in compliance (for example, bump tested and calibrated), to flash a green light every second (the factory setting default) from the top visual alarm indicator. In Fleet Manager II, use the IntelliFlash Interval option to change how often the detector flashes.

Set up Reverse IntelliFlash

IntelliFlash flashes a green light when the detector is in compliance, but Reverse IntelliFlash® flashes an amber light when the detector is not in compliance (a bump test or calibration is overdue, or a sensor is not working and has been overridden). Use Fleet Manager II to change how often the detector flashes for Reverse IntelliFlash. IntelliFlash and Reverse IntelliFlash can be configured in one of four scenarios:

Set up the Confidence and Compliance Beep interval

The Confidence Compliance Beep is a sound that tells the user the detector is in compliance (for example, bump tested and calibrated). In Fleet Manager II use the Confidence/Compliance Beep option to change how often the detector beeps for Confidence Compliance Beep.

Configure the Detector

You can configure the Honeywell BW™ Ultra detector's device and sensor using Fleet Manager II.

Necessary to configure detector settings:

For device settings, a startup message can be added; Confidence Compliance Beep can be activated, bump tests can be forced, stealth mode can be enabled, etc. For sensor settings, the calibration gas type and frequency can be changed, the bump test interval and alarm setpoints can be set, STEL and TWA can be selected, etc. You can perform some other calibrations with Fleet Manager II.

Note: When the operator configures the Honeywell BW™ Ultra using FleetManager II, Honeywell strongly recommends reviewing the detector's settings before the operation to ensure that settings were applied successfully and comply with performance requirements.

Custom configuration created in Fleet Manager II can be used to configure detector settings. Example: Five detectors must have the same bump test reminders and alarm setpoints. Each detector could be configured separately or Fleet Manager II can be utilized to create a custom settings configuration. This configuration can then be loaded on each detector. This saves time and allows settings to be managed from one location.

Connect With IntelliDoX

If a detector's calibration is overdue and the forced calibration feature is enabled, the calibration can be performed with an IntelliDoX docking station or through the calibration option from the detector's main menu.

Connect With an IR Link

The detector can be paired with an IR link, aka dongle. There is an IR connection at the bottom of the detector that allows FleetManager II configurations to be efficiently transferred to multiple detectors. The IR link will also allow new firmware to be transferred to detectors or data/event logs to be transferred to FleetManager II.

Note: You should have the IR Connectivity Kit (sold separately) to transfer the data from a computer to the detector.

Select an operation mode

The detector can be used in one of three modes: Essentials mode, Hole Watch mode, and Inert mode.

Note: Honeywell BW™ Ultra always monitors gas levels, regardless of the operating mode. If the detector detects a sudden gas exposure, it will flash, vibrate, and produce a loud siren noise. A gas level alarm takes precedence over all of the detector's other functions.

Bluetooth Pairing

The user can pair the Honeywell BW™ Ultra to a mobile device via built in Bluetooth Low Energy (BLE). The Honeywell Safety Communicator app, installed on the mobile phone, can then show gas readings and alarms from the Honeywell BW™ Ultra unit that is connected. Readings and alarms can then be sent to Honeywell's remote monitoring software.

  1. On the mobile device, turn on the Bluetooth connection and look for available detectors. On the Honeywell BW™ Ultra, the Bluetooth connection is on by default.
  2. On the mobile device, select the detector and then enter the last six digits from the detector's serial number.

Note: Pairing is not allowed at start up, during calibration, or bump test.

Warning: Wireless communication and infrastructure should be used as informational monitoring only.

Replace a Sensor

Use only sensors designed by Honeywell for the Honeywell BW™ Ultra detectors. Replace the sensors in a non-hazardous location. The Honeywell BW™ Ultra detector can be configured for a maximum of 5 gases and may contain dummy sensors.

Review Logs

Many of the detector events are logged and can be reviewed via IntelliDoX or BLE. Typical logged events would be:

Maintenance

Maintenance

Perform the following tasks to maintain the detector in good operating condition:

Gas Cylinder Guidelines

Charge the Battery

You can charge the battery using the provided charging adapter, that is certified for SELV/LVLC (isolated) with an output Um of 6.3 V. It can take up to 8 hours to get the battery fully charged in a temperature range from 5 °C to 35 °C.

Note: If you charge with the power on, charging may not be completed within 8 hours.

Maintain the Battery

Lithium-ion batteries do not respond well to cycles of full discharge followed by a full charging cycle. Recharge the battery before it is exhausted. Do not charge the battery at low or elevated temperatures. 30°C (86°F) is considered an elevated temperature and should be avoided whenever possible. A rechargeable battery's runtime decreases approximately 20% over a two-year period of typical use.

