3M™ Prevena™ 125 Therapy Unit Indicators and Alerts
Quick Guide
This is a quick guide; further details are in the 3M™ Prevena™ Therapy Clinician Guide.
General Indicators:
⚠️ Caution Indicator? Audible Alarm (muted)? Battery Level
Alerts and How to Fix
Alert Type | How to Fix |
---|---|
Leak Alert One beep, one solid yellow light. | To ensure proper application of the 3M™ Prevena™ Incision Dressing, the Prevena 125 Therapy Unit provides the Visicheck™ Feature. Double press ON/OFF button.
|
Canister Full Alert Two beeps, one solid yellow light. | Check fluid level in the canister:
|
Low Battery Alert Low: One slow beep, one solid yellow light. Critical: One beep, repeating rapidly, increasing volume, two solid yellow lights. | This therapy unit comes with three AA batteries.
|
System Error Alert One beep, repeating rapidly, increasing in volume, two solid yellow lights. |
|
Device Lifecycle Expired Three lights show life cycle is up. There will be a repeated beep for 15 seconds and then the unit will shut off. | Lifecycle: 192 hours (eight days). |
Blockage Alert Audible blockage alert will sound two beeps repeating every 15 seconds. | Check for a blockage:
|
Leak Alert Leak Alert will sound two beeps repeating every 15 seconds. | Ensure canister is security locked onto the therapy unit. When canister is installed, a distinct click will be heard. Check tubing connectors to ensure they are fully engaged and locked. Press down through drape onto the adhesive seal all the way around to ensure seal. Use fingers to smooth out any creases or wrinkles. Use 3M™ Prevena™ Patch Strips or excess drape along the outer edge of drape to seal leaks. |
Low Battery Alert Alert will sound two beeps repeating every four minutes. | Power Connected Status:
|
System Fault Alert All LEDs flashing and two beeps every 15 seconds. | Turn therapy unit off then on. If the therapy is still not running, call 1-800-668-5403. |
Alert Mute Button | Press and hold center of the alert mute button for three seconds during an alert condition to silence the audible alert for two minutes. When pressed, the alert mute button will illuminate to indicate mute has been selected. The alert will re-occur after two minutes unless the alert condition has been corrected. |
Therapy Ended | A solid yellow LED at the top of the Therapy Life Indicator will turn on.
|
Contact and Important Information
For assistance, call 1-800-668-5403.
Available in Canada from your authorized 3M-KCI distributors.
Company Information:
KCI Medical Canada Inc.
75 Courtneypark Dr W, Unit 4
Mississauga, ON L5W 0E3
KCI USA, Inc., a 3M Company
KCI USA, INC.
12930 IH 10 West
San Antonio, TX 78249
Note: Specific indications, limitations, contraindications, warnings, precautions, and safety information exist for these products and therapies. Please consult a clinician and product Instructions For Use prior to application. This material is intended for healthcare professionals.
3M and Prevena are trademarks of 3M. Used under license in Canada. © 2022, 3M. All rights reserved. Other marks shown are marks and/or registered marks of their respective owners. Unauthorized use prohibited. 2204-23874 E.