Hach Polymetron 9582 Dissolved Oxygen System with HART Communications, 24 V DC
Product #: 9582.99.75P4
AED Price: Contact Hach
Overview
The Hach Polymetron 9582 Dissolved Oxygen System with HART Communications is an integral part of a comprehensive water analytics system for the Power industry. Hach offers a wide range of products designed to work together, providing flexible solutions for unique needs. This approach saves time on design, installation, training, maintenance, and operation.
- Save Time on Design: A single design source and product platform streamline the process of finding design files and configuring components, allowing for the creation and reuse of optimal design templates.
- Accelerate Your Installation: Interchangeable electronic components, a common user interface, and a unified support team simplify installation. User settings can be quickly transferred between dissolved oxygen loops.
- Reduce Training Complexity: A single platform minimizes the time required to learn product operations, enabling faster deployment of new systems.
- Simplify Maintenance and Operation: Common menu guides and step-by-step procedures for maintenance and calibration reduce variability. Standard visual alerts notify operators of required troubleshooting. Start-up and maintenance are expedited with pre-mounted membrane caps and factory pre-conditioned sensors.
Specifications
- Analogue output functional mode:
- Linear, Logarithmic, Bi-linear, PID
- Cable length:
- 10 m (33 ft)
- Calibration method:
- Zero: Electrically or with oxygen free water, Slope: in air or against a laboratory measurement
- Communication capabilities:
- Hart
- Communication: digital:
- MODBUS RS232/RS485, PROFIBUS DPV1, HART optional
- Conduit openings:
- 1/2" NPT Conduit
- Connection drain line:
- 8 mm tubing
- Connections:
- 1/4" NPT thread (6mm or 1/4" tubing advised)
- Detection limit:
- < 1 ppb
- Display:
- Graphic dot matrix LCD with LED backlighting, transreflective
- Display resolution:
- 240 x 160 pixels
- Display size:
- 48 x 68 mm
- Electrical Certifications:
- EMC
- CE compliant for conducted and radiated emissions:
- CISPR 11 (Class A limits), EMC Immunity EN 61326-1 (Industrial limits)
- Safety:
- CAN/CSA C22.2 No. 61010-1
- Enclosure waterproof rating:
- IP66/NEMA 4X
- Flow rate:
- 66-166
- Maintenance interval:
- Membrane Lifetime: 6 months depending on sample
- Manual languages:
- Bulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
- Material:
- Polycarbonate, Aluminium (powder coated), Stainless Steel
- Measurements:
- 0 to 2000 ppb (0-2 ppm)
- Measuring range:
- 0 - 2000 ppb (0-2 ppm)
- Operating temperature range:
- -20 - 60 °C at 0 - 95% RH (non-condensing)
- Power requirements (Voltage):
- 24 V DC
- Relay functions:
- Scheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control, and Warning
- Relay: Operational mode:
- Primary or secondary measurement, calculated value (dual channel only) or timer/scheduler
- Relays:
- Four electromechanical SPDT (Form C) contacts, 1200 W, 5 A
- Repeatability:
- ± 0.5 ppb or ±5%, whichever is greater
- Reproducibility:
- ± 0.5 ppb or ± 2% whichever is greater
- Response time:
- For step change 1-40 ppb: <30s
- Security levels:
- 2
- Storage conditions:
- -20 °C - 70 °C
- Temperature compensation:
- Automatic in the range of 0 - 45 °C (32 - 113 °F)
- Units:
- mg/L, ppm, µg/L, ppb, mbar, hPa, inch Hg, mmHg
- Weight:
- 3.2 kg
What's included?
Controller, mounting hardware, oxygen sensor, temperature sensor, cable, flow-chamber, basic user manual.