Real Time Clock Display

The real time clock is displayed in the upper left corner of the detector's screen. It is configurable through FleetManager II in 12- or 24-hour formats. The date display can also be configured in several formats through FleetManager II. The time/date information is retained even when the detector's battery is being changed.

Languages

Honeywell BW™ Ultra supports eight languages: English, French, German, Portuguese, Spanish, Simplified Chinese, Russian, Italian, Dutch, Slovak, Czech, Polish, Norwegian, Danish, Swedish, Finnish, Turkish, and Arabic. These are configurable through FleetManager II. Custom startup text can be entered in all languages except Simplified Chinese through FleetManager II.

Clean the Detector

Clean the exterior of the detector with a soft, damp cloth. Use only water-based (non-alcohol) cleaners. Do not use soaps, solvents, or polishes.

Upgrade the Firmware

Upgrade the Firmware via IR Link using the Fleet Manager II Software.

Before you begin:

  1. Turn on the detector.
  2. Start Fleet Manager II application.
    1. Expand Administration from the left pane.
    2. Click Login/Logout.
    3. Type in the default password: Admin.
    4. Click OK to continue.
  3. In the left pane select Devices > Configure device via IR link. In the Device Selection window:
    1. Select Honeywell BW™ Ultra.
    2. Click OK.
  4. In the Honeywell BW™ Ultra configuration window, click Bootloader to select the binary file. In the Honeywell BW™ Ultra Bootloader window, click Choose File.
  5. In the Choose Firmware File to Upload window, select the downloaded file, and then click Open.
  6. Connect the Honeywell BW™ Ultra detector to the computer using the IR Link connector.
  7. Click Send to initiate the file transfer to the gas detector. After the file transfer is complete, the Bootload process will start. During Bootload, the display will go blank and the detector will beep several times.
  8. The Programming Succeeded message is displayed. Press the button to finish the procedure, and then disconnect the detector from the computer.

Service

Replace the Printed Circuit Board (PCB)

  1. Turn the instrument off.
  2. Remove the front shell:
    1. Turn the instrument face down and unscrew on the battery pack.
    2. Remove the battery pack and the four screws in the battery compartment.
    3. Remove the two screws at the top of the instrument and pull the back shell apart from the front shell.
    4. Remove the two screws holding the PCBA in place.
    5. Remove the PCB assembly from the front shell.
  3. Remove the LCD:
    1. Disengage the two snaps on the upper sides of the LCD carrier.
    2. Swing LCD forward watching for the ZIF connector.
    3. Lift the latch on the ZIF connector, and then pull the LCD cable forward and remove LCD assembly.
  4. Disengage the two snaps holding the manifold carrier in place and carefully pull up the manifold carrier off the PCB.
  5. Lift the latch on the 4R+ ZIF connector and pull out the 4R+ cable assembly to disconnect the pump connector from the PCB.
  6. Retrieve the new PCB.
  7. Connect the pump to the PCB.
  8. Align the snaps on the manifold carrier and push down to engage the snaps to the PCB.
  9. Insert the 4R+ ribbon cable into the ZIF connector and press down on the 4R+ ZIF latch to engage the cable.
  10. Re-connect the 4R+ sensor cable.
  11. Attach the LCD:
    1. Insert the LCD ribbon cable into the ZIF connector and press down on the ZIF connector latch to engage the cable.
    2. Lever the LCD assembly back toward the PCB, then press down until the side snaps are engaged.
  12. Attach the shells:
    1. On the front shell, ensure the LCD gasket is aligned with the pins. The frame should be facing up.
    2. Place the PCB into the front shell and insert the two screws torque 0.34-0.45 N-m (3-4 in-lbs) to fasten both pieces.
    3. Replace the back shell and install the four screws back into the battery compartment torque 0.45-0.56 N-m (4-5 in-lbs).
    4. Reinstall the two screws at the top of the instrument torque 0.45-0.56 N-m (4-5 in-lbs).
  13. Place the battery pack:
    1. Engage hooks on the bottom of the battery pack.
    2. Push the battery pack into place.
    3. Tighten the battery retention screw torque 0.45-0.56 N-m (4-5 in-lbs).
  14. Turn on the instrument and allow for the sensors to stabilize. Calibrate sensors.

Replace the LCD

  1. Turn the instrument off.
  2. Remove the battery:
    1. Turn the instrument face down and unscrew on the battery pack.
    2. Remove the battery pack and the four screws in the battery compartment.
  3. Remove the front shell:
    1. Remove the two screws at the top of the instrument and pull the back shell apart from the front shell.
    2. Remove the two screws holding the PCB in place.
    3. Remove the PCB assembly from the front shell.
  4. Remove the LCD:
    1. Disengage the two snaps on the upper sides of the LCD carrier.
    2. Swing LCD forward watching for the ZIF connector.
    3. Lift the latch on the ZIF connector, pull the LCD cable forward and remove LCD assembly to remove the LCD assembly.
    4. Pull the LCD cable forward and remove the LCD assembly.
  5. Retrieve the new LCD.
  6. Attach the LCD:
    1. Insert the LCD ribbon cable into the ZIF connector and press down on the ZIF connector latch to engage the cable.
    2. Lever the LCD assembly back towards the PCB, then press down until the side snaps are engaged.
  7. Attach the front shell:
    1. On the front shell, ensure the LCD gasket is aligned with the pins. The frame should be facing up.
    2. Place the PCB into the front shell and insert the two screws torque 0.34-0.45 N-m (3-4 in-lbs) to fasten both pieces.
    3. Replace the back shell and install the four screws back into the battery compartment torque 0.45-0.56 N-m (4-5 in-lbs).
    4. Reinstall the 2 screws at the top of the instrument torque 0.45-0.56 N-m (4-5 in-lbs).
  8. Attach the battery:
    1. To place the battery pack correctly, first engage hooks on the bottom of the battery pack.
    2. Push the battery pack into place and tighten the battery retention screw torque 0.45-0.56 N-m (4-5 in-lbs).
  9. Turn on the instrument and allow for the sensors to stabilize.

Replace the 1-Series Sensors

  1. Turn the instrument off.
  2. Remove the battery:
    1. Turn the instrument face down and unscrew on the battery pack.
    2. Remove the battery pack and the four screws in the battery compartment.
  3. Remove the sensor:
    1. Remove the two screws at the top of the instrument and pull the back shell apart from the front shell.
    2. Remove the screw from the manifold's corner.
    3. Pull back the two retaining clips on the manifold.
  4. Lift the manifold sideways, farthest from the pump.
  5. Pull the sensor manifold out from the pump manifold.
  6. Remove the desired sensor.
  7. Install the new sensor:
    1. Mind the orientation key, and then push the sensor down into manifold.
    2. Insert manifold inlet into pump manifold.
    3. Push the manifold down and as the two latches engage, listen for two snapping sounds.
  8. Replace the back shell and install the four screws back into the battery compartment torque 0.45-0.56 N-m (4-5 in-lbs).
  9. Reinstall the 2 screws at the top of the instrument torque 0.45-0.56 N-m (4-5 in-lbs).
  10. Attach the battery:
    1. To place the battery pack correctly, first engage hooks on the bottom of the battery pack.
    2. Push the battery pack into place and tighten the battery retention screw torque 0.45-0.56 N-m (4-5 in-lbs).
  11. Turn on the instrument and allow for the sensors to stabilize.

Replace the 4R+ Sensors

  1. Turn the instrument off.
  2. Remove the battery:
    1. Turn the instrument face down and unscrew on the battery pack.
    2. Remove the battery pack and the 4 screws in the battery compartment.
  3. Remove the sensor:
    1. Remove the two screws at the top of the instrument and pull the back shell apart from the front shell.
    2. Disconnect the 4R+ cable from the sensor.
    3. Push at the PRESS marking in the manifold toward the top of the instrument.
    4. Use the tabs on the sensor to pull it out of the manifold.
  4. Install the new sensor:
    1. Retrieve the new sensor and insert it into the manifold aligning the sensor guides to the instrument's slots.
    2. Push sensor forward until it stops moving.
    3. Re-connect the 4R+ sensor cable.
  5. Replace the back shell and install the four screws back into the battery compartment torque 0.45-0.56 N-m (4-5 in-lbs).
  6. Reinstall the two screws at the top of the instrument (torque 4-5 in-lbs).
  7. To place the battery pack correctly, first engage hooks on the bottom of the battery pack.
  8. Push the battery pack into place and tighten the battery retention screw torque 0.45-0.56 N-m (4-5 in-lbs).
  9. Turn on the instrument and allow for the sensors to stabilize.

Replace the Pump

  1. Turn the instrument off.
  2. Remove the battery:
    1. Turn the instrument face down and unscrew on the battery pack.
    2. Remove the battery pack and the four screws in the battery compartment.
  3. Remove the two screws at the top of the instrument and pull the back shell apart from the front casing.
  4. Remove the pump:
    1. Pull the pump out of the manifold assembly.
    2. Release the pump connector retainer clip and pull the pump connector out.
  5. Install the new pump:
    1. Retrieve the new pump and insert the pump connector.
    2. Fold the pump wires across the bottom of the pump motor.
  6. Insert the pump into the manifold assembly.
  7. Replace the back shell and install the 4 screws back into the battery compartment torque 0.45-0.56 N-m (4-5 in-lbs).
  8. Reinstall the two screws at the top of the instrument (torque 4-5 in-lbs).
  9. To place the battery pack correctly, first engage hooks on the bottom of the battery pack.
  10. Push the battery pack into place and tighten the battery retention screw torque 0.45-0.56 N-m (4-5 in-lbs).

Replace the Battery

  1. Turn the instrument off.
  2. Turn the instrument face down and unscrew on the battery pack.
  3. Remove the battery pack.
  4. To place the new battery pack correctly, first engage hooks on the bottom of the battery pack.
  5. Push the battery pack into place and tighten the battery retention screw torque 0.45-0.56 N-m (4-5 in-lbs).

Note: Use only the HU-BAT Honeywell battery pack.

Replace the Pump Inlet Filter

  1. Turn the instrument off.
  2. Unthread the screw on the pump inlet cover.
  3. Rotate the cover counter-clockwise.
  4. Remove the cover.
  5. Remove both the particulate and hydrophobic filter.
  6. First, Install both the hydrophobic and then the particulate filters.
  7. Place the filter cover and rotate the filter cover clockwise until stop.
  8. Tighten the screw torque to 0.34-0.45 N-m (3-4 in-lbs)

Replace the Pump Filter

  1. Turn the instrument off.
  2. Remove the battery:
    1. Turn the instrument face down and unscrew on the battery pack.
    2. Remove the battery pack and the four screws in the battery compartment.
  3. Remove the sensors manifold:
    1. Remove the two screws at the top of the instrument and pull the back shell apart from the front casing.
    2. Remove the screw from the manifold's corner.
    3. Pull back the two retaining clips on the manifold.
    4. Lift the manifold sideways, farthest from the pump, and then pull the sensor manifold out from the pump manifold.
  4. Gently detach the pump filter from the red restrictor.
  5. Pull up the new pump filter and remove it.
  6. Install the new pump filter:
    1. Place the new pump filter in the instrument.
    2. Gently attach the pump filter to the red restrictor.
  7. Insert manifold inlet into pump manifold.
  8. Push the manifold down and as the two latches engage, listen for two snapping sounds, and then reinstall the screw to the manifold's corner.
  9. Replace the back shell and install the four screws back into the battery compartment torque 0.45-0.56 N-m (4-5 in-lbs).
  10. Reinstall the two screws at the top of the instrument torque 0.45-0.56 N-m (4-5 in-lbs).
  11. Attach the battery:
    1. To place the battery pack correctly, first engage hooks on the bottom of the battery pack.
    2. Push the battery pack into place and tighten the battery retention screw torque 0.45-0.56 N-m (4-5 in-lbs).
  12. Turn on the instrument and allow for the sensors to stabilize.

Appendices

Auto detect gas

While the gas is applied during calibration, the detector will wait for up to 300 seconds to allow the gas to stabilize. If the gas has not stabilized by then, the detector will display a Gas unstable message. If the gas stabilizes within 300 seconds, it will be automatically detected and will not have to be selected from a menu. The name of the gas and a Span calibration in progress message will be displayed. If a quad gas mixture was used, the detector will display the names of all four gases.

Gas Alarm Setpoints

Gas alarms are activated when detected gas concentrations are above or below the user-defined setpoints. The gas alarms are described as follows.

Alarm Condition
Low Toxics and combustibles: Ambient gas level above low alarm setpoint. Oxygen: Ambient gas level may be set above or below 20.9% (or 20.8%).
High Toxics and combustibles: Ambient gas level above high alarm setpoint. Oxygen: Ambient gas level may be set above or below 20.9% (or 20.8%).
TWA Toxics only: Accumulated value above the TWA alarm setpoint.
STEL Toxics only: Accumulated value above the STEL alarm setpoint.
Multi-gas Two or more gas alarm conditions simultaneously.
Over Limit (OL) OL or -OL displays when readings are above or below the sensor detection range, respectively.

Sample Factory Alarm Setpoints

Following table lists alarm setpoints as defined by the US Occupational Safety and Health Association (OSHA).

Gas TWA STEL Low High
H₂S 10 ppm 15 ppm 10 ppm 15 ppm
CO 35 ppm 50 ppm 35 ppm 200 ppm
O₂ N/A N/A 19.5% vol. 23.5% vol.
LEL N/A N/A 10% LEL 20% LEL

Note: To disable an alarm, set the alarm setpoint to 0 (zero) in Fleet Manager II.

Specifications

Detector dimensions: 8.1 x 14.6 x 5.1cm
Mass: 444.2 g
Operating temperatures: -4°F to +122°F (-20°C to +50°C)
Battery operating times: 10 hours.
Rechargeable battery: 8 hours in a temperature range from 5 °C to 35 °C.
Storage temperature: -40°F to +122°F (-40°C to +50°C)
Operating humidity: 0% to 95% relative humidity (non-condensing)
Detection range:
H₂S: 0 – 100 ppm (1 / 0.1 ppm increments)
CO: 0 – 500 ppm (1 ppm increments)
O₂: 0 – 30.0% vol. (0.1% vol. increments)
Combustible (LEL): 0% to 100% LEL (1% LEL increments) or 0.0% to 5.0% v/v methane
Sensor type: Alarm conditions: STEALTH, TWA alarm, STEL alarm, low alarm, high alarm, multi-gas alarm, low battery alarm, confidence beep, automatic deactivation alarm
Audible alarm: 95 dB at 30 cm (1 ft.) (100 dB typical) variable pulsed beeper
Visual alarm: Red light-emitting diodes (LED)
Display: Alphanumeric liquid crystal display (LCD)
Screen resolution: 160X240 pixels.
Backlight: Activates when the pushbutton is pressed and deactivates after 5 seconds; also activates during an alarm condition
Self-test: Initiated upon activation.
Calibration: Automatic zero and automatic span.

Standards and Certifications

The Honeywell BW™ Ultra gas detector is in conformity with the following standards and certifications:

Approvals: Approved by UL to both U.S. and Canadian Standards UL 913, 8th Edition UL 60079-0, 6th Edition UL 60079-1, 7th Edition. UL 60079-11, 6th Edition ANSI/ISA 60079-29-1 (12.13.01) - 2013 CSA C22.2 No.152-M1984 (R2016) CSA C22.2 No. 60079-0:19 CSA C22.2 No. 60079-11:14 CSA C22.2 No. 60079-1:16

UL: E480011
Class I, Division I, Group A, B, C and D, Temperature code T4, -40 ≤ Tamb ≤ +50°C
Class I, Zone 0, AEx ia IIC T4 Ga, -40 ≤ Tamb ≤ +50°C (without LEL and IR sensor installed)
Class I, Zone 0, AEx da ia IIC T4 Ga, -40 ≤ Tamb ≤ +50°C (with LEL sensor installed and without IR sensor installed)
Class I, Zone 0, AEx ia IIC T4 Ga, -20°C ≤ Tamb ≤ +50°C (with IR sensor installed and without LEL sensor installed)
Class I, Zone 0, AEx da ia IIC T4 Ga, -20°C ≤ Tamb ≤ +50°C (with LEL and IR sensor installed)

CSA: E480011
Class I, Division I, Group A, B, C and D, Temperature code T4, -40 ≤ Tamb ≤ +50°C
Ex ia IIC T4 Ga, -40 ≤ Tamb ≤ +50°C (without LEL and IR sensor installed)
Ex da ia IIC T4 Ga, -40 ≤ Tamb ≤ +50°C (with LEL sensor installed and without IR sensor installed)
Ex ia IIC T4 Ga, -20°C ≤ Tamb ≤ +50°C (with IR sensor installed and without LEL sensor installed)
Ex da ia IIC T4 Ga, -20°C ≤ Tamb ≤ +50°C (with LEL and IR sensor installed)

ATEX: DEMKO 18 ATEX 1833X
EN IEC 60079-0:2018
EN 60079-11:2012
EN 60079-1:2014
EN 60079-26:2015
I M1 Ex ia I Ma, -20°C ≤ Tamb ≤ +50°C
II 1 G Ex ia IIC T4 Ga, -40°C ≤ Tamb ≤ +50°C (without LEL and IR sensor installed)
I M1 Ex da ia I Ma, -20°C ≤ Tamb ≤ +50°C
II 1 G Ex da ia IIC T4 Ga, -40°C ≤ Tamb ≤ +50°C (with LEL sensor installed and without IR sensor installed)
I M1 Ex db ia I Ma, -20°C ≤ Tamb ≤ +50°C
II 2 G Ex db ia IIC T4 Gb, -20°C ≤ Tamb ≤ +50°C (with IR sensor installed)

IECEx: UL 18.0061X
IEC 60079-0:2017
IEC 60079-11:2011
IEC 60079-1:2014
IEC 60079-26:2014
Ex ia I Ma, -20°C ≤ Tamb ≤ +50°C
Ex ia IIC T4 Ga, -40°C ≤ Tamb ≤ +50°C (without LEL and IR sensor installed)
Ex da ia I Ma, -20°C ≤ Tamb ≤ +50°C
Ex da ia IIC T4 Ga, -40°C ≤ Tamb ≤ +50°C (with LEL sensor installed and without IR sensor installed)
Ex db ia I Ma, -20°C ≤ Tamb ≤ +50°C
Ex db ia IIC T4 Gb, -20°C ≤ Tamb ≤ +50°C (with IR sensor installed)

The following additional previous editions of Standards noted under the "Standards" section of this Certificate were applied to integral Components as itemized below. There are no significant safety related changes between these previous editions and the editions noted under the "Standards" section.

Product Certificate Num Standards
Dynament Ltd. Gas Sensors Type MSH2ia*** IECEx FTZU 15.0002U IEC 60079-0 Edition 2011
City Technology Limited, Miniature Combustible Gas Sensor - 1 LEL 75 IECEx ULD 16.0016U IEC 60079-0 Edition 2011

JPEx
DEK21.0013X
JNIOSH-TR-46-1: 2015
JNIOSH-TR-46-6: 2015
Ex ia IIC T4 Ga, -40°C ≤ Tamb ≤ +50°C (without LEL and IR sensor installed)

FCC Compliance
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter. This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance.

NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

CAUTION Changes or modifications not expressly approved by the manufacturer responsible for compliance could void the user's authority to operate the equipment.

This device complies with FCC radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines. This device has very low levels of RF energy that it is deemed to comply without maximum permissive exposure evaluation (MPE).

RED Compliance
Honeywell Analytics Asia Pacific Co., Ltd. hereby declares that this gas detector, Honeywell BW™ Ultra, is in compliance with the essential requirements and other relevant provisions of Directive 2014/53/EU.

Canada, Industry Canada (IC) Notices
This device complies with Industry Canada license-exempt RSS. Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
This device complies with FCC/ISED radiation exposure limits set forth for an uncontrolled environment and meets the FCC/ISED radio frequency (RF) Exposure Guidelines. This device has very low levels of RF energy that it is deemed to comply without maximum permissive exposure evaluation (MPE).

Canada, avis d'Industrie Canada (IC)
Cet appareil est conforme aux normes RSS exemptes de licence d'Industrie Canada. Son utilisation est soumise aux deux conditions suivantes : (1) cet appareil n'engendre pas d'interférences, et (2) cet appareil doit tolérer tout type d'interférences, notamment les interférences pouvant provoquer une utilisation non désirée de l'appareil.
Cet équipement est conforme aux limites établies par FCC/ Industrie Canada en matière d'exposition aux radiations dans un environnement non contrôlé. Cet équipement ne doivent pas être colocalisés ou fonctionner en conjonction avec tout autre antenne ou émetteur.

In terms of North America flammable gas performance approval:
Honeywell BW™ Ultra is approved for ISA 60079-29-1 and CSA C22.2 No.152.
Only Honeywell BW™ Ultra catalytic bead flammable sensor was evaluated for CSA C22.2 No.152 and ISA 60079-29-1.
The evaluation is valid only with the pumping flow rate 300 ml/min, 3 m length tube and CH4 (Methane) gas.
The other options are not the scope of CSA C22.2 No.152 and ISA 60079-29-1.

For the compliance of CSA C22.2 No.152 and ISA 60079-29-1, the adjustable alarm point shall not exceed 60 % LEL and the highest alarm shall be configured as latching alarm.
In ISA 60079-29-1, Honeywell BW™ Ultra was tested only for IP54. Other IP ratings are not the scope of ISA 60079-29-1. Honeywell BW™ Ultra was pressure tested for 80, 100 and 120 kPa in ISA 60079-29-1. Outside of 80 - 120 kPa is NOT the scope of ISA 60079-29-1.

Label Information

BW Ultra Battery label

Various certification labels (cULus, ATEX, IECEx, JPEx) are shown for the battery. Each label includes markings related to safety standards, model information, and manufacturer details. A breakdown of the 2D barcode for serial number is provided, showing digits for sequence number, year of week, year of manufacture, revision number, plant code, and manufacturer address.

BW Ultra Instrument label

Various certification labels (cULus, ATEX, IECEx, JPEx) are shown for the instrument. Each label includes markings related to safety standards, class/zone protection, FCC/IC certificates, part number, and serial number. A detailed explanation of the markings on the cULus label is provided, including certificate marking, class division/zone protection markings for different sensor configurations, North America flammable performance approval, FCC and IC certificate numbers, part number, and serial number.

Part Number Format

BW Ultra's explosion-proof protection type varies per the sensor configuration. The user can notice the sensor configuration with the part number format.

B. (5 mm X 5 mm) 2D barcode for Part Number: 14 digits

The format includes:

C. (5 mm X 5 mm) 2D barcode for Serial Number: 18 digits

The format includes:

Sensor Replacement

A table lists sensor nomenclature, part numbers, and manufacturers.

Nomenclature Part Number Sensor
Sensor 1 Oxygen X1 10X, manufactured by City
AAW85-07WA-HON
2023B1099 Empty sensor, manufactured by City
Sensor 2_LEL Catalytic bead W1 1 LEL 75, manufactured by City
PM979-600
W2 1 LEL 75C, manufactured by City
PM989-600
W3 1 LEL 75M, manufactured by City
PM999-600
00 Empty sensor, manufactured by City
2023B1099
Sensor 3 H₂S H1 1HS, manufactured by City
AC400-R00A-HON
00 Empty sensor, manufactured by City
2023B1099
Sensor 4 CO M1 1HS, manufactured by City
AB010-R01A-HON
00 Empty sensor, manufactured by City
2023B1099
Sensor 5 Q1 C03-0912-103, manufactured by RAE Systems
C03-0912-103
W4 MSH2ia-LS/HC/5/V/P/F (Range 0-5 or 0-100 %Vol), manufactured by Dynament
50122982-084
B1 MSH2ia-LS/CO2/5/V/P/F (Range 0-5 or 0-100 %Vol), manufactured by Dynament
50122982-085
S1 2112B2015R, manufactured by City
50122982-086
A1 SensoriC NH3 3E 100 SE, manufactured by City
50122982-087
R1 4HYT, manufactured by City
50122982-088
C1 032-0121-000, manufactured by RAE Systems
50122982-089
D1 032-0112-000, manufactured by RAE Systems
50122982-090
Z1 4HN, manufactured by City
50122982-091
N1 032-0111-000, manufactured by City
50122982-092
M2 CO-AX, manufactured by Alphasense
50122982-210
00 Empty sensor, manufactured by RAE Systems
50122982-203

Troubleshooting

This section provides solutions for common problems encountered with the detector.

Problem Probable Cause Solution
Detector does not display normal gas reading after startup sequence Sensor not stabilized
Sensor requires calibration
Target gas is present
Used sensor: wait 60 seconds.
New sensor: wait 5 minutes.
Calibrate the detector.
Detector is operating properly. Use caution in suspect areas.
Detector does not respond to button Battery is in critical low battery state or is completely depleted
Detector is performing operations that do not require user input
Replace the battery.
Button operation restores automatically when the operation ends.
Detector does not accurately measure gas Sensor requires calibration
Detector is colder/hotter than gas temperature
Sensor filter is blocked
Calibrate the detector.
Allow the detector to acquire ambient temperature before use.
Clean the sensor filter.
Detector does not enter alarm Alarm setpoint(s) are set incorrectly
Alarm setpoint(s) set to zero
Detector is in calibration mode
Reset the alarm setpoints.
Reset the alarm setpoints.
Complete calibration.
Detector intermittently enters alarm without reason Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of hazardous gas Detector is operating normally. Use caution in suspect areas. Check MAX gas exposure reading.
Features and options are not operating as expected Alarm setpoints are set incorrectly
Detector requires calibration
Missing or faulty sensor
Changes in Fleet Manager II
Reset the alarm setpoints.
Calibrate the detector.
Change the sensor.
Verify the setup in Fleet Manager II.
Battery has been charging for 6 hours. Charging indicator on LCD shows the battery is still charging Battery is trickle charging Verify the charger is properly connected to the AC outlet.
Battery indicator does not display when charging Battery is depleted below normal levels Replace the battery.
Battery does not charge Replace the battery.
Sensor fails to zero during startup self-test Depleted battery.
Damaged detector.
Change the sensor.
Replace the battery.
Contact Honeywell.
The detector does not activate.
Detector automatically deactivates. Automatic deactivation due to critical low battery.
Lockout on Self-Test Error is enabled and a sensor(s) has failed the startup self-test.
Sensor(s) require calibration.
Replace the battery.
Change the sensor.
Calibrate the detector.

Glossary

ACGIH: The ACGIH method is defined as the infinite (total) accumulated average, whether it is 2 hours or 8 hours.

BLE: Bluetooth Low Energy.

Bump test: A compliance test that confirms the detector's ability to respond to target gases by exposing the detector to a known gas concentration. Other procedures that are specified to occur automatically when the detector is inserted into a docking module may be performed in conjunction with the bump test.

Calibration: A two-step compliance test that determines the measurement scale for the detector's response to gas. In the first step, a baseline reading is taken in a clean, uncontaminated environment. In the second step, the sensors are exposed to known concentrations of gas. The detector uses the baseline and known gas concentrations to determine the measurement scale.

Datalog: A datalog is a file that contains detailed, date-stamped records related to detector operations and configuration settings. The datalog is continuously updated. Records that span the operating life of the detector are retained.

Event log: An event log is a file that contains detailed, time-stamped records data related to gas events, and compliance tests. The event log is updated when an event occurs. A specified number records for the most recent events are retained.

Fleet Manager: A proprietary, Windows-based software developed by Honeywell to configure and manage docking modules, calibration, bump tests, and data logs. Fleet Manager II is available for download from www.honeywellanalytics.com

Gang: A group of two to five connected IntelliDoX modules. Modules that are connected share power, network and gas connections.

IntelliDoX Docking Module: An automatic bump test and calibration docking station for use with portable gas detectors manufactured by Honeywell.

IR: Infrared. IR is an invisible radiant energy that can be used for short- range wireless communications between enabled devices.

LCD: Liquid crystal display. LCD is a technology that is commonly used for display screens on mobile digital devices.

Normal atmosphere: A fresh air environment with 20.9% v/v oxygen (O₂) that is free of hazardous gas

Operating Life: The period of operational use required to attain the specified operating limit. Operating life includes normal operating time, alarm time, and all types of idle time.

OSHA: The US OSHA method is defined as a moving average that accumulates over an 8-hour average. If the worker is in the field longer, the oldest accumulated values (first hour) are replaced by the newest values (ninth hour). This continues for the duration of the work shift until the detector is deactivated.

PPM: Parts per million, a measure of concentration.

Reboot: Restart the operating system for the module

Service Life: The expected lifetime of a product, as specified by the manufacturer.

Station: An area or zone dedicated to a specific activity. A compliance testing station may contain several IntelliDoX modules and gangs of connected modules.

Stealth mode: When enabled, the backlight, visual alarms, and audible alarms are disabled. During an alarm, the vibrator activates and readings are displayed on the LCD.

STEL: The Short-Term exposure Limit is the maximum permissible gas concentration a worker can be safely exposed to for short periods of time (5-15 minutes maximum).

TWA Alarm: The time-weighted average (TWA) is a safety measure used to calculate accumulated averages of gases. Using the US Occupational Safety and Health Administration (OSHA) method or the American Conference of Governmental Hygienists (ACGIH) method, an average is calculated to ensure the detector alarms when the TWA has accumulated.

V/V: Volume percent concentration.

Contact Honeywell

Corporate Headquarters
Honeywell Analytics
Suite 110, 4411-6 St SE
Calgary, Alberta
Canada T2G 4E8
Toll free: 1-888-749-8878

United States
Honeywell Analytics
405 Barclay Boulevard
Lincolnshire, Illinois
USA 60069
Toll free: 1-888-749-8878

Asia
Honeywell Analytics Asia Pacific
7F SangAm IT Tower,
434 Worldcup Buk-ro, Mapo-gu,
Seoul 03922, Republic of Korea
Tel: +82 (0) 2 6909 0300
Analytics.ap@honeywell.com

Europe
Honeywell Analytics
Javastrasse 2
8604 Hegnau
Switzerland
Toll free: 00800-333-22244
Other countries, toll free: 1-403-248-9226
Bwa.customerservice@honeywell.com

sps.honeywell.com

Models: BW Ultra, Portable Five-gas Detector with Internal Pump, BW Ultra Portable Five-gas Detector with Internal Pump

File Info : application/pdf, 38 Pages, 1.69MB

PDF preview unavailable. Download the PDF instead.

sps-his-50122982-265-en-revd-bw-ultra-user-manual ?download=false Adobe PDF Library 16.0 Adobe InDesign 16.4 (Windows)

Related Documents

Preview Honeywell BWTM Ultra Portable Gas Detector User Manual - Operation, Maintenance, Safety
Detailed user manual for the Honeywell BWTM Ultra portable multi-gas detector. Covers setup, calibration, bump testing, maintenance, sensor replacement, troubleshooting, and safety guidelines for industrial gas monitoring.
Preview Honeywell Portable Gas Detection Price List 2021
Official 2021 price list for Honeywell's portable gas detectors including BW Clip Series, BW Solo, BW Icon, BW MicroClip, GasAlert Quattro, BW Max XT II, GasAlertMicro 5 Series, BW Ultra, plus IntelliDoX and Safety Suite systems.
Preview Honeywell BW Ultra Gas Detector: Quick Reference Guide & Safety
Get essential information on the Honeywell BW Ultra portable gas detector. This quick reference guide covers setup, safety precautions, operation, and alarms for reliable gas detection.
Preview Honeywell BW Solo User Manual: Portable Single-Gas Detector Guide
This user manual provides comprehensive instructions for operating, maintaining, and calibrating the Honeywell BW Solo portable single-gas detector. Learn about its features, safety protocols, and troubleshooting.
Preview Honeywell BW Solo Portable Single-Gas Detector User Manual
Comprehensive user manual for the Honeywell BW Solo portable single-gas detector, covering operation, maintenance, safety, specifications, and troubleshooting.
Preview Honeywell BW™ Icon & BW™ Icon+ Portable Gas Detector Quick Reference Guide
A quick reference guide for operating, testing, and calibrating the Honeywell BW™ Icon and BW™ Icon+ portable gas detectors, covering device setup, user interface, common operations, and safety information.
Preview Honeywell BW Solo User Manual: Portable Single-Gas Detector Guide
This user manual provides comprehensive guidance for the Honeywell BW Solo portable single-gas detector, covering its operation, safety features, maintenance, calibration procedures, and technical specifications.
Preview Honeywell BW™ Flex Series Gas Detector User Manual
Comprehensive user manual for the Honeywell BW™ Flex Series gas detector, covering setup, operation, maintenance, and safety information. Learn about its features, certifications, and troubleshooting